Effect of ME Collimator Characteristic, Energy Window Width, and Reconstruction Algorithm Selection on Imaging Performance of Yttrium-90: Simulation Study

  • Original Article
  • Published:
Nuclear Medicine and Molecular Imaging Aims and scope Submit manuscript

Abstract

Purpose

The treatment efficiency of 90Y and providing reliable estimates of activity are evaluated by SPECT imaging of bremsstrahlung radiation released during beta therapy. In this technique, the resulting spectrum from 90Y is very complex and continuous, which creates difficulties on the imaging protocol. Moreover, collimator geometry has an impressive effect on the spatial resolution, system sensitivity, image contrast, and the signal-to-noise ratio (SNR), which should be optimized.

Methods

We evaluated the effect of energy window width, reconstruction algorithms, and different geometries of a medium-energy (ME) parallel-hole collimator on the image contrast and SNR of 90Y SPECT images. The Siemens E.Cam gamma camera equipped with a ME collimator and a digital Jaszczak phantom were simulated by SIMIND Monte Carlo program to generate the 90Y bremsstrahlung SPECT images.

Results

Our results showed that optimal image quality can be acquired by the reconstruction algorithm of OS-EM in the energy window width of 60 to 400 keV for 90Y bremsstrahlung SPECT imaging. Furthermore, the optimal values of the hole diameter and hole length of a ME collimator were obtained 0.235 and 4.4 cm, respectively.

Conclusions

The acquired optimal ME collimator and energy window along with using a suitable reconstruction algorithm lead to improved contrast and SNR of 90Y bremsstrahlung images of hot spheres of the digital Jaszczak phantom. This can improve the accuracy and precision of the 90Y activity distribution estimation after radioembolization in targeted radionuclide therapy.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Minarik D, Gleisner KS, Ljungberg M. Evaluation of quantitative 90-Y SPECT based on experimental phantom studies. Phys Med Biol. 2008;53:5689–703.

    Article  CAS  Google Scholar 

  2. Rong X, Frey EC. A collimator optimization method for quantitative imaging: application to Y-90 bremsstrahlung SPECT. Med Phys. 2013;40:1–9.

    Article  Google Scholar 

  3. Rong X, Ghaly M, Frey EC. Optimization of energy window for 90Y bremsstrahlung SPECT imaging for detection tasks using the ideal observer with model-mismatch. Med Phys. 2013;40:062502.

    Article  Google Scholar 

  4. Rong X, Du Y, Ljungberg M, Rault E, Vandenberghe S, Frey EC. Development and evaluation of an improved quantitative 90-Y bremsstrahlung SPECT method. Med Phys. 2012;39:2346–58.

    Article  CAS  Google Scholar 

  5. Walrand S, Hesse M, Wojcik R, Lhommel R, Jamar F. Optimal design of Anger camera for bremsstrahlung imaging: Monte Carlo evaluation. Front Oncol. 2014;4:1–7.

    Article  Google Scholar 

  6. McQuaid SJ, Southekal S, Kijewski MF, Moore SC. Joint optimization of collimator and reconstruction parameters in SPECT imaging for lesion quantification. Phys Med Biol. 2011;56:6983–7000.

    Article  Google Scholar 

  7. Chery SR, Sorenson JA, Phelps ME. Physics in Nuclear Medicine. Elsevier/Saunders. Philadelphia; 2012.

  8. Bonutti F, Avolio M, Magro G, Cecotti A, Della Schiava E, Del Dò E, et al. Optimization of the image contrast in SPECT-CT bremsstrahlung imaging for selective internal radiation therapy of liver malignancies with Y-90 microspheres. arXiv preprint arXiv:1509.08857. 2015;1-13.

  9. Rezaei Roshan H, Azarm AR, Mahmoudian B, Gharepapagh E, Islamian J. Collimator and energy window optimization for Y-90 bremsstrahlung SPECT imaging: a SIMIND Monte Carlo study. Appl Radiat Isot. 2016;108:124–8.

    Article  Google Scholar 

  10. Sadremomtaz A, Taherparvar P. The influence of filters on the SPECT image of Carlson phantom. J Biomed Sci Eng. 2013;6:291–7.

    Article  Google Scholar 

  11. Bruyant PP. Analytic and iterative reconstruction algorithms in SPECT. J Nucl Med. 2002;43:1343–58.

    PubMed  Google Scholar 

  12. Hatton RL, Hutton BF, Angelides S, Choong KK, Larcos G. Improved tolerance to missing data in myocardial perfusion SPET using OS-EM reconstruction. Eur J Nucl Med Mol Imaging. 2004;31:857–61.

