Abstract
Purpose
The treatment efficiency of 90Y and providing reliable estimates of activity are evaluated by SPECT imaging of bremsstrahlung radiation released during beta therapy. In this technique, the resulting spectrum from 90Y is very complex and continuous, which creates difficulties on the imaging protocol. Moreover, collimator geometry has an impressive effect on the spatial resolution, system sensitivity, image contrast, and the signal-to-noise ratio (SNR), which should be optimized.
Methods
We evaluated the effect of energy window width, reconstruction algorithms, and different geometries of a medium-energy (ME) parallel-hole collimator on the image contrast and SNR of 90Y SPECT images. The Siemens E.Cam gamma camera equipped with a ME collimator and a digital Jaszczak phantom were simulated by SIMIND Monte Carlo program to generate the 90Y bremsstrahlung SPECT images.
Results
Our results showed that optimal image quality can be acquired by the reconstruction algorithm of OS-EM in the energy window width of 60 to 400 keV for 90Y bremsstrahlung SPECT imaging. Furthermore, the optimal values of the hole diameter and hole length of a ME collimator were obtained 0.235 and 4.4 cm, respectively.
Conclusions
The acquired optimal ME collimator and energy window along with using a suitable reconstruction algorithm lead to improved contrast and SNR of 90Y bremsstrahlung images of hot spheres of the digital Jaszczak phantom. This can improve the accuracy and precision of the 90Y activity distribution estimation after radioembolization in targeted radionuclide therapy.
Similar content being viewed by others
References
Minarik D, Gleisner KS, Ljungberg M. Evaluation of quantitative 90-Y SPECT based on experimental phantom studies. Phys Med Biol. 2008;53:5689–703.
Rong X, Frey EC. A collimator optimization method for quantitative imaging: application to Y-90 bremsstrahlung SPECT. Med Phys. 2013;40:1–9.
Rong X, Ghaly M, Frey EC. Optimization of energy window for 90Y bremsstrahlung SPECT imaging for detection tasks using the ideal observer with model-mismatch. Med Phys. 2013;40:062502.
Rong X, Du Y, Ljungberg M, Rault E, Vandenberghe S, Frey EC. Development and evaluation of an improved quantitative 90-Y bremsstrahlung SPECT method. Med Phys. 2012;39:2346–58.
Walrand S, Hesse M, Wojcik R, Lhommel R, Jamar F. Optimal design of Anger camera for bremsstrahlung imaging: Monte Carlo evaluation. Front Oncol. 2014;4:1–7.
McQuaid SJ, Southekal S, Kijewski MF, Moore SC. Joint optimization of collimator and reconstruction parameters in SPECT imaging for lesion quantification. Phys Med Biol. 2011;56:6983–7000.
Chery SR, Sorenson JA, Phelps ME. Physics in Nuclear Medicine. Elsevier/Saunders. Philadelphia; 2012.
Bonutti F, Avolio M, Magro G, Cecotti A, Della Schiava E, Del Dò E, et al. Optimization of the image contrast in SPECT-CT bremsstrahlung imaging for selective internal radiation therapy of liver malignancies with Y-90 microspheres. arXiv preprint arXiv:1509.08857. 2015;1-13.
Rezaei Roshan H, Azarm AR, Mahmoudian B, Gharepapagh E, Islamian J. Collimator and energy window optimization for Y-90 bremsstrahlung SPECT imaging: a SIMIND Monte Carlo study. Appl Radiat Isot. 2016;108:124–8.
Sadremomtaz A, Taherparvar P. The influence of filters on the SPECT image of Carlson phantom. J Biomed Sci Eng. 2013;6:291–7.
Bruyant PP. Analytic and iterative reconstruction algorithms in SPECT. J Nucl Med. 2002;43:1343–58.
Hatton RL, Hutton BF, Angelides S, Choong KK, Larcos G. Improved tolerance to missing data in myocardial perfusion SPET using OS-EM reconstruction. Eur J Nucl Med Mol Imaging. 2004;31:857–61.
Leong LK, Kruger RL, O’Connor MK. A comparison of the uniformity requirements for SPECT image reconstruction using FBP and OS-EM techniques. J Nucl Med Technol. 2001;29:79–83.
Ahmadzadehfar H, Muckle M, Sabet A, Wilhelm K, Kuhl C, Biermann K, et al. The significance of bremsstrahlung SPECT/CT after yttrium-90 radioembolization treatment in the prediction of extrahepatic side effects. Eur J Nucl Med Mol Imaging. 2012;39:309–15.
Rong X, Du Y, Frey EC. A method for energy window optimization for quantitative tasks that includes the effects of model-mismatch on bias: application to Y-90 bremsstrahlung SPECT imaging. Phys Med Biol. 2012;57:3711–25.
Rezaei Roshan H, Azarm AR, Mahmoudian B, Islamian J. Advances in SPECT for optimizing the liver tumors radioembolization using yttrium-90 microspheres. World J Nucl Med. 2015;14:75–80.
Kim M, Bae JK, Hong BH, Kim KM, Lee W. Effects of collimator on imaging performance of yttrium-90 bremsstrahlung photons: Monte Carlo simulation. Nucl Eng Technol. 2019;51:539–45.
Ljungberg M. The SIMIND Monte Carlo program. Monte Carlo Calculations in Nuclear Medicine: Applications in Diagnostic Imaging. 2012. pp. 111-28.
Toossi MB, Islamian JP, Momennezhad M, Ljungberg M, Naseri S. SIMIND Monte Carlo simulation of a single photon emission CT. J Med Phys/Assoc Med Phys India. 2010;35:42.
Taherparvar P, Sadremomtaz A. Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT. Australas Phys Eng Sci Med. 2018;41:31–9.
Rajabi R, Taherparvar P. Monte Carlo dosimetry for a new 32P brachytherapy source using FLUKA code. J Contemp Brachyther. 2019;11:76–90.
Fardi Z, Taherparvar P. A Monte Carlo investigation of the dose distribution for new I-125 low dose rate brachytherapy source in water and in different media. Polish J Med Phys Eng. 2019;25:15–22.
http://depts.washington.edu/simset/html/simset_main.html. Accessed 18 May 2019.
Islamian JP, Bahreyni Toossi MT, Momennezhad M, Naseri S, Ljungberg M. Simulation of a quality control Jaszczak Phantom with SIMIND monte carlo and adding the phantom as an accessory to the program. Iran J Med Phys. 2012;9:135–40.
Smith MF, Majewski S, Weisenberger AG. Optimizing pinhole and parallel hole collimation for scintimammography with compact pixellated detectors. Nucl Sci IEEE Trans. 2003;50:321–6.
Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging. 2007;26:375–85.
Acknowledgments
The authors would like to thank Professor Michael Ljungberg for providing the valuable advices.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of Interest
Payvand Taherparvar declares that he has no conflict of interest. Nazila Shahmari declares that she has no conflict of interest.
Ethical Approval
This article does not contain any studies with human participants or animals performed by any of the authors.
Informed Consent
The institutional review board of our institute approved this retrospective study, and the requirement to obtain informed consent was waived.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Taherparvar, P., Shahmari, N. Effect of ME Collimator Characteristic, Energy Window Width, and Reconstruction Algorithm Selection on Imaging Performance of Yttrium-90: Simulation Study. Nucl Med Mol Imaging 53, 414–422 (2019). https://doi.org/10.1007/s13139-019-00619-5
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13139-019-00619-5