A Review of Electronic Devices to Assess Inhaler Technique

  • Allergies and the Environment (M Hernandez, Section Editor)
  • Published:
Current Allergy and Asthma Reports Aims and scope Submit manuscript

Abstract

Purpose of Review

Multiple electronic devices exist that provide feedback on the accuracy of patient inhaler technique. Our purpose is to describe the inhaler technique feedback provided by these devices, including specific technique steps measured, how feedback is displayed, target of feedback (patient, provider, researcher), and compatibility with inhaler type (metered-dose inhaler [MDI], diskus, etc.).

Recent Findings

We identified eight devices that provide feedback on inhaler technique. Only one device assessed all evidence-based MDI technique steps. Most devices provide limited real-time feedback to patients, if any feedback at all.

Summary

Technologies to assess inhaler technique are advancing and hold great potential for improving patient inhaler technique. Many devices are limited in their ability to detect all evidence-based technique steps and provide real-time user-friendly feedback to patients and providers. Usability tests with patients and providers could identify ways to improve these devices to improve their utility in clinical settings.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

Papers of particular interest, published recently, have been highlighted as: •• Of major importance

  1. Masoli M, Fabian D, Holt S, Beasley R. The global burden of asthma: executive summary of the GINA dissemination committee report. Allergy. 2004;59(5):469–78. doi:10.1111/j.1398-9995.2004.00526.x.

    Article  PubMed  Google Scholar 

  2. Akinbami LJ, Moorman JE, Bailey C, Zahran HS, King M, Johnson CA, et al. Trends in asthma prevalence, health care use, and mortality in the United States, 2001–2010. NCHS data brief, no 94. Hyattsville, MD: National Center for Health Statistics; 2012.

    Google Scholar 

  3. Centers for Disease Control and Prevention. Asthma's impact on the nation—data from the CDC National Asthma Control Program. 2014. p. 1–4.

  4. Sullivan PW, Ghushchyan VH, Campbell JD, Globe G, Bender B, Magid DJ. Measuring the cost of poor asthma control and exacerbations. J Asthma. 2017;54(1):24–31. doi:10.1080/02770903.2016.1194430.

    Article  PubMed  Google Scholar 

  5. Bahadori K, Doyle-Waters MM, Marra C, Lynd L, Alasaly K, Swiston J, et al. Economic burden of asthma: a systematic review. BMC Pulm Med. 2009;9(1):24. doi:10.1186/1471-2466-9-24.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Bartlett SJ, Lukk P, Butz A, Lampros-Klein F, Rand CS. Enhancing medication adherence among inner-city children with asthma: results from pilot studies. J Asthma. 2002;39(1):47–54. doi:10.1081/JAS-120000806.

    Article  PubMed  Google Scholar 

  7. Bauman LJ, Wright E, Leickly FE, Crain E, Kruszon-Moran D, Wade SL, et al. Relationship of adherence to pediatric asthma morbidity among inner-city children. Pediatrics. 2002;110(1):e6. doi:10.1542/peds.110.1.e6.

    Article  PubMed  Google Scholar 

  8. Zeiger RS, Hay JW, Contreras R, Chen W, Quinn VP, Seal B, et al. Asthma costs and utilization in a managed care organization. J Allergy Clin Immunol. 2008;121(4):885–92.e5. doi:10.1016/j.jaci.2007.12.1178.

    Article  PubMed  Google Scholar 

  9. •• Kikidis D, Konstantinos V, Tzovaras D, Usmani OS. The digital asthma patient: the history and future of inhaler based health monitoring devices. J Aerosol Med Pulm Drug Deliv. 2016;29(3):219–32. doi:10.1089/jamp.2015.1267. This article reviewed inhaler-based monitoring devices and documented that several digital inhaler monitoring devices are available to assess inhaler technique. It shows that digital devices can validly and reliably assess patient inhaler technique, and we extend upon this research.

    Article  PubMed  Google Scholar 

  10. Clark NM, Gong M, Kaciroti N. A model of self-regulation for control of chronic disease. Health Educ Behav. 2001;28(6):769–82. doi:10.1177/109019810102800608.

    Article  CAS  PubMed  Google Scholar 

  11. Al-Jahdali H, Ahmed A, Al-Harbi A, Khan M, Baharoon S, Bin Salih S, et al. Improper inhaler technique is associated with poor asthma control and frequent emergency department visits. Allergy Asthma Clin Immunol. 2013;9(1):1–7. doi:10.1186/1710-1492-9-8.

    Article  Google Scholar 

  12. Giraud V, Roche N. Misuse of corticosteroid metered-dose inhaler is associated with decreased asthma stability. Eur Respir J. 2002;19(2):246–51. doi:10.1183/09031936.02.00218402.

    Article  CAS  PubMed  Google Scholar 

  13. Basheti IA, Reddel HK, Armour CL, Bosnic-Anticevich SZ. Improved asthma outcomes with a simple inhaler technique intervention by community pharmacists. J Allergy Clin Immunol. 2007;119(6):1537–8. doi:10.1016/j.jaci.2007.02.037.

