Abstract
Early and successful extubation prevents several morbidities in preterm newborns. Several secondary non-invasive respiratory modalities exist but with their merits and demerits. Given the benefits of nasal high-frequency oscillatory ventilation (nHFOV), we tried to examine whether nHFOV could reduce reintubation rates compared to nasal intermittent positive pressure ventilation (NIPPV) during the post-extubation phase in preterm infants. Stratified randomisation based on gestational age was done for 86 mechanically ventilated preterm infants between 26 and 36+6 weeks of gestation within 2 weeks of age to receive either nHFOV or NIPPV post-extubation. The main objective was to compare extubation failure within 72 h following extubation and secondarily feed intolerance, intraventricular haemorrhage (IVH) (> grade 3), composite bronchopulmonary dysplasia (BPD)/mortality, composite duration of oxygen supplementation/ventilation support and SpO2/FiO2 ratio. No statistical difference was noted for primary outcome (RR 0.8, 95% CI: 0.23 to 2.78; p = 1.00) and secondary outcomes. However, nHFOV appeared possibly better in respect to feed tolerance rates and pCO2 washout.
Conclusion: Extubation failure within 72 h in infants less than 37 weeks of gestation did not differ between the two groups. However, nHFOV seems promising in reducing enteral feeding issues and pCO2 elimination. Larger multicentre studies are required for exploring benefits of nHFOV.
Trial registration: www.ctri.nic.in id CTRI/2019/07/020055, registration date July 5, 2019
What is Known: • NIPPV is superior to nCPAP as a secondary mode of respiratory support. • Synchronisation is preferred for optimum ventilation. | |
What is New: • nHFOV, a novel non-invasive respiratory modality without need for synchronisation, appears promising as a secondary mode subject to further trials. • It seems promising in reducing enteral feeding issues and pCO2 elimination. |
Similar content being viewed by others
Introduction
The transition of intrauterine to extra-uterine environment is marked by complex pulmonary and haemodynamic changes which occur smoothly and uneventfully in most infants [1]. Disturbed adaptation to extra-uterine life leading to respiratory distress occurs in about 7% of neonates [2]. Preterm neonates are at a greater risk for developing respiratory distress due to myriad of causes. Most of these infants therefore require some form of respiratory support to aid their breathing effort. With the advancement of medical research and science, it has been proved that mechanical ventilation, though the gold standard and probably the best mode of ventilation, is crippled with long-term respiratory morbidities. Therefore, in the quest for various non-invasive methods, the present day has seen a balanced arsenal of tools starting from synchronized non-invasive positive pressure ventilation (SNIPPV) to nasal continuous positive airway pressure (nCPAP), each of which has its pros and cons. The idea of this study is to highlight another brick in the wall, the nasal high-frequency oscillatory ventilation (nHFOV).
Time and again, it has been proved that the lesser the time an infant spends on invasive ventilation, the lesser the risk of chronic lung injury. Moreover, preterm infants need to tolerate extubation successfully. To date, nasal IPPV is considered the best modality post-extubation to enable infants to have a smooth transition from invasive to non-invasive modes. However, synchronization which is difficult to achieve in most cases, may provide better results [3]. The beauty of nHFOV lies primarily in the fact that it does not need synchronization [4]. Theoretically, nasal HFOV combines the benefits of both invasive high-frequency oscillatory ventilation (HFOV) and nasal continuous positive airway pressure (nCPAP) [4]. De Luca et al. suggested working parameters for nHFOV in different clinical scenarios, which need verification in adequately powered studies [4].
Few crossover and randomized control trials using nHFOV as a primary mode in RDS have been undertaken, but large-scale data is still lacking [5,6,7,8,9]. In most of these studies, nHFOV was compared with nCPAP. nHFOV is being presently practised in some European countries, Canada and China. However, a worldwide acclamation of this novel method is yet to happen [10]. Only two studies have evaluated the efficacy of nHFOV post-extubation [11, 12]. SNIPPV might be the most effective non-invasive respiratory support modality in the immediate post-extubation phase but is not readily available [13]. Currently, a large multicentric trial is being conducted in China which aims to select the superior secondary mode of non-invasive ventilation amongst nCPAP, NIPPV and nHFOV [14].
