The sensitivity of new color systems in blood-flow diagnosis

The maximum entropy method and angio-color-comparative in vitro flow measurements to determine sensitivity

  • Published:
Surgical Endoscopy Aims and scope Submit manuscript

An Erratum to this article was published on 01 November 1997

Abstract

Background: Two new blood-flow-diagnosis techniques have recently been developed as supplements to the established color techniques: the MEM (maximum entropy method) technique and color flow amplitude (power Doppler). These are capable of representing blood flow in distinctly more slowly flowing areas than is possible with the conventional Doppler technique.

Methods: Both methods make use of the Doppler technique in part, yet analyze the reflected signal in a different manner, in so doing optimizing the relation between the noise and the signal. Measurements were obtained on two anatomic flow models to test the sensitivity of both techniques under slow flow conditions.

Results: The slowest flow the MEM technique was capable of recording was 0.5 mm/s, whereas the angiotechnique was able to detect 0.4 mm/s, albeit utilizing a 5-MHz transducer for the MEM technique and a 10-MHz transducer for the angio technique. One may thus assume that the MEM technique would be still more sensitive when utilizing a 10-MHz transducer. The advantage of the MEM technique is its real-time flow representation: The angio technique requires a few seconds of acquisition time. This could have serious consequences during clinical utilization. Doppler sonography was merely capable of detecting a minimum flow velocity of 15 mm/s. The angio technique is less dependent on the angle during flow representation than the MEM technique.

Conclusions: The distinctly higher sensitivity of these two new color techniques offers new possibilities in the clinical sector as far as the perfusion diagnosis of organs and tumors is concerned.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 9 December 1996/Accepted: 11 March 1997

An erratum to this article is available at http://dx.doi.org/10.1007/s004649900554.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Sohn, C., Weskott, H. The sensitivity of new color systems in blood-flow diagnosis . Surg Endosc 11, 1040–1044 (1997). https://doi.org/10.1007/s004649900522

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s004649900522

Navigation