Temporal averaging of two-dimensional correlation functions for velocity vector imaging of cardiac blood flow

  • Original Article
  • Published:
Journal of Medical Ultrasonics Aims and scope Submit manuscript

Abstract

Purpose

Ultrasonic imaging of blood flow in the cardiac lumen is a very useful tool to evaluate the pumping function of the human heart. The speckle tracking technique makes it possible to estimate the blood velocity vector. However, a stable estimation of the velocity vector of blood flow is difficult because signal-to-noise ratios of echoes from tiny blood particles are low. In this study, the speckle tracking technique with averaging of multiple two-dimensional correlation functions was employed for stable estimation of the blood velocity vector.

Methods

Multiple two-dimensional correlation functions can be averaged during a very short period by using the echo data acquired by high-frame-rate echocardiography with diverging beam transmission. A steady flow experiment using blood-mimicking fluid (mean fluid velocity 0.2 m/s, flow angle 56° from the transducer surface) was implemented to investigate the effect of the averaging of two-dimensional correlation functions at a frame rate of 6024 Hz.

Results

First, to examine the averaging duration required for stable estimation of the flow velocity vector, the accuracies of vector estimates were evaluated at different durations for averaging of two-dimensional correlation functions in the steady flow measurement. It was found that the proposed averaging process with an averaging duration of over 8 frames could reduce the directional error in vector estimation to almost half that of the conventional speckle tracking technique. In subsequent experiments, the averaging duration was set at 12 frames corresponding to 2 ms. Measurements of steady flow at higher velocities were further implemented. The steady flow measurements with higher flow velocities of 0.4 and 0.6 m/s were simulated by changing the frame interval of the echo data at a flow velocity of 0.2 m/s. Although the averaging duration was a mere 2 ms, directional errors at mean flow velocities of 0.2, 0.4, and 0.6 m/s were reduced significantly. In an in vivo experiment of the healthy human heart, to produce a fine B-mode image, the diverging wave transmissions with different steered angles for compounding were interleaved in the transmission sequence. From the in vivo experimental result, the blood velocity vector of the left ventricular cavity showed the flow getting into/out of the cavity in ejection and early diastolic phases. Furthermore, estimated flow directions revealed rotating flow in the cavity in mid-diastole.

Conclusion

Our proposed method has the feasibility to visualize the vortex flow by velocity vector mapping without a contrast agent.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. de Korte CL, Nillesen MM, Saris AECM, et al. New developments in paediatric cardiac functional ultrasound imaging. J Med Ultrason. 2014;41:279–90.

    Google Scholar 

  2. Hasegawa H, Kanai H. Simultaneous imaging of artery-wall strain and blood flow by high frame rate acquisition of RF signals. IEEE Trans Ultrason Ferroelectr Freq Control. 2008;55:2626–39.

    Article  PubMed  Google Scholar 

  3. Bercoff J, Montaldo G, Loupas T, et al. Ultrafast compounding Doppler imaging: providing full blood flow characterization. IEEE Trans Ultrason Ferroelectr Freq Control. 2011;58:134–47.

    Article  PubMed  Google Scholar 

  4. Hasegawa H, Hongo K, Kanai H. Measurement of regional pulse wave velocity using very high frame rate ultrasound. J Med Ultrason. 2013;40:91–8.

    Article  Google Scholar 

  5. Kruizinga P, Mastik F, van den Oord SCH, et al. High-definition imaging of carotid artery wall dynamics. Ultrasound Med Biol. 2014;40:2392–403.

    Article  PubMed  Google Scholar 

  6. Ohtsuki S, Tanaka M. The flow velocity distribution from the Doppler information on a plane in three-dimensional flow. J Vis. 2006;9:69–82.

    Article  Google Scholar 

  7. Garcia D, Alamo JCD, Tanne D, et al. Two-dimensional intraventricular flow mapping by digital processing conventional color-Doppler echocardiography images. IEEE Trans Med Imag. 2010;29:1701–12.

    Article  Google Scholar 

  8. Itatani K, Uejima T, Tanaka T, et al. Intraventricular flow velocity vector visualization based on the continuity equation and measurements of vorticity and wall shear stress. Jpn J Appl Phys. 2013;52:07HF16-1-6.