We hypothesized that using nHFOV as a post-extubation mode will enable easy weaning of an infant from the ventilator and reduce the need for reintubation. Comparison of nHFOV with NIPPV is still in its early stages and more studies are required to reach consensus statement. Therefore, we planned to compare nHFOV versus NIPPV as a post-extubation respiratory mode in preterm neonates between 26 and 36+6 weeks gestational age admitted in a tertiary care centre. To the best of our knowledge, such a study has not yet been carried out. Given the advantages of nHFOV over NIPPV, this modality can provide an added advantage in a future newborn respiratory care.
Material and methods
Trial design, settings and participants
This randomized control trial was conducted in level III neonatal intensive care unit (NICU) in a tertiary care hospital in Kolkata, India from July 2019 to September 2020. All preterm newborns (26–36+6 weeks) with respiratory distress, presenting within 15 days of life requiring invasive ventilatory support for at least 12 h were enrolled in the study. Small for gestational age infants were identified according to Fenton’s preterm growth chart [15]. Infants with major congenital anomalies or known/suspected chromosomal anomalies, upper airway anomalies, severe perinatal asphyxia or born outside the institute were excluded.
Intervention
Informed consent from parents was taken. Infants were intubated and put on Drager Babylog 8000 plus ventilator (Lϋbeck, Germany) using synchronised intermittent positive pressure ventilation with volume guarantee mode primarily for absent, weak, or intermittent spontaneous effort , frequent (> 6 events/h) or severe apnoea requiring positive pressure ventilation, marked retractions, severe tachypnea > 100/min, Silverman Anderson score > 7, pH < 7.2 and not improving, pCO2 > 65 on days 0–3, > 70 beyond day 3, shock requiring inotropic support. Extubation was done when working Pmean < 8 cm H2O, FiO2 < 0.3 and tidal volume < 4 ml/kg. Post-extubation respiratory support was provided in intervention (nHFOV) and comparator (NIPPV) groups, after randomization by simple online randomization done at the time of intubation. Blinding was not possible because of the nature of the study. nHFOV was provided by SLE 6000 ventilator (Surrey, UK) via Fisher Paykel FlexiTrunkTM interface. NIPPV was given via Dragger Babylog 8000 plus ventilator (Lϋbeck, Germany) via Fisher Paykel FlexiTrunkTM interface. Cycling of prongs and masks was done every 4 h. Chin straps were not used. Prong or mask size was chosen to have a snug fitting as per recommendations of Fisher Paykel FlexiTrunkTM interface. Caffeine was given to infants with birth weight less than 1250 g and continued till 5 days post-weaning from respiratory support. The oxygen saturation probe placement was standardized on pre-ductal location and was monitored continuously. Arterial blood gas was obtained at 12 h post-intervention. Infant’s respiratory condition was monitored with Silverman Anderson scoring every 2 h. Haemodynamic condition was assessed every 2 h in form of saturation, perfusion, and capillary refill time. Assessment for feed tolerance was done every 3 h. Echocardiographic assessment of haemodynamics and patent ductus arteriosus and transcranial ultrasound for intraventricular haemorrhage was done as per unit protocol. Sepsis workup was done on clinical basis and based on risk factors as and when needed. Respiratory morbidity was assessed at appropriate time frame. Infants were followed up as specified in the foregoing until discharge from unit or death. Pre-extubation mode of invasive ventilation was at the treating physician’s discretion. Settings used in the two arms are depicted in Table 1.
Outcomes
Primary outcome: Extubation failure within 72 h.
