Background

Bexarotene (brand name: Targretin) [1] is an antineoplastic (anticancer) agent approved by the US Food and Drug Administration (FDA) (in late 1999) and the European Medicines Agency (EMA) (early 2001) for use as a treatment for cutaneous T cell lymphoma (CTCL) (Fig. 1) [2, 3]. It is a third-generation retinoid. The retinoic acid receptors (RARs) regulate cell differentiation and proliferation whereas RXRs regulate apoptosis [4]. LC-MS-based method that utilized both RPLC and HILIC separations was carried out [1,2,3,4], followed by multivariate data analysis to discriminate the global urine profiles of BC patients and healthy controls [1, 5]. The purpose of this study was to identify a potential biomarker pattern in urine using metabonomics to aid noninvasive BC detection using complementary chromatographic techniques [6, 7].

Fig. 1
figure 1

Chemical structure of bexarotene. Formula: C24H28O2. Molar mass: 348.478 g/mol

Methods

A few methods are available in literature [3, 8,8,9,10,11,12,13,14,15,16]. A new bioanalytical LC-MS/MS method was performed on the LC-MS/MS (API 4000) [6, 7, 17], consisting of binary gradient pump UV detector (LC-20 AD) employed for analysis, and rheodyne injector with 20 μl fixed loop was used for the present study. Bexarotene was eluted with a flow rate of 1 ml/min using a mobile phase of acetonitrile: buffer 1(90:10, v/v). The retention time of bexarotene analyte is 2.75 ± 0.3 min.

Extraction procedure: acetone-M: 10 mM ammonium format as extraction solvent

Spiked plasma samples were vortexed to ensure complete mixing of contents; 50 μl of internal standard (1 μl/ml of bexarotene D4) solution was added into all respectively labeled empty RIA vials except blank. Five hundred microliters of plasma samples was added to the respective labeled RIA vials containing internal standard solution and vortexed. Two hundred microliters of buffer 1 was added to all the samples and vortexed. 2.5 ml of extraction solvent was added to all the samples and capped. Samples were vibramaxed at 2000 rpm for 10 min. Then, the samples were centrifuged at 3500 rpm for 5 min in a refrigerated centrifuge between 2 and 8 °C. Two milliliters of supernatant was transferred into respective labeled RIA vials. All the samples were dried at 40 °C and 15 psi using LV evaporator. The dried residues get reconstituted with 300 μl of mobile phase and vortexed. The phospholipid removal cartridges get conditioned with 1 ml of acetone-M followed by 1 ml of conditioning solution. The samples were loaded into cartridges and eluted into RIA vials. The samples were transferred into respective labeled auto injector vials and loaded into LC-MS/MS [3].

Results

Method validation

The method was validated according to ICH Guidelines Q2 (R1) with respect to system suitability, linearity, accuracy, precision, matrix effect, auto sampler carryover test, and recovery.

System suitability

Aqueous standard or extracted standard equivalent to middle level of CC standard concentration with internal standard was prepared. Six replicates from the same vial were injected into the chromatographic device. Mean, standard deviation, and percentage coefficient of variation for the retention time and area/area ratio were calculated.

Linearity

Different serial dilutions were repeated, and fresh aqueous standards (for CCs) were prepared. An appropriate regression model with minimal or no weighing (1/x or 1/x2) was used. The standards were run in the LC-MS/MS, and linearity was evaluated.

Selectivity/specificity

This is to check whether there is an interference in peak. Two sets of six normal lots of plasma and one hemolyzed were taken. The aqueous LLOQ dilution was prepared and was spiked in one set of six normal lots of plasma, one hemolyzed lot to achieve LLOQ concentration for analyte, and the specificity sample was processed. The internal standard dilution was prepared, and only 50 μL of internal standard dilution was added to another set of six normal lots of plasma, one hemolyzed was processed for specificity samples. Selectivity samples were prepared in the presence of both analyte and internal standard using the six normal blank plasmas and one hemolyzed.

