Introduction

The efficiency of an activated sludge plant depends on the ability of the biomass to mineralize biodegradable organics at a desire rate so as to meet effluent standards. The oxygen uptake, in this regard, reflects the ability of the biomass to metabolize wastewater components (Hill et al. 1998). Respirometry, which is widely used for the monitoring of oxygen uptake in different biochemical processes provides a procedure to assess the state of microbial activity and is used for the determination of the parameters of microbial kinetic models including the characterization of wastewater and activated sludge (Marsili and Tabani 2002; Young and Cowan 2004).

Respirometric technique was initiated in studying the biodegradability of wastewater as early as in 1920s. The instrument designed for the purpose is known as Barcroft-Warburg device. It represents a simple system consisting of a liquid reactor connected with a manometer (Jowett 1928; Perkins 1943). The biodegradation process takes place in the liquid phase of the reactor with the consumption of oxygen. The oxygen is transferred from the gas phase to the liquid, resulting in a loss of pressure in the gas phase. This pressure-loss is monitored by the manometer.

Respiration can be considered as a general measure of the microbial activity. It can provide a reliable and scientifically sound assessment of biodegradation criteria. For this reason, respirometry (oxygen uptake rate and/or carbon dioxide evolution rate) has been widely used as an important tool in biomonitoring fields. A wide range of respirometric protocols has been reported based either on carbon dioxide production, oxygen uptake or release of heat. The most common methods are those based on oxygen uptake (Ghosh et al. 1985; Mitz and Newman 1988). Through the years, different kinds of oxygen uptake measuring technique have become available and the respirometers have been upgraded. Commercially available or user-built respirometer systems usually fall into two major categories: (1) dissolve oxygen (DO) depletion respirometer and (2) headspace gas respirometer (Young and Cowan 2004). The first type is based on measuring DO concentration in the liquid phase by an electrochemical DO-sensor and a DO mass balance is made over the liquid phase only. The second type is based on measuring oxygen concentration in the gas phase and the oxygen balance is made over the whole gas–liquid system. Gaseous oxygen is measured by physical methods such as the paramagnetic, manometric and electrolytic methods. As the biodegradation process takes place in the liquid phase, it may seem more advantageous for analysis to acquire oxygen concentration data in the liquid phase. Measurement of oxygen in the liquid phase, however, is associated with interference from contaminants common in the liquid phase (e.g. formation of sludge film on the sensor) (Vanrolleghem 2002). For this reason, the measurement in the gas phase, to some extent, is still the preferred option. In respirometry, however, emphasis is given on the oxygen consumption measurement, the amount of carbon dioxide generated as by-product during degradation process plays an important role in overall bioactivity monitoring test. Moreover, in headspace gas respirometry, solubility of carbon dioxide in water or its absorption by the caustic scrubber significantly affects the respirometric measurement. However, mixing rates in biodegradation test that are sufficient for meeting oxygen transfer limits certainly are sufficient to allow rapid equilibrium between dissolved and headspace gases carbon dioxide (Young and Cowan 2004).

Development of instrumentation and automation has also widened the application possibilities of respirometers in terms of the optimization and operation of wastewater treatment systems, waste management, biodegradation, and kinetic parameter studies (Logan and Wagenseller 1993; Mahendraker and Viraraghavan 1995; Wu et al. 2004; Kim and Park 2005; Roppola 2009). The respirometers can provide an effective method for quickly acquiring data for toxicity screening (Tzoris and Hall 2006), organic removal rate (Edgardo et al. 2008), sludge activity and biological process modeling (Ince et al. 2008). Thus, the technique can be a useful tool for rapid characterization of wastewater and activated sludge in every small or large activated sludge plant. In spite of the large opportunities that the respirometric technique provides, the BOD5 and COD are still the most widely used parameters for the monitoring of influent and effluent wastewater quality of a treatment plant in the developing country (DOE 2008; Khan et al. 2009; Panda 2011). In fact, these parameters are also the measures of oxygen consumption by the wastewater, but unlike respirometric technique they do not provide information about the biodegradation kinetics. The measurement of BOD5 is a time-consuming method and the 5-day time had been accepted rather somewhat arbitrarily and without any well-argued scientific background (Min et al. 2004). The COD-measurement, on the other hand, is rapid, but its preciseness and reproducibility depend much on the procedure followed and also on the operator-skill. The respirometric method with its opportunities for controlling biodegradation kinetics at any time has got advantages over traditional method of BOD5 and COD control, but still did not find much application in the laboratories associated with activated sludge plants. The probable reasons may be (1) expensive apparatus required for the application of respirometric method, (2) necessity of skilled and efficient operating personnel to interpret the acquired data and (3) too little data for comparison between respirometric and BOD5/COD method to convince someone about the advantages of the first on the later (Wenxiang et al. 2008; Rahman and Islam 2013).

