Abstract
Objectives
To ascertain the effects of exposure parameters (tube current and tube voltage) and the gutta-percha cone (GPC) size on root fracture-like artifacts obtained with cone-beam computed tomography (CBCT).
Methods
Fracture-like artifacts appearing on CBCT images of nine extracted human mandibular premolars filled with GPCs of size #50 or #80 were analyzed using six exposure factors: two tube voltages (80 kV and 110 kV); and three tube currents (4 mA, 7 mA, and 10 mA). On axial images, the gray value (GV) was recorded at three points: the mesiobuccal portion (MBP) as the sound dentin, the mesial portion (MP) as the artifact line, and the water area (WA). The rate of decrease in the GV (RDGV) of the artifact line was calculated using the formula: RDGV (%) = (GV of MBP − GV of MP) × 100/(GV of MBP − GV of WA).
Results
Comparison of the #80 group and the #50 group with equal tube voltages and tube currents shows that artifact lines in the #80 group were more obvious than those in the #50 group. The artifact lines with 80 kV were markedly more visible than those with 110 kV for each tube current and GPC size. Tube current changes did not affect the artifact line for any tube voltage or GPC size.
Conclusions
For the reduction of artifacts, we recommend selection of higher tube voltages and lower tube currents when taking CBCT images of teeth with each GPC size.
Similar content being viewed by others
References
Zhang Y, Zhang I, Zhu XR, Lee AK, Chambers M, Dong L. Reduction metal artefacts in cone-beam CT images by preprocessing projection data. Int J Radiat Oncol Biol Phys. 2007;67:924–32.
Hassan B, Metska ME, Ozok AR, van der Stelt P, Wesselink PR. Detection of vertical root fractures in endodontically treated teeth by a cone beam computed tomography scan. J Endod. 2009;35:719–22.
Brito-Júnior M, Santos LA, Faria-e-Silva AL, Pereira RD, Sousa-Neto MD. Ex vivo evaluation of artifacts mimicking fracture lines on cone-beam computed tomography produced by different root canal seals. Int Endod J. 2014;47:26–31.
Iikubo M, Nishioka T, Okura S, Kobayashi K, Sano T, Akitoshi K, et al. Influence of voxel size and scan field of view on fracture-like artifacts from gutta-percha obturated endodontically treated teeth on cone-beam computed tomography images. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;122:631–7.
Pauwels R, Araki K, Siewerdsen JH, Thongvigitmanee SS. Technical aspects of dental CBCT: state of the art. Dentomaxillofac Radiol. 2015;44:20140224.
Ludlow JB, Timothy R, Walker C, Hunter R, Benavides E, Samuelson DB, et al. Correction to effective dose of dental CBCT – a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol. 2015;44:20159003.
Iikubo M, Osano T, Sano T, Katsumata A, Ariji E, Kobayashi K, et al. Root canal filling materials spread pattern mimicking root fractures in dental CBCT images. Oral Surg Oral Med Oral Pathol Oral Radiol. 2015;120:521–7.
Schulze R, Heil U, Gross D, Bruellman DD, Dranischnikow E, Schwanecke U, et al. Artifacts in CBCT: a review. Dentomaxillofac Radiol. 2011;40:265–73.
Chen CY, Chuang KS, Wu J, Lin HR, Li MJ. Beam hardening correction for computed tomography images using a postreconstruction method and equivalent tissue concept. J Digit Imaging. 2001;14:54–61.
Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;106:930–8.
Iikubo M, Kobayashi K, Mishima A, Shimoda S, Daimaruya T, Igarashi C, et al. Accuracy of intraoral radiography, multidetector helical CT, and limited cone-beam CT for the detection of horizontal tooth root fracture. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;108:e70–4.
Khedmat S, Rouhi N, Drage N, Shokouhinejad N, Nekoofar MH. Evaluation of three imaging techniques for the detection of vertical root fractures in the absence and presence of gutta-percha root fillings. Int Endod J. 2012;45:1004–9.
Panjnoush M, Kheirandish Y, Kashani PM, Fakhar HB, Younesi F, Mallahi M. Effect of exposure parameters on metal artifacts in cone beam computed tomography. J Dent (Tehran). 2016;13:143–50.
Acknowledgements
This work was supported in part by a Grant-in-Aid for Project Research from the Japanese Association for Dental Science (no. 16K15810) and Research Center for Biomedical Engineering (2035).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
All authors have nothing to disclose.
Human rights statements and informed consent
All procedures followed were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the Helsinki Declaration of 1964 and later versions. Informed consent was obtained from all patients for being included in the study.
Animal rights statements
This study did not use any animal.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Iikubo, M., Kagawa, T., Fujisawa, J. et al. Effect of exposure parameters and gutta-percha cone size on fracture-like artifacts in endodontically treated teeth on cone-beam computed tomography images. Oral Radiol 36, 344–348 (2020). https://doi.org/10.1007/s11282-019-00411-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11282-019-00411-3