    Article  Google Scholar 

  9. Capineri L, Scabia M, Masotti LA. Doppler system for dynamic vector velocity maps. Ultrasound Med Biol. 2002;28:237–48.

    Article  PubMed  Google Scholar 

  10. Kripfgans OD, Rubin JM, Hall AL, et al. Vector Doppler imaging of a spinning disc ultrasound Doppler phantom. Ultrasound Med Biol. 2006;32:1037–46.

    Article  PubMed  Google Scholar 

  11. Ekroll IK, Swillens A, Segers P, et al. Simultaneous quantification of flow and tissue velocities based on multi-angle plane wave imaging. IEEE Trans Ultrason Ferroelectr Freq Control. 2013;60:727–38.

    Article  PubMed  Google Scholar 

  12. Hong GR, Pedrizzetti G, Giovanni Tonti, et al. Characterization and quantification of vortex flow in the human left ventricle by contrast echocardiography using vector particle image velocimetry. J Am Coll Cardiol Imaging. 2008;1:705–17.

    Article  Google Scholar 

  13. Faludi R, Szulik M, D’hooge J, et al. Left ventricular flow patterns in healthy subjects and patients with prosthetic mitral valves: an in vivo study using echocardiographic particle image velocimetry. J Thorac Cardiovasc Surg. 2010;139:1501–10.

    Article  PubMed  Google Scholar 

  14. Løvstakken L, Bjærum S, Martens D, et al. Blood flow imaging-a new real-time, 2-D flow imaging technique. IEEE Trans Ultrason Ferroelectr Freq Control. 2006;53:289–99.

    Article  PubMed  Google Scholar 

  15. Hasegawa H, Kanai H. Blood flow stream line imaging by direct visualization of echo trajectories. Proc IEEE Ultrasonics Symp. 2010;1319–22.

  16. Takahashi H, Hasegawa H, Kanai H. Speckle-enhanced cardiac blood flow imaging with high frame rate ultrasound. Proc 2013 Joint UFFC, EFTF and PFM symposium. 2013;2030–33.

  17. Takahashi H, Hasegawa H, Kanai H. Echo speckle imaging of blood particles with high frame rate echocardiography. Jpn J Appl Phys. 2014;53:07KF08-1-7.

    Article  Google Scholar 

  18. Bohs LN, Geiman BJ, Anderson ME, et al. Speckle tracking for multi-dimensional flow estimation. Ultrasonics. 2000;38:369–75.

    Article  CAS  PubMed  Google Scholar 

  19. Hasegawa H, Kanai H. High-frame-rate echocardiography using diverging transmit beams and parallel receive beamforming. J Med Ultrason. 2011;38:129–40.

    Article  CAS  Google Scholar 

  20. Hasegawa H, Kanai H. High-frame-rate echocardiography with reduced sidelobe level. IEEE Trans Ultrason Ferroelectr Freq Control. 2012;59:2569–75.

    PubMed  Google Scholar 

  21. Cespedes I, Huang Y, Ophir J, et al. Methods for estimation of subsample time delays of digitized echo signals. Ultrason Imaging. 1995;17:142–71.

    Article  CAS  PubMed  Google Scholar 

  22. Ebbers T. Flow imaging: cardiac applications of 3D cine phase-contrast MRI. Curr Cardiovasc Imaging Rep. 2011;4:127–33.

    Article  Google Scholar 

  23. Prinz C, Faludi R, Walker A, et al. Can echocardiographic particle image velocimetry correctly detect motion patterns as they occur in blood inside heart chambers? A validation study using moving phantoms. Cardiovasc Ultrasound. 2012;10:24–33.

    Article  PubMed Central  PubMed  Google Scholar 

Download references

Acknowledgments

This work has been supported by the JSPS Research Fellowship (25-5479) and the International Advanced Research and Education Organization of Tohoku University.

Conflict of interest

There are no financial or other relations that could lead to a conflict of interest.

Ethical standard

The in vivo experiment in this study was approved by the institutional ethical committee. The subject gave informed consent to this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hiroki Takahashi.

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Takahashi, H., Hasegawa, H. & Kanai, H. Temporal averaging of two-dimensional correlation functions for velocity vector imaging of cardiac blood flow. J Med Ultrasonics 42, 323–330 (2015). https://doi.org/10.1007/s10396-015-0620-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10396-015-0620-x

Keywords

Navigation