In the past decades developing nations, specially Bangladesh, have made much progress in building their industrial base and are installing biological treatment units (DOE 2008; Sultana et al. 2013). For proper maintenance and successful operation of the plants, the associated laboratories are to be equipped with inexpensive facilities with low maintenance cost. The purpose of the present paper is to present a low-cost headspace gas respirometric device for measuring the oxygen and carbon dioxide concentration in the gas phase by manometric method. However, there are a number of commercial respirometers available in the market having different features and functional objectives. The present one may be considered as a cost-effective simplest form appropriate for the developing countries. A comparative description of the major features of some common batch respirometers is given in Table 1. The performance of the respirometer has been tested with a standard glucose-glutamic acid (GGA) solution and the oxygen uptake data acquired from the biodegradation of the solution have been compared with those reported in the literature (Lee and Oswald 1954; Kessick 1976; Min et al. 2004). The theoretical ground of the measurement of oxygen depletion in the system has been discussed in a comprehensive manner and the calculation method has been summarized in a sequential order. Similarly, a concise calculation method for carbon dioxide generation is also presented in the paper. Efforts have been made to make a parallel between the traditional BOD concept and the oxygen consumption measured by the present respirometer. The method of determination of the Monod kinetic parameters based on the respirometric data has also been illustrated.

Table 1 Features of some commonly available batch respirometers

Although batch type closed respirometer has been widely used for treatment plant monitoring and controls for its simplicity and low-cost (Kim et al. 2001; Young and Cowan 2004; Tzoris and Hall 2006; Roppola 2009), due to its operation under oxygen limiting condition, it does not fully resemble the actual plant situation that usually exists in a suspended growth system in form of activated sludge. In that case, sensor-based open and/or continuous respirometry may provide good approximation which is not achievable through the present respirometer. However, proper scaling-up of the laboratory findings obtained from this respirometer may give satisfactory results for large-scale treatment plant considering the kinetic trends, thermodynamics and hydrodynamics of fluid flow properly (Doble and Gummadi 2007).

Description of the apparatus

The present apparatus, named as ‘BIOSUST multi-reactor respirometric system (BMRS), is a specially designed device based on the principle of headspace gas respiromerty with an option of measuring oxygen consumption and indirect estimation of carbon dioxide evolved during biodegradation process. The BMRS consists of six respirometric units (RU), each of which can operate independently. The oxygen consumption of the blank (consists of oxygen saturated dilution water along with seeding biomass) and the sample solutions with different dilutions can be studied simultaneously with the BMRS. Each RU comprises a constantly stirred batch reactor (CSBR) and an open tube manometric unit (Fig. 1). The reactor contains the sample solution stirred by a magnetic bar. The carbon dioxide scrubber is held in a hanging position. The gas space is connected with the manometer.

Fig. 1
figure 1

Schematic diagram of a single respirometric unit (RU). 1 Water bath, 2 Constantly stirred batch reactor (CSBR), 3 Magnetic stirrer with stir-bar, 4 CO2 absorber, 5 Thermometer, 6 Connecting tubing, 7 Control valve, 8 Measuring arm of the manometer and 9 Manometric liquid reservoir

The CSBR of all the six respirometric units are placed in a common water bath, the temperature of which is maintained constant by a temperature control system. The body structure of the BMRS is organized in two sections (Fig. 2): (1) reaction section (situated on the rear side of the apparatus) containing the six reactors with all accessories and (2) measuring section (situated on the front side of the apparatus) containing the six manometers.

Fig. 2
figure 2

The BMRS with six measuring units at the front side and the reaction section (partially shown) at the rear side

Reaction section

The body structure of this section represents a rectangular box of 18-gauge metal sheet with a polished white coating surface. A 5-mm thick glass reservoir is placed within the unit with an automatic aquarium water heater of 50 W to serve as a temperature control unit for the CSBRs. The magnetic stirrer is made of six small cooling fans of 12 volt, and powerful magnets are placed at the bottom part of the unit. A power adaptor is connected to the back outer surface of the unit to supply DC current to the magnetic stirrers from a 220 AC source. Six conical flasks fitted with rubber stoppers and carbon dioxide scrubbers are used as experimental reaction cell. The reactors are made airtight so that no gases are leaked out from the system during the study.

Measuring section

This section constitutes a vertical plastic panel sheet holding the manometers in the vertical position. The two arms of the manometers are fixed at the two faces of the panel. The measuring arm of the well-type manometers is constructed with a 2-mm laboratory pipette (located at the front side of the panel sheet) and is connected with a 5-ml plastic syringe as manometric liquid reservoir (located at the rear side of the panel sheet). Changes in the capillary height in the measuring arm do not make appreciable changes in the liquid-height of the reservoir arm. The open end of the measuring arm (volumetric pipette) is attached to the rubber stopper of the reaction flask through a 5-mm rubber tube. The oxygen uptake is read from the manometer arms at the front side. The technical specification of the BMRS is presented in Table 2.

Table 2 Technical specification of the BMRS

Theoretical ground of oxygen consumption measurement

Assumptions for the treatment of the experimental data

The change in volume is monitored in the manometer tube as a function of time and the data is treated under the following assumptions:

  1. 1.

    The components of the gases in the headspace system follow ideal gas laws.

  2. 2.

    The partial pressure of the water vapor in the headspace is equal to its saturated vapor pressure, psw, at the working temperature throughout the experimentation.

  3. 3.