Precision and accuracy

The precision was determined by calculating percentage %CV at each concentration level of QC sample, and the accuracy was determined by calculating the percentage of nominal value at each concentration level of QC samples.

Ruggedness

One P&A batch was performed by employing the same instrument with different analysts and alternatively performed on different instruments of same make.

Recovery

The overall mean recovery, SD, and %CV were calculated. The recovery experiment was carried out by inject the six replicates of unextracted low, medium, and high QC samples, along with freshly processed CC set and QCs (6 LOC, 6 MQC, and 6 HQ).

Stability

Evaluation of stability should be carried out to ensure that every step taken during sample preparation and sample analysis, as well as the storage conditions used, do not affect the concentration of the analyte. The stability tests conducted in method validation are as follows:

  1. 1.

    Stock solution stability: short-term stock solution stability, long-term stock solution stability

  2. 2.

    Stability in biological matrix: bench top stability, freeze-thaw stability, long-term stability, blood stability

Discussion

Method development

The LCMS/MS procedure was optimized for the estimation of bexarotene with the mobile phase of acetonitrile: buffer 1 (90:10, v/v); the optimum flowrate was 1 ml/min with a column oven temperature and autosampler temperature of 40 °C and 10 °C respectively. Retention time of analyte is 2.75 ± 0.3 min, and IS is 2.73 ± 0.3 min.

Specificity and selectivity

Selectivity was evaluated by analyzing a total of nine lots on the instrument [17] obtained from independent sources (Table 1). No significant interferences were observed at the retention times of analyte and internal standard (see Table 3).

Table 1 Specificity and selectivity for bexarotene and internal standard (MTR-BA-LC-MS/MS-05)

Signal-to-noise (S/N) ratio

The signal-to-noise ratio was determined for bexarotene at LLOQ concentrations in nine independent lots of K2EDTA human normal plasma including one lot of hemolyzed plasma, one lot of heparin plasma, and one lot of lipemic plasma [8] demonstrating acceptable S/N intensity.

Carryover test

Carryover is calculated as the percentage peak area observed in a processed blank plasma injected immediately after a processed ULOQ calibration standard, which was used from PA-01 batch sample analysis. No significant carryover was observed for bexarotene and internal standard (see in Table 2).

Table 2 Carry over test

Matrix effect and matrix factor

Matrix factor and matrix effect were calculated, and results are given in Tables 3 and 4.

Table 3 Matrix effect and matrix factor for bexarotene at LQC level
Table 4 Matrix effect and matrix factor for bexarotene at LQC level (MTR-BA-LC-MS/MS-05)

Linearity

Linearity established [9] by preparing an eight-point standard calibration curve in K2EDTA human plasma covering the bexarotene concentration ranges from 1.0440 to 351.9320 μg/mL using bexarotene D4 as internal standard. The calibration curve was shown to be linear for bexarotene as shown in Fig. 2; the results are seen in Table 5.

Fig. 2
figure 2

Calibration curve

Table 5 Back-calculated concentrations of bexarotene for calibration curve standards

Weighting scheme

The absolute values of residuals of the back-calculated bexarotene calibration standards for the curve were tabulated, and the sum of the absolute values of the residuals was calculated for each weighting factor. The weighting factor of 1/X2 provided the least sum value with residuals of calibration curve standards. Hence, 1/X2 was selected to use for this validation. See the results in Table 6.

Table 6 Weighting scheme

Sensitivity

The sensitivity for bexarotene at LLOQ level in K2EDTA human plasma determined based on the analysis of six replicates of LLOQ (1.0440 ng/mL) samples was prepared and analyzed against calibration curve standards. See the results in Table 7.

Table 7 Sensitivity

Intra-batch precision and accuracy of bexarotene

See the results in Table 8.

Table 8 Intra-batch precision and accuracy of bexarotene

Ruggedness

Accuracy, assay precision, and accuracy value for ruggedness batch (PA-03) were determined by analyzing six replicates each of LOQQC, LQC, INTQC, MQC, and HQC samples using different instrument (MTR-BA-LC-MS/MS-05) of the same make and model (UPLC with Triple Quad API 4000), different analytical column (BAC-0644), and different analyst (Table 9).

