Abstract
Purpose
Difficulty can be encountered during advancement of the tracheal tube (TT) over the bronchoscope after successful endotracheal bronchoscopy due to impingement on laryngeal structures. A new TT, the Parker Flex-Tip (PFT), has been shown to be superior to polyvinylchloride (PVC) TTs in anesthetized, paralyzed patients with normal airways. However, no study to date has shown the superiority of the new tapered tip design in patients with difficult airways during awake fibreoptic intubations (AFOI). The purpose of this study was to compare the PFT with PVC TTs for AFOI in patients with difficult airways or unstable c-spines. Clinical features: In this prospective observational study, 1 1 1 patients with predicted or documented difficult airways, or unstable c-spines were assessed for ease of TT advancement during AFOI. First attempt success rates were 91 % for PFT TTs and 84% for PVC TTs (P = NS). Resistance to TT advancement was none to mild and similar in both groups. Advancement without the need to rotate the TT 180° was also similar in both groups (57% vs 53%). Conclusion: For AFOI in patients with difficult airways, the PFT is not superior to conventional PVC TTs.
Objectif
Il peut être difficile d’avances le tube endotrachéal (TET) le long du bronchoscope lors d’une intubation fibroscopique à cause d’un contact avec des structures laryngées. Un nouveau TET, le Parker Flex-Tip (PFT), s’est révélé supérieur aux TET de polychlorure de vinyle (PCV) chez les patients anesthésiés et paralysés aux voies aériennes normales. Mais aucune étude à ce jour n’a montré la supériorité de ce tube à pointe effilée dans les cas de problèmes de voies aériennes pen-dant l’intubation fibroscopique vigile (IFOV). Notre étude compare le PFT avec les TET de PCV pour l’IFOV en cas d’anomalies des voies aériennes ou d’instabilité de la colonne cervicale. Éléments cliniques : Notre étude observationnelle prospective a évalué 111 patients avec des problèmes de voies aériennes prévus ou documentés ou une instabilité de la colonne cervicale, quant à la faci-lité d’avancer le TET pendant l’IFOV. Les succès du premier essai ont été de 91 % pour le TET PFT et de 84 % pour le TET de PCV (P = NS). La résistance à l’avancement du TET a été de nulle à légère dans les deux groupes. L’avancement sans nécessité de rotation de 180° du TET a aussi été comparable dans les deux groupes (57 % vs 53 %).
Conclusion
Pour l’IFOV chez les patients qui ont des problèmes de voies aériennes, le PFT n’est pas supérieur au TET de PCV traditionnel.
Article PDF
Similar content being viewed by others
References
Katsnelson T, Frost EA, Farcon E, Goldiner PL. When the endotracheal tube will not pass over the flexible fiberoptic bronchoscope (Letter). Anesthesiology 1992; 76:151–2.
Greer JR, Smith SP, Strang T. A comparison of tracheal tube tip designs on the passage of an endotracheal tube during oral fiberoptic intubation. Anesthesiology 2001; 94:729–31.
Hakala P, Randell T, Valli H. Comparison between tracheal tubes for orotracheal fibreoptic intubation. Br J Anaesth 1999; 82:135–6.
Lucas DN, Yentis SM. A comparison of the intubating laryngeal mask tracheal tube with a standard tracheal tube for fibreoptic intubation. Anaesthesia 2000; 55:358–61.
Kristensen MS. The Parker Flex-Tip tube versus a standard tube for fiberoptic orotracheal intubation: a randomized double-blind study. Anesthesiology 2003; 98:354–8.
Connelly NR, Kyle R, Gotta J, et al. Comparison of wire reinforced tubes with warmed standard tubes to facilitate fiberoptic intubation. J Clin Anesth 2001; 13:3–5.
Jones HE, Pearce AC, Moore P. Fibreoptic intubation. Influence of tracheal tube tip design. Anaesthesia 1993; 48:672–4.
Benumof JL. Management of the difficult adult airway. With special emphasis on awake tracheal intubation. Anesthesiology 1991; 75:1087–110.
Hakala P, Randell T. Comparison between two fibrescopes with different diameter insertion cords for fibreoptic intubation. Anaesthesia 1995; 50:735–7.
Makino H, Katoh T, Kobayashi S, Bito H, Sato S. The effects of tracheal tube tip design and tube thickness on laryngeal pass ability during oral tube exchange with an introducer. Anesth Analg 2003; 97:285–8.
Author information
Authors and Affiliations
Corresponding author
Additional information
Assessed June 22, 2004.
Rights and permissions
About this article
Cite this article
Joo, H.S., Naik, V.N. & Savoldelli, G.L. Parker Flex-Tip™ are not superior to polyvinylchloride tracheal tubes for awake fibreoptic intubations. Can J Anesth 52, 297–301 (2005). https://doi.org/10.1007/BF03016067