    The oxygen consumption from the liquid phase is compensated by the transfer of oxygen from the liquid phase at such a rate that the oxygen concentration in the gas and the liquid phase are always in equilibrium and follow Henry law of gas solubility in liquid.

  4. 4.

    Carbon dioxide evolved in the process is totally absorbed by the scrubber.

  5. 5.

    No loss of gases are occurred during experimentation.

Formulae for the calculation of oxygen depletion

The changes in the amount of oxygen in the gas phase (correspondingly the partial pressure of oxygen) is calculated based on the volume and pressure variation data. Then the oxygen concentration in the liquid phase at every moment is calculated based on the assumption 3. Finally, the total amount of oxygen in gas and liquid phase altogether is calculated for every moment during the test course.

Gas phase

Referring to Fig. 1, the total number of moles of all the gases and vapors present at any time t in the gas phase of the reactor is given by the following relation:

$$n_{\text{s(g)}} = n_{\text{A(g)}} + n_{\text{w(g)}} + n_{\text{I(g)}}$$
(1)

where n stands for the number of moles and the subscripts s, A, w and I stand, respectively, for sum-total of the number of moles of all species, oxygen, water vapor and all other components of the gas mixture. The subscript A is used to denote ’Oxygen’ instead of easily noticeable symbol O2 simply to avoid clumsiness of the notations. The subscript g indicates, gas phase’.

Initially (at t = 0), the corresponding number of moles are given by the following relations in accordance with ideal and Dalton’s gas law:

$$n_{\text{s(g)}}^{0} = P^{0} V_{{ ( {\text{g)}}}}^{0} /(RT),\,\,\,n_{\text{A(g)}}^{0} = y_{\text{A}}^{0} n_{\text{s(g)}}^{0} ,\,\,\,n_{\text{w(g)}}^{0} = a_{\text{w}} n_{\text{s(g)}}^{0} \,\,\,\,{\text{and}}\,\,\,\,n_{\text{I(g)}}^{0} = n_{\text{s(g)}}^{0} (1 - y_{\text{A}}^{0} - a_{\text{w}} )$$
(2-5)
$${\text{With }}\,\,\,\,\,a_{\text{w}} = p_{\text{sw}} /P^{0}$$
(5a)

where the superscript 0 indicates parameter at t = 0, V0(g) is the initial volume of the gas phase in the headspace (m3), R is the universal gas constant, J/mol.K and T is the working temperature, K. P0 is the atmospheric pressure (assumed to be equal to 1.013 × 105 Pa) and y0A is the initial mole composition of oxygen in the air and is assumed to be equal to 0.21.

Assuming that the liquid level in the manometric liquid reservoir (see Fig. 1) does not change appreciably, at a given time t, the total number of moles in the gas phase, ns(g), is given by the following relation:

$$n_{\text{s(g)}} = n_{\text{s(g)}}^{0} (1 - a_{\text{h}} \Delta v)(1 - a_{\text{v}} \Delta v)\,\,\,\,\,\,{\text{with}}\,\,\,\,a_{\text{h}} = \rho\text{g}/(S_{\text{m}} P^{0} )\,\,\,\,{\text{and }}\,\,a_{\text{v}} = 1/V_{{ ( {\text{g)}}}}^{ 0}$$
(6)

where g is the acceleration due to gravity (m/s2), Sm is the cross-sectional area of the manometer tube (m2), ρ is the density of the manometer-liquid (kg/m3) and Δv is the change in volume in the gas phase measured by the manometric tube (m3).

Expanding Eq. (6) and neglecting the term containing square and higher order of Δv, the Eq. (6) is reduced to

$$n_{\text{s(g)}} \approx n_{\text{s(g)}}^{ 0} \left[ {1 - (a_{\text{h}} + a_{\text{v}} )\Delta v} \right]$$
(6a)

The number of moles of water vapor, inert gases and oxygen (A) in the gas phase at any time t are given by the following relations:

$$n_{\text{w(g)}} = n_{\text{w(g)}}^{ 0} (1 - a_{\text{v}} \Delta v) = a_{\text{w}} n_{\text{s(g)}}^{ 0} (1 - a_{\text{v}} \Delta v)\,\,\,\,{\text{and}}\,\,\,\,n_{\text{I(g)}} = n_{\text{I(g)}}^{ 0}$$
(7, 8)

and

$$n_{\text{A(g)}} = n_{\text{s(g)}} - n_{\text{w(g)}} - n_{\text{I(g)}} = n_{\text{s(g)}}^{ 0} (y_{\text{A}}^{ 0} - a_{\text{n}} \Delta v)\,\,\,\,{\text{with}}\,\,\,\,a_{\text{n}} = a_{\text{h}} + a_{\text{v}} - a_{\text{w}} a_{\text{v}}$$
(9)

The number of moles, ΔnA(g), (and the corresponding mass ΔmA(g)) of oxygen depleted in the gas phase are given by the following relations:

$$\Delta n_{\text{A(g)}} = n_{\text{s(g)}}^{ 0} a_{\text{n}} \Delta v\,\, \Rightarrow \Delta m_{\text{A(g)}} = a_{\text{m(g)}} \Delta v\,\,\,\,\,\,\,{\text{with}}\,\,\,\,a_{\text{m(g)}} = n_{\text{s(g)}}^{ 0} a_{\text{n}} M_{\text{A}}$$
(10, 11)

where MA is the molecular mass of oxygen (= 32 × 10−3 kg/mol).

