Intrarater reliability of goniometry and hand-held dynamometry for shoulder and elbow examinations in female team handball athletes and asymptomatic volunteers

  • Arthroscopy and Sports Medicine
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction

To evaluate the intrarater reliability for examining active range of motion (ROM) and isometric strength of the shoulder and elbow among asymptomatic female team handball athletes and a control group using a manual goniometer and hand-held dynamometry (HHD).

Materials and methods

22 female team handball athletes (age: 21.0 ± 3.7 years) and 25 volunteers (13 female, 12 male, age: 21.9 ± 1.24 years) participated to determine bilateral ROM for shoulder rotation and elbow flexion/extension, as well as isometric shoulder rotation and elbow flexion/extension strength. Subjects were assessed on two separate test sessions with 7 days between sessions. Relative (intraclass correlation coefficients (ICC) and standard error of measurement (SEM) reliability were calculated.

Results

Reliability for ROM and strength were good to excellent for both shoulders and groups (athletes: ICC = 0.94–0.97, SEM 1.07°–4.76 N, controls: ICC = 0.96–1.00, SEM = 0.00 N–4.48 N). Elbow measurements for both groups also showed good-to-excellent reliability (athletes: ICC = 0.79–0.97, SEM = 0.98°–5.94 N, controls: ICC = 0.87–1.00, SEM = 0.00 N–5.43 N).

Conclusions

It is important to be able to reliably reproduce active ROM and isometric strength evaluations. Using a standardized testing position, goniometry and HHD are reliable instruments in the assessment of shoulder and elbow joint performance testing. We showed good-to-excellent reproducible results for male and female control subjects and female handball athletes, although the single parameters in ROM and strength were different for each group and between the shoulders and elbows.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Kolber MJ, Hanney WJ (2012) The reliability and concurrent validity of shoulder mobility measurements using a digital inclinometer and goniometer: a technical report. Int J Sports Phys Ther 7:306–313

    PubMed Central  PubMed  Google Scholar 

  2. Berryman Reese N (2012) Muscle and sensory testing, 3rd edn. Saunders, St. Louis

    Google Scholar 

  3. Quin L, Gordon J (2010) Documentation for rehabilitation: a guide to clinical decision making, 2nd edn. Saunders, Maryland Heights

    Google Scholar 

  4. Kolber MJ, Beekhuizen K, Cheng MS, Fiebert IM (2007) The reliability of hand-held dynamometry in measuring isometric strength of the shoulder internal and external rotator musculature using a stabilization device. Physiother Theory Pract 23:119–124

    Article  PubMed  Google Scholar 

  5. Byram IR, Bushnell BD, Dugger K, Charron K, Harrell FE Jr, Noonan TJ (2010) Preseason shoulder strength measurements in professional baseball pitchers: identifying players at risk for injury. Am J Sports Med 38:1375–1382

    Article  PubMed  Google Scholar 

  6. Cools AM, De Wilde L, Van Tongel A, Ceyssens C, Ryckewaert R, Cambier DC (2014) Measuring shoulder external and internal rotation strength and range of motion: comprehensive intra-rater and inter-rater reliability study of several testing protocols. J Shoulder Elbow Surg 23:1454–1461

    Article  PubMed  Google Scholar 

  7. Donatelli R, Ellenbecker TS, Ekedahl SR, Wilkes JS, Kocher K, Adam J (2000) Assessment of shoulder strength in professional baseball pitchers. J Orthop Sports Phys Ther 30:544–551

    Article  CAS  PubMed  Google Scholar 

  8. Fieseler G, Jungermann P, Koke A, Irlenbusch L, Delank KS, Schwesig R (2015) Range of motion and isometric strength in shoulders of team handball sport athletes during playing season, Part II: changes after mid-season. J Shoulder Elbow Surg 24:391–398

    Article  PubMed  Google Scholar 

  9. Fieseler G, Jungermann P, Koke A, Irlenbusch L, Delank K-S, Schwesig R (2015) Glenohumeral range of motion (ROM) and isometric strength of professional team handball athletes, part III: changes over the playing season. Arch Orthop Trauma Surg. doi:10.1007/s00402-015-2308-5

    Google Scholar 

  10. Hurd WJ, Kaplan KM, El Attrache NS, Jobe FW, Morrey BF, Kaufman KR (2011) A profile of glenohumeral internal and external rotation motion in the uninjured high school baseball pitcher, part I: motion. J Athl Train 46:282–288

    PubMed Central  PubMed  Google Scholar 

  11. Shanley E, Thigpen CA, Clark JC, Wyland DJ, Hawkins RJ, Noonan TJ, Kissenberth MJ (2012) Changes in passive range of motion and development of glenohumeral internal rotation deficit (GIRD) in the professional pitching shoulder between spring training in two consecutive years. J Shoulder Elbow Surg 21:1605–1612

    Article  PubMed  Google Scholar 

  12. Mullaney MJ, McHugh MP, Johnson CP, Tyler TF (2010) Reliability of shoulder range of motion comparing a goniometer to a digital level. Physiother Theory Pract 26:327–333

    Article  PubMed  Google Scholar 

  13. Muir SW, Corea CL, Beaupre L (2010) Evaluating change in clinical status: reliability and measures of agreement for the assessment of glenohumeral range of motion. N Am J Sports Phys Ther 5:98–110