    Article  PubMed  Google Scholar 

  14. Sleath B, Ayala GX, Gillette C, Williams D, Davis S, Tudor G, et al. Provider demonstration and assessment of child device technique during pediatric asthma visits. Pediatrics. 2011;127(4):642–8. doi:10.1542/peds.2010-1206.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Pedersen S. Inhaler use in children with asthma. Dan Med Bull. 1987;34(5):234–49.

    CAS  PubMed  Google Scholar 

  16. Pedersen SK, Frost L, Arnfred T. Errors in inhalation technique and efficacy of inhaler use in asthmatic children. Allergy. 1986;41(2):118–24. doi:10.1111/j.1398-9995.1986.tb00287.x.

    Article  CAS  PubMed  Google Scholar 

  17. Toumas-Shehata M, Price D, Basheti IA, Bosnic-Anticevich S. Exploring the role of quantitative feedback in inhaler technique education: a cluster-randomised, two-arm, parallel-group, repeated-measures study. NPJ Prim Care Respir Med. 2014;24:14071. doi:10.1038/npjpcrm.2014.71.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Crane MA, Jenkins CR, Goeman DP, Douglass JA. Inhaler device technique can be improved in older adults through tailored education: findings from a randomised controlled trial. NPJ Primary Care Respiratory Medicine. 2014;24:14034. doi:10.1038/npjpcrm.2014.34.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Dantic DE. A critical review of the effectiveness of ‘teach-back’technique in teaching COPD patients self-management using respiratory inhalers. Health Educ J. 2014;73(1):41–50. doi:10.1177/0017896912469575.

    Article  Google Scholar 

  20. Sleath B, Carpenter DM, Ayala GX, Williams D, Davis S, Tudor G, et al. Communication during pediatric asthma visits and child asthma medication device technique 1 month later. J Asthma. 2012;49(9):918–25. doi:10.3109/02770903.2012.719250.

    Article  PubMed  Google Scholar 

  21. Lavorini F, Levy ML, Corrigan C, Crompton G. The ADMIT series-issues in inhalation therapy. 6 training tools for inhalation devices. Prim Care Respir J. 2010;19(4):335–41. doi:10.4104/pcrj.2010.00065.

    Article  PubMed  Google Scholar 

  22. Sanchis J, Gich I, Pedersen S. Systematic review of errors in inhaler use: has patient technique improved over time? Chest. 2016;150(2):394–406. doi:10.1016/j.chest. 2016.03.041.

    Article  PubMed  Google Scholar 

  23. D'Arcy S, MacHale E, Seheult J, Holmes MS, Hughes C, Sulaiman I, et al. A method to assess adherence in inhaler use through analysis of acoustic recordings of inhaler events. PLoS One. 2014;9(6):e98701. doi:10.1371/journal.pone.0098701.

    Article  PubMed  PubMed Central  Google Scholar 

  24. O’Dwyer SM, MacHale E, Sulaiman I, Holmes M, Hughes C, D’Arcy S, et al. The effect of providing feedback on inhaler technique and adherence from an electronic audio recording device, INCA®, in a community pharmacy setting: study protocol for a randomised controlled trial. Trials. 2016;17(1):226. doi:10.1186/s13063-016-1362-9.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Inhaler Compliance Assessment. INCA. Available from: http://www.incadevice.com/.

  26. The Innovation Academy. Martin Holmes—improving inhaler user technique: Youtube; 2013. Available from: https://www.youtube.com/watch?v=azj-6QCdvsw.

  27. Ditcham W, Murdzoska J, Zhang G, Roller C, von Hollen D, Nikander K, et al. Lung deposition of 99mTc-radiolabeled albuterol delivered through a pressurized metered dose inhaler and spacer with facemask or mouthpiece in children with asthma. J Aerosol Med Pulm Drug Deliv. 2014;27(Suppl 1):S63–75. doi:10.1089/jamp.2014.1139.

    Article  PubMed  Google Scholar 

  28. Vitalograph. AIM: for effective inhaler training on dry powder and metered dose inhalers. 2017. Available from: https://vitalograph.com/product/162430/aim.

  29. Cambridge Consultants. T-Haler—new asthma inhaler training device: YouTube; 2012. Available from: https://www.youtube.com/watch?v=I_2ThiO09Sk.

  30. Cambridge Consultants. T-haler inhaler training device. 2017. Available from: https://www.cambridgeconsultants.com/projects/t-haler-inhaler-training-device.

  31. Julius SM, Sherman JM, Hendeles L. Accuracy of three electronic monitors for metered-dose inhalers. Chest. 2002;121(3):871–6. doi:10.1378/chest.121.3.871.