Liquid phase

The concentration of oxygen, CA(L) (kg/m3) in the aqueous phase (denoted by the subscript L) initially and at any time t are related to its partial pressure by Henry law as follows:

$$C_{\text{A(L)}}^{\text{o}} = P_{\text{A}}^{ 0} \,/H\,\,\,\,{\text{and }}\,\,\,\,C_{\text{A}} = P_{\text{A}} /H$$
(12, 13)

where H (Pa.m3/kg) is the Henry constant at the temperature T, and PA is the partial pressure of oxygen.

The partial pressure of oxygen (A) initially and at any time t are given by the following relations:

$$P_{\text{A}}^{ 0} = y_{\text{A}}^{ 0} P^{ 0} \,\,\,{\text{and}}\,\,\,\,\,\,P_{\text{A(g)}} = \frac{{n_{\text{A(g)}} RT}}{{V_{{ ( {\text{g)}}}}^{ 0} - \Delta v}} = P^{0} (y_{\text{A}}^{ 0} - a_{\text{n}} \Delta v)(1 - a_{\text{v}} \Delta v)^{ - 1} \,\,$$
(14, 15)

Expanding the last term of the Eq. 15 and neglecting square and higher terms of Δv, we obtain

$$P_{\text{A(g)}} = P^{0} (y_{\text{A}}^{ 0} - a_{\text{n}} \Delta v)(1 - a_{\text{v}} \Delta v)^{ - 1} \,\, \approx P^{0} (y_{\text{A}}^{ 0} - a_{\text{p}} \Delta v)\,\,\,\,\,\,{\text{with }}\,\,\,a_{\text{p}} = a_{\text{n}} - y_{\text{A}}^{ 0} a_{\text{v}}$$
(16)

The change in the concentration of oxygen in the liquid phase, ΔCA(L) (kg/m3) is expressed as follows:

$$\Delta C_{\text{A(L)}} = C_{\text{A(L)}}^{ 0} \,- C_{\text{A(L)}} = \Delta P_{\text{A}} /H = (P^{0} a_{\text{p}} \,/H)\Delta v$$
(17)

The oxygen-depletion in the liquid phase (expressed in kg) is calculated by the following relation:

$$\Delta m_{\text{A(L)}} = \Delta C_{\text{A(L)}} \times V_{\text{L}} = a_{\text{m(L)}} \Delta v\,\,\,\,\,\,{\text{with }}\,\,\,a_{\text{m(L)}} = P^{0} a_{\text{p}} V_{\text{L}} \,/H$$
(18)

Finally, the oxygen-depletion (in kg) from the whole gas–liquid system, i.e. oxygen demand (OD) is given by the following relation:

$${\text{OD}} = \Delta m_{\text{A(g,L)}} = a_{\text{m(g,L)}} \Delta v\,\,\;\quad \,\,{\text{with}}\,\,\,\,a_{\text{m(g,L)}} = a_{\text{m(g)}} + a_{\text{m(L)}} \,\,$$
(19)

Summary of the data treatment procedure in working form

The first task is to calculate the coefficient for the transformation of the volume change Δv data to the mass change, ΔmA data for the oxygen in the system. For the purpose, the following calculations (Steps 1 through 9) are to be done sequentially. This is a preparatory step and can be done before conducting the experiment.

  • Step 1: Record V0(g), VL, Sm and the temperature T

  • Step 2: Assume P0 = 1.013 × 105 Pa and y0A = 0.21.

  • Step 3: Find psw (Pa) and Henry constant H at the temperature T from some reference book

  • Step 4: Calculate n0s(g) (Eq. 2), aw (Eq. 5a), and ah and av (Eq. 6)

  • Step 5: Calculate an (Eq. 9)

  • Step 6: Calculate am(g) (Eq. 11)

  • Step 7: Calculate ap (Eq. 16)

  • Step 8: Calculate am(L) (Eq. 18)

  • Step 9: Calculate am(g,L) (Eq. 19)

Knowing the value of am(g,L), the Δv vs. t data can readily be transformed into OD vs. t data.

Indirect estimation of carbon dioxide from manometric reading

This particular headspace gas respirometer is designed to measure the amount of oxygen consumed by the biomass under the assumption that the evolved carbon dioxide will be completely absorbed by the caustic scrubber placed in the scrubber tube within the reactor. Although these measurements usually over-simplify the actual situation as they do not consider the effect of liquid-phase pH and oxygen and carbon dioxide transfer limitations. However, there is an option of indirect estimation of carbon dioxide (approximately) generated in the biodegradation process. To perform this type of analysis, the apparatus needs to orient in a particular fashion where three types of wastewater samples with two replicas from each group are to be placed in the six CSBRs of the respirometer. One sample from each group is placed in a reactor having carbon dioxide scrubbing system while the other is placed in the second reactor without the scrubber. Then the new positions of the manometer fluid after a certain interval are recorded to find out the amount of carbon dioxide generated during the biodegradation phase.

