Abstract
The determination of a cyclist’s lactate threshold (LT) has become an important test performed in sports science laboratories around the world. A limitation of such testing is that it is relatively expensive and invasive, requiring multiple blood samples. The purpose of the current study was to evaluate a commercially available, wearable lactate threshold sensor (WLT) that uses near infrared LED technology to measure gastrocnemius muscle oxygenation and predict LT. The WLT was compared to four traditional calculations of determining LT following an incremental exercise test. Ten male and five female recreational to elite cyclists (mean ± SD; age 24 ± 8, body mass 69.7 ± 7.3 kg, VO2max; 59.7 ± 9.9 ml kg−1 min−1) performed an incremental cycling test to exhaustion. Blood lactate samples were taken at the end of each 3-min stage during the test to determine lactate threshold using four traditional methods (TradLT, Dmax, mDmax, OBLA). Traditional methods were then compared against the WLT predicted value. The correlation between the WLT and TradLT, Dmax, mDmax and OBLA were all >r = 0.96. The highest level of agreement for the WLT was with the Dmax method (95 % limits of agreement: ±17 W, TEE = 8.6 W, 4.4 %). The 95 % level of agreement between the WLT and all other traditional methods was <±40 W (TEE <18 W, 8 %). In summary, the WLT is practical, easy to use and exhibits an acceptable level of agreement with four of the traditionally accepted methods of determining lactate threshold.
References
Bentley DJ, McNaughton LR, Thompson D, Vleck VE, Batterham AM (2001) Peak power output, the lactate threshold, and time trial performance in cyclists. Med Sci Sports Exerc 33(12):2077–2081
Bishop D, Jenkins DG, Mackinnon LT (1998) The relationship between plasma lactate parameters, Wpeak and 1-h cycling performance in women. Med Sci Sports Exerc 30(8):1270–1275
Coyle EF, Coggan AR, Hopper M, Walters TJ (1988) Determinants of endurance in well-trained cyclists. J Appl Physiol 64(6):2622–2630
McGehee JC, Tanner CJ, Houmard JA (2005) A comparison of methods for estimating the lactate threshold. J Strength Cond Res 19(3):553–558
Larsson P, Olofsson P, Jakobsson E, Burlin L, Henriksson-Larsén K (2002) Physiological predictors of performance in cross-country skiing from treadmill tests in male and female subjects. Scand J Med Sci Sports 12(6):347–353
Fell JW (2008) The modified D-max is a valid lactate threshold measurement in veteran cyclists. J Sci Med Sport 11(5):460–463
Pyne DB, Boston T, Martin DT, Logan A (2000) Evaluation of the Lactate Pro blood lactate analyser. Eur J Appl Physiol 82(1–2):112–116
Tanner RK, Fuller KL, Ross MLR (2010) Evaluation of three portable blood lactate analysers: lactate Pro, Lactate Scout and Lactate Plus. Eur J Appl Physiol 109(3):551–559
Dascombe BJ, Reaburn P, Sirotic A, Coutts AJ (2007) The reliability of the i-STAT clinical portable analyser. J Sci Med Sport 10(3):135–140
Fell J, Rayfield J, Gulbin J, Gaffney P (1998) Evaluation of the accusport lactate analyser. Int J Sports Med 19(3):199–204
Vachon JA, Bassett DR, Clarke S (1999) Validity of the heart rate deflection point as a predictor of lactate threshold during running. J Appl Physiol 87(1):452–459
Plato P, McNulty M, Crunk S, Ergun AT (2008) Predicting lactate threshold using ventilatory threshold. Int J Sports Med 29:732–737
Beaver WL, Wasserman K, Whipp BJ (1986) A new method for detecting anaerobic threshold by gas exchange. J Appl Physiol 60(6):2020–2027
Fabre N, Mourot L, Zerbini L, Pellegrini B, Bortolan L, Schena F (2013) A novel approach for lactate threshold assessment based on rating of perceived exertion. Int J Sports Physiol Perform 8(3):263–270
Weltman A, Snead D, Seip R, Schurrer R, Levine S, Rutt R, Reilly T, Weltman J, Rogol A (1987) Prediction of lactate threshold and fixed blood lactate concentrations from 3200-m running performance in male runners. Int J Sports Med 8(6):401–406
Van Schuylenbergh R, Vanden EB, Hespel P (2004) Correlations between lactate and ventilatory thresholds and the maximal lactate steady state in elite cyclists. Int J Sports Med 25(6):403–408
Borges NR, Driller MW (In Press) Wearable lactate threshold predicting device is valid and reliable in runners. J Strength Cond Res
Driller MW, Halson SL (2013) The effects of wearing lower body compression garments during a cycling performance test. Int J Sports Physiol Perform 8:300–306
Newell J, Higgins D, Madden N, Cruickshank J, Einbeck J, McMillan K, McDonald R (2007) Software for calculating blood lactate endurance markers. J Sports Sci 25(12):1403–1409
Atkinson G, Nevill AM (1998) Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med 26(4):217–238
Altman DG, Bland JM (1983) Measurement in medicine: the analysis of method comparison studies. Statistician. pp 307–317
Hopkins W (2000) Analysis of validity by linear regression (Excel spreadsheet). A new view of statistics sportsci org: Internet Society for Sport Science
Seiler S, Tønnessen E (2009) Intervals, thresholds, and long slow distance: the role of intensity and duration in endurance training. Sport Sci 13:32–53
Coyle E, Feltner ME, Kautz S, Hamilton M, Montain S, Baylor A, Abraham L, Petrek G (1991) Physiological and biomechanical factors associated with elite endurance cycling performance. Med Sci Sports Exerc 23(1):93–107
Miller FR, Manfredi TG (1987) Physiological and anthropometrical predictors of 15-kilometer time trial cycling performance time. Res Q Exerc Sport 58(3):250–254
Wang B, Xu G, Tian Q, Sun J, Sun B, Zhang L, Luo Q, Gong H (2012) Differences between the vastus lateralis and gastrocnemius lateralis in the assessment ability of breakpoints of muscle oxygenation for aerobic capacity indices during an incremental cycling exercise. J Sport Sci Med 11(4):606
Bhambhani YN, Buckley SM, Susaki T (1997) Detection of ventilatory threshold using near infrared spectroscopy in men and women. Med Sci Sports Exerc 29(3):402–409
Bellotti C, Calabria E, Capelli C, Pogliaghi S (2013) Determination of maximal lactate steady state in healthy adults: can NIRS help. Med Sci Sports Exerc 45(6):1208–1216
Wang L, Yoshikawa T, Hara T, Nakao H, Suzuki T, Fujimoto S (2006) Which common NIRS variable reflects muscle estimated lactate threshold most closely? Appl Physiol Nutr Metab 31(5):612–620
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Driller, M., Borges, N. & Plews, D. Evaluating a new wearable lactate threshold sensor in recreational to highly trained cyclists. Sports Eng 19, 229–235 (2016). https://doi.org/10.1007/s12283-016-0198-6
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12283-016-0198-6