    PubMed Central  PubMed  Google Scholar 

  14. Clarkson HM (2005) Joint motion and function assessment: a research based practical guide. Lippincott Williams & Wilkins, Philadelphia

    Google Scholar 

  15. Gajdosik RL, Bohannon RW (1987) Clinical measurement of range of motion. Review of goniometry emphasizing reliability and validity. Phys Ther 67:1867–1872

    CAS  PubMed  Google Scholar 

  16. Wilk KE, Reinold MM, Macrina LC, Porterfield R, Devine KM, Suarez K, Andrews JR (2009) Glenohumeral internal rotation measurements differ depending on stabilization techniques. Sports Health 1:131–136

    Article  PubMed Central  PubMed  Google Scholar 

  17. Ginn KA, Cohen ML, Herbert RD (2006) Does hand-behind-back range of motion accurately reflect shoulder internal rotation? J Shoulder Elbow Surg 15:311–314

    Article  PubMed  Google Scholar 

  18. Roddey TS, Cook KF, O’Malley KJ, Gartsman GM (2005) The relationship among strength and mobility measures and self-report outcome scores in persons after rotator cuff repair surgery: impairment measures are not enough. J Shoulder Elbow Surg 14:95S–98S

    Article  PubMed  Google Scholar 

  19. Rudiger HA, Fuchs B, von Campe A, Gerber C (2008) Measurements of shoulder mobility by patient and surgeon correlate poorly: a prospective study. J Shoulder Elbow Surg 17:255–260

    Article  PubMed  Google Scholar 

  20. Schrama PP, Stenneberg MS, Lucas C, van Trijffel E (2014) Intraexaminer reliability of hand-held dynamometry in the upper extremity: a systematic review. Arch Phys Med Rehabil 95:2444–2469

    Article  PubMed  Google Scholar 

  21. Stark T, Walker B, Phillips JK, Fejer R, Beck R (2011) Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: a systematic review. PM R 3:472–479

    Article  PubMed  Google Scholar 

  22. Bohannon RW (2005) Manual muscle testing: does it meet the standards of an adequate screening test? Clin Rehabil 19:662–667

    Article  PubMed  Google Scholar 

  23. Schwartz S, CohenME Herbison GJ, Shah A (1992) Relationship between two measures of upper extremity strength: manual muscle test compared to hand-held myometry. Arch Phys Med Rehabil 73:1063–1068

    CAS  PubMed  Google Scholar 

  24. Bohannon RW (1999) Inter-tester reliability of hand-held dynamometry: a concise summary of published research. Percept Mot Skills 88:899–902

    Article  CAS  PubMed  Google Scholar 

  25. Wang CY, Olson SL, Protas EJ (2002) Test-retest strength reliability: hand-held dynamometry in community-dwelling elderly fallers. Arch Phys Med Rehabil 83:811–815

    Article  PubMed  Google Scholar 

  26. Portney LG, Watkins MP (2009) Foundations of clinical research: applications to practice, 3rd edn. Pearson Prentice Hall, Upper Saddle River

    Google Scholar 

  27. Weir JP (2005) Quantifying test–retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res 19:231–240

    PubMed  Google Scholar 

  28. Bland JM, Altman DG (1986) Statistical methods for assessing agreements between two methods of clinical measurement. Lancet 8:307–310

    Article  Google Scholar 

  29. Kolber MJ, Fuller C, Marshall J, Wright A, Hanney WJ (2012) The reliability and concurrent validity of scapular plane shoulder elevation measurements using a digital inclinometer and goniometer. Physiother Theory Pract 28:161–168

    Article  PubMed  Google Scholar 

  30. Gajdosik CG (2005) Ability of very young children to produce reliable isometric force measurements. Pediatr Phys Ther 17:251–257

    Article  PubMed  Google Scholar 

  31. Rex Wong YC, Cameron D, Bohannon RW (1998) Elbow and hand muscle strength are not affected by head-neck position. Isokinet Exerc Sci 7:43–47

    Google Scholar 

  32. Vermeulen HM, de Bock GH, van Houwelingen HC, van der Meer RL, Mol MC, Plus BT, Rozing PM, Vlieland TV (2005) A comparison of two portable dynamometers in the assessment of shoulder and elbow strength. Physiotherapy 91:101–112

    Article  Google Scholar 

Download references

Acknowledgments

We are grateful to the volunteers and the team of SV Union Halle-Neustadt (2nd national female team handball league), who made this study possible.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Georg Fieseler.

Ethics declarations

The study has been performed without any source of support in the form of grants, financial donations or other items.

All participants provided written consent.

The work is part of a study that was approved by the Research Ethics Committee of the University of Medicine of the Federal University of Halle—Wittenberg, Germany (study 2013–13).

Additional information

The investigation is part of the dissertation work by Thomas Molitor.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Fieseler, G., Molitor, T., Irlenbusch, L. et al. Intrarater reliability of goniometry and hand-held dynamometry for shoulder and elbow examinations in female team handball athletes and asymptomatic volunteers. Arch Orthop Trauma Surg 135, 1719–1726 (2015). https://doi.org/10.1007/s00402-015-2331-6

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-015-2331-6

Keywords

Navigation