    Article  CAS  Google Scholar 

  13. Leong LK, Kruger RL, O’Connor MK. A comparison of the uniformity requirements for SPECT image reconstruction using FBP and OS-EM techniques. J Nucl Med Technol. 2001;29:79–83.

    CAS  PubMed  Google Scholar 

  14. Ahmadzadehfar H, Muckle M, Sabet A, Wilhelm K, Kuhl C, Biermann K, et al. The significance of bremsstrahlung SPECT/CT after yttrium-90 radioembolization treatment in the prediction of extrahepatic side effects. Eur J Nucl Med Mol Imaging. 2012;39:309–15.

    Article  CAS  Google Scholar 

  15. Rong X, Du Y, Frey EC. A method for energy window optimization for quantitative tasks that includes the effects of model-mismatch on bias: application to Y-90 bremsstrahlung SPECT imaging. Phys Med Biol. 2012;57:3711–25.

    Article  Google Scholar 

  16. Rezaei Roshan H, Azarm AR, Mahmoudian B, Islamian J. Advances in SPECT for optimizing the liver tumors radioembolization using yttrium-90 microspheres. World J Nucl Med. 2015;14:75–80.

    Article  Google Scholar 

  17. Kim M, Bae JK, Hong BH, Kim KM, Lee W. Effects of collimator on imaging performance of yttrium-90 bremsstrahlung photons: Monte Carlo simulation. Nucl Eng Technol. 2019;51:539–45.

    Article  CAS  Google Scholar 

  18. Ljungberg M. The SIMIND Monte Carlo program. Monte Carlo Calculations in Nuclear Medicine: Applications in Diagnostic Imaging. 2012. pp. 111-28.

    Chapter  Google Scholar 

  19. Toossi MB, Islamian JP, Momennezhad M, Ljungberg M, Naseri S. SIMIND Monte Carlo simulation of a single photon emission CT. J Med Phys/Assoc Med Phys India. 2010;35:42.

    Google Scholar 

  20. Taherparvar P, Sadremomtaz A. Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT. Australas Phys Eng Sci Med. 2018;41:31–9.

    Article  Google Scholar 

  21. Rajabi R, Taherparvar P. Monte Carlo dosimetry for a new 32P brachytherapy source using FLUKA code. J Contemp Brachyther. 2019;11:76–90.

    Article  Google Scholar 

  22. Fardi Z, Taherparvar P. A Monte Carlo investigation of the dose distribution for new I-125 low dose rate brachytherapy source in water and in different media. Polish J Med Phys Eng. 2019;25:15–22.

    Article  Google Scholar 

  23. http://depts.washington.edu/simset/html/simset_main.html. Accessed 18 May 2019.

  24. Islamian JP, Bahreyni Toossi MT, Momennezhad M, Naseri S, Ljungberg M. Simulation of a quality control Jaszczak Phantom with SIMIND monte carlo and adding the phantom as an accessory to the program. Iran J Med Phys. 2012;9:135–40.

    Google Scholar 

  25. Smith MF, Majewski S, Weisenberger AG. Optimizing pinhole and parallel hole collimation for scintimammography with compact pixellated detectors. Nucl Sci IEEE Trans. 2003;50:321–6.

    Article  Google Scholar 

  26. Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging. 2007;26:375–85.

    Article  Google Scholar 

Download references

Acknowledgments

The authors would like to thank Professor Michael Ljungberg for providing the valuable advices.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Payvand Taherparvar.

Ethics declarations

Conflict of Interest

Payvand Taherparvar declares that he has no conflict of interest. Nazila Shahmari declares that she has no conflict of interest.

Ethical Approval

This article does not contain any studies with human participants or animals performed by any of the authors.

Informed Consent

The institutional review board of our institute approved this retrospective study, and the requirement to obtain informed consent was waived.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Taherparvar, P., Shahmari, N. Effect of ME Collimator Characteristic, Energy Window Width, and Reconstruction Algorithm Selection on Imaging Performance of Yttrium-90: Simulation Study. Nucl Med Mol Imaging 53, 414–422 (2019). https://doi.org/10.1007/s13139-019-00619-5

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13139-019-00619-5

Keywords

Navigation