    Article  PubMed  Google Scholar 

  32. Farr SJ, Rowe AM, Rubsamen R, Taylor G. Aerosol deposition in the human lung following administration from a microprocessor controlled pressurised metered dose inhaler. Thorax. 1995;50(6):639–44. doi:10.1136/thx.50.6.639.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  33. Food and Drug Administration. MDILog, 510(k)Number:K9703441997. Available at http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn_template.cfm?id=k970344.

  34. Aslam T. Inhaler spacer. Google Patents. Central Manchester University Hospitals NHS Foundation Trust Patents. US20150059739A1; 2015.

  35. Fyne Dynamics. Inhaler trainer: Mag-Flo. 2016. Available from: http://www.fyne-dynamics.com/magflo_inhaler_trainer.htm.

  36. National Heart Lung and Blood Institute. Expert panel report 3: guidelines for the diagnosis and management of asthma. Bethesda, MD: National Institutes of Health; National Heart, Lung, and Blood Institute, 2007 Publication Number: 08–5846.

  37. Kamps AWA, van Ewijk B, Roorda RJ, Brand PLP. Poor inhalation technique, even after inhalation instructions, in children with asthma. Pediatr Pulmonol. 2000;29(1):39–42. doi:10.1002/(sici)1099-0496(200001)29:1<39::aid-ppul7>3.0.co;2-g.

    Article  CAS  PubMed  Google Scholar 

  38. Carpenter DM, Lee C, Blalock SJ, Weaver M, Reuland D, Coyne-Beasley T, et al. Using videos to teach children inhaler technique: a pilot randomized controlled trial. J Asthma. 2015;52(1):81–7. doi:10.3109/02770903.2014.944983.

    Article  PubMed  Google Scholar 

  39. Noar SM, Benac CN, Harris MS. Does tailoring matter? Meta-analytic review of tailored print health behavior change interventions. Psychol Bull. 2007;133(4):673–93. doi:10.1037/0033-2909.133.4.673.

    Article  PubMed  Google Scholar 

  40. Skinner CS, Campbell MK, Rimer BK, Curry S, Prochaska JO. How effective is tailored print communication? Ann Behav Med. 1999;21(4):290–8. doi:10.1007/BF02895960.

    Article  CAS  PubMed  Google Scholar 

  41. Skinner CS, Strecher VJ, Hospers H. Physicians' recommendations for mammography: do tailored messages make a difference? Am J Public Health. 1994;84(1):43–9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  42. Carpenter DM, Alexander DS, Elio A, DeWalt D, Lee C, Sleath B. Using tailored videos to teach inhaler technique to children with asthma: results from a school nurse-led pilot study. J Pediatr Nurs. 2016;31(4):380–9. doi:10.1016/j.pedn.2016.02.005.

    Article  PubMed  Google Scholar 

  43. Bandura A. Human agency in social cognitive theory. Am Psychol. 1989;44(9):1175–84.

    Article  CAS  PubMed  Google Scholar 

  44. Bandura A. Social cognitive theory: an agentic perspective. Annu Rev Psychol. 2001;52:1–26. doi:10.1146/annurev.psych.52.1.1.

    Article  CAS  PubMed  Google Scholar 

  45. DIS I. 9241-210: 2010. Ergonomics of human system interaction—part 210: human-centred design for interactive systems. International Standardization Organization (ISO) Switzerland. 2009.

Download references

Acknowledgements

This study was financially supported in part from the PharmAlliance, a three-way partnership between the schools of pharmacy at UNC, UCL, and Monash.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Delesha M. Carpenter.

Ethics declarations

Conflict of Interest

Mr. Sage Ms. Roberts, and Dr. Carpenter declare no conflicts of interest relevant to this manuscript.

Dr. George has received in-kind support from Vitalograph in the form of devices (COPD-6; ASMA-1; AIM) for pilot testing.

Dr. Horne was supported by the National Institute for Health Research (NIHR) Collaboration for Leadership in Applied Health Research and Care (CLAHRC) North Thames at Bart’s Health NHS Trust; has undertaken speaker engagements with honoraria with the following companies: Abbvie, Amgen, Biogen Idec, Gilead Sciences, GlaxoSmithKline, Janssen, Pfizer, Roche, Shire Pharmaceuticals, MSD, Astellas, Astrazeneca, DRSU, Erasmus, and Novartis; and is founder and shareholder of a UCL business spin out company (Spoonful of Sugar) providing consultancy on medication-related behaviors to healthcare policy makers, providers, and industry.

Human and Animal Rights and Informed Consent

This article does not contain any studies with human or animal subjects performed by any of the authors.

Additional information

This article is part of the Topical Collection on Allergies and the Environment

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Carpenter, D.M., Roberts, C.A., Sage, A.J. et al. A Review of Electronic Devices to Assess Inhaler Technique. Curr Allergy Asthma Rep 17, 17 (2017). https://doi.org/10.1007/s11882-017-0684-3

Download citation

  • Published:

  • DOI: https://doi.org/10.1007/s11882-017-0684-3

Keywords

Navigation