Introduction
Intragastric balloons (IGB) are an effective non-surgical weight loss option for patients with obesity (body mass index 30–35 kg/m2) who are not surgical candidates [1]. There are four FDA-approved IGBs; however, only two—Spatz3 (Spatz FGIA, Great Neck, NY) and Orbera (Apollo Endosurgery, Austin, TX, USA)—are currently available within the USA. This study characterizes device issues and patient complications involving these IGBs, using post-marketing surveillance data from the FDA Manufacturer and User Facility Device Experience (MAUDE) database between December 2020 and October 2023. Our analysis provides an update to the 2021 Ramai et al. publication [2] on IGB adverse events reported to MAUDE and presents new data on Spatz3, not FDA approved until October 2021.
Methods
The MAUDE database (https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/search.cfm) tracks adverse event reports involving FDA-approved medical devices. Reports contain information on the device, manufacture, and event details, including report date, device problems, patient complications, and free-texted narratives.
MAUDE was queried for entries on Orbera and Spatz3 IGBs submitted between December 2020 and September 2023. Search terms included Spatz3, Spatz FGIA, and Orbera, Apollo Endosurgery.
Results
During the study period, 728 cases (Orbera = 354, Spatz3 = 374) with 1099 device issues and 1021 patient complications were reported to MAUDE.
Orbera IGBs, FDA approved in 2005, had 651 device issues between December 2020 to September 2023 with reports increasing each year since 2021 (Fig. 1a–c). Spatz3 IGBs had 448 device issues reported between November 2021 to September 2023. A similar number of device issues were reported during the first year after FDA approval in October 2021 (N = 231) and the subsequent 11 months (N = 217).
Balloon deflation accounted for 49% and 63% of all Orbera and Spatz3 device issues, respectively. Leak (N = 140, 22%) and balloon inflation problems (N = 64, 10%) for Orbera IGBs and patient-device incompatibility (N = 45, 10%) and device migration or expulsion (N = 41, 9%) for Spatz3 IGBs, were the next most reported device issues (Table 1a). Report narratives described balloon hyperinflation (N = 29, 64%), gastric ulcer formation (N = 6, 13%), and symptomatic intolerance (N = 5, 11%) as leading reasons for patient-device incompatibility. Among this group, 10 patients required hospitalization and 22 required device removal or replacement. Hyperinflation accounted for 50% of 24 patient-device incompatibility issues, with 6 noting device removal or replacement was required.
A total of 599 patient complications involving Orbera and 422 involving Spatz3 IGBs were reported (Table 1b). Failure of implant (N = 256, 43%), vomiting (N = 83, 14%), pain (N = 77, 13%), nausea (N = 32, 5%), obstruction (N = 21, 4%), and perforation (N = 20, 3%) were leading Orbera-related patient complications. Report narratives detail IGB leak or deflation (N = 142, 55%), hyperinflation (N = 48, 19%), and migration or expulsion (N = 30, 12%) as primary causes of Orbera implant failure (Table 1c). For Spatz3, appropriate clinical signs, symptoms, conditions term or code not available (N = 265, 63%) was the most frequently indicated patient complication, followed by pain (N = 86, 20%), vomiting (N = 22, 5%), and nausea (N = 17, 4%). IGB leak or deflation (N = 215, 81%) is the leading device issue underlying “Appropriate Clinical Signs/Code Not Available” according to narrative reports. Ulcer formation and pancreatitis each accounted for ≤ 1% of all patient complications for either IGB. Six deaths were reported during the study period (Orbera = 5, Spatz3 = 1). Five were related to perforation (Orbera = 4, Spatz3 = 1), with few additional details provided, and one did not disclose a cause.
Conclusion
This study provides an accounting of FDA approved, market-available IGB device issues and patient-related complications. Randomized controlled trials of Orbera [3] and Spatz3 [4] found high rates of nausea (Orbera = 86.9%, Spatz3 = 90.4%), vomiting (Orbera = 75.6%, Spatz3 = 71.7%), and abdominal pain (Orbera = 75.6%, Spatz3 = 60.4%) after IGB placement. Collectively, these were also the most frequently reported symptoms to MAUDE for either IGB; however, each accounted for a smaller than expected proportion of overall reports when considering prior study findings. These differences may highlight reporting bias resulting from event underreporting due to voluntary reporting requirements.
