Abstract
We investigated a spectral-contrast-angle (θ) method to determine whether mass spectra of structural isomers are the same or significantly different. This method represents collisionally activated dissociation (CAD) spectra as vectors in space. Mass spectra of different isomers are represented as different vectors, having characteristic lengths and direction. The derived spectral contrast angle, which is a measure of the angle between two vectors corresponding to two closely related spectra, is a measure of whether the mass spectra are the same or significantly different. We compare this method with the similarity index (SI) method and show that the spectral contrast angle method is superior and can differentiate between very similar spectra in cases where the SI cannot. Both methods can be implemented simply in situations where the analyst is called on to decide, on the basis of mass or product-ion spectra, whether reference and unknown compounds are the same or to evaluate the reproducibility of spectra comprised of many peaks.
Article PDF
Similar content being viewed by others
References
Lay, J. O., Jr.; Gross, M. L.; Zwinselman, J. J.; Nibbering, N. M. M. A Field Ionization and Collisionally Activated Dissociation/Charge Stripping Study of Some [C9H10]+. Ions. Org. Mass Spectrom. 1983, 18, 16–21.
Leek, M. R.; Dorman, M. F.; Summerfield, Q. Minimum Spectral Contrast for Vowel Identification by Normal-Hearing and Hearing-Impaired Listeners. J. Acoust. Soc. Am. 1987, 81, 148–154.
Lucey, P. G.; Blewett, D. T.; Taylor, G. J.; Hawke, B. R. Imaging of Lunar Surface Maturity. J. Geophys. Res. [Planets] 2000, 105(E8), 20377–20386.
Wald, A. E.; Kaufman, Y. J.; Tanre, D.; Gao, B.-C. Daytime and Nighttime Detection of Mineral Dust Over Desert Using Infrared Spectral Contrast. J. Geophys. Res. [Atmos.] 1998, 103(D24), 32307–32313.
Swartz, M. E.; Brown, P. R. Use of Mathematically Enhanced Spectral Analysis and Spectral Contrast Techniques for the Liquid Chromatographic and Capillary Electrophoretic Detection and Identification of Pharmaceutical Compounds. Chirality 1996, 8(1), 67–76.
Warren, W. J.; Stanick, W. A.; Gorenstein, M. V.; Young, P. M. HPLC Analysis of Synthetic Oligonucleotides Using Spectral Contrast Techniques. BioTechniques 1995, 18(2), 282–287.
Young, P. M.; Gorenstein, M. V. Tryptic Mapping by Reversed-Phase HPLC with Photodiode-Array Detection Incorporating the Spectral-Contrast Technique. LC-GC 1994, 12(11), 832–838.
Maga, J. A.; Johnson, D. L.; Morini, G. Major Volatiles in Toasted Sesame Seed Oil. J. Food Lipids 1995, 4(2), 259–268.
Gatlin, C. L.; Eng, J. K.; Cross, S. T.; Detter, J. C.; Yates, J. R., III. Automated Identification of Amino Acid Sequence Variations in Proteins by HPLC/Microspray Tandem Mass Spectrometry. Anal. Chem. 2000, 72(4), 757–763.
Stein, S. E.; Scott, D. R. Optimization and Testing of Mass Spectral Library Search Algorithms for Compound Identification. J. Am. Soc. Mass Spectrom. 1994, 5, 859–866.
Rosenthal, D; Bursey, J. T. 20th Annual Conference on Mass Spectrometry and Allied Topics; Dallas, TX 1972; Paper T4, p 419.
Drahos, L.; Vekey, K. Quantification of Isomeric Differences in Mass Spectra. Rapid Commun. Mass Spectrom. 1996, 10, 1309–1315.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Wan, K.X., Vidavsky, I. & Gross, M.L. Comparing similar spectra: From similarity index to spectral contrast angle. J Am Soc Mass Spectrom 13, 85–88 (2002). https://doi.org/10.1016/S1044-0305(01)00327-0
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1016/S1044-0305(01)00327-0