Optimization of iterative reconstruction parameters with 3-dimensional resolution recovery, scatter and attenuation correction in 123I-FP-CIT SPECT

  • Original Article
  • Published:
Annals of Nuclear Medicine Aims and scope Submit manuscript

Abstract

Objective

The aim of this study was to determine the optimal reconstruction parameters on ordered-subset expectation maximization iterative reconstruction with resolution recovery, scatter and attenuation correction (OSEMRRSCAC) on 123I-FP-CIT SPECT in terms of the image quality, quantification and diagnostic ability.

Methods

We evaluated the quality and quantification in 123I-FP-CIT SPECT images obtained by different reconstruction parameters using the anthropomorphic striatal phantom. The phantom images were acquired using a SPECT/CT system equipped with a low- and medium-energy general-purpose collimator and then were reconstructed using OSEMRRSCAC with various update numbers and the full width at half maximum (FWHM) of the Gaussian filter. The count ratio of the striatum, the coefficient of variation (CV) on the background and the specific binding ratio (SBR) were calculated to determine the optimal reconstruction parameters. Then, 42 consecutive patients who underwent 123I-FP-CIT SPECT were selected for clinical study. The patients were grouped into potentially decreasing group (20 with Parkinson’s disease, 5 with parkinsonian syndrome and 5 with dementia with Lewy bodies) and potentially normal binding group (4 with Alzheimer disease and 8 with essential tremor). Clinical images were reconstructed using OSEMRRSCAC and FBP with Chang’s AC (FBPAC). The performance of OSEMRRSCAC with the optimal reconstruction parameters was evaluated according to the receiver operating characteristic analysis and visual assessment.

Results

The count ratio increased with an increasing update number and converged uniformly with an update number over 90. The CV continuously increased with the update number. The SBR also increased with an increasing update number and converged uniformly with an update number of 90 or larger. The FWHM of the Gaussian filter influenced the image quality and quantification. In the clinical study, the area under curve (AUC) for the OSEMRRSCAC and FBPAC were 0.9812 and 0.9759, respectively. The quality of the SPECT images of OSEMRRSCAC (3.8 ± 0.8) was significantly superior to that of FBPAC (2.2 ± 0.7).

Conclusions

We determined that the optimal reconstruction parameter for OSEMRRSCAC was 90 update numbers with 6.6 mm FWHM of the Gaussian filter. Our results suggested that the optimal reconstruction parameters have a potential to improve the performance and the image quality of 123I-FP-CIT SPECT in comparison with the FBP reconstruction.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Booij J, Tissingh G, Boer GJ, Speelman JD, Stoof JC, Janssen AG, et al. 123I FP-CIT SPECT shows a pronounced decline of striatal dopamine transporter labelling in early and advanced Parkinson’s disease. J Neurol Neurosurg Psychiatry. 1997;62:133–40.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  2. Booij J, Speelman JD, Horstink MW, Wolters EC. The clinical benefit of imaging striatal dopamine transporters with [123I] FP-CIT SPET in differentiating patients with presynaptic parkinsonism from those with other forms of parkinsonism. Eur J Nucl Med. 2001;28:266–72.

    Article  CAS  PubMed  Google Scholar 

  3. Schwarz J, Linke R, Kerner M, Mozley PD, Trenkwalder C, Gasser T, et al. Striatal dopamine transporter binding assessed by [I-123] IPT and single photon emission computed tomography in patients with early Parkinson’s disease: implications for a preclinical diagnosis. Arch Neurol. 2000;57:205–8.

    Article  CAS  PubMed  Google Scholar 

  4. Benamer HT, Patterson J, Grosset DG, Booij J, De Bruin K, Van Royen E, et al. Accurate differentiation of parkinsonism and essential tremor using visual assessment of [123I]-FP-CIT SPECT imaging: the [123I]-FP-CIT study group. Mov Disord. 2000;15:503–10.

    Article  CAS  PubMed  Google Scholar 

  5. Djang DS, Janssen MJ, Bohnen N, Booij J, Henderson TA, Herholz K, et al. SNM practice guideline for dopamine transporter imaging with 123I-ioflupane SPECT 1.0. J Nucl Med. 2012;53:154–63.

