Abstract
Felt strips are widely used for reinforcement of the aortic stump in surgery for aortic dissection (AD). Postoperative hemolytic anemia (HA) due to an inverted internal felt strip at the aortic stump fixation for AD is extremely rare. A 70-year-old woman underwent ascending aorta replacement for acute type A AD, where both proximal and distal anastomotic sites were reinforced with Teflon felt strips. A week later, macroscopic hematuria and HA emerged. Three-dimensional transesophageal echocardiography (3D-TEE) demonstrated that the proximal inner felt strip turned up and protruded into the aortic inner lumen. At redo surgery, which was performed 2 weeks after the initial surgery, the findings of 3D-TEE were confirmed, and the inverted internal felt strip was replaced with a bovine pericardial strip. The findings of HA disappeared immediately after the second surgery. 3D-TEE is a very informative, valuable modality for accurate diagnosis that leads to a safe surgery.
Similar content being viewed by others
References
Sekine Y, Yamamoto S, Fujikawa T, et al. Haemolytic anaemia resulting from the surgical repair of acute type A aortic dissection. Interact Cardiovasc Surg. 2014;18:230–3.
Nakamura Y, Ogino H, Matsuda H, et al. Hemolytic anemia after operation for aortic dissection using Teflon felt strips. Ann Thorac Surg. 2008;85:1784–7.
Shiga T, Wajima Z, Apfel CC, et al. Diagnostic accuracy of transesophageal echocardiography, helical computed tomography, and magnetic resonance imaging for suspected thoracic aortic dissection. Arch Intern Med. 2006;166:1350–6.
Scohy TV, Geniets B, McGhie J, et al. Feasibility of real-time three-dimensional transesophageal echocardiography in type A aortic dissection. Interact Cardiovasc Thorac Surg. 2010;11:112–3.
Hata M, Yoshitake I, Wakui S, et al. Postoperative early hemolytic anemia due to inverted teflon felt strip after emergency repair for type A dissection. Thorac Cardiovasc Surg. 2012;60:482–4.
Kang H, Choe JW, Cho DY, et al. Hemolytic anemia case caused by an inverted inner felt after Bentall operation. J Korean Med Sci. 2013;28:1827–9.
Floten HS, Ravichandran PS, Furnary AP, et al. Adventitial inversion technique in repair of aortic dissection. Ann Thorac Surg. 1995;59:771–2.
Acknowledgements
The authors are grateful to Yoko Motomura for her assistance in manuscript preparation.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
None declared.
Ethical statements
Additional informed consent was obtained from the patient for whom identifying information is included in this article.