Secondary outcomes:
-
(i)
Reintubation rate
-
(ii)
Invasive ventilator free days [16]
-
(iii)
SpO2/FiO2 ratio
-
(iv)
Composite duration of oxygen supplementation/ventilation support
-
(v)
pCO2 and pH 12 h post-intervention
-
(vi)
IVH (above grade 3) [17]
-
(vii)
Composite bronchopulmonary dysplasia/mortality [18]
-
(viii)
Rate of feed intolerance [19]
-
(ix)
Time taken to full enteral feeds
-
(x)
Pulmonary air leaks
Sample size
NIPPV failure rate in our NICU ranges between 25% and 35%. From the Cochrane review by Lemyre et al. [3], we identified 5 studies with the use of non-synchronous NIPPV which had varied heterogeneity amongst the study populations in respect to gestational age but all of them belonged uniformly to preterm gestation. Kirpalani et al. studied the maximum number of infants in this context [20]. However, in their study, NIPPV was supposed to be delivered via mixed devices and due to unavailability of any Food and Drug Administration (FDA)–approved synchronised devices, as acknowledged by the author, it can be considered that non-synchronised machines were used mostly. From these studies, the failure rate of NIPPV was cumulatively around 33% which, along with our internal NICU data, prompted us to consider a 35% failure rate for secondary mode of NIPPV. We, therefore, chose to study 43 subjects in each group in order to have 80% power to reduce the extubation failure rate from this baseline rate of 35 to 10% with a significance level of 0.05% (two-tailed), using uncorrected chi-square test to evaluate the null hypothesis. Sample size was calculated using PS: Power and Sample Size Calculation (Version 3.1.6, Department of Biostatistics, Vanderbilt University School of Medicine, Nashville, USA).
Randomization
Randomization was done by computer-generated random sequence number (Research Randomizer (Version 4.0)); further stratification was done based on gestational age into two subgroups 26–31+6weeks and 32–36+6 weeks. The allocation ratio was 1:1 and concealment was done by using a serially numbered opaque sealed envelope. The generation of random numbers and assignment was done by a person not involved in the study. The infants and personnel could not be blinded due to the nature of intervention; however, the outcome assessor was blinded.
Statistical analysis
Analysis was done using GraphPad Prism version 7.0.0 for Windows, (GraphPad Software, San Diego, CA, USA), MedCalc for Windows, version 19.4 (MedCalc Software, Ostend, Belgium). Data was summarized by routine descriptive statistics, median and interquartile range for numerical variables, and counts and percentages for categorical variables. Numerical variables were compared between subgroups by Student’s independent samples t test if normally distributed, or by Mann-Whitney U test if otherwise. Fisher’s exact test was employed along with calculation of relative risk (RR) and 95% confidence interval (CI) for inter-group comparison of categorical variables. Analyses were two-tailed and statistical significance level was set at p < 0.05 for all comparisons.
Ethics
This study was carried out in accordance with the Declaration of Helsinki and was approved by the institutional ethics committee of the Institute of Post Graduate Medical Education and Research, Kolkata, India (IPGME&R/IEC/2019/434). Written informed consent was obtained from all legal guardians before participation in the study. This trial was registered in Clinical trial registry of India (Registration number CTRI/2019/07/020055).
Results
Out of 650 total preterm live births during the study period, 112 infants were assessed for eligibility. After exclusion of 26, a total of 86 infants were subjected to stratified randomisation into one of the two groups of nHFOV and NIPPV, stratification done based on gestational age in two subgroups 26–31+6weeks and 32–36+6 weeks. The flow of participants in the study is summarised in Fig. 1. Both groups were comparable with the baseline variables (Table 2). The median gestation age of the infants was 31.5 weeks with a median birth weight of 1500 g. The duration of invasive ventilation before being extubated to the respective intervention arms was a median of 29 h for nHFOV and 27 h for NIPPV. The results of the primary and secondary outcomes are depicted in Tables 3, 4 and 5.
There was no reduction in need of reintubation at 72 h between the nHFOV group (9.3%) and the NIPPV group (11.6%) (p = 1.000). In a subgroup analysis also, statistical significance was not found for the primary outcome. Overall reintubation rate was reduced but not statistically significant (16.2 vs. 18.6%). Composite duration of oxygen supplementation/ventilation support and Spo2/FiO2 ratio was similar between the two groups.. There was a possible trend towards better pCO2 elimination (p = 0.097) and pH optimisation (p = 0.073) 12 h after the start of intervention. No differences were observed between the two groups as well as sub groups for IVH (> grade 3), air leaks, composite BPD/ mortality before discharge. There was a possible trend towards lesser feed intolerance in the nHFOV group (37.2 vs. 58.13%, p = 0.084). However, time to achieve full enteral feeds was similar amongst the groups as well as subgroups. We did not find any difference in the number of ventilator free days between the two groups.
