Abstract
In this study, we developed a new catheter-mounted micro-axial flow blood pump (MFBP) using a new miniature magnetic fluid shaft seal (MFSS). The prototype of the catheter-mounted MFBP had a maximum diameter of 8 mm and a length of 50 mm. The new MFSS composed a neodymium magnet ring, an iron ring, and a magnetic fluid particularly designed for the MFSS. The new MFSS had outer and inner diameters of 4.0 mm and 2.6 mm, respectively, and a length of 3.0 mm. The sealing pressure of the MFSS was calculated to be 432 mmHg using FEM (Finite Element Method) result; therefore, the MFSS had sufficient sealing pressure for the catheter-mounted MFBP. The friction loss of the MFSS included the friction owing to the viscosity of the magnetic fluid and the magnetic force between the iron ring and ring magnet. The total friction loss of the MFSS was 0.08–0.09 W in the pump operational speed range from 22,000 to 35,000 rpm. From the in vitro experimental results, the catheter-mounted MFBP using the MFSS had a pump output of 3 L/min. against a differential pressure of 60 mmHg, and the pump characteristics of the MFBP were almost the same as those of Impella 5.0.
Similar content being viewed by others
References
Lauten A, Engstrom AE, Jung C, Empen K, Erne P, Cook S, Windecker S, Bergmann MW, Klingenberg R, Luscher TF, Haude M, Rulands D, Butter C, Ullman B, Hellgren L, Modena MG, Pedrazzini G, Henriques JPS, Figulla HR, Ferrari M. Percutaneous left-ventricular support with the Impella-2.5 assist device in acute cardiogenic shock—results of the Impella Euroshock Registry. Circ Heart Fail. 2013;6:23–30.
Ramzy D, Soltesz E, Anderson M. New surgical circulatory support system outcomes. ASAIO J. 2020;66:746–52.
Abiomed Inc. Homepage http://www.abiomed.com/ Accessed 7 Jan 2021
Attached document of Impella 5.0 by Pharmaceuticals and Medical Devices Agency https://www.info.pmda.go.jp/ygo/pack/580889/22800BZI00032000_B_01_08/22800BZI00032000_B_01_08?view=body Accessed 30 Apr 2021 (in Japanese)
Lemaire A, Anderson MB, Lee LY, Scholz P, Prendergast T, Goodman A, Lozano AM, Spotnitz A, Batsides G. The impella device for acute mechanical circulatory support in patients in cardiogenic shock. Ann Thorac Surg. 2014;97:133–8.
Khalid N, Rogers T, Shlofmitz E, Chen Y, Musallam A, Khan JM, Iantorno M, Gajanana D, Hahim H, Torguson R, Bernardo N, Waksman R. Adverse events and modes of failure related to Impella RP: Insights form the manufacture and user facility device experience (MAUDE) database. Cardio REevascular Med. 2019;20:503–6.
Chung JS, Emerson D, Ramzy D, Akhmerov A, Megna D, Esmailian F, Kobashiwaga J, Cole RM, Moriguch J, Trento A. A new paradigm in mechanical circulatory support:100-patient experience. Ann Thorac Srug. 2020;109:1370–7.
Mitamura Y, Sekine K, Asakawa M, Yozu R, kawada S, Okamoto E. A durable, non power consumptive, simple seal for rotary blood pumps. ASAIO J. 2001;47:392–6.
Sekine K, Mitamura Y, Murabayashi S, Nishimura I, Yozu R, Kim DW. Development of a magnetic fluid seal for an axial flow blood pump. Artifi Organs. 2003;27:892–6.
Mitamura Y, Takahashi S, Kano K, Okamoto E, Murabayashi S, Nishimura I, Higuchi T. Sealing performance of a magnetic fluid seal for rotary blood pumps. Artifi Organs. 2009;33:770–3.
Mitamura Y, Takahashi S, Amari S, Okamoto E, Murabayashi S, Nishimura I. A magnetic fluid seal for rotary blood pumps: effects of seal structure on long-term performance in liquid. J Artifi Organs. 2011;14:23–30.
Mitamura Y, Durst CA. Miniature magnetic fluid seal working in liquid environments. J Magn Magn Mater. 2017;431:285–8.
Mitamura Y, Sekine K, Okamoto E. Magnetic fluid seals working in liquid environments: Factor limiting their life and solution methods. J Magn Magn Mater. 2020;500:1–5.
Ramzy D, Soltesz E, Anderson M. New surgical circulatory support system outcome. ASAIO J. 2020;66:746–52.
Siess T, Nix C, Menzler F. From a lab type to a product: a retrospective view on Impella’s assist technology. Artif Organs. 2001;25:414–21.
Berkovsky BM, Medvedev VF, Krakov MS. Magnetic fluids: engineering applications. Oxford: Oxford University Press; 1993. p. 137.
Jianhui H, Meng Z, Lu W, Jibin Z, Li Y. Loss of torque on magnetic fluid seals with rotating–shafts. J Magn. 2017;22:286–90.
Sekine K, Mitamura Y. Evaluation of the cytotoxicity of oil-based magnetic fluid based on cell proliferation study. J Chin Soc Mech Eng. 2020;41:647–52.
Acknowledgements
This study was supported by KAKENHI (Grant-in-Aid for Scientific Research-C (No. 19K12792)), and supported by research program of Institute of Development, Aging and Cancer, Tohoku University (No. 2021_23).
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Okamoto, E., Yano, T., Sekine, K. et al. Development and initial performance of a miniature axial flow blood pump using magnetic fluid shaft seal. J Artif Organs 26, 12–16 (2023). https://doi.org/10.1007/s10047-022-01330-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10047-022-01330-7