    Article  CAS  PubMed  Google Scholar 

  6. Darcourt J, Booij J, Tatsch K, Varrone A, Vander BT, Kapucu ÖL, et al. EANM procedure guidelines for brain neurotransmission SPECT using 123I-labelled dopamine transporter ligands, version 2. Eur J Nucl Med Mol Imaging. 2010;37:443–50.

    Article  CAS  PubMed  Google Scholar 

  7. Koch W, Hamann C, Welsch J, Pöpperl G, Radau PE, Tatsch K. Is iterative reconstruction an alternative to filtered backprojection in routine processing of dopamine transporter SPECT studies? J Nucl Med. 2005;46:1804–11.

    PubMed  Google Scholar 

  8. Bienkiewicz M, Siennicki J, Plachcinska A, Gorska-Chrzastek M, Kusmierek J, Gajos A, et al. Impact of CT based attenuation correction on quantitative assessment of DaTSCAN (123 I-Ioflupane) imaging in diagnosis of extrapyramidal diseases. Nucl Med Rev. 2008;11:53–8.

    Google Scholar 

  9. Pareto D, Cot A, Pavía J, Falcón C, Juvells I, Lomeña F, et al. Iterative reconstruction with correction of the spatially variant fan-beam collimator response in neurotransmission SPET imaging. Eur J Nucl Med Mol Imaging. 2003;30:1322–9.

    Article  PubMed  Google Scholar 

  10. Tossici-Bolt L, Hoffmann SM, Kemp PM, Mehta RL, Fleming JS. Quantification of [123I] FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio. Eur J Nucl Med Mol Imaging. 2006;33:1491–9.

    Article  PubMed  Google Scholar 

  11. Dickson JC, Tossici-Bolt L, Sera T, Erlandsson K, Varrone A, Tatsch K. The impact of reconstruction method on the quantification of DaTSCAN images. Eur J Nucl Med Mol Imaging. 2010;37:23–35.

    Article  PubMed  Google Scholar 

  12. Davidsson A, Georgiopoulos C, Dizdar N, Granerus G, Zachrisson H. Comparison between visual assessment of dopaminergic degeneration pattern and semi-quantitative ratio calculations in patients with Parkinson’s disease and Atypical Parkinsonian syndromes using DaTSCAN® SPECT. Ann Nucl Med. 2014;28:851–9.

    Article  CAS  PubMed  Google Scholar 

  13. Winz OH, Hellwig S, Mix M, Weber WA, Mottaghy FM, Schäfer WM, et al. Image quality and data quantification in dopamine transporter SPECT: advantage of 3-dimensional OSEM reconstruction? Clin Nucl Med. 2012;37:866–71.

    Article  PubMed  Google Scholar 

  14. Hughes T, Celler A. A multivendor phantom study comparing the image quality produced from three state-of-the-art SPECT-CT systems. Nucl Med Commun. 2012;33:663–70.

    Article  PubMed  Google Scholar 

  15. Onishi H, Motomura N, Fujino K, Natsume T, Haramoto Y. Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom models. Radiol Phys Technol. 2013;6:42–53.

    Article  PubMed  Google Scholar 

  16. Seret A, Forthomme J. Comparison of different types of commercial filtered backprojection and ordered-subset expectation maximization SPECT reconstruction software. J Nucl Med Technol. 2009;37:179–87.

    Article  PubMed  Google Scholar 

  17. Lavalaye J, Booij J, Reneman L, Habraken JB, van Royen EA. Effect of age and gender on dopamine transporter imaging with [123I] FP-CIT SPET in healthy volunteers. Eur J Nucl Med. 2000;37:867–9.

    Article  Google Scholar 

Download references

Conflict of interest

Authors declare no conflict of interest in this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Norikazu Matsutomo.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Matsutomo, N., Nagaki, A., Yamao, F. et al. Optimization of iterative reconstruction parameters with 3-dimensional resolution recovery, scatter and attenuation correction in 123I-FP-CIT SPECT. Ann Nucl Med 29, 636–642 (2015). https://doi.org/10.1007/s12149-015-0987-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12149-015-0987-1

Keywords

Navigation