Referring to Fig. 3, let the net volume changed in time t considering full carbon dioxide absorption in the reactor by the scrubber is ∆V1 (measured by the manometer tube) and the net volume changed in the same duration t considering no absorption of carbon dioxide is ∆V2 (it may be positive or negative).

Fig. 3
figure 3

Schematic diagram of the respirometer setup for CO2 measurement: R1 Reactor with CO2 absorber, R2 Reactor without CO2 absorber, M1 Manometer linked with R1 reactor showing net volume change as ∆V1 and M2 Manometer linked with R2 reactor showing the net volume change as ∆V2 (similar arrangement will be made for the remaining reactors)

Since in both cases, same amount of carbon dioxide get dissolved in the liquid phase, considering the volume and pressure variation only in the gas phase the amount of carbon dioxide evolved (CE) in the system can be approximated from the following relation:

$${\text{CE}} = \frac{{P^{0} M_{{{\text{CO}}_{ 2} }} }}{RT}\left( {1 + \frac{{\rho gV^{0}_{{ ( {\text{g)}}}} }}{{S_{\text{m}} P^{0} }}} \right)\left( {\Delta V_{1} \pm \Delta V_{2} } \right)$$
(20)

where, \(M_{{\rm CO}_2}\) is molecular mass of the carbon dioxide (kg/mol).

From the last two equations, i.e. 19 and 20, an important dimensionless ratio could be calculated which is called respiratory quotient (RQ) and can be expressed as

$${\text{RQ = }}\frac{\text{CE}}{\text{OD}}{ = }\frac{{{\text{CO}}_{ 2} \,{\text{Evolved}}}}{{{\text{O}}_{ 2} \,{\text{Demand}}\; ( {\text{i}} . {\text{e}} .\;{\text{consumed)}}}}$$
(21)

Demonstration of the performance of the device

The performance of a RU of BMRS is tested by monitoring the oxygen uptake during the biodegradation of a glucose-glutamic acid standard solution. The data have been compared with those available in literature (Lee and Oswald 1954; Kessick 1976; Min et al. 2004). Also the biodegradation of the wastewater from an industrial plant has been monitored through oxygen uptake measurement in the five RUs of BMRS at different dilutions and the method of the determination of the bio-kinetic coefficients has been illustrated with the acquired data. Finally, the respiratory quotients of the effluent samples collected from three different effluent treatment plants (ETP) have also been determined.

Check with standard glucose-glutamic acid (GGA) solution

The oxidative potential of a given amount of seed inoculum (biomass agent) as well as the reliability of a BOD measuring instrument are checked periodically by monitoring the oxygen uptake data of some pure organic compounds for which BOD is known or determinable. One of the most commonly used such compounds is a mixture of glucose (C6H12O6) and glutamic acid (C5H9NO4), and a mixture of 150 mg/L of each is recommended in 1 L aqueous solution (APHA 2005). Pure glucose has an exceptionally high oxidation rate with relatively simple seeds. When used with glutamic acid, the oxidation rate is stabilized and is similar to that of most municipal wastewater (Ramalho 1977). Stoichiometrically, ThOD for 150 mg of Glucose is 160 mg and that for the 150 mg of Glutamic acid is 212.2 mg. Therefore, the theoretical chemical oxygen demand for the prepared GGA solution is 372.2 mg/L. For biodegradation studies, a small volume, vs (ml), of the GGA solution is diluted to a total test volume Vtest (standard volume = 300 ml) of oxygen-saturated aqueous solution containing the seed and the oxygen depletion is monitored for the diluted solution.

The oxygen consumption data from a biodegradation experiment are conventionally presented by a concept called oxygen uptake (OU) (mg/L) which is related to the oxygen demand (OD) as follows:

$${\text{OU}} = \frac{\text{OD}}{{v_{\text{s}} }} \times 10^{9} \, = \frac{{{\text{OD}} \times \,{\text{DF}}}}{{V_{\text{test}} }} \times 10^{9} \,\,\,\,\quad \,{\text{with}}\;\quad \,{\text{Dilution}}{\kern 1pt} \,{\text{factor}}\;{\text{(DF)}} = V_{\text{test}} /v_{\text{s}} \,$$
(22)

For studying the biodegradation of the GGA solution in the present work, a special type of commercial microbial blend named as, Probios plus’, product of Jayson Agrovet, Bangladesh, was used as the seed. The seed solution was prepared by rehydrating 10 g dry probios in 100 ml distilled water with aeration for 1 h followed by 10-min-long settling for the separation of bran material from the seeded supernatant. The solution in the RU contained 5 ml of GGA standard solution, 10 ml probios and 285 ml oxygen saturated distilled water. The oxygen uptake profile of the GGA solution is presented in Fig. 4. Similar data available in the literature (Lee and Oswald 1954; Kessick 1976; Min et al. 2004) have also been presented in the Figure for comparison.

