High-resolution small field-of-view magnetic resonance image acquisition system using a small planar coil and a pneumatic manipulator in an open MRI scanner

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Background

Low-field open magnetic resonance imaging (MRI) is frequently used for performing image-guided neurosurgical procedures. Intraoperative magnetic resonance (MR) images are useful for tracking brain shifts and verifying residual tumors. However, it is difficult to precisely determine the boundary of the brain tumors and normal brain tissues because the MR image resolution is low, especially when using a low-field open MRI scanner. To overcome this problem, a high-resolution MR image acquisition system was developed and tested.

Methods

An MR-compatible manipulator with pneumatic actuators containing an MR signal receiver with a small radiofrequency (RF) coil was developed. The manipulator had five degrees of freedom for position and orientation control of the RF coil. An 8-mm planar RF coil with resistance and inductance of 2.04 \(\Omega \) and 1.00 \(\upmu \mathrm{H}\) was attached to the MR signal receiver at the distal end of the probe. MR images of phantom test devices were acquired using the MR signal receiver and normal head coil for signal-to-noise ratio (SNR) testing.

Results

The SNR of MR images acquired using the MR signal receiver was 8.0 times greater than that of MR images acquired using the normal head coil. The RF coil was moved by the manipulator, and local MR images of a phantom with a 2-mm grid were acquired using the MR signal receiver. A wide field-of-view MR image was generated from a montage of local MR images.

Conclusion

A small field-of-view RF system with a pneumatic manipulator was integrated in a low-field MRI scanner to allow acquisition of both wide field-of-view and high-resolution MR images. This system is promising for image-guided neurosurgery as it may allow brain tumors to be observed more clearly and removed precisely.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11
Fig. 12
Fig. 13
Fig. 14
Fig. 15
Fig. 16
Fig. 17
Fig. 18
Fig. 19

Similar content being viewed by others

References

  1. Iseki H, Muragaki Y, Taira T, Kawamata T, Maruyama T, Naemura K, Nambu K, Sugiura M, Hirai N, Hori T, Takakura K (2002) New possibilities for stereotaxis. Stereotact Funct Neurosurg 76(3–4):159–167

    Google Scholar 

  2. McDannold N, Clement G, Black P, Jolesz F, Hynynen K (2010) Transcranial MRI-guided focused ultrasound surgery of brain tumors. Initial findings in three patients. Neurosurgery 66(2):323–332

    Article  PubMed Central  PubMed  Google Scholar 

  3. Massin C, Eroglu S, Vincent F, Gimi BS, Besse P-A, Magin RL, Popovic RS (2003) Planar microcoil-based magnetic resonance imaging of cells. In: 12th International Conference on Solid-state Sensors, Actuators and Microsystems, Boston, MA, vol 2, pp 967–970

  4. Takahashi H, Dohi T, Matsumoto K, Shimoyama I (2007) A micro planar coil for local high resolution magnetic resonance imaging. In: IEEE 20th international conference on micro electro mechanical system, pp 549–552

  5. Badilita V, Kratt K, Baxan N, Mohmmadzadeh M, Burger T, Weber H, Elverfeldt DV, Hennig J, Korvink JG, Wallrabe U (2010) On-chip three dimensional microcoils for MRI at the microscale. Lab Chip 10:1387–1390

    Article  CAS  PubMed  Google Scholar 

  6. Inamura T, Dohi T (2013) Cone-shaped micro coil for magnetic resonance imaging. In: 26th IEEE international conference on micro electro mechanical system, pp 335–338

  7. Dave UR, Williams AD, Wilson JA, Amin Z, Gilderdale DJ, Larkman DJ, Thursz MR, Taylor-Robinson SD, deSouza NM (2004) Esophageal cancer staging with endoscopic MR imaging pilot study. Radiology 230(1):281–286

    Article  PubMed  Google Scholar 

  8. Masamune K, Kobayashi E, Masutani Y, Suzuki M, Dohi T, Iseki H, Takakura K (1995) Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery. J Image Guided Surg 1(4):242–248

    Article  CAS  Google Scholar 

  9. Sutherland GR, Latour I, Greer AD (2008) Integrating an image-guided robot with intraoperative MRI. Eng Med Biol Mag 27(3):59–65

  10. Fischer GS, Iordachita II, Csoma C, Tokuda J, Dimaio SP, Tempany CM, Hata N, Fichtinger G (2008) MRI-compatible pneumatic robot for transperineal prostate needle placement. IEEE/ASME Trans Mechatron 13(3):295–305

    Article  Google Scholar 

  11. Seifabadi R, Song S-E, Krieger A, Cho NB, Tokuda J, Fichtinger G, Iordachita II (2012) Robotic system for MRI-guided prostate biopsy: feasibility of teleoperated needle insertion and ex vivo phantom study. Int J Comput Assist Radiol Surg 7:181–190

    Article  PubMed Central  PubMed  Google Scholar 

  12. Stoianovici D, Song D, Petrisor D, Ursu D, Mazilu D, Mutener M, Schar M, Patriciu A (2007) MRI stealth robot for prostate interventions. Minim Invasive Ther Allied Technol 16(4):241–248

    Article  PubMed Central  PubMed  Google Scholar 

  13. Yang B, Tan U-X, McMillan AB, Gullapalli R, Desai JP (2011) Design and implementation of a pneumatically-actuated robot for breast biopsy under continuous MRI. In: IEEE international conference on robotics and automation, pp 674–679

  14. Hashizume M (2007) MRI-guided laparoscopic and robotic surgery for malignancies. Int J Clin Oncol 12(2):94–98

    Article  PubMed  Google Scholar 

  15. Koseki Y, Mizuhara K, Chinzei K (2008) Registration for image integration of intraoperative MRI and endoscope, using MRI-compatible manipulator. In: 20th International Conference of the Society for Medical Innovation and Technology, pp 28–30

  16. Mylonas N, Damianou C (2013) MR compatible positioning device for guiding a focused ultrasound system for the treatment of brain diseases. Int J Med Robot Comput Assist Surg 10(1):1–10

    Article  Google Scholar 

  17. Stoianovici D, Patriciu A, Petrisor D, Mazilu D, Kavoussi L (2007) A new type of motor: pneumatic step motor. IEEE/ASME Trans Mechatron 12(1):98–106

    Article  Google Scholar 

  18. Sajima H, Kamiuchi H, Kuwana K, Dohi T, Masamune K (2012) MR-safe pneumatic rotation stepping actuator. J Robot Mechatron 24(5):820–827

    Google Scholar 

  19. Su H, Cardona DC, Shang W, Camilo A, Cole GA, Rucker DC, Webster RJ, Fischer GS (2012) A MRI-guided concentric tube continuum robot with piezoelectric actuation: a feasibility study. In: IEEE International Conference on Robotics and Automation, pp 1939–1945

  20. American Society for Testing and Materials (ASTM) International, Designation (2005) F2503–05. Standard practice for marking medical devices and other items for safety in the magnetic resonance environment, ASTM International, West Conshohocken, PA

  21. Wheeler HA (1928) Simple inductance formulas for radio coils. Proc Inst Radio Eng 16(10):1398–1400

    Google Scholar 

Download references

Conflict of interest

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kohei Miki.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Miki, K., Masamune, K. High-resolution small field-of-view magnetic resonance image acquisition system using a small planar coil and a pneumatic manipulator in an open MRI scanner. Int J CARS 10, 1687–1697 (2015). https://doi.org/10.1007/s11548-014-1136-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-014-1136-3

Keywords

Navigation