Abstract
Purpose
Aortic dissection is a complex condition with high morbidity and mortality. Endovascular treatments including percutaneous fenestration can be used to manage branch vessel ischaemia or risk of aortic rupture. A variety of techniques for aortic fenestration have been described. We describe the novel use of the TruePath Chronic Total Occlusion (CTO) device for aortic intimal fenestration to achieve side branch reconnection.
Materials and Methods
We present three cases of aortic dissection presenting with symptoms of aortic side branch occlusion and end organ malperfusion, treated with aortic fenestration using the TruePath CTO device via trans-brachial and trans-femoral approaches.
Results
Technical success was achieved in all three cases. No complications were encountered. Flow was restored in compromised visceral branches. Branches remained patent on follow-up CT angiography over a minimum 2.5-year follow-up period.
Conclusion
Percutaneous aortic fenestration techniques enable a minimally invasive approach to treat visceral branch malperfusion associated with aortic dissection. The TruePath CTO device improves the control of the fenestration procedure with the potential to improve efficacy and safety.
References
Golledge J, Eagle KA. Acute aortic dissection. Lancet. 2008;372(9632):55–66.
Oderich GS, et al. Aortic dissection with aortic side branch compromise: impact of malperfusion on patient outcome. Perspect Vasc Surg Endovasc Ther. 2008;20(2):190–200.
Vendrell A, et al. Acquired cardiovascular disease: aortic dissection: Aortic dissection with acute malperfusion syndrome: Endovascular fenestration via the funnel technique. J Thorac Cardiovasc Surg. 2015;150:108–15.
Chavan A, et al. Endoluminal treatment of aortic dissection. Eur Radiol. 2003;13(11):2521–34.
Rousseau H et al. Endovascular treatment of the complications of aortic dissection: fenestration and stenting. 2006;247.
Australasian Vascular Audit Report. 2016, Australian and New Zealand Society for vascular surgery; 2016. p. 37.
Hartnell GG, Gates J. Aortic fenestration: A why, when, and how-to guide. RadioGraphics. 2005;25(1):175–89.
Elefteriades JA, Escalon JC. Cardiac surgery: aortic fenestration for dissection. Oper Tech Thorac Cardiovasc Surg. 2009;14:12–22.
Wolfschmidt F, et al. Aortic dissection: accurate subintimal flap fenestration by using a reentry catheter with fluoroscopic guidance-initial single-institution experience. Radiology. 2015;276(3):862–72.
Midulla M, et al. Acquired cardiovascular disease: endovascular fenestration in aortic dissection with acute malperfusion syndrome: Immediate and late follow-up. J Thorac Cardiovasc Surg. 2011;142:66–72.
International, B.S. TruePath CTO. [pdf] 2012 [cited 2017 10/06/2017]; 01:[Instruction Manual for use of TruePath CTO Device]. Available from: https://www.bostonscientific.com/content/dam/Manuals/us/current-rev-en/90741285-01A_TruePathCTO_dfu_us_S.pdf.
Patel HJ, et al. Acquired cardiovascular disease: long-term results of percutaneous management of malperfusion in acute type B aortic dissection: Implications for thoracic aortic endovascular repair. J Thorac Cardiovasc Surg. 2009;138:300–8.
Bosiers M, et al. Revascularization of lower extremity chronic total occlusions with a novel intraluminal recanalization device: results of the ReOpen study. J Endovasc Ther. 2014;21(1):61–70.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
We declare that we have no conflict of interest.
Rights and permissions
About this article
Cite this article
Ghali, R., Maingard, J., Kok, H. et al. Catheter-Directed Fenestration for Branch Vessel Reconnection in Aortic Dissection Using a Novel Diamond-Tipped Chronic Total Occlusion Drilling Device: A Technical Report. Cardiovasc Intervent Radiol 42, 608–614 (2019). https://doi.org/10.1007/s00270-018-2124-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-018-2124-8