Using The Morphology of Photoplethysmogram Peaks to Detect Changes in Posture

  • Published:
Journal of Clinical Monitoring and Computing Aims and scope Submit manuscript

Abstract

The morphology of the pulsatile component of the photoplethysmogram (PPG) has been shown to vary with physiology, but changes in the morphology caused by the baroreflex response to orthostatic stress have not been investigated.

Using two FDA approved Nonin® pulse oximeters placed on the finger and ear, we monitored 11 subjects, for three trials each, as they stood from a supine position. Each cardiac cycle was automatically extracted from the PPG waveform and characterized using statistics corresponding to normalized peak width, instantaneous heart rate, and amplitude of the pulsatile component of the ear PPG. A nonparametric Wilcoxon rank sum test was then used to detect in real-time changes in these features with p < 0.01.

In all 33 trials, the standing event was detected as an abrupt change in at least two of these features, with only one false alarm. In 26 trials, an abrupt change was detected in all three features, with no false alarms. An increase in the normalize peak width was detected before an increase in heart rate, and in 21 trials a peak in the feature was detected before or as standing commenced. During standing, the pulse rate always increases, and then amplitude of the ear PPG constricts by a factor of two or more.

We hypothesis that the baroreflex first reduces the percentage of time blood flow is stagnant during the cardiac cycle, then increases the hear rate, and finally vasoconstricts the peripheral tissue in order to reestablishing a nominal blood pressure. These three features therefore can be used as a detector of the baroreflex response to changes in posture or other forms of blood volume sequestration.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Cook LB. Extracting arterial flow waveforms from pulse oximeter waveforms apparatus. Anaesthesia 2001;56(6):551–555.

    Article  PubMed  CAS  Google Scholar 

  2. Wisely NA, Cook LB. Arterial flow waveforms from pulse oximetry compared with measured Doppler flow waveforms apparatus. Anaesthesia 2001;56(6):556–561.

    Article  PubMed  CAS  Google Scholar 

  3. Murray WB, Foster PA. The peripheral pulse wave: Information overlooked. J Clin Monit 1996;12:365–377.

    Article  PubMed  CAS  Google Scholar 

  4. Rusch TL, Sankar R, Scharf JE. Signal processing methods for pulse oximetry. Comput. Bol. Med. 1996;26(2):143–159.

    Article  CAS  Google Scholar 

  5. Shelley K, Shelley S. Pulse Oximeter Waveform: Photoelectric Plethysmography. In: Carol Lake, Hines R, Blitt C (eds.), Clinical Monitoring: W.B. Saunders Company 2001:420–428.

  6. Johnston WS, Mendelson Y. Extracting Breathing Rate Information from a Wearable Reflectance Pulse Oximeter. Paper presented at: International Conference of the IEEE EMBS, 2004; San Francisco, CA, USA.

  7. Budinger TF. Biomonitoring with wireless communications. Annu Rev Biomed Eng 2003;5:383–412.

    Article  PubMed  CAS  Google Scholar 

  8. Montgomery K, Mundt C, Thonier G, Tellier A, Udoh U, Barker V, et al. Lifeguard- A Personal Physiological Monitor For Extreme Environments. Paper presented at: International Conference of the IEEE EMBS, 2004; San Francisco, CA, USA.

  9. Anliker U, Ward JA, Lukowicz P, Tröster G, Dolveck F, Baer M, et al. AMON: A Wearable Multiparameter Medical Monitoring and Alert System. IEEE Trans Inf Technol Biomed 2004;8(4):415–427.

    Article  PubMed  Google Scholar 

  10. Cooke WH, Ryan KL. Convertino VA. Lower body negative pressure as a model to study progression to acute hemorrhagic shock in humans. J Appl Physiol 2004;98:1249–1261.

    Article  Google Scholar 

  11. Olsen H, Vernersson E, Länne T. Cardiovascular response to acute hypovolemia in relation to age. Implications for orthostasis and hemorrhage. AJP- Heart and Circulatory Physiology 2000;278:H222–H232.