Fig. 4
figure 4

Oxygen Uptake (OU) profile for GGA solution

The biodegradation rate is seed-specific, which results in different oxygen consumption rate for the same effluent (Ramalho 1977; Rahman et al. 2013). But still three of the curves (excluding that of Lee and Oswald 1954) have shown satisfactorily similar oxygen uptake profile for a biodegradation period of 120 h, with the oxygen uptake value lying in the range of 250–260 mg/L. Therefore, it can be concluded that the BMRS can successfully and reliably be used to monitor biodegradation process and to determine the biokinetic coefficients in wastewater treatment processes.

Parallel Respirometric and BOD test using dilution series

To make a comparative analysis, the respirometric and conventional BOD tests were also performed with the same dilution series. The object of biodegradation study was chosen arbitrarily, and it represents a wastewater sample collected from the outlet enriched with food processing industrial discharge. The characteristic features of the test sample are presented in the Table 3. The sample was rich in microorganism and no seed was used from outside. For respirometric study, five reactors were run at a time with different dilutions of the sample while sixth reactor was used to monitor the manometric observation corresponds to the thermobarometric change during the experiment. For BOD tests, five 300 mL BOD bottles were charged with the wastewater with the same dilutions. Both the tests were conducted at 30 °C for a period of 120 h. The oxygen uptake profile of the demonstration test is presented in Fig. 5.

Table 3 Characteristics of the test sample for parallel respirometric and BOD test
Fig. 5
figure 5

Oxygen uptake profiles of a real wastewater sample run simultaneously with five different dilutions

As seen from the Fig. 5, the oxygen uptake increases with the progress of the biodegradation process, but the rate gradually decreases showing a trend to become zero as the uptake approaches some limiting value. The initial uptake rate is the highest for the highest dilution. Such saturate type of curve could easily be described by an empirical equation of the type (Metcalf and Eddy 2005):

$${\text{BOD}}_{\text{t}} = {\text{BOD}}_{\infty } (1 - e^{ - kt} )$$
(23)

where BOD is the biochemical oxygen demand (equivalent to oxygen uptake, OU) up to time t, BOD (equivalent to the ultimate oxygen uptake, OU) is the amount of oxygen consumed for t →  (the whole biodegradable mass is removed), and k is the rate constant of a pseudo first-order biodegradation process. The OU-data of simultaneously operating RU units have been fitted to the Eq. (23), and the fitted parameters have been estimated. These parameters together with the oxygen uptake value at t = 120 h, OU120 (read from the averaged curves) are presented in Table 4 along with the measured BOD5 for comparison.

Table 4 Different parameters obtained from respirometric and conventional BOD methods

The OU120-values obtained from different dilutions are uniformly distributed around the averaged value (86 ± 4 mg/L), whereas the equivalent BOD5-values are much scattered around its averaged value (109 ± 33 mg/L). The BOD5-values are much affected by the dilution factors (Young and Cowan 2004). The average BOD5-range as found in this work encompasses the average OU120-range. Thus the average OU120-value may be accepted as the BOD5-value of the wastewater with some reservation as the average OU120-value lies near the lower limit of the BOD5-range. This might be an effect from the assumption in the respirometric analysis that the concentration of the oxygen in the liquid and gas phases are always in equilibrium (Kuyen and Yuebo 1996). The resistance to the mass-transfer of the gaseous to the liquid phase might be high enough that the oxygen concentration in the liquid is maintained at a level lower than its equilibrium value. In BOD test, the oxygen concentration is measured in the liquid phase where the biodegradation process actually takes place. For more precise respirometric analysis, the oxygen concentration in the aqueous phase should also be monitored simultaneously with the oxygen concentration in the gaseous phase. In spite of the bit lower OU120-values as compared to the corresponding BOD5-values, the advantage of the respirometric method is that it provides information about the biodegradation kinetics as well as OU-values. This parallel study has categorically demonstrated that the RU units of the present device along with the developed procedure of OD-calculation can successfully replace the traditional BOD tests and provides more information about the biodegradation process than the BOD does.

Calculation of respiratory quotients of the effluent samples collected from different ETPs

Measurement of the rate of oxygen utilization and carbon dioxide production by activated sludge–sewage mixtures has become prominent in recent years as a criterion of sludge activity and other characteristics of the activated sludge methods of sewage treatment (Sawyer and Nichols 1939; Genc and Yonsel 2007). The respiratory quotient (RQ), in this regard, plays an important role in aerobic respirometric investigations as its value depends on the substrate type, microorganisms present and on the environmental conditions too (Dilly 2001; ASAP 2012). This value can indicate the presence of different metabolic pathways, e.g. nitrification, denitrification, and anaerobic metabolic routes (Dilly 2003). Moreover, RQ can be used as a control parameter to supply a process with an optimum concentration of oxygen (Zeng et al. 1994; Franzén et al. 1996) or to control the feed rate (Aiba and Furuse 1990). Therefore, handling and operation procedures and appropriate treatment strategy could be selected considering the RQ values of the wastewater and activated sludge samples. In this case three wastewater samples collected from three different effluent treatment plants were selected to use and a high-rate batch test protocol was followed. The characteristic features of the test sample and experimental outcome are presented in the Table 5. All the tests were conducted at 30 °C for a period of 6 h only.