Discussion
In our randomised controlled trial, there was no difference in reintubation rates within 72 h in nHFOV group, compared with NIPPV group. The consideration of higher NIPPV failure rates based on western literature and the desired reduction led to relatively smaller sample size estimation than ideally required for a statistical significant reduction of primary outcome. Need for mechanical ventilation was significantly reduced in many studies and two meta-analysis using nHFOV as a primary mode; however, all of them used nCPAP as the comparator arm [5, 6, 21]. nCPAP is still considered the standard of care post-extubation, while NIPPV now appears to be the better modality, with its own pitfalls [13]. nHFOV is adjudged in benchmark studies as a better respiratory support modality [4]. One RCT by Chen et al. with 206 infants showed lesser reintubation and pCO2 in 6 h time with nHFOV post-extubation against nCPAP [11]. Malakian et al., however, did not find any difference in rate of intubation within 72 h [9]. To the best of our knowledge, this is the first randomised controlled trial comparing nHFOV versus NIPPV as a post-extubation modality in preterm infants worldwide. The only trial (NCT02543125) comparing nHFOV with NIPPV as a post-extubation modality was initiated in China in the year 2016 and is yet to be completed. The composite duration of oxygen supplementation/ventilation support, SpO2/FiO2 ratio during the intervention in both groups was also similar in the study. These, along with no difference in primary outcome can be explained by the fact that both the interventions were working at similar mean airway pressures which is unlikely to bring about a difference in oxygenation.
Assessment of extubation readiness should be done with lung mechanics, pressure time index, minute ventilation test apart from clinical tests and assessment of dynamics of biological signals. One of the key factors for successful extubation depends on the mode of respiratory support provided post-extubation, so as to keep the lungs open. Synchronised form of NIPPV seems to be the best choice but unavailability seems to be an issue. The pre-extubation mean airway pressure (MAP), FiO2 and tidal volume in volume ventilation play key roles in the success of extubation. Though failure rates increase with decreasing gestation and weight, secondary measures can be taken to sustain a successful extubation. Another factor which might help prevent extubation failure is efficacious clearance of pCO2. Our study showed possible trend towards better clearance of pCO2 and normalisation of blood gases but failed to achieve statistical significance. Mukherjee et al., Colaizy et al. and Czernik et al. also showed similar significant reductions in pCO2 [12, 22, 23]. There was a time-dependent variation in pCO2 levels in many studies probably because of different amplitudes used. In our study, we used a mean airway pressure of 10.9 ± 2.06 and a median amplitude of 15 which was similar to most studies done with nHFOV [4, 5, 11, 21, 24]. The MAP was similar in the two groups and this was an advantage since it allowed us to compare the effect of the pressure waveform and the active expiration as these seem to be the only real difference between groups. On the contrary, low MAPs in nHFOV probably failed to recruit the lung effectively.
Added advantages of nHFOV include reduced episodes of bradycardia and desaturation [25], enhanced alveolar ventilation due to better alveolar recruitment and improved functional residual capacity (FRC), theoretically lesser ventilator-induced lung injury (VILI) [26], and reduced gastroesophageal reflux [27]. Various physiological and benchmark studies have demonstrated positive results in favour of nHFOV, viz. the feasibility of nHFOV in extreme low birth weight (ELBW) infants [28], effectiveness of different interfaces delivering nHFOV [24, 29, 30], the efficiency in eliminating carbon dioxide (CO2) [12, 21, 26], the effect of different parameters and leak on CO2 removal [24, 29, 31, 32], the transmission of oscillation and tidal volume delivery in the airways [31].
In this study, we found that infants in the nHFOV arm seemed to have better feed tolerance rates and earlier full enteral feeds by 1 day. Given the advantage of nHFOV over NIPPV in the aspect of no need for synchronisation [4], no glottic constrictions during breaths [33], and an active expiration, it is not surprising that nHFOV seems to lower feed intolerance rates. Because of lack of synchronised machines from this part of the world, we need to strike a fine balance between non-invasive ventilation and feed intolerance issues, and in this regard nHFOV seems promising.
As for secondary outcomes like IVH, more than grade 3, air leaks, ventilator free days, composite BPD/mortality, composite oxygen supplementation/ventilation support, S/F ratio did not differ much between the groups as well as subgroups. Interestingly, most studies done on nHFOV including the two meta-analysis failed to show any difference in BPD, air leaks, IVH and mortality [5,6,7, 9, 11, 21]. Though the literature suggests viscid secretions interfering with efficacy of nHFOV [10], we did not find this alarming side effect probably because of good nursing care and maintenance of oral hygiene while on ventilation. We cycled between short binasal prongs and mask every 4-hourly, both of which have been shown to be efficacious [24, 29,30,31]. However, none of the previous studies have used both simultaneously. Mask-delivered nHFOV has more dampening effect, therefore, theoretically would require more aggressive ventilatory parameters [30]. In our case, cycling of the interfaces was done for the comfort of the infants and whether it caused a drop in efficacy of nHFOV cannot be ascertained. It will require more studies to evaluate this aspect.
