Mechanical Removal of Migrated Onyx Due to Microcatheter Rupture During AVM Embolization: A Technical Case Report

  • CASE REPORT
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

We report a case where migrated Onyx could be removed safely with a mechanical thrombectomy device after withdrawal of the damaged microcatheter. A microcatheter was retained in the superior cerebellar artery with attached Onyx fragments on its tip during embolization of a bled cerebellar AVM. During retrieval maneuvers microcatheter shaft was ruptured and Onyx fragments were released into the vertebrobasilar system causing occlusion. Finally, Onyx fragment on tip of the microcatheter could be detached with dimethyl sulfoxide injection and the microcatheter could be withdrawn. A Merci retriever device (Concentric Medical, Mountain View, CA, USA) was successfully used to retrieve most of the dislodged Onyx material and recanalize the vertebrobasilar circulation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Murayama Y, Viñuela F, Ulhoa A, et al. Nonadhesive liquid embolic agent for cerebral arteriovenous malformations: preliminary histopathological studies in swine rete mirabile. Neurosurgery. 1998;43:1164–75.

    Article  CAS  PubMed  Google Scholar 

  2. Walcott BP, Gerrard JL, Nogueira RG, Nahed BV, Terry AR, Ogilvy CS. Microsurgical retrieval of an endovascular microcatheter trapped during Onyx embolization of a cerebral arteriovenous malformation. J Neurointerv Surg. 2011;3:77–9.

    Article  PubMed  Google Scholar 

  3. Kelly ME, Turner R 4th, Gonugunta V, Rasmussen PA, Woo HH, Fiorella D. Monorail snare technique for the retrieval of an adherent microcatheter from an onyx cast: technical case report. Neurosurgery. 2008;63(1 Suppl 1):ONSE89. doi:10.1227/01.neu.0000335018.68369.e8.

    PubMed  Google Scholar 

  4. Newman CB, Park MS, Kerber CW, Levy ML, Barr JD, Pakbaz RS. Over the catheter retrieval of a retained microcatheter following Onyx embolization: a technical report. J Neurointerv Surg. 2012;. doi:10.1136/neurintsurg-2011-010040.

    PubMed  Google Scholar 

  5. Saatci I, Geyik S, Yavuz K, Cekirge HS. Endovascular treatment of brain arteriovenous malformations with prolonged intranidal Onyx injection technique: long-term results in 350 consecutive patients with completed endovascular treatment course. J Neurosurg. 2011;115(1):78–88.

    Article  PubMed  Google Scholar 

  6. Maimon S, Strauss I, Frolov V, Margalit N, Ram Z. Brain arteriovenous malformation treatment using a combination of Onyx and a new detachable tip microcatheter, SONIC: short-term results. Am J Neuroradiol. 2010;31:947–54.

    Article  CAS  PubMed  Google Scholar 

  7. Alamri A, Hyodo A, Suzuki K, et al. Retrieving microcatheters from Onyx casts in a series of brain arteriovenous malformations: a technical report. Neuroradiology. 2012;54:1237–40.

    Article  CAS  PubMed  Google Scholar 

  8. Santillan A, Zink W, Knopman J, Riina H, Gobin YP. Balloon-assisted technique for trapped microcatheter retrieval following onyx embolization. A case report. Interv Neuroradiol. 2009;15:453–5.

    PubMed Central  CAS  PubMed  Google Scholar 

  9. Zoarski GH, Lilly MP, Sperling JS, et al. Surgically confirmed incorporation of a chronically retained neurointerventional microcatheter in the carotid artery. Am J Neuroradiol. 1999;20:177–8.

    CAS  PubMed  Google Scholar 

  10. Michael SG, Swarnkar AS, Latorre JGS, Ramachandran TS, lodi YM. Revascularization of Onyx induced intra-operative occlusion of vertebrobasilar artery using the Merci device. Neurocrit Care. 2010;12:269–71.

    Article  PubMed  Google Scholar 

Download references

Conflict of interest

I declare that I have no conflict of interest.

Statement of Human and Animal Rights

Does not apply to this article.

Statement of Informed Consent

Informed consent was obtained from all individual participants included in the study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Cagin Senturk.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Senturk, C. Mechanical Removal of Migrated Onyx Due to Microcatheter Rupture During AVM Embolization: A Technical Case Report. Cardiovasc Intervent Radiol 38, 1654–1657 (2015). https://doi.org/10.1007/s00270-015-1095-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-015-1095-2

Keywords

Navigation