Abstract
Purpose
We assessed the accuracy and safety of spine stereotactic body radiation therapy (SBRT) using helical tomotherapy (HT) via evaluating intrafractional patient movement.
Methods
From July 2009 to April 2011, 22 patients with spine lesions received SBRT using HT, with a total of 61 fractions. To evaluate intrafractional movement, we compared post-treatment megavoltage CT scans with planning CT images and obtained translational [lateral (X), craniocaudal (Y), anterioposterior (Z)] offsets and total displacements (R). We analyzed the correlation of intrafractional motion with patient and treatment characteristics. We also analyzed dosimetric change to the target and spinal cord, resulting from intrafractional movement, in the three patients that showed the greatest R values.
Results
Intrafractional movements were 0.7 ± 0.6 mm (X), 1.1 ± 0.7 mm (Y), 0.9 ± 0.6 mm (Z), and 1.8 ± 0.6 mm (R). This movement did not correlate with age, pain score, treatment time, or treatment site. Only patients with lower BMIs have a tendency to move more during treatment. Patient immobilization using wrapping form (thermoplastic mask and BodyFIX® system) resulted in less lateral movement and total displacement than others (0.498 ± 0.409 vs. 1.138 ± 0.637 mm, P < 0.001 for X; and 1.638 ± 0.691 vs. 1.976 ± 0.495 mm, P = 0.032 for R). However, this intrafractional motion did not affect the dose delivery to the target and spinal cord.
Conclusion
SBRT using HT can be a safe treatment modality for spine metastasis with enhanced targeting accuracy.
Similar content being viewed by others
References
Agazaryan N, Tenn SE, Desalles AA, Selch MT (2008) Image-guided radiosurgery for spinal tumors: methods, accuracy and patient intrafraction motion. Phys Med Biol 53:1715–1727. doi:10.1088/0031-9155/53/6/015
Chan M, Yang J, Song Y, Burman C, Chan P, Li S (2011) Evaluation of imaging performance of major image guidance systems. Biomed Imaging Interv J 7:e11. doi:10.2349/biij.7.2.e11
Chang EL, Shiu AS, Mendel E, Mathews LA, Mahajan A, Allen PK, Weinberg JS, Brown BW, Wang XS, Woo SY, Cleeland C, Maor MH, Rhines LD (2007) Phase I/II study of stereotactic body radiotherapy for spinal metastasis and its pattern of failure. J Neurosurg Spine 7:151–160. doi:10.3171/SPI-07/08/151
Engelsman M, Rosenthal SJ, Michaud SL, Adams JA, Schneider RJ, Bradley SG, Flanz JB, Kooy HM (2005) Intra- and interfractional patient motion for a variety of immobilization devices. Med Phys 32:3468–3474. doi:10.1118/1.2089507
Forrest LJ, Mackie TR, Ruchala K, Turek M, Kapatoes J, Jaradat H, Hui S, Balog J, Vail DM, Mehta MP (2004) The utility of megavoltage computed tomography images from a helical tomotherapy system for setup verification purposes. Int J Radiat Oncol Biol Phys 60:1639–1644. doi:10.1016/j.ijrobp.2004.08.016
Furweger C, Drexler C, Kufeld M, Muacevic A, Wowra B, Schlaefer A (2010) Patient motion and targeting accuracy in robotic spinal radiosurgery: 260 single-fraction fiducial-free cases. Int J Radiat Oncol Biol Phys 78:937–945. doi:10.1016/j.ijrobp.2009.11.030
Guckenberger M, Meyer J, Wilbert J, Baier K, Bratengeier K, Vordermark D, Flentje M (2007) Precision required for dose-escalated treatment of spinal metastases and implications for image-guided radiation therapy (IGRT). Radiother Oncol 84:56–63. doi:10.1016/j.radonc.2007.05.021
Hoogeman MS, Nuyttens JJ, Levendag PC, Heijmen BJ (2008) Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging. Int J Radiat Oncol Biol Phys 70:609–618. doi:10.1016/j.ijrobp.2007.08.066
Hui SK, Lusczek E, DeFor T, Dusenbery K, Levitt S (2012) Three-dimensional patient setup errors at different treatment sites measured by the Tomotherapy megavoltage CT. Strahlenther Onkol 188:346–352. doi:10.1007/s00066-011-0066-z