    PubMed  CAS  Google Scholar 

  12. Shamir M, Eidelman LA, Floman Y, Kaplan L, Pizov R. Pulse oximetry plethysmographic waveform during changes in blood volume. Br J Anaesth 1999;82(2):178–181.

    PubMed  CAS  Google Scholar 

  13. Klabunde RE. Cardiovascular Physiology Concepts. Lippincott Williams and Wilkins, Baltimore, 2005.

    Google Scholar 

  14. Mohrman DE, Heller LJ. Cardiovascular Physiology 5 ed. New York: McGraw-Hill/Appleton & Lange; 2003.

    Google Scholar 

  15. Yasuma F, Hayano J-I. Respiratory sinus arrhythmia: Why does the heartbeat synchronize with respiratory rhythm? Chest 2004;125(2):683–690.

    Article  PubMed  Google Scholar 

  16. Nilsson L, Johansson A, Kalman S. Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique. J Clin Monit Comput 2000;16:309–315.

    Article  PubMed  CAS  Google Scholar 

  17. Nilsson L, Johansson A, Kalman S. Macrocirculation is not the sole determinant of respiratory induced variations in the reflection mode photoplethysmographic signal. Physiol Meas 2003;24(4):925–937.

    Article  PubMed  CAS  Google Scholar 

  18. Angelone A, Jr. NAC. Respiratory sinus arrhythmia: A frequency dependent phenomenon. J Appl Physiol 1964;19:479–482.

    PubMed  Google Scholar 

  19. Bootsma M, Swenne CA, Bruschke AV. Heart rate variability during repeated incremental head-up tilt discloses time dependence of individual autonomic dynamics. Clin Cardiol 1996;19(1):62–68.

    Article  PubMed  CAS  Google Scholar 

  20. Gratze G, Fortin J, Holler A, Grasenick K, Pfurtscheller G, Wach P, et al. A software package for non-invasive, real-time beat-to-beat monitoring of stroke volume, blood pressure, total peripheral resistance and for assessment of autonomic function. Computers in Biology & Medicine 1998;28(2):121–142.

    Article  CAS  Google Scholar 

  21. anonymous. Heart rate variability: Standards of measurement, physiological interpretation and clinical use. Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Circulation 1996;93(5):1043–1065.

  22. Johansson A. Neural network for photoplethysmographic respiratory rate monitoring. Medical & Biomedical Engineering & Computing 2003;41:242–248.

    Article  CAS  Google Scholar 

  23. Shelley KH, Tamai D, Jablonka D, Gesquiere M, Stout RG, Silverman DG. The effect of venous pulsation on the forehead pulse oximeter wave form as a possible source of error in Spo2 calculation. Anesth Analg 2005;10:743–747.

    Article  Google Scholar 

  24. Partridge BL. Use of pulse oximetry as a noninvasive indicator of intravascular volume status. J Clin Monit 1987;3(4): 263–268.

    PubMed  CAS  Google Scholar 

  25. Schell C, Linder SP, Zeider JR. Tracking highly maneuverable targets with unknown behavior. Proceedings of the IEEE 2004;92(3):558–574.

    Google Scholar 

  26. Awad AA, Ghobashy MA, Ouda W, Stout RG, Silverman DG, Shelley KH. Different responses of ear and finger pulse oximeter wave form to cold pressor test. Anesth Analg 2001;92(6):1483–1486.

    Article  PubMed  CAS  Google Scholar 

  27. Jablonka DH, Awad AA, Tamai DN, Stout RG, Silverman DG, Shelley KH. The effect of Vasopressin on ear pulse plethysmography and its implications on cerebral blood flow regulation. Paper presented at: ASA Annual Meeting, Las Vegas, Nevada 2004.

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Stephen P. Linder.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Linder, S.P., Wendelken, S.M., Wei, E. et al. Using The Morphology of Photoplethysmogram Peaks to Detect Changes in Posture. J Clin Monit Comput 20, 151–158 (2006). https://doi.org/10.1007/s10877-006-9015-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10877-006-9015-2

Key words

Navigation