Abstract
Purpose
Ultrasonic imaging of blood flow in the cardiac lumen is a very useful tool to evaluate the pumping function of the human heart. The speckle tracking technique makes it possible to estimate the blood velocity vector. However, a stable estimation of the velocity vector of blood flow is difficult because signal-to-noise ratios of echoes from tiny blood particles are low. In this study, the speckle tracking technique with averaging of multiple two-dimensional correlation functions was employed for stable estimation of the blood velocity vector.
Methods
Multiple two-dimensional correlation functions can be averaged during a very short period by using the echo data acquired by high-frame-rate echocardiography with diverging beam transmission. A steady flow experiment using blood-mimicking fluid (mean fluid velocity 0.2 m/s, flow angle 56° from the transducer surface) was implemented to investigate the effect of the averaging of two-dimensional correlation functions at a frame rate of 6024 Hz.
Results
First, to examine the averaging duration required for stable estimation of the flow velocity vector, the accuracies of vector estimates were evaluated at different durations for averaging of two-dimensional correlation functions in the steady flow measurement. It was found that the proposed averaging process with an averaging duration of over 8 frames could reduce the directional error in vector estimation to almost half that of the conventional speckle tracking technique. In subsequent experiments, the averaging duration was set at 12 frames corresponding to 2 ms. Measurements of steady flow at higher velocities were further implemented. The steady flow measurements with higher flow velocities of 0.4 and 0.6 m/s were simulated by changing the frame interval of the echo data at a flow velocity of 0.2 m/s. Although the averaging duration was a mere 2 ms, directional errors at mean flow velocities of 0.2, 0.4, and 0.6 m/s were reduced significantly. In an in vivo experiment of the healthy human heart, to produce a fine B-mode image, the diverging wave transmissions with different steered angles for compounding were interleaved in the transmission sequence. From the in vivo experimental result, the blood velocity vector of the left ventricular cavity showed the flow getting into/out of the cavity in ejection and early diastolic phases. Furthermore, estimated flow directions revealed rotating flow in the cavity in mid-diastole.
Conclusion
Our proposed method has the feasibility to visualize the vortex flow by velocity vector mapping without a contrast agent.
Similar content being viewed by others
References
de Korte CL, Nillesen MM, Saris AECM, et al. New developments in paediatric cardiac functional ultrasound imaging. J Med Ultrason. 2014;41:279–90.
Hasegawa H, Kanai H. Simultaneous imaging of artery-wall strain and blood flow by high frame rate acquisition of RF signals. IEEE Trans Ultrason Ferroelectr Freq Control. 2008;55:2626–39.
Bercoff J, Montaldo G, Loupas T, et al. Ultrafast compounding Doppler imaging: providing full blood flow characterization. IEEE Trans Ultrason Ferroelectr Freq Control. 2011;58:134–47.
Hasegawa H, Hongo K, Kanai H. Measurement of regional pulse wave velocity using very high frame rate ultrasound. J Med Ultrason. 2013;40:91–8.
Kruizinga P, Mastik F, van den Oord SCH, et al. High-definition imaging of carotid artery wall dynamics. Ultrasound Med Biol. 2014;40:2392–403.
Ohtsuki S, Tanaka M. The flow velocity distribution from the Doppler information on a plane in three-dimensional flow. J Vis. 2006;9:69–82.
Garcia D, Alamo JCD, Tanne D, et al. Two-dimensional intraventricular flow mapping by digital processing conventional color-Doppler echocardiography images. IEEE Trans Med Imag. 2010;29:1701–12.
Itatani K, Uejima T, Tanaka T, et al. Intraventricular flow velocity vector visualization based on the continuity equation and measurements of vorticity and wall shear stress. Jpn J Appl Phys. 2013;52:07HF16-1-6.
Capineri L, Scabia M, Masotti LA. Doppler system for dynamic vector velocity maps. Ultrasound Med Biol. 2002;28:237–48.
Kripfgans OD, Rubin JM, Hall AL, et al. Vector Doppler imaging of a spinning disc ultrasound Doppler phantom. Ultrasound Med Biol. 2006;32:1037–46.
Ekroll IK, Swillens A, Segers P, et al. Simultaneous quantification of flow and tissue velocities based on multi-angle plane wave imaging. IEEE Trans Ultrason Ferroelectr Freq Control. 2013;60:727–38.
Hong GR, Pedrizzetti G, Giovanni Tonti, et al. Characterization and quantification of vortex flow in the human left ventricle by contrast echocardiography using vector particle image velocimetry. J Am Coll Cardiol Imaging. 2008;1:705–17.
Faludi R, Szulik M, D’hooge J, et al. Left ventricular flow patterns in healthy subjects and patients with prosthetic mitral valves: an in vivo study using echocardiographic particle image velocimetry. J Thorac Cardiovasc Surg. 2010;139:1501–10.
Løvstakken L, Bjærum S, Martens D, et al. Blood flow imaging-a new real-time, 2-D flow imaging technique. IEEE Trans Ultrason Ferroelectr Freq Control. 2006;53:289–99.
Hasegawa H, Kanai H. Blood flow stream line imaging by direct visualization of echo trajectories. Proc IEEE Ultrasonics Symp. 2010;1319–22.
Takahashi H, Hasegawa H, Kanai H. Speckle-enhanced cardiac blood flow imaging with high frame rate ultrasound. Proc 2013 Joint UFFC, EFTF and PFM symposium. 2013;2030–33.
Takahashi H, Hasegawa H, Kanai H. Echo speckle imaging of blood particles with high frame rate echocardiography. Jpn J Appl Phys. 2014;53:07KF08-1-7.
Bohs LN, Geiman BJ, Anderson ME, et al. Speckle tracking for multi-dimensional flow estimation. Ultrasonics. 2000;38:369–75.
Hasegawa H, Kanai H. High-frame-rate echocardiography using diverging transmit beams and parallel receive beamforming. J Med Ultrason. 2011;38:129–40.
Hasegawa H, Kanai H. High-frame-rate echocardiography with reduced sidelobe level. IEEE Trans Ultrason Ferroelectr Freq Control. 2012;59:2569–75.
Cespedes I, Huang Y, Ophir J, et al. Methods for estimation of subsample time delays of digitized echo signals. Ultrason Imaging. 1995;17:142–71.
Ebbers T. Flow imaging: cardiac applications of 3D cine phase-contrast MRI. Curr Cardiovasc Imaging Rep. 2011;4:127–33.
Prinz C, Faludi R, Walker A, et al. Can echocardiographic particle image velocimetry correctly detect motion patterns as they occur in blood inside heart chambers? A validation study using moving phantoms. Cardiovasc Ultrasound. 2012;10:24–33.
Acknowledgments
This work has been supported by the JSPS Research Fellowship (25-5479) and the International Advanced Research and Education Organization of Tohoku University.
Conflict of interest
There are no financial or other relations that could lead to a conflict of interest.
Ethical standard
The in vivo experiment in this study was approved by the institutional ethical committee. The subject gave informed consent to this study.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Takahashi, H., Hasegawa, H. & Kanai, H. Temporal averaging of two-dimensional correlation functions for velocity vector imaging of cardiac blood flow. J Med Ultrasonics 42, 323–330 (2015). https://doi.org/10.1007/s10396-015-0620-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10396-015-0620-x