Table 9 Ruggedness (MTR-BA-LC-MS/MS-05)

Intercept = 0.0019, Slope = 0.0118, r2 = 0.9996

Precision

The precision of the assay was measured by the percentage co-efficient of variation over the concentration range of LOQQC, LQC, INTQC, MQC, and HQC samples of bexarotene during the course of partial validation. See the results in Table 10.

Table 10 Precision

Recovery of bexarotene and IS

The recovery of bexarotene was determined by comparing the detector response of analyte at three distinct levels of extracted low-, medium-, and high-quality control samples of PA-01 with detector response obtained from unextracted aqueous quality control samples at low, medium, and high level respectively. See the results in Tables 11 and 12.

Table 11 Recovery of bexarotene
Table 12 Recovery of internal standard

IS recovery = 95.72%

Stability

Freeze-thaw stability

Six replicates of bexarotene samples at LQC and HQC concentration in K2EDTA human plasma were analyzed after four freeze-thaw (FT4) cycles. See the results in Table 13.

Table 13 Freeze-thaw stability for bexarotene (at − 70 °C ± 15 °C and − 30 °C ± 10)

Bench top stability

Bench top stability of bexarotene in K2EDTA human plasma was evaluated at room temperature. Six replicates of LQC and HQC samples were processed after keeping the samples on bench for about 12.30 h. See the results in Table 14.

Table 14 Bench top stability for bexarotene

Auto sampler stability for bexarotene

Six replicates of LQC and HQC samples were processed and kept stored in auto sampler at 10 °C for 96.80 h [11]. See the results in Tables 15 and 16.

Table 15 Auto sampler stability for bexarotene
Table 16 Auto sampler stability for internal standard

Long-term stock solution stability for bexarotene

Stock solution bexarotene with concentration of 975.6417 μg/mL was kept in the refrigerator for 14 days [12]. A fresh stock of 986.1296 μg/mL was prepared on the day of analysis. Both stocks were diluted to LQC and HQC equivalent concentration of 0.1317 μg/mL and 0.1331 μg/mL and 13.1712 and 13.3127 μg/mL for stored and fresh stock respectively. The area ratios of stability stock solution at LQC and HQC level were compared against freshly prepared stock solution LQC and HQC level. See the results in Tables 17 and 18.

Table 17 Long-term stock solution stability for analyte LQC
Table 18 Long-term stock solution stability for internal standard LQC

Limit of detection

From LLOQ sample (1.0540 ng/mL), four different lower concentrations (0.8440, 0.6340, 0.4240, and 0.2120 ng/mL) including five times the lower concentration (LOD dilution) were prepared, and six replicates of these samples were analyzed. So the selected LLOQ (approx. 1.0540 ng/mL) was more suitable to quantify bexarotene in plasma using LC-MS/MS.

Reinjection reproducibility [14,15,16]

CC standards, LQC, and HQC samples of PA-02 were reinjected after 08.95 h. Percentage nominal for LQC and HQC for bexarotene was 98.99 and 100.30%, respectively. The percentage CV for LQC and HQC for bexarotene was 3.66 and 1.97%, respectively. See the results in Table 19.

Table 19 Reinjection reproducibility for bexarotene

Conclusion

Bioanalytical method is developed and validated as per ICH guidelines for the estimation of bexarotene in human plasma by using LC-MS/MS. The mobile phase was selected after trying various combinations of polar solvents. The proportion of solvents and variation of buffers were found to be quite critical as slight variation in it adversely affected the resolution of peaks. Considering all the facts, the validation parameter was finally fixed for this method. The bioanalytical method, a selective and sensitive liquid chromatography-mass spectrometry method to quantitate bexarotene in K2EDTA human plasma over the concentration range from 1.0440 to 351.9320 ng/mL, was successfully validated. This method is suitable for sample analysis to support bioequivalence/bioavailability and/or pharmacokinetic studies involving formulations of bexarotene.