Age-corrected reference values for the Heidelberg multi-color anomaloscope

  • Retinal Disorders
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

Background

To determine reference values for the HMC anomaloscope (Heidelberg multi-color anomaloscope) of healthy subjects.

Methods

One hundred and thirteen healthy subjects were divided into four age groups: <20 years of age (ten female, five male), 20–39 years of age (23 female, 15 male), 40–59 years of age (23 female, ten male) and >60 years of age (nine female, 18 male). Match midpoint, matching range (MR) and anomaly quotient (AQ), according to the Moreland equation [blue (436 nm) + blue–green (490 nm) = cyan (480 nm) + yellow (589 nm)] and according to the Rayleigh equation [green (546 nm) + red (671 nm) = yellow (589 nm)] were determined. The neutral adaptation was done showing white light every 5 seconds in absolute mode and every 15 seconds in relative mode.

Results

The mean match midpoint according to the Rayleigh equation was 43.9 ± 2.6 scale units in absolute mode. It was highest between 20–39 years (45.2 ± 2.2) and lowest in subjects >60 years of age (42.2 ± 2.2). The mean MR in absolute mode was 3.1 ± 3.5 scale units with a maximum >60 years (4.4 ± 4.4). The MR in relative mode was between 1.6 ± 1.9 (20–39 years) and 4.4 ± 3.8 (>60 years). The resulting mean AQ was 1.01 ± 0.15 in both modes. The mean match midpoint of the Moreland equation was 51.0 ± 5.2 scale units in absolute mode. It was highest between 20–39 years (52.5 ± 5.7), and lowest in subjects >60 years of age (48.7 ± 3.6). The mean MR according to the Moreland equation was lower in absolute mode (13.4 ± 15.6) than in relative mode (16.2 ± 15.2). The mean resulting AQ was 1.02 ± 0.21 in both modes.

Conclusion

The values of this study can be used as references for the diagnosis of red–green and blue perception impairment with the HMC anomaloscope.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Shin YJ, Park KH, Hwang JM, Wee WR, Lee JH (2007) A new color vision test to differentiate congenital and acquired color vision defects. Ophthalmology 114:1341–1347

    Article  PubMed  Google Scholar 

  2. Connolly DM, Barbur JL, Hosking SL, Moorhead IR (2008) Mild hypoxia impairs chromatic sensitivity in the mesopic range. Invest Ophthalmol Vis Sci 49:820–827

    Article  PubMed  Google Scholar 

  3. Erb C, Fahle M (2006) Farbensehen und erworbene Farbsinnstörungen. Teil 1: Grundlagen. Ophthalmologe 103:349–360

    Article  PubMed  CAS  Google Scholar 

  4. Marré M, Marré E (1978) Different types of acquired colour vision deficiencies on the base of CVM patterns in dependence upon the fixation mode of the diseased eye. Mod Probl Ophthalmol 19:248–252

    PubMed  Google Scholar 

  5. Kurtenbach A, Flögel W, Erb C (2002) Anomaloscope matches in patients with diabetes mellitus. Graefes Arch Clin Exp Ophthalmol 240:79–84

    Article  PubMed  Google Scholar 

  6. Sommerhalder J, Baglivo E, Barbey C, Hirschel B, Roth A, Pelizzone M (1998) Colour vision in AIDS patients without HIV retinopathy. Vision Res 38:3441–3446

    Article  PubMed  CAS  Google Scholar 

  7. Nolan JM, Kenny R, O’Regan C, Cronin H, Loughman J, Connolly EE, Kearney P, Loane E, Beatty S (2010) Macular pigment optical density in an ageing Irish population: the Irish Longitudinal Study on Ageing. Ophthalmic Res 44:131–139

    Article  PubMed  Google Scholar 

  8. Robson AG, Moreland JD, Pauleikhoff D, Morrissey T, Holder GE, Fitzke FW, Bird AC, van Kuijk FJ (2003) Macular pigment density and distribution: comparison of fundus autofluorescence with minimum motion photometry. Vision Res 43:1765–1775

    Article  PubMed  Google Scholar 

  9. Stillman JA (2008) Psychophysical influences on the validity of anomaloscopic assessments of color vision. Percept Psychophys 70:1243–1247

    Article  PubMed  Google Scholar 

  10. Moreland JD, Torczynski E, Tripathi R (1991) Rayleigh and Moreland matches in the ageing eye. In: Drum B, Moreland JD, Serra A (eds) Color vision deficiencies X. Kluwer Academic, Dordrecht, pp 347–352