Hyde D, Lochray F, Korol R, Davidson M, Wong CS, Ma L, Sahgal A (2012) Spine stereotactic body radiotherapy utilizing cone-beam CT image-guidance with a robotic couch: intrafraction motion analysis accounting for all six degrees of freedom. Int J Radiat Oncol Biol Phys 82:e555–e562. doi:10.1016/j.ijrobp.2011.06.1980
Jin JY, Ryu S, Rock J, Faber K, Chen Q, Ajlouni M, Movsas B (2008) Evaluation of residual patient position variation for spinal radiosurgery using the Novalis image guided system. Med Phys 35:1087–1093. doi:10.1118/1.2839097
Kim S, Jin H, Yang H, Amdur RJ (2009) A study on target positioning error and its impact on dose variation in image-guided stereotactic body radiotherapy for the spine. Int J Radiat Oncol Biol Phys 73:1574–1579. doi:10.1016/j.ijrobp.2008.12.023
Li W, Sahgal A, Foote M, Millar BA, Jaffray DA, Letourneau D (2012) Impact of immobilization on intrafraction motion for spine stereotactic body radiotherapy using cone beam computed tomography. Int J Radiat Oncol Biol Phys. doi:10.1016/j.ijrobp.2011.12.039
Ma L, Sahgal A, Hossain S, Chuang C, Descovich M, Huang K, Gottschalk A, Larson DA (2009) Nonrandom intrafraction target motions and general strategy for correction of spine stereotactic body radiotherapy. Int J Radiat Oncol Biol Phys 75:1261–1265. doi:10.1016/j.ijrobp.2009.04.027
Mackie TR, Holmes T, Swerdloff S, Reckwerdt P, Deasy JO, Yang J, Paliwal B, Kinsella T (1993) Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. Med Phys 20:1709–1719. doi:10.1118/1.596958
Meeks SL, Harmon JF Jr, Langen KM, Willoughby TR, Wagner TH, Kupelian PA (2005) Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit. Med Phys 32:2673–2681. doi:10.1118/1.1990289
Ruchala KJ, Olivera GH, Schloesser EA, Mackie TR (1999) Megavoltage CT on a tomotherapy system. Phys Med Biol 44:2597–2621. doi:10.1088/0031-9155/44/10/316
Ruchala KJ, Olivera GH, Kapatoes JM, Schloesser EA, Reckwerdt PJ, Mackie TR (2000a) Megavoltage CT image reconstruction during tomotherapy treatments. Phys Med Biol 45:3545–3562
Ruchala KJ, Oliverat GH, Kapatoest JM, Schloesser EA, Reckwerdt PJ, Mackie TR (2000b) Megavoltage CT imaging as a by-product of multileaf collimator leakage. Phys Med Biol 45:N61–N70
Ryu S, Fang Yin F, Rock J, Zhu J, Chu A, Kagan E, Rogers L, Ajlouni M, Rosenblum M, Kim JH (2003) Image-guided and intensity-modulated radiosurgery for patients with spinal metastasis. Cancer 97:2013–2018. doi:10.1002/cncr.11296
Sahgal A, Larson DA, Chang EL (2008) Stereotactic body radiosurgery for spinal metastases: a critical review. Int J Radiat Oncol Biol Phys 71:652–665. doi:10.1016/j.ijrobp.2008.02.060
Sterzing F, Schubert K, Sroka-Perez G, Kalz J, Debus J, Herfarth K (2008) Helical tomotherapy. Experiences of the first 150 patients in Heidelberg. Strahlenther Onkol 184:8–14. doi:10.1007/s00066-008-1778-6
Zhou J, Uhl B, Dewit K, Young M, Taylor B, Fei DY, Lo YC (2010) Analysis of daily setup variation with tomotherapy megavoltage computed tomography. Med Dosim 35:31–37. doi:10.1016/j.meddos.2009.01.005
Acknowledgments
This study was supported by a grant from the Korea Healthcare Technology R&D Project, the Ministry for Health, Welfare, and Family Affairs, and the Republic of Korea (A084120).
Conflict of interest
There are no actual or potential conflicts of interest to disclosure for any of the authors.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Chung, Y., Yoon, H.I., Kim, J.H. et al. Is helical tomotherapy accurate and safe enough for spine stereotactic body radiotherapy?. J Cancer Res Clin Oncol 139, 243–248 (2013). https://doi.org/10.1007/s00432-012-1321-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00432-012-1321-0