While clinical trials involving Orbera and Spatz3 found low rates of spontaneous deflation between 0 and 1.66% [3, 5], over half of all reported device issues in this study was related to balloon deflation or leaking. A majority of deflations/leaks were detected by the patient due to change in urine color or during device retrieval; few were identified because of suboptimal weight loss. Extending IGB use beyond recommendations or non-adherence with proton pump inhibitors can increase the risk of deflation [6] and may contribute to deflation reports. Balloon hyperinflation, another rarely cited IGB complication [7], was frequently described within narrative reports for patient-device incompatibility issues for both IGBs (Orbera = 64%, Spatz3 = 50%). While most reports lack manufacturer verification or root cause analysis, these findings suggest that spontaneous deflation and hyperinflation may be more common than previously described.
Study limitations relate to the quality and completeness of MAUDE reports. Nevertheless, MAUDE data provides insights on complications encountered in clinical practice, playing an important role in monitoring devices, such as the recently approved Spatz3. Findings from this study may inform patient-provider discussions of IGB complications and have identified spontaneous deflation and hyperinflation as potential areas for further study.
Data Availability
The data that support the findings of this study are openly available at MAUDE - Manufacturer and User Facility Device Experience: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/search.cfm.
References
Sullivan S, Edmundowicz SA, Thompson CC. Endoscopic bariatric and metabolic therapies: new and emerging technologies. Gastroenterology. 2017;152(7):1791–801.
Ramai D, Bhandari P, Facciorusso A, et al. Real-world experience of intragastric balloons for obesity: insights from the FDA manufacturer and user facility device experience (MAUDE) Database. Obes Surg. 2021;31(7):3360–4.
Courcoulas A, Abu Dayyeh BK, Eaton L, et al. Intragastric balloon as an adjunct to lifestyle intervention: a randomized controlled trial. Int J Obes. 2017;41(3):427–33.
Abu Dayyeh BK, Maselli DB, Rapaka B, et al. Adjustable intragastric balloon for treatment of obesity: a multicentre, open-label, randomised clinical trial. Lancet. 2021;398(10315):1965–73.
Neto MG, Silva LB, Grecco E, et al. Brazilian Intragastric Balloon Consensus Statement (BIBC): practical guidelines based on experience of over 40,000 cases. Surg Obes Relat Dis. 2018;14(2):151–9.
Laing P, Pham T, Taylor LJ, et al. Filling the void: a review of intragastric balloons for obesity. Dig Dis Sci. 2017;62(6):1399–408.
Maisel W. The FDA alerts health care providers about potential risks with liquid-filled intragastric balloons [Internet]. The United States Food and Drug Administration. 2017 [cited 2023 Nov 3]. Available from: https://www.fda.gov/medical-devices/letters-health-care-providers/fda-alerts-health-care-providers-about-potential-risks-liquid-filled-intragastric-balloons.
Funding
Open access funding provided by SCELC, Statewide California Electronic Library Consortium
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Ethical Approval
This article does not contain any studies with human participants or animals performed by any of the authors.
Informed Consent
Informed consent does not apply.
Conflict of Interest
The authors—Morgan C Goodman MD, Patrick Chang MD, William Minteer MD, Denis Nguyen MD, Kalpana Gopalkrishnan, Jennifer Phan MD—declare that they have no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Key points
• Nausea, vomiting, and abdominal pain are common IGB complications in clinical practice and in MAUDE reports.
• Spontaneous deflation was by far the most reported device issue.
• Severe or life-threatening complications were rarely reported to MAUDE.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Goodman, M.C., Chang, P., Minteer, W. et al. Patient Complications and Device Issues Associated With FDA-Approved Intragastric Balloons Available in the USA: A Maude Database Study. OBES SURG 34, 1971–1974 (2024). https://doi.org/10.1007/s11695-024-07128-1
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11695-024-07128-1