    Chapter  Google Scholar 

  11. Bailey JE, Neitz M, Tait DM, Neitz J (2004) Evaluation of an updated HRR color vision test. Vis Neurosci 21:431–436

    Article  PubMed  Google Scholar 

  12. Cole BL, Lian KY, Lakkis C (2006) The new Richmond HRR pseudoisochromatic test for colour vision is better than the Ishihara test. Clin Exp Optom 89:73–80

    Article  PubMed  Google Scholar 

  13. Kuchenbecker J, Röhl FW, Wesselburg A, Bernarding J, Behrens-Baumann W (2007) Untersuchungen zur Validität eines webbasierten Farbsehtests für Screeninguntersuchungen des Farbensehens. Ophthalmologe 104:47–53

    Article  PubMed  CAS  Google Scholar 

  14. Good GW, Schepler A, Nichols JJ (2005) The reliability of the Lanthony Desaturated D-15 test. Optom Vis Sci 82:1054–1059

    Article  PubMed  Google Scholar 

  15. Erb C, Adler M, Stübiger N, Wohlrab M, Zrenner E, Thiel HJ (1998) Colour vision in normal subjects tested by the colour arrangement test ‘Roth 28-hue desaturated’. Vision Res 38:3467–3471

    Article  PubMed  CAS  Google Scholar 

  16. Papaconstantinou D, Georgalas I, Kalantzis G, Karmiris E, Koutsandrea C, Diagourtas A, Ladas I, Georgopoulos G (2009) Acquired color vision and visual field defects in patients with ocular hypertension and early glaucoma. Clin Ophthalmol 3:251–257

    PubMed  Google Scholar 

  17. Nguyen NX, Korth M, Wisse M, Jünemann A (1994) Anwendung eines neuen Anomaloskop-Tests in der Glaukomdiagnostik. Klin Monatsbl Augenheilkd 204:149–154

    Article  PubMed  CAS  Google Scholar 

  18. Muftuoglu O, Karel F, Duman R (2007) Effect of a yellow intraocular lens on scotopic vision, glare disability, and blue color perception. J Cataract Refract Surg 33:658–666

    Article  PubMed  Google Scholar 

  19. Moreland JD (2004) Moreland match revisited. Vis Neurosci 21:471–476

    Article  PubMed  CAS  Google Scholar 

  20. Van de Kraats J, Van Norren D (2007) Optical density of the aging human ocular media in the visible and the UV. J Opt Soc Am A 24:1842–1857

    Article  Google Scholar 

  21. Erb C, Nicaeus T, Adler M, Isensee J, Zrenner E, Thiel HJ (1999) Colour vision disturbances in chronic smokers. Graefes Arch Clin Exp Ophthalmol 237:377–380

    Article  PubMed  CAS  Google Scholar 

  22. Schröder A, Erb C, Falk S, Schwartze G, Radermacher J, Winter R (2002) Farbsinnstörungen bei Patienten mit einer arteriellen Hypertonie. Ophthalmologe 99:375–379

    Article  PubMed  Google Scholar 

  23. Kinnear PR, Sahraie A (2002) New Farnsworth–Munsell 100 hue test norms of normal observers for each year of age 5–22 and for age decades 30–70. Br J Ophthalmol 86:1408–1411

    Article  PubMed  CAS  Google Scholar 

  24. Verriest G, Van LJ, Uvijls A (1982) A new assessment of the normal ranges of the Farnsworth–Munsell 100-hue test scores. Am J Ophthalmol 93:635–642

    PubMed  CAS  Google Scholar 

  25. Schröder A, Meinrenken J, Gockeln R, Erb C (2002) Positiver Einfluss von Farberfahrung auf das Ergebnis von Farbsinnprüfungen mit dem Farbpigmentprobentest Roth 28-hue (E) desaturiert — Klinische Studie an 44 Probanden. Klin Monatsbl Augenheilkd 219:33–36

    Article  PubMed  Google Scholar 

  26. Sommerhalder J, Pelizzone M, Roth A (1997) Paramètres de stimulation pour l’examen automatisé de la vision colorée. Klin Monatsbl Augenheilkd 210:319–324

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Florian Rüfer.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Rüfer, F., Sauter, B., Klettner, A. et al. Age-corrected reference values for the Heidelberg multi-color anomaloscope. Graefes Arch Clin Exp Ophthalmol 250, 1267–1273 (2012). https://doi.org/10.1007/s00417-012-1949-0

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00417-012-1949-0

Keywords

Navigation