Table 5 Characteristic features of the test samples collected from ETPs and their respirometric and conventional BOD results

As like previous discussion, here also a clear difference was observed between conventional BOD and respirometric oxygen consumption. However, in this case, incubation time extends for a short duration (6 h) under a high rate test protocol against strong biomass agents collected from a reaction tank of an activated sludge plant. This type of experiment yields extant kinetics of a test sample (Young and Cowan 2004). Sample collected from municipal wastewater treatment plant showing a RQ value of 0.55 indicates the presence of a diverse type of mix substrates while the confectionery effluent having a RQ value of 1.19 indicates the presence of higher amount of carbohydrate groups. Since the sources of the samples and the experimental conditions are different from those found in the literature, it is quite difficult to make a significant comparison between these values. However, the findings of the current study are consistent with those of Kovács et al. (2007) who found 1.2 of RQ during biodegradation tests using glucose and sewage sludge as feed which closely resembles the confectionery effluent that contains high amount of carbohydrate ingredients.  The RQ value determined by Sawyer and Nichols (1939) falls between the limits of 0.51 and 1.18 for activated sludge-sewage mixture which is quite similar to municipal effluent in nature and the current finding (i.e. 0.55) also lies within the range. On the other hand, a RQ value of 0.84 was found for the effluent collected from the recycled paper industry. However, the value found for paper sludge was 0.92 by Atkinson et al. (1997) which is quite close to the present results.

From what has been said above, it will be clear that the determination of the RQ is important as an indicator of the type of substrates involved and the nature of some of the various phases of the respiration process in the sample. However, the interpretation of result from this measurement should be done with some reservations. In many cases, more than one substrate may be utilized at the same time or more than one chain of reaction may be involved in the degradation of the substrate. Since respiration may signify the simultaneous utilization of a mixture of several different substrates, the calculated RQ represents an overall or average gaseous exchange for the combustion of the entire complex of materials. However, ETP personnel could make a rough idea about the type of influent wastewater and subsequent follow-up steps could be formulated and maintained measuring the extant kinetics of the raw sample through continuous monitoring of oxygen uptake and respiratory quotient of the representing test samples collected from the plant.

Analysis of growth kinetics of wastewater

Design, monitoring and performance-evaluation of a biological treatment plant exclusively depend on the microbial growth parameters of the wastewater. Due to the lack of proper growth kinetics data, most of the biological reactors, in most cases, are found to have poor design dimension and uneven treatment results (Kovar and Egli 1998; Metcalf and Eddy 2005). To overcome this situation oxygen uptake measurement can be a good alternative for the designer, investor and plant operators. Unlike BOD test, the respirometric method provides adequate data for such analysis. In the following sections, model-based discussion has been made on the relationship between the oxygen uptake rate and the microbial growth kinetics, and also the procedure of the determination of the biokinetic coefficients from the oxygen uptake data obtained from BMRS has been illustrated. Thus the data already presented in Fig. 5 undergo further treatment to determine biokinetic coefficients.

Relation between the microbial growth kinetics and oxygen uptake rate

Biological growth kinetics can be expressed by the well-known Monod model. The relationship between the growth rate and limited substrate concentration in a batch culture process can be expressed by the following equation (Metcalf and Eddy 2005):

$$\frac{{{\text{d}}X}}{{{\text{d}}t}} = \left( {\mu_{\text{m}} \frac{S}{{K_{\text{s}} + S}} - k_{\text{d}} } \right)X$$
(24)

where X (mg/L) is the biomass concentration at any time t, S (mg/L) is the substrate concentration, μm (h−1) is the maximum specific growth rate, Ks (mg/L) is the half saturation constant and kd (h−1) is the decay coefficient. Expressing the substrate concentration S in terms of COD (Yoong et al. 2000), the Eq. (24) is reduced to

$$\frac{{{\text{d}}X}}{{{\text{d}}t}} = \left( {\mu_{\text{m}} \frac{\text{COD}}{{K_{\text{s}} + {\text{COD}}}} - k_{\text{d}} } \right)X$$
(25)

To use respirometric data for the analysis of growth kinetics, the oxygen uptake is to be related to X and COD. For the simplicity of the analysis, the oxygen uptake data are related to the biomass growth and the COD consumption by the following linear relationship (Rozich and Gaudy 1992)

$$X = X_{0} + \beta {\text{OU}}\,\,\,\,\,\,{\text{with}}\,\,\,\,\,\beta = \frac{Y}{{1 - YO_{x} }}$$
(26)

where Y is the biomass yield coefficient and Ox is the oxidative potential of the biomass. Y can be determined from individual batch kinetic tests for ΔX vs. ΔCOD data using a laboratory fermentor. A rough estimation of Y by Eq. (27), however, seems satisfactory for respirometric analysis of biodegradation assuming an average value of Ox in the range of 1.42–1.48 (Contreras et al. 2002; Chiil et al. 2008).

$$Y = (1 - \Delta {\text{OU}}/\Delta {\text{COD}})/O_{x}$$
(27)

In that case, the COD is to be measured at two points (initially and immediately after the completion of the experiment) and Y can be estimated from Eq. (27) (Choung et al. 1999). To estimate the biomass concentration X at any time t, the initial biomass concentration X0 should be measured separately and then OU vs. t data from respirometric measurements can be transformed into X vs. t data by Eq. (26).

