Abstract
Object
Proton resonance frequency shift thermometry is sensitive to breathing motion that leads to incorrect phase differences. In this work, a novel velocity-sensitive navigator technique for triggering MR thermometry image acquisition is presented.
Materials and methods
A segmented echo planar imaging pulse sequence was modified for velocity-triggered temperature mapping. Trigger events were generated when the estimated velocity value was less than 0.2 cm/s during the slowdown phase in parallel to the velocity-encoding direction. To remove remaining high-frequency spikes from pulsation in real time, a Kalman filter was applied to the velocity navigator data. A phantom experiment with heating and an initial volunteer experiment without heating were performed to show the applicability of this technique. Additionally, a breath-hold experiment was conducted for comparison.
Results
A temperature rise of ΔT = +37.3°C was seen in the phantom experiment, and a root mean square error (RMSE) outside the heated region of 2.3°C could be obtained for periodic motion. In the volunteer experiment, a RMSE of 2.7°C/2.9°C (triggered vs. breath hold) was measured.
Conclusion
A novel velocity navigator with Kalman filter postprocessing in real time significantly improves the temperature accuracy over non-triggered acquisitions and suggests being comparable to a breath-held acquisition. The proposed technique might be clinically applied for monitoring of thermal ablations in abdominal organs.
Similar content being viewed by others
References
Kennedy JE, ter Haar GR, Cranston D (2003) High intensity focused ultrasound: surgery of the future?. Brit J Radiol 76: 590–599
Chapman A, ter Haar GR (2007) Thermal ablation of uterine fibroids using MR-guided focused ultrasound-a truly non-invasive treatment modality. Eur Radiol 17(10): 2505–2511
Vogl TJ, Naguib NNN, Eichler K, Lehnert T, Ackermann H, Mack MG (2008) Volumetric evaluation of liver metastases after thermal ablation: long-term results following MR-guided laser-induced thermotherapy. Radiology 249(3): 865–871
Rieke V, Pauly KB (2008) MR thermometry. J Magn Reson Imaging 27(2): 376–390
de Poorter J, de Wagter C, de Deene Y, Thomsen C, Ståhlberg F, Achten E (1995) Noninvasive MRI thermometry with the proton resonance frequency (PRF) method: in vivo results in human muscle. Magn Reson Med 33(1): 74–81
Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K, Suzuki Y (1995) A precise and fast temperature mapping using water proton chemical shift. Magn Reson Med 34(6): 814–823
Vigen KK, Daniel BL, Pauly JM, Butts K (2003) Triggered, navigated, multi-baseline method for proton resonance frequency temperature mapping with respiratory motion. Magn Reson Med 50(5): 1003–1010
Rieke V, Vigen KK, Sommer G, Daniel BL, Pauly JM, Butts K (2004) Referenceless PRF shift thermometry. Magn Reson Med 51(6): 1223–1231
de Zwart JA, Vimeux FC, Palussière J, Salomir R, Quesson B, Delalande C, Moonen CTW (2001) On-line correction and visualization of motion during MRI-controlled hyperthermia. Magn Reson Med 45(1): 128–137
Lepetit-Coiffé M, Quesson B, Seror O, Dumont E, Bail BL, Moonen CTW, Trillaud H (2006) Real-time monitoring of radiofrequency ablation of rabbit liver by respiratory-gated quantitative temperature MRI. J Magn Reson Imaging 24(1): 152–159
Morikawa S, Inubushi T, Kurumi Y, Naka S, Sato K, Demura K, Tani T, Haque HA (2004) Feasibility of respiratory triggering for MR-guided microwave ablation of liver tumors under general anesthesia. Cardiovasc Inter Rad 27(4): 370–373
Kalman RE (1960) A new approach to linear filtering and prediction problems. J Basic Eng 82: 35–45
Spincemaille P, Nguyen TD, Prince MR, Wang Y (2008) Kalman filtering for real-time navigator processing. Magn Reson Med 60(1): 158–168
Moran PR (1982) A flow velocity zeugmatographic interlace for NMR imaging in humans. Magn Reson Imaging 1(4): 197–203
Maier F, Krafft AJ, Jenne JW, Semmler W, Bock M (2009) TAM–a thermal ablation monitoring tool: in vivo evaluation. IFMBE Proc 25(VI): 247–250
Sapareto SA, Dewey WC (1984) Thermal dose determination in cancer therapy. Int J Radiat Oncol 10(6): 787–800
Ries M, de Senneville BD, Roujol S, Berber Y, Quesson B, Moonen CTW (2010) Real-time 3D target tracking in MRI guided focused ultrasound ablations in moving tissues. Magn Reson Med 64(6): 1704–1712
Hey S, Maclair G, de Senneville BD, Lepetit-Coiffe M, Berber Y, Köhler MO, Quesson B, Moonen CTW, Ries M (2009) Online correction of respiratory-induced field disturbances for continuous MR-thermometry in the breast. Magn Reson Med 61(6): 1494–1499
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Maier, F., Krafft, A.J., Yung, J.P. et al. Velocity navigator for motion compensated thermometry. Magn Reson Mater Phy 25, 15–22 (2012). https://doi.org/10.1007/s10334-011-0245-z
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10334-011-0245-z