Reliability of Optic Nerve Ultrasound for the Evaluation of Patients with Spontaneous Intracranial Hemorrhage

  • Take Notice Technology
  • Published:
Neurocritical Care Aims and scope Submit manuscript

Abstract

Introduction

The aim of our study is to confirm the reliability of optic nerve ultrasound as a method to detect intracranial hypertension in patients with spontaneous intracranial hemorrhage, to assess the reproducibility of the measurement of the optic nerve sheath diameter (ONSD), and to verify that ONSD changes concurrently with intracranial pressure (ICP) variations.

Methods

Sixty-three adult patients with subarachnoid hemorrhage (n = 34) or primary intracerebral hemorrhage (n = 29) requiring sedation and invasive ICP monitoring were enrolled in a 10-bed multivalent ICU. ONSD was measured 3 mm behind the globe through a 7.5-MHz ultrasound probe. Mean binocular ONSD was used for statistical analysis. ICP values were registered simultaneously to ultrasonography. Twenty-eight ONSDs were measured consecutively by two different observers, and interobserver differences were calculated. Twelve coupled measurements were taken before and within 1 min after cerebrospinal fluid (CSF) drainage to control elevated ICP.

Results

Ninety-four ONSD measurements were analyzed. 5.2 mm proved to be the optimal ONSD cut-off point to predict raised ICP (>20 mmHg) with 93.1% sensitivity (95% CI: 77.2–99%) and 73.85% specificity (95% CI: 61.5–84%). ONSD–ICP correlation coefficient was 0.7042 (95% CI for r = 0.5850–0.7936). The median interobserver ONSD difference was 0.25 mm. CSF drainage to control elevated ICP caused a rapid and significant reduction of ONSD (from 5.89 ± 0.61 to 5 ± 0.33 mm, P < 0.01).

Conclusion

Our investigation confirms the reliability of optic nerve ultrasound as a non-invasive method to detect elevated ICP in intracranial hemorrhage patients. ONSD measurements proved to have a good reproducibility. ONSD changes almost concurrently with CSF pressure variations.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Hayreh SS. Pathogenesis of oedema of the optic disc (papilloedema). Br J Opthalmol. 1964;48:522–43.

    Article  CAS  Google Scholar 

  2. Liu D, Kahn M. Measurement and relationship of subarachnoid pressure of the optic nerve to intracranial pressures in fresh cadavers. Am J Opthalmol. 1993;116:548–56.

    CAS  Google Scholar 

  3. Helmke K, Hansen HC. Fundamentals of transorbital sonographic evaluation of optic nerve sheath expansion under intracranial hypertension. I. Experimental study. Pediatr Radiol. 1996;26:701–5.

    Article  CAS  PubMed  Google Scholar 

  4. Helmke K, Hansen HC. Fundamentals of transorbital sonographic evaluation of optic nerve sheath expansion under intracranial hypertension. II. Patient study. Pediatr Radiol. 1996;26:706–10.

    Article  CAS  PubMed  Google Scholar 

  5. Helmke K, Burdelski M, Hansen H. Detection and monitoring of intracranial pressure dysregulation in liver failure by ultrasound. Transplantation. 2000;70(2):392–5.

    Article  CAS  PubMed  Google Scholar 

  6. Newman WD, Hollman AS, Dutton GN, Carachi R. Measurement of optic nerve sheath diameter by ultrasound: a means of detecting acute raised intracranial pressure in hydrocephalus. Br J Opthalmol. 2002;86:1109–13.

    Article  CAS  Google Scholar 

  7. Tayal VS, Neulander M, Norton HJ, Foster T, Saunders T, Blavias M. Emergency department sonographic measurement of optic nerve sheath diameter to detect findings of increased intracranial pressure in adult head injury patients. Ann Emerg Med. 2007;49:508–14.

    Article  PubMed  Google Scholar 

  8. Blavias M, Theodoro DL, Sierzenski P. Elevated intracranial pressure detected by bedside emergency ultrasound of the optic nerve sheath. Acad Emerg Med. 2003;10:376–81.

    Article  Google Scholar 

  9. Goel RS, Goyal NK, Dharap SB, Kumar M, Gore MA. Utility of optic nerve ultrasonography in head injury. Injury. 2008;39(5):519–24.

    Article  PubMed  Google Scholar 

  10. Hansen HC, Helmke K. Validation of the optic nerve sheath response to changing cerebrospinal fluid pressure: ultrasound findings during intrathecal infusion tests. J Neurosurg. 1997;87:34–40.

    Article  CAS  PubMed  Google Scholar 

  11. Geeraerts T, Launey Y, Martin L, Kumar M, Gore MA. Ultrasonography of the optic nerve sheath may be useful for detecting raised intracranial pressure after severe brain injury. Intensive Care Med. 2007;33:1704–11.

    Article  PubMed  Google Scholar 

  12. Harbison Kimberly H, Shah S, Marill K, Noble V. Correlation of optic nerve sheath with direct measurement of intracranial pressure. Acad Emerg Med. 2008;15:201–4.

    Article  Google Scholar 

  13. Geeraerts T, Merceron S, Benhamou D, Vigué B, Duranteau J. Non-invasive assessment of intracranial pressure using ocular sonography in neurocritical care patients. Intensive Care Med. 2008;34(11):2062–7.

    Article  PubMed  Google Scholar 

  14. Soldatos T, Karakistos D, Chatzimichail K, Papathanasiou M, Gouliamos A, Karabinis A. Optic nerve sonography in the diagnostic evaluation of adult brain injury. Crit Care. 2008;12(3):150.

    Article  Google Scholar 

  15. Moretti R, Pizzi B. Optic nerve ultrasound for detection of intracranial hypertension in intracranial hemorrhage patients. J Neurosurg Anesthesiol. 2009;21(1):16–20.

    Article  PubMed  Google Scholar 

  16. Copetti R, Cattarossi L. Optic nerve ultrasound: artifacts and real images. Intensive Care Med [On-line] 2009. doi:10.1007/s00134-009-1494-4.

  17. Romagnulo L, Tayal V, Tomazewski C, Saunders T, Northon J. Optic nerve sheath does not change with patient position. Am J Emerg Med. 2005;23:686–8.

    Article  Google Scholar 

  18. Lichtenstein D. Optic nerve and intracranial hypertension. In: Lichtenstein D, editor. General ultrasound in the critically ill. Berlin: Springer; 2007. p. 152–4.

    Google Scholar 

  19. Geeraerts T, Bergès O, Merceron S, et al. Reply to Copetti and Cattarossi. Intensive Care Med [On-line] 2009. doi:10.1007/s00134-009-1495-3.

  20. Geeraerts T, Newcombe VFJ, Coles JP, Abate MJ, Perkes IE, Hutchinson PJA, et al. Use of T2-weighed magnetic resonance imaging of the optic nerve sheath to detect raised intracranial pressure. Crit Care. 2008;12:R114.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Riccardo Moretti.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Moretti, R., Pizzi, B., Cassini, F. et al. Reliability of Optic Nerve Ultrasound for the Evaluation of Patients with Spontaneous Intracranial Hemorrhage. Neurocrit Care 11, 406–410 (2009). https://doi.org/10.1007/s12028-009-9250-8