Illustration of the method for the determination of biokinetic coefficients

As discussed in the previous section, to determine the biokinetic coefficients μm, Ks and kd, the value of Y and O x have to be determined first. For the purpose, the COD and the VSS were measured for the undiluted sample and the corresponding initial COD and the initial X in each RU were calculated based on the dilution factor. The final COD was measured for a single sample with DF = 10 after the experiment was over (t = 120 h), and the value of Y was calculated for this sample from Eq. (27) assuming that O x  = 1.45. The value of Y in this demonstration experiment was found to be 0.61. Such set of value for O x  = 1.45 and Y = 0.61 are satisfactorily in the same order as those found by previous authors (Park et al. 1997; Choung et al. 1999; Chiil et al. 2008). With known (or assumed) value of Y, O x and X0 for each dilution, the OU-data are quite adequate to determine the biokinetic coefficients μm, Ks and kd.

For the determination of the biokinetic coefficients, the Eq. (25) is rewritten as Eq. (28):

$$\frac{{{\text{d}}\ln X}}{{{\text{d}}t}} = \mu_{\text{m}} \frac{\text{COD}}{{K_{\text{s}} + {\text{COD}}}} - k_{\text{d}}$$
(28)

The experimentally obtained OU vs. t data are transformed into X vs. t through the Eq. (26). Then ln X vs. t plot can be drawn. The COD being a decreasing function of time in any RU, the slope of the curve described by the right-hand side term of the Eq. (28) is also decreasing. The change in COD value is not monitored and hence the slope at arbitrary point cannot be used for analysis. At t = 0, however, the COD is known and thus the initial slope will be useful for further analysis. Thus,

$$\left( {\frac{{{\text{d}}\ln X}}{{{\text{d}}t}}} \right)_{t \to 0} = m_{0} = \mu_{\text{m}} \frac{{{\text{COD}}_{ 0} }}{{K_{\text{s}} + {\text{COD}}_{ 0} }} - k_{\text{d}}$$
(29)

where m0 is the initial slope of the ln X vs. t plot. Now the m0 vs. COD0 data can be plotted and fitted to the Eq. (29) by a program like KaleidaGraph to give the value of the biokinetic coefficients μm, Ks and kd. Due to the specific nature of the relation in Eq. (29), few m0 vs. COD0 data (only fine points in our case) might show large ranges of the three biokinetic coefficients satisfying the Eq. (29). Better result is expected if the effect of kd in the process is ignored and thus, the Eq. (29) is approximated as

$$m_{0} \approx \mu_{\text{m}} \frac{{{\text{COD}}_{ 0} }}{{K_{\text{s}} + {\text{COD}}_{ 0} }}\,\, \Rightarrow \,\,\frac{1}{{m_{0} }} = \frac{{K_{\text{s}} }}{{\mu_{\text{m}} }} \times \frac{ 1}{{{\text{COD}}_{ 0} }} + \frac{1}{{\mu_{\text{m}} }}$$
(30)

Thus, a 1/m0 vs. 1/COD0 plot will be a straight line and from the slope and the intercept of the straight line, the fitted value of μm and Ks can be determined.

The data from the Fig. 5 have been treated by Eq. (26) and are presented in Fig. 6 as a ln X vs. t plot. As predicted by Eq. (29), the slope of the curves gradually decreases. The slope, m0, of the curves at t → 0 were estimated for further treatment. Finally, 1/m0 vs. 1/COD0 plot is presented in Fig. 7 and the data are fitted to the straight line described by the Eq. (30). The fitted value of μm and Ks are found to be 0.048 h−1 and 96 mg/L.

Fig. 6
figure 6

ln X vs. t plot for different dilutions (Each curve corresponds to that in Fig. 5 with the same DF)

Fig. 7
figure 7

1/m0 vs. 1/COD0 plot

This illustration shows the further opportunity of using the data acquired by the present device. The OU-data can be routinely treated to evaluate the state of the activated sludge in an operating plant and the operating parameters can be regulated for better performances. Actual biokinetic coefficients can be determined on the spot and be used (instead of using literature data acquired under different conditions) in the design of the treatment plant, and the design parameters are expected to be more appropriate for functioning.

Conclusions

A low cost respirometric device called ‘BMRS’ has been designed and fabricated, which can be used for different purposes such as (1) monitoring of biodegradation process in wastewater, (2) alternative to BOD test providing equivalent information (3) calculating respiratory quotient and (4) study of microbial growth kinetics and routine determination of biokinetic coefficients. Data treatment for oxygen consumption and carbon dioxide generation in a biodegradation process has been presented by a simple calculation method. The device would provide more precise information about oxygen uptake in a system than the conventional BOD method does. The performance of the device is quite satisfactory and it is concluded that the present respirometer is a simple and cost-effective non-digital wastewater monitoring device, which could be used in the academic research as well as in industrial applications in developing countries.