Abstract
Objective
The aim of this study was to determine the optimal reconstruction parameters on ordered-subset expectation maximization iterative reconstruction with resolution recovery, scatter and attenuation correction (OSEMRRSCAC) on 123I-FP-CIT SPECT in terms of the image quality, quantification and diagnostic ability.
Methods
We evaluated the quality and quantification in 123I-FP-CIT SPECT images obtained by different reconstruction parameters using the anthropomorphic striatal phantom. The phantom images were acquired using a SPECT/CT system equipped with a low- and medium-energy general-purpose collimator and then were reconstructed using OSEMRRSCAC with various update numbers and the full width at half maximum (FWHM) of the Gaussian filter. The count ratio of the striatum, the coefficient of variation (CV) on the background and the specific binding ratio (SBR) were calculated to determine the optimal reconstruction parameters. Then, 42 consecutive patients who underwent 123I-FP-CIT SPECT were selected for clinical study. The patients were grouped into potentially decreasing group (20 with Parkinson’s disease, 5 with parkinsonian syndrome and 5 with dementia with Lewy bodies) and potentially normal binding group (4 with Alzheimer disease and 8 with essential tremor). Clinical images were reconstructed using OSEMRRSCAC and FBP with Chang’s AC (FBPAC). The performance of OSEMRRSCAC with the optimal reconstruction parameters was evaluated according to the receiver operating characteristic analysis and visual assessment.
Results
The count ratio increased with an increasing update number and converged uniformly with an update number over 90. The CV continuously increased with the update number. The SBR also increased with an increasing update number and converged uniformly with an update number of 90 or larger. The FWHM of the Gaussian filter influenced the image quality and quantification. In the clinical study, the area under curve (AUC) for the OSEMRRSCAC and FBPAC were 0.9812 and 0.9759, respectively. The quality of the SPECT images of OSEMRRSCAC (3.8 ± 0.8) was significantly superior to that of FBPAC (2.2 ± 0.7).
Conclusions
We determined that the optimal reconstruction parameter for OSEMRRSCAC was 90 update numbers with 6.6 mm FWHM of the Gaussian filter. Our results suggested that the optimal reconstruction parameters have a potential to improve the performance and the image quality of 123I-FP-CIT SPECT in comparison with the FBP reconstruction.
Similar content being viewed by others
References
Booij J, Tissingh G, Boer GJ, Speelman JD, Stoof JC, Janssen AG, et al. 123I FP-CIT SPECT shows a pronounced decline of striatal dopamine transporter labelling in early and advanced Parkinson’s disease. J Neurol Neurosurg Psychiatry. 1997;62:133–40.
Booij J, Speelman JD, Horstink MW, Wolters EC. The clinical benefit of imaging striatal dopamine transporters with [123I] FP-CIT SPET in differentiating patients with presynaptic parkinsonism from those with other forms of parkinsonism. Eur J Nucl Med. 2001;28:266–72.
Schwarz J, Linke R, Kerner M, Mozley PD, Trenkwalder C, Gasser T, et al. Striatal dopamine transporter binding assessed by [I-123] IPT and single photon emission computed tomography in patients with early Parkinson’s disease: implications for a preclinical diagnosis. Arch Neurol. 2000;57:205–8.
Benamer HT, Patterson J, Grosset DG, Booij J, De Bruin K, Van Royen E, et al. Accurate differentiation of parkinsonism and essential tremor using visual assessment of [123I]-FP-CIT SPECT imaging: the [123I]-FP-CIT study group. Mov Disord. 2000;15:503–10.
Djang DS, Janssen MJ, Bohnen N, Booij J, Henderson TA, Herholz K, et al. SNM practice guideline for dopamine transporter imaging with 123I-ioflupane SPECT 1.0. J Nucl Med. 2012;53:154–63.
Darcourt J, Booij J, Tatsch K, Varrone A, Vander BT, Kapucu ÖL, et al. EANM procedure guidelines for brain neurotransmission SPECT using 123I-labelled dopamine transporter ligands, version 2. Eur J Nucl Med Mol Imaging. 2010;37:443–50.
Koch W, Hamann C, Welsch J, Pöpperl G, Radau PE, Tatsch K. Is iterative reconstruction an alternative to filtered backprojection in routine processing of dopamine transporter SPECT studies? J Nucl Med. 2005;46:1804–11.
Bienkiewicz M, Siennicki J, Plachcinska A, Gorska-Chrzastek M, Kusmierek J, Gajos A, et al. Impact of CT based attenuation correction on quantitative assessment of DaTSCAN (123 I-Ioflupane) imaging in diagnosis of extrapyramidal diseases. Nucl Med Rev. 2008;11:53–8.
Pareto D, Cot A, Pavía J, Falcón C, Juvells I, Lomeña F, et al. Iterative reconstruction with correction of the spatially variant fan-beam collimator response in neurotransmission SPET imaging. Eur J Nucl Med Mol Imaging. 2003;30:1322–9.
Tossici-Bolt L, Hoffmann SM, Kemp PM, Mehta RL, Fleming JS. Quantification of [123I] FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio. Eur J Nucl Med Mol Imaging. 2006;33:1491–9.
Dickson JC, Tossici-Bolt L, Sera T, Erlandsson K, Varrone A, Tatsch K. The impact of reconstruction method on the quantification of DaTSCAN images. Eur J Nucl Med Mol Imaging. 2010;37:23–35.
Davidsson A, Georgiopoulos C, Dizdar N, Granerus G, Zachrisson H. Comparison between visual assessment of dopaminergic degeneration pattern and semi-quantitative ratio calculations in patients with Parkinson’s disease and Atypical Parkinsonian syndromes using DaTSCAN® SPECT. Ann Nucl Med. 2014;28:851–9.
Winz OH, Hellwig S, Mix M, Weber WA, Mottaghy FM, Schäfer WM, et al. Image quality and data quantification in dopamine transporter SPECT: advantage of 3-dimensional OSEM reconstruction? Clin Nucl Med. 2012;37:866–71.
Hughes T, Celler A. A multivendor phantom study comparing the image quality produced from three state-of-the-art SPECT-CT systems. Nucl Med Commun. 2012;33:663–70.
Onishi H, Motomura N, Fujino K, Natsume T, Haramoto Y. Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom models. Radiol Phys Technol. 2013;6:42–53.
Seret A, Forthomme J. Comparison of different types of commercial filtered backprojection and ordered-subset expectation maximization SPECT reconstruction software. J Nucl Med Technol. 2009;37:179–87.
Lavalaye J, Booij J, Reneman L, Habraken JB, van Royen EA. Effect of age and gender on dopamine transporter imaging with [123I] FP-CIT SPET in healthy volunteers. Eur J Nucl Med. 2000;37:867–9.
Conflict of interest
Authors declare no conflict of interest in this study.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Matsutomo, N., Nagaki, A., Yamao, F. et al. Optimization of iterative reconstruction parameters with 3-dimensional resolution recovery, scatter and attenuation correction in 123I-FP-CIT SPECT. Ann Nucl Med 29, 636–642 (2015). https://doi.org/10.1007/s12149-015-0987-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12149-015-0987-1