Clinical validation of synthetic brain MRI in children: initial experience

  • Paediatric Neuroradiology
  • Published:
Neuroradiology Aims and scope Submit manuscript

Abstract

Introduction

The purpose of this study was to determine the diagnostic accuracy of synthetic MR sequences generated through post-acquisition processing of a single sequence measuring inherent R1, R2, and PD tissue properties compared with sequences acquired conventionally as part of a routine clinical pediatric brain MR exam.

Methods

Thirty-two patients underwent routine clinical brain MRI with conventional and synthetic sequences acquired (22 abnormal). Synthetic axial T1, T2, and T2 fluid attenuation inversion recovery or proton density-weighted sequences were made to match the comparable clinical sequences. Two exams for each patient were de-identified. Four blinded reviewers reviewed eight patients and were asked to generate clinical reports on each exam (synthetic or conventional) at two different time points separated by a mean of 33 days. Exams were rated for overall and specific finding agreement (synthetic/conventional and compared to gold standard consensus review by two senior reviewers with knowledge of clinical report), quality, and diagnostic confidence.

Results

Overall agreement between conventional and synthetic exams was 97%. Agreement with consensus readings was 84% (conventional) and 81% (synthetic), p = 0.61. There were no significant differences in sensitivity, specificity, or accuracy for specific imaging findings involving the ventricles, CSF, brain parenchyma, or vasculature between synthetic or conventional exams (p > 0.05). No significant difference in exam quality, diagnostic confidence, or noise/artifacts was noted comparing studies with synthetic or conventional sequences.

Conclusions

Diagnostic accuracy and quality of synthetically generated sequences are comparable to conventionally acquired sequences as part of a standard pediatric brain exam. Further confirmation in a larger study is warranted.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Warntjes JB, Engström M, Tisell A, Lundberg P (2013) Brain characterization using normalized quantitative magnetic resonance imaging. PLoSOne 8(8):e70864

    Article  CAS  Google Scholar 

  2. Riederer SJ, Suddarth SA, Bobman SA, Lee JN, Wang HZ, MacFall JR (1984) Automated MR image synthesis: feasibility studies. Radiology 153(1):203–206

    Article  CAS  PubMed  Google Scholar 

  3. Bobman SA, Riederer SJ, Lee JN, Suddarth SA, Wang HZ, Drayer BP, MacFall JR (1985) Cerebral magnetic resonance image synthesis. AJNR Am J Neuroradiol 6(2):265–269

    CAS  PubMed  Google Scholar 

  4. Deoni SC, Peters TM, Rutt BK (2005) High-resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DESPOT2. Magn Reson Med 53(1):237–241

    Article  PubMed  Google Scholar 

  5. Gulani V, Schmitt P, Giswold MA, Webb AG, Jakob PM (2004) Towards a single-sequence neurologic magnetic resonance imaging examination: multiple-contrast images from an IR TrueFISP experiment. Investig Radiol 39(12):767–774

    Article  Google Scholar 

  6. European Society of Radiology (2015) Magnetic resonance fingerprinting—a promising new approach to obtain standardized imaging biomarkers from MRI. Insights Imaging 6(2):163–165

    Article  Google Scholar 

  7. Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, Griswold MA (2013) Magnetic resonance fingerprinting. Nature 495(7440):187–192

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Warntjes JBM, Dahlqvist O, West J, Lundberg P (2008) Rapid magnetic resonance quantification on the brain: optimization for clinical usage. Magn Reson Med 60(2):320–329

    Article  CAS  PubMed  Google Scholar 

  9. Warntjes MJ, Kihlberj EJ (2010) Rapid T1 quantification based on 3D phase sensitive inversion recovery. BMC Med Imaging 10:90

    Article  Google Scholar 

  10. West J, Warntjes JB, Lundberg P (2012) Novel whole brain segmentation and volume estimation using quantitative MRI. Eur Radiol 22(5):998–1007

    Article  CAS  PubMed  Google Scholar 

  11. Blystad I, Warntjes JBM, Smedby O, Landtblom A-M, Lundberg P, Larsson E-M (2012) Synthetic MRI of the brain in a clinical setting. Acta Radiol 53(10):1158–1163

    Article  CAS  PubMed  Google Scholar 

  12. Betts AM, Leach JL, Jones BV, Zhang B, Serai S (2016) Brain imaging with synthetic MR in children: clinical quality assessment. Neuroradiology PMID: 27438803

  13. Ing C, Wall MM, DiMaggio CJ, Whitehouse AJ, Hegarty MK, Sun M, von Ungern-Sternber BS, Li G, Sun LS (2016) Latent class analysis of neurodevelopmental deficit after exposure to anesthesia in early childhood. J Neurosurg Anesthesiol PMID: 27077892

  14. Ambarki K, Lindquvist T, Wåhlin A, Petterson E, Warntjes MJ, Birgander R, Malm J, Eklund A (2012) Evaluation of automatic measurement of the intracranial volume based on quantitative MR imaging. AJNR Am J Neuroradiol 33(10):1951–1956

    Article  CAS  PubMed  Google Scholar 

  15. Ambarki K, Wåhlin A, Birgander R, Eklund A, Malm J (2011) MR imaging of brain volumes: evaluation of a fully automatic software. AJNR Am J Neuroradiol 32(2):408–412

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hollie West.

Ethics declarations

We declare that all human and animal studies have been approved by the institutional review board and have therefore been performed in accordance with the ethical standards laid down in the 1964 Declaration of Helsinki and its later amendments. We declare that all patients gave informed consent prior to inclusion in this study.

Conflict of interest

We declare that we have no conflict of interest.

Electronic supplementary material

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

West, H., Leach, J.L., Jones, B.V. et al. Clinical validation of synthetic brain MRI in children: initial experience. Neuroradiology 59, 43–50 (2017). https://doi.org/10.1007/s00234-016-1765-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00234-016-1765-z

Keywords

Navigation