Abstract
Three cases of shunt valve fracture with separation are reported. Fracture with separation of the shunt valve of a cerebrospinal fluid shunt is an infrequent occurrence. Because a portion of the shunt valve is radiolucent on a shunt survey radiograph, this problem is easily overlooked if it occurs. The diagnosis can be made by comparing old and new shunt survey radiographs and noting the increased distance between radiopaque portions of the shunt on the new X-ray films.
Similar content being viewed by others
References
Forrest DM, Cooper DGW (1968) Complications of ventriculo-atrial shunts. J Neurosurg 29:506–512
Jakoby RK (1963) Bony spicule piercing the pumping device for shunting cerebrospinal fluid. J Neurosurg 20:523
Liesegang J, Strahl EW, Streicher HR (1980) Complications following shunt operations in children. In: Grote W, Brock M, Clar HE, Klinger M, Nau HE (eds) Advances in neurosurgery, vol 8, Springer, Berlin Heidelberg New York, pp 222–226
Martínez-Lage JF, Poza M, Esteban JA (1992) Mechanical complications of the reservoirs and flushing devices in ventricular shunt systems. Br J Neurosurg 6:321–326
Metzemaekers JDM, Beks JWF, Popta JS van (1987) Cerebrospinal fluid shunting for hydrocephalus: a retrospective analysis. Acta Neurochir (Wien) 88:75–78
Pedersen PH, Waaler PE, Wester K (1992) Dysfunction of Holter mini elliptical valves caused by minor head trauma. Dev Med Child Neurol 34:66–68
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Hellbusch, L.C. Cerebrospinal fluid shunt valve fracture with separation: report of three cases. Child's Nerv Syst 12, 142–144 (1996). https://doi.org/10.1007/BF00266816
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00266816