Skip to main content
Log in

Intra-abdominal Breakage of Laparoscopic Needle Holder Tip and Its Retrieval Under Fluoroscopic Guidance

  • Instrumentation and Techniques
  • Published:
The Journal of Obstetrics and Gynecology of India Aims and scope Submit manuscript

Abstract

Background

Intra-operative breakage of laparoscopy instruments is a rare occurrence. It entails a risk of the foreign body being retained in the abdomen and may lead to medico-legal implications. The foreign body migrates due to bowel movements. Therefore, the retrieval of such foreign bodies can be challenging even for a highly skilled and experienced surgeon. Fluoroscopy can guide in locating the missing foreign body. The only preventive measure is the vigilance of the operating team and thorough inspection of all the instruments prior to completion of surgery.

Case Report

We present a case report of a 50-year-old lady who underwent total laparoscopic hysterectomy with bilateral salpingoopherectomy at a private nursing home. During the closure of the vault, the tip of the needle holder broke inside the abdomen and the primary surgeon was unable to retrieve it. The missing tip of the needle holder was removed from the right hypochondriac region, by a second laparoscopic surgery by an advanced laparoscopic surgeon under fluoroscopic guidance.

Conclusion

Intra-operative breakage of laparoscopy instruments can occur, demanding a great deal of expertise for the retrieval of such foreign bodies. Fluoroscopy can guide in locating the missing foreign body which can be retrieved by an expert laparoscopic surgeon.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Katara AN, Bhandarkar DS, Shah RS, Udwadia TE. Breakage of fascial closure device during laparoscopic surgery. J Mini Access Surg. 2005;1(2):79.

    Article  CAS  Google Scholar 

  2. Sang AX, Lal R, August A, Danzer E, Buckingham B, Mueller CM. Fluoroscopic-assisted laparoscopic retrieval of retained glucose sensor wire from the omentum. Clin Case Rep. 2019;7(9):1717–20.

    Article  Google Scholar 

  3. Rani BD, Jarugulla S. Retained surgical instrument in minimally invasive surgery. J Obstet Gynecol India. 2019; 1–2.

  4. Lynch CM, Powers AK. Management of a broken needle at the time of laparoscopic burch. JSLS J Soc Laparoendosc Surg. 2000;4(3):275.

    CAS  Google Scholar 

  5. Duhan N, Malik R, Hooda MR. A broken cautery tip lost in the abdomen at laparoscopic ovarian cystectomy. Open J Clin Med Case Rep. 2016;2(2):1067.

    Google Scholar 

  6. Kandioler-Eckersberger D, Niederle B, Herbst F, Wenzl E. A magnetic probe to retrieve broken metallic parts of instruments during laparoscopic procedures. Surg Endos Other Interv Tech. 2002;16(1):208–9.

    Article  CAS  Google Scholar 

  7. Yasuhara H, Fukatsu K, Komatsu T, Murakoshi S, Saito Y, Uetera Y. Occult risk of broken instruments for endoscopy-assisted surgery. World J Surg. 2014;38(11):3015–22.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Yagnesh V. Patel.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interests.

Informed Consent

Patient’s identity is not revealed as only X-ray and fluroscopic images are being sent for publication.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Sanjay Brahmbhatt M.D., Amrita Makhija M.S., Jayna Brahmbhatt M.D., Yagnesh V Patel D.N.B.: Department of Gynecological Endoscopy, Himalaya Hospital, Anand, India.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Brahmbhatt, S., Makhija, A., Brahmbhatt, J. et al. Intra-abdominal Breakage of Laparoscopic Needle Holder Tip and Its Retrieval Under Fluoroscopic Guidance. J Obstet Gynecol India 71, 452–455 (2021). https://doi.org/10.1007/s13224-020-01394-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13224-020-01394-6

Keywords

Navigation