The major limitations of this study are firstly, a small sample size to derive any statistical significance. The higher NIPPV failure rate consideration at the onset of the study probably led to a relatively smaller sample size estimation. Further, comfort level of infants were not quantified objectively, neither GER could be documented. Also, whether intermittent use of mask led to lesser efficacy needs to be investigated further. Moreover, long-term side effects and neurodevelopmental outcome are something to look forward to. As far as blinding was concerned, neither the infants nor the personnel were blinded. However, the person performing the final analysis was blinded to the intervention. In this regard, we acknowledge that the data compilation could have been blinded. The Pmean pressures used in nHFOV arm was possibly sub-optimal without any effective alveolar recruitment as per available evidences [4]. Unfortunately, this was done as per our pre-planned study protocol. Finally, a multicentre study is the need of the hour for formulating standard operating protocols as well as better utilisation of this novel respiratory modality.
In summary, among preterm ventilated infants, nHFOV did not bring about a significant reduction in reintubation rates within 72 h of post-extubation. It seemed promising in reducing feed intolerance and optimising arterial blood pH and pCO2. Most secondary outcomes, however, were similar between the two groups. Further multicentric studies need to be planned to explore further benefits of this novel respiratory support as well as formulate standard operating protocols.
Data and materials availability
Available for viewing at https://drive.google.com/file/d/1ZH7jElbBkzTvwrpdQLi_AgnJ3413Aq73/view?usp=sharing
Code availability
N/A.
Abbreviations
- BPD:
-
Bronchopulmonary dysplasia
- CI:
-
Confidence interval
- CO2 :
-
Carbon dioxide
- ELBW:
-
Extreme low birth weight
- FiO2:
-
Fraction of inspired oxygen
- FRC:
-
Functional residual capacity
- GER:
-
Gastroesophageal reflux
- HFOV:
-
High-frequency oscillatory ventilation
- IQR:
-
Interquartile range
- IVH:
-
Intraventricular haemorrhage
- IVH Gr3+:
-
Intraventricular haemorrhage more than grade 3
- MV:
-
Mechanical ventilation
- nCPAP:
-
Nasal continuous positive airway pressure
- nHFOV:
-
Nasal high-frequency oscillatory ventilation
- NICU:
-
Neonatal intensive care unit
- NIPPV:
-
Nasal intermittent positive pressure ventilation
- pCO2:
-
Partial pressure of carbon dioxide
- PIP:
-
Peak inspiratory pressure
- PEEP:
-
Positive end-expiratory pressure
- RDS:
-
Respiratory distress syndrome
- RR:
-
Relative risk
- S/F ratio:
-
Saturation to fraction of inspired oxygen ratio
- SNIPPV:
-
Synchronised nasal intermittent positive pressure ventilation
- VILI:
-
Ventilator-induced lung injury
References
Martin RJ, Crowley MA (2013) Respiratory problems. In: Fanaroff AA, Fanaroff JM (eds) Care of the high risk neonate, 6th edn. Elsevier Saunders, Philadelphia, p 244
Agarwal R, Deorari A, Paul V, Sankar MJ, Sachdeva A (2019) AIIMS Protocols in Neonatology, 2nd edn. Noble Vision, Delhi
Lemyre B, Davis PG, De Paoli AG, Kirpalani H (2014) Nasal intermittent positive pressure ventilation (NIPPV) versus nasal continuous positive airway pressure (NCPAP) for preterm neonates after extubation. Cochrane Database Syst Rev (9):CD003212
De Luca D, Dell’Orto V (2016) Non-invasive high-frequency oscillatory ventilation in neonates: review of physiology, biology and clinical data. Arch Dis Child Fetal Neonatal Ed 101:F565–F570
Haidar Shehadeh AM (2019) Non-invasive high flow oscillatory ventilation in comparison with nasal continuous positive pressure ventilation for respiratory distress syndrome, a literature review. J Matern Fetal Neonatal Med 7:1–10
