Background

Radiotherapy brings the opportunity to control tumoral remnants at the post-surgical bed of oral cancers, and also, it is sometimes the only choice for definitive treatment of unresectable cancers of the oral cavity. It is estimated that 1–4% of radiation therapy patients have a prosthetic device that may interfere with dose delivery, plan design, and verification [1]. An increasing number of patients have modern dental restorations that are composed of biocompatible metals such as titanium. Zirconium is increasingly considered in new dental implants for its known properties of minimizing bacterial colonization and reducing plaque formation. It has induced less inflammation and hypersensitivity reactions and also has shown high stability [2, 3]. Mass densities of P-F-M crowns of implant systems and zirconium are higher than titanium, and therefore, their artifacts in imaging, interaction with higher energy photons in radiotherapy, and the amount of backscattered dose can be considerable [3,4,5]. These materials cause beam hardening and streaking artifacts in CT images and therefore conceal volumetric information and boundaries of radiotherapy targets. Additionally, Hounsfield unit (HU)-to-electron density conversion curves that are incorporated in Treatment Planning Systems (TPS) do not produce reliable outputs in high Z materials [6,7,8], so they misrepresent density information and alter global dose calculations. Moreover, attenuated beam penetrating the high Z material and lateral scatter cause dose inhomogeneities in Planning Target Volume (PTV) in head and neck cancer radiotherapy that can affect Tumor Control Probability (TCP) [9]. Imaging quality may be improved by metal artifact reduction (MAR) algorithms and extending the CT scale; however, these solutions have not always shown efficacy [10].

Modern collapsed cone convolution/superposition algorithms (CCCS) used for photon dose calculations pose the advantage of lateral scatter calculations in commercial Treatment Planning Systems (TPS) have weaknesses in calculations at metal interfaces which their clinical significance is undetermined [1]. Density correction metal kernel solutions are algorithms that can account for the attenuation of devices and are incorporated in a number of TPSs. According to Physicists in Medicine’s Task Group 63 recommendations ignoring the presence of the prosthesis by setting their density to water, or modifying beam orientations are suggested approaches that have been used in radiotherapy plans to reduce inevitable calculation errors in presence of metal devices in the hip [6, 7]. Meanwhile, because there are no universal practical recommendations for radiotherapy planning when beams penetrate a dental implant, and these devices are not removed before radiotherapy, in this in vitro study a simplified oral cavity phantom was used for point-to-point POI and ROI dose value and distribution comparison of calculated and measured film dosimetry when an implant replaced a normal tooth. Radiochromic films are a prime candidate for dosimetry when demanding high spatial resolution and conditions of charged particle disequilibrium [11]. Therefore, volumes of compositions of the implant system were contoured and calculations were compared to film dosimetry measurements. Relative dose homogeneity at a 2-dimensional level of crowns that represented the soft tissue of the oral cavity was also assessed. We aimed to answer questions of, (1) How does the titanium, zirconium, P-F-M dental implant system affect dose distribution upstream and downstream to the implant in a simplified field, and (2) Do specifying material for contoured components of an implant and applying metal kernels improve dose calculations?

Materials and methods

Phantom

A modeled coronal section of an oral cavity phantom made of Perspex (the equivalent of tissue), Teflon (the equivalent of mandible), and teeth were used for our study as described by Jabbari et. al [12]. Roots of three molar teeth were inserted inside the screwed modeled jaw. The teeth were then fixed and surrounded by soft tissue-equivalent bolus pellets (Orfit, Wijnegem, Belgium) (Fig. 1). The middle tooth was later replaced by a Z1410 dental implant that was composted of a titanium core and a zirconium collar, and abutment (TBR Implants Group, Toulouse, France). The implant system carries a porcelain-fused-to-metal (PFM) crown that includes an alloy of mainly CO and Cr. Two Perspex layers of 4.0 × 4.0 ×1.0 cm and one layer of 4.0 × 4.0 × 0.5 cm represented the soft tissue of the oral cavity including the tongue. Location of Gafchromic EBT-3 films (Ashland ISP Advanced Materials, NJ, USA) was set for two areas at different depths in the horizontal plane between Perspex layers. One film was placed at the level of the roots of the teeth that was also the level of isocenter and the titanium screw where the isocenter was set, and the other one was placed 1 cm above the isocenter, at the uppermost level of crowns (UL) with the same dimension as Perspex layers. A third film (3×4 cm) was placed at a slot on the concave side of the phantom.

