Abstract
Purpose
Assessment of hip internal rotation (IR) is a critical element in the evaluation of a painful hip, yet there currently exists no consistent method of determining this measurement. The purpose of this study is to report the intra-rater and inter-rater reliability of active and passive hip IR measurement using a digital inclinometer.
Methods
Twenty-five subjects (8 men, 17 women) with asymptomatic hip joints, and no known pathology between the ages of 22 and 42 participated. Two examiners measured hip IR for each subject using a digital inclinometer while visually monitoring pelvic motion. Measurements included passive IR, active IR, and active IR while internally rotating both legs simultaneously. Statistical analysis was performed including intra-rater and inter-rater reliability.
Results
There was no significant difference in measured IR values. Mean range of motion values for all techniques and both examiners ranged from 30.7° to 32.8°. Intra-rater reliability for examiner 1 ranged from 0.84 to 0.92. Inter-rater reliability ranged from 0.89 to 0.93. Of the methods utilized, active IR of a single leg was the most responsive technique with a minimal detectable change value of 5.4°.
Conclusions
Clinical assessment of hip IR range of motion can be performed efficiently and reliably with a digital inclinometer along with visual monitoring of the pelvis for unwanted motion. Findings from this study support unilateral active hip IR in a seated position as a reliable and responsive technique for obtaining hip IR measurements.
Level of evidence
III.
Similar content being viewed by others
References
Altman R, Alarcón G, Appelrouth D, Bloch D, Borenstein D, Brandt K, Brown C, Cooke TD, Daniel W, Feldman D, Greenwald R, Hochberg M, Howell D, Ike R, Kapila P, Kaplan D, Koopman W, Marino C, McDonald E, McShane DJ, Medsger T, Michel B, Murphy WA, Osial T, Ramsey-Goldman R, Rothschild B, Wolfe F (1991) The American college of rheumatology criteria for the classification and reporting of osteoarthritis of the hip. Arthritis Rheum 34:505–514
Audenaert EA, Mahieu P, Pattyn C (2011) Three-dimensional assessment of cam engagement in femoroacetabular impingement. Arthroscopy 27:167–171
Bierma-Zeinstra S, Bohnen A, Ramial R, Ridderikhoff J, Verhaar J, Prins A (1998) Comparison between two devices for measuring hip joint motions. Clin Rehabil 12:497–505
Bland JM, Altman DG (1990) A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. Comput Biol Med 20:337–340
Croft PR, Nahit ES, Macfarlane GJ, Silman AJ (1996) Interobserver reliability in measuring flexion, internal rotation, and external rotation of the hip using a plurimeter. Ann Rheum Dis 55:320–323
Ecker TMMD, Tannast MMD, Puls MM, Siebenrock KAMD, Murphy SBMD (2007) Pathomorphologic alterations predict presence or absence of hip osteoarthrosis. Clin Orthop Relat Res 465:46–52
Hollman J, Burgess B, Bokermann J (2003) Passive hip rotation range of motion: effects of testing position and age in runners and non-runners. Physiother Theory Pract 19:77–86
Kelly BT, Bedi A, Robertson CM, Dela Torre K, Giveans MR, Larson CM (2012) Alterations in internal rotation and alpha angles are associated with arthroscopic cam decompression in the hip. Am J Sports Med 40:1107–1112
Kolber MJ, Hanney WJ (2012) The reliability and concurrent validity of shoulder mobility measurements using a digital inclinometer and goniometer: a technical report. Int J Sports Phys Ther 7:306–313
Krause DA, Levy BA, Shah JP, Stuart MJ, Hollman MJ, Dahm DL (2013) Reliability of the dial test using a handheld inclinometer. Knee Surg Sports Traumatol Arthrosc 21:1011–1016
Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174
Norkin C, White D (2009) Measurement of joint motion: a guide to goniometry, 4th edn. Philadelphia, FA, Davis
Nussbaumer S, Leunig M, Glatthorn JF, Stauffacher S, Gerber H, Maffiuletti NA (2010) Validity and test–retest reliability of manual goniometers for measuring passive hip range of motion in femoroacetabular impingement patients. BMC Musculoskelet Disord 11:194
Pua Y-H, Wrigley TW, Cowan SM, Bennell KL (2008) Intrarater test–retest reliability of hip range of motion and hip muscle strength measurements in persons with hip osteoarthritis. Arch Phys Med Rehabil 89:1146–1154
Reichenbach S, Jüni P, Nüesch E, Frey F, Ganz R, Leunig M (2010) An examination chair to measure internal rotation of the hip in routine settings: a validation study. Osteoarthritis Cartilage 18:365–371
Roach KE, Miles TP (1991) Normal hip and knee active range of motion: the relationship to age. Phys Ther 71:656–665
Shindle MK, Voos JE, Heyworth BE, Mintz DN, Moya LE, Buly RL, Kelly BT (2007) Hip arthroscopy in the athletic patient: current techniques and spectrum of disease. J Bone Joint Surg Am 89(Suppl 3):29–43
Shrout P, Fleiss J (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428
Simoneau GG, Hoenig KJ, Lepley JE, Papanek PE (1998) Influence of hip position and gender on active hip internal and external rotation. J Orthop Sports Phys Ther 28:158–164
Yuan BJ, Bartelt RB, Levy BA, Bond JR, Trousdale RT, Sierra RJ (2013) Decreased range of motion is associated with structural hip deformity in asymptomatic adolescent athletes. Am J Sports Med 41:1519–1525
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Krause, D.A., Hollman, J.H., Krych, A.J. et al. Reliability of hip internal rotation range of motion measurement using a digital inclinometer. Knee Surg Sports Traumatol Arthrosc 23, 2562–2567 (2015). https://doi.org/10.1007/s00167-014-3096-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00167-014-3096-0