Abstract
A patient underwent aortic valve closure for de novo aortic insufficiency that had deteriorated to severe insufficiency during six months of support with a continuous flow left ventricular assist device (cf-LVAD). Aortic insufficiency was initially noted one month after LVAD implantation, and then deterioration quickly developed. Right heart catheterization revealed that when the rotational speed of the cf-LVAD was increased, the cardiac index was decreased by an increase in regurgitant volume, as shown by echocardiography. During surgery, fusion and shortening of the aortic leaflets as well as left coronary ostial occlusion were observed. Direct aortic closure improved hemodynamics. Thrombus formation on the aortic valve shown by echocardiography in the early postoperative period may be a trigger of aortic insufficiency. Control of the cf-LVAD rotational speed is likely required to prevent aortic insufficiency.
Similar content being viewed by others
References
Toda K, Fujita T, Domae K, Shimahara Y, Kobayashi J, Nakatani T. Late aortic insufficiency related to poor prognosis during left ventricular assist device support. Ann Thorac Surg. 2011;92:929–34.
Adamson RM, Dembitsky WP, Baradarian S, Chammas J, May-Newman K, Chillcott S, Stahovich M, McCalmont V, Ortiz K, Hoagland P, Jaski B. Aortic valve closure associated with HeartMate left ventricular device support: technical considerations and long-term results. J Heart Lung Transplant. 2011;30:576–82.
Cohn WE, Demirozu ZT, Frazier OH. Surgical closure of left ventricular outflow tract after left ventricular assist device implantation in patients with aortic valve pathology. J Heart Lung Transplant. 2011;30:59–63.
Goda A, Takayama H, Pak SW, Uriel N, Mancini D, Naka Y, Jorde UP. Aortic valve procedures at the time of ventricular assist device placement. Ann Thorac Surg. 2011;91:750–4.
Pak SW, Uriel N, Takayama H, Cappleman S, Song R, Colombo PC, Charles S, Mancini D, Gillam L, Naka Y, Jorde UP. Prevalence of de novo aortic insufficiency during long-term support with left ventricular assist devices. J Heart Lung Transplant. 2010;29(10):1172–6 (Epub 2010 Jul 8).
Mudd JO, Cuda JD, Halushka M, Soderlund KA, Conte JV, Russell SD. Fusion of aortic valve commissures in patients supported by a continuous axial flow left ventricular assist device. J Heart Lung Transplant. 2008;27:1269–74.
Abbate R, Cioni G, Ricci I, Miranda M, Gori AM. Thrombosis and acute coronary syndrome. Thromb Res. 2012;129:235–40.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Fujita, T., Kobayashi, J., Hata, H. et al. Aortic valve closure for rapidly deteriorated aortic insufficiency after continuous flow left ventricular assist device implantation. J Artif Organs 16, 98–100 (2013). https://doi.org/10.1007/s10047-012-0661-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10047-012-0661-5