Zhu XW, Zhao JN, Tang SF, Yan J, Shi Y (2017) Noninvasive high-frequency oscillatory ventilation versus nasal continuous positive airway pressure in preterm infants with moderate-severe respiratory distress syndrome: a preliminary report. PediatrPulmonol. 52(8):1038–1042
Iranpour R, Armanian AM, Abedi AR, Farajzadegan Z (2019) Nasal high-frequency oscillatory ventilation (nHFOV) versus nasal continuous positive airway pressure (NCPAP) as an initial therapy for respiratory distress syndrome (RDS) in preterm and near-term infants. BMJ Paediatr Open 3(1):e000443
Klotz D, Schneider H, Schumann S, Mayer B, Fuchs H (2018) Non-invasive high-frequency oscillatory ventilation in preterm infants: a randomised controlled cross-over trial. Arch Dis Child Fetal Neonatal Ed 103(4):F1–F5
Malakian A, Bashirnezhadkhabaz S, Aramesh MR, Dehdashtian M (2020) Noninvasive high-frequency oscillatory ventilation versus nasal continuous positive airway pressure in preterm infants with respiratory distress syndrome: a randomized controlled trial. J Matern Fetal Neonatal Med 33(15):2601–2607
Fischer HS, Bohlin K, Bührer C, Schmalisch G, Cremer M, Reiss I, Czernik C (2015) Nasal high-frequency oscillation ventilation in neonates: a survey in five European countries. Eur J Pediatr 174(4):465–471
Chen L, Wang L, Ma J, Feng Z, Li J, Shi Y (2019) Nasal high-frequency oscillatory ventilation in preterm infants with respiratory distress syndrome and ARDS after extubation: a randomized controlled trial. Chest. 155(4):740–748
Mukerji A, Singh B, Helou SE, Fusch C, Dunn M, Belik J, Shah V (2015) Use of noninvasive high-frequency ventilation in the neonatal intensive care unit: a retrospective review. Am J Perinatol 30:171–176
Ramaswamy VV, Bandyopadhyay T, Nanda D, Bandiya P, More K, Oommen VI, Gupta A (2020) Efficacy of noninvasive respiratory support modes as postextubation respiratory support in preterm neonates: a systematic review and network meta-analysis. Pediatr Pulmonol 55:2924–2939
Shi Y, De Luca D, (2019) NASalOscillatioN post-Extubation (NASONE) study group. Continuous positive airway pressure (CPAP) vs. noninvasive positive pressure ventilation (NIPPV) vs. noninvasive high frequency oscillation ventilation (nHFOV) as post-extubation support in preterm neonates: protocol for an assessor-blinded, multicenter, randomized controlled trial. BMC Pediatr.;19(1):256
Fenton TR, Nasser R, Eliasziw M, Kim JH, Bilan D, Sauve R (2013) Validating the weight gain of preterm infants between the reference growth curve of the fetus and the term infant. BMC Pediatr 13(1):92
Dell’Orto V, Raschetti R, Centorrino R, Montane A, Tissieres P, Yousef N, De Luca D (2019) Short- and long-term respiratory outcomes in neonates with ventilator-associated pneumonia. Pediatr Pulmonol 54:1982–1988
Papile LA, Burstein J, Burstein R, Koffler H (1978) Incidence and evolution of subependymal and intraventricular hemorrhage: a study of infants with birth weights less than 1,500 gm. J Pediatr 92(4):529–534
Jobe AH, Bancalari E (2001) Bronchopulmonary dysplasia: NICHD/NHLBI/ORD workshop summary. Am J Respir Crit Care Med 163:1723–1729
Moore TA, Wilson ME (2011) Feeding intolerance: a concept analysis. Adv Neonatal Care 11(3):149–154
Kirpalani H, Millar D, Lemyre B, Yoder BA, Chiu A, Roberts RS, NIPPV Study Group (2013) A trial comparing noninvasive ventilation strategies in preterm infants. N Engl J Med 369(7):611–620
Li J, Li X, Huang X, Zhang Z (2019) Noninvasive high-frequency oscillatory ventilation as respiratory support in preterm infants: a meta-analysis of randomized controlled trials. Respir Res 20:58
Colaizy TT, Younis UM, Bell EF, Klein JM (2008) Nasal high-frequency ventilation for premature infants. Acta Paediatr 97(11):1518–1522