Fig. 1
figure 1

Oral cavity phantom made of Perspex, Teflon, and teeth. The placement of EBT-3 films at two levels is shown. The red arrow shows the direction of the radiotherapy beam

Equipment

The dose was delivered using a compact linear accelerator (Elekta Ltd., Stockholm, Sweden) equipped with 6 MV photons. CT images were acquired on a syngo CT VC40 (Siemens, Shanghai Medical Equipment Ltd.) with a slice thickness of 1 mm and an in-plane resolution of 512×512 pixels. The dose calculation was accomplished by CCCS algorithm version 3.5 implemented in RayPlan version 7.0.0.19 (RaySearch Laboratories, Stockholm, Sweden) treatment planning system (TPS) that uses energy corrected effective density water to calculate dose-to-different-density-water [13]. The dose grid size was set to 1 mm. The density of the region of interest (ROI) can be manually set to a specific mass density ranging from a value of 10 -3 to 22.57 g/cm3.

Calibration of dosimetry films

Films were cut to 2 ×2 cm dimensions and exposed by doses of 0.5, 1, 1.5, 2, 3, and 4 Gy for calibration. Monitor unit (MU) values for all exposures were calculated. A 30×30×30 cm3 phantom was created in TPS and then MUs of different dose values were calculated at a depth of 5 cm of the central axis and field size of 10×10 cm2. For calibration, a 30×30×30 cm3 Perspex phantom was used with the same setup. For calibration validity, a 0.6 cc Farmer-type ionization chamber (PTW model TM 30013) was used at a 5-cm depth. Exposures were repeated 3 times for reproducibility. Mean doses to the central area of 0.014 cm2 of the films were read. All calibration films were scanned by ScanMaker 9800XL (Microtech CO., Hsinchu, Taiwan) after 48 h. The central area of 0.014 cm2 was selected for each film in ImageJ software (v1.52a, USA), the mean value of optical density (ODn) in the area was read and it was used to calculate NetOD = LOG10 (ODn – OD0) where the OD0 is the optical density of non-irradiated control film. Dose values were calculated by using a dose-optical density calibration histogram.

Planning

CT planning images of non-implanted and implanted phantoms were taken with a 1-mm thickness in a reclining position in which the row of teeth was along Z axis. Margins of films and tooth implants were contoured. The crown, abutment, and screw were separately contoured based on their dimensions https://www.medicalexpo.com/prod/tbr-implants-group/product-74246-724926.html. The 6 MV lateral beam, perpendicular to the row of teeth, was planned with SSD=95.1 and field size=10×10 cm2. The plan was run 3 times when densities of the material were not set for the implant components, when they were set to water 1 g/cm3, and when the densities of the components were set manually in TPS software. Hence, mass densities of the crown were set to 8.8 g/cm3, zirconium 6.51 g/cm3, and titanium 4.54 g/cm3. The rationale was the CT-to-electron density conversion of the planning system may not be reliable when high Z material is used [6]. The PTV of this plan was the representative tongue soft tissue and floor of the mouth.

Dosimetry

The films were put inside the phantom for radiation. The measurement was repeated 3 times for both setups of the phantom, i.e., with and without the implanted tooth. Dose profiles and mean doses of EBT-3 films were calculated using an in-house code in MATLAB R2019a (Math Works, MA, USA). Homogeneity indices (HI) were calculated using D2 -D98 /Dp×100 in which Dp is the prescribed dose, D2 is the value of dose in 2% volume, and D98 is the value of dose in 98% volume. The lower value indicates a more homogenous dose distribution within the volume.

Results

The HU value of the implant system saturated at 3071 uniformly. Consequently, visual discrimination of implant components was unattainable and we had to rely on the manufacturer’s dimensions from the catalog. After planning, a single 6 MV photon field when 200 cGy was prescribed to the isocenter. Calculations based on density conversion curves for normal teeth were not significantly affecting the MU calculations, as it was observed that the TPS calculated MU of the machine 238 for the non-implanted phantom, identical to the MU when the implant was set to water (W-plan). Noteworthy, M-plan possessed the highest MU that is explained by the beam attenuation effect of the implant. These values were 241 for the implanted phantom when implant material was not manually set or the default plan (D-plan), and 258 for implanted phantom when implant materials were set manually (M-plan). The relative dose uncertainties were calculated ±0.3% in measured and calculated dose, and additionally, the gray level value and location of some points (penumbra and center axis of the beam) in UL and isocenter area were estimated. The dose value of points was calculated by using a dose calibration plot in the penumbra and center axis of each area and compared with TPS as described in [14]. By using this method, the gamma index was equal to 1.8 at the level of isocenter and 2.1 at the level of crowns which both acceptably show the validity of measurement and calculated dose.