Czernik C, Schmalisch G, Bührer C, Proquitté H (2012) Weaning of neonates from mechanical ventilation by use of nasopharyngeal high-frequency oscillatory ventilation: a preliminary study. J Matern Fetal Neonatal Med 25(4):374–378
De Luca D, Piastra M, Pietrini D, Conti G (2012) Effect of amplitude and inspiratory time in a bench model of non-invasive HFOV through nasal prongs. Pediatr Pulmonol 47:1012–1018
Rüegger CM, Lorenz L, Kamlin OC, Manley BJ, Owen LS, Bassler D et al (2018) The effect of noninvasive high-frequency oscillatory ventilation on desaturations and Bradycardia in very preterm infants: a randomized crossover trial. J Pediatr 201:269–273.e2
Bottino R, Pontiggia F, Ricci C, Gambacorta A, Paladini A, Chijenas V et al (2018) Nasal high-frequency oscillatory ventilation and CO2 removal: a randomized controlled crossover trial. Pediatr Pulmonol 53(9):1245–1251
Cantin D, Djeddi D, Samson N, Nadeau C, Praud JP (2018) Nasal high-frequency oscillatory ventilation inhibits gastroesophageal reflux in the neonatal period. Respir Physiol Neurobiol 251:28–33
Aktas S, Unal S, Aksu M, Ozcan E, Ergenekon E, Turkyilmaz C, Hirfanoglu I, Atalay Y (2016) Nasal HFOV with binasal cannula appears effective and feasible in ELBW newborns. J Trop Pediatr 62(2):165–168
De Luca D, Carnielli VP, Conti G, Piastra M (2010) Noninvasive high frequency oscillatory ventilation through nasal prongs: bench evaluation of efficacy and mechanics. Intensive Care Med 36(12):2094–2100
Centorrino R, Dell’Orto V, Gitto E, Conti G, De Luca D (2019) Mechanics of nasal mask-delivered HFOV in neonates: a physiologic study. Pediatr Pulmonol 54(8):1304–1310
De Luca D, Costa R, Visconti F, Piastra M, Conti G (2016) Oscillation transmission and volume delivery during face mask-delivered HFOV in infants: bench and in vivo study. Pediatr Pulmonol 51:705–712
Klotz D, Schaefer C, Stavropoulou D, Fuchs H, Schumann S (2017) Leakage in nasal high-frequency oscillatory ventilation improves carbon dioxide clearance—a bench study. Pediatr Pulmonol 52(3):367–372
Hadj-Ahmed MA, Samson N, Nadeau C, Boudaa N, Praud JP (2015) Laryngeal muscle activity during nasal high-frequency oscillatory ventilation in non sedated newborn lambs. Neonatology. 107(3):199–205
Acknowledgements
We would like to thank Dr. Bhupendra Kumar Gupta for generation of random sequence and assigning participants to intervention.
Author information
Authors and Affiliations
Contributions
Dr. Soutrik Seth conceptualized and designed the study, developed the protocol, patient management and prepared the first draft. Dr. Bijan Saha helped in protocol development, coordinated, supervised data collection, reviewed and revised the manuscript at all stages of its production. Dr. Anindya Kumar Saha critically reviewed and revised the manuscript. Dr. Suchandra Mukherjee helped in protocol development and critically reviewed the manuscript for improving the content. Dr. Avijit Hazra performed the statistical analysis of the data. All the authors approved the final manuscript as submitted and agree to be accountable for all aspect of the work.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Informed consent was obtained from the legal guardian of all included participants.
Consent for publication
Informed consent was obtained from the legal guardian of all included participants.
Conflict of interest
The authors declare no competing interests.
Additional information
Communicated by Daniele De Luca
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
ESM 1
(DOCX 13 kb)
Rights and permissions
About this article
Cite this article
Seth, S., Saha, B., Saha, A.K. et al. Nasal HFOV versus nasal IPPV as a post-extubation respiratory support in preterm infants—a randomised controlled trial. Eur J Pediatr 180, 3151–3160 (2021). https://doi.org/10.1007/s00431-021-04084-1
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00431-021-04084-1