D1 (Max dose to 1% of the volume of the film) at the UL was 110.5% for W-plan, 119% for M-plan, and 122% for D-plan. The maximum calculated dose in the M-plan in the whole phantom was 123.5%. Average calculated dose to all films were higher in the implanted phantom in M-plan compared to the non-implanted phantom, particularly, this was 215.93 vs 192.25 at the level of crowns. In this plane areas with 115-119% maximum relative dose emerged at both sides of a ‘shadow’ streak with the implanted phantom (Fig. 2). Dose distribution at the level of crowns was more homogeneous (HI= 91.06%) for W-plan compared to D-plan and M-plan that HI were 97.96 and 98.66 respectively. Dose Volume Histograms (DVH)s of the film at the UL are shown in Fig. 3 which indicates that difference between histograms of D-plan and M-plan is minimal. Measured film dosimetry profiles at the UL in 3×5 cm areas and mean and Standard Deviation of doses calculated by MATLAB software are shown in Fig. 4. Table 1 shows calculated and measured doses of selected points of interest in 2 areas as examples.

Fig. 2
figure 2

Comparing dose distributions normalized to isocenter between 3 different settings for implant material at the level of crowns. The red arrow shows the direction of the photon beam. Areas with a solid color show isodose levels above 100%. Magenta 100%, rosewood 105%, lavender 110%, and fuchsia 115%

Fig. 3
figure 3

Comparing DVH of the whole film at the level of crowns in different plans when the implant system

Fig. 4
figure 4

EBT-3 relative dose profiles of the area of 3cm × 5 cm within the field at the level of crowns a with and b without dental implant

Table 1 Absolute and relative (percentage) calculated and measured dose by film dosimetry for point U

The concave area showed highest backscatter dose with D-plan (D1-2 = 125%), the second was M-plan (D1-2= 120%), and the least was W-plan (D1-2=114%). D1-2 for the non-implanted phantom was 115% at the concave area. As illustrated in Fig. 5, Dosimetry showed that an area of 0.81 cm2 received as high as 125-127% on the concave area when the implant was placed that was roughly 3-5% higher than max dose to the non-implanted phantom. Measured dose by EBT-3 films in implanted phantom was validated by calculated dose in implanted phantom when set to default setting in TPS. In this study there was 0.3-0.4% relative dose discrepancy between dosimetry and Rayplan calculations with density corrections for hot areas and the off-center POI in comparison to non-implanted phantom with three repeats of dosimetry. This level deemed acceptable with respect to statistical and systematic uncertainties of Gafchromic film dosimetry that is in order of 0.45% for relative dosimetry provided it is repeated five times [11].

Fig. 5
figure 5

EBT-3 Measured dose distribution at the concave area that represents the buccal surface. The legend shows a dose percentage relative to dose at isocenter a without and b with an implant. The Y-axis is along the concaveness and the X-axis is along the row of teeth

Discussion

The main findings of point and profile dosimetry of the representative oral soft tissue at the level of crowns were applying metal compositions (M-plan) to the planning system reduced hot areas of PTV and made the plan more resembling film dosimetry compared to (D-plan) that is when dose was calculated with HU-to-electron density conversion curves. In this regard, despite similar DVHs of D and M-plan, relative selected point dose (U) of the planar level of crowns (UL) in ‘implant shadow’ was lower in both M-plan and dosimetry compared to D-plan and W-plan. All these findings show that when inputs of metal kernels are not available, erroneous calculations mislead plan verification.

The results further showed that the backscattered electrons to the concave buccal surface caused an increase in dose to 125-127% (3% higher than baseline) measured with film dosimetry that was close to M-plan calculations. This was predictable as high-density materials induce backscatter electrons. backscatter dose calculations are not correctly predicted in modern CCCS TPSs [10]. However, simplified CCCS are supposed to underestimate the backscatter dose enhancement at proximal interface of an implant but overestimate the dose directly downstream to it [15, 16]. Interestingly, the backscatter dose was the highest in D-plan when compared to M-plan and dosimetry results. This means that the direction of scatter may not have been predicted correctly without metal kernels. Another finding was, relative dose to U was higher both in M-plan (0.3% higher) and measured by film dosimetry (0.2% higher), than the non-implanted phantom. This result shows that the implant had made a global dose increase because of extra scatter even in the shadow area and suggest that PTV near dental implants may unexpectedly receive higher mean dose. Also, areas lateral to the shadow streak showed overdose in M-plan, and as measured by film dosimetry. According to a Tumor Control Probability (TCP) prediction curves, small areas of overdose in the tumor volume does not improve TCP so has no added benefit. Instead, increased scatter dose can adversely increase dose to normal structures such as mucosa of oral cavity, and induce higher mucositis, or increase dose to parotid glands resulting in more frequent xerostomia [17]. However, very small volumes of under-dose can have an enormous effect on TCP [18]. This should be taken into account as non-homogenous dose distribution can lead to underdosed areas when dose levels are adjusted.

Moreover, in our study HU of planning CT saturated at 3071 in all components of implant system and its interfaces, and therefore, the contouring of components of the implant was not feasible before aided by catalogue of the manufacturer. Artifacts can inflate the size of the device and overestimate beam attenuation [10]. This is important as implant geometry can be concealed particularly in devices with different components. Several solutions for planning with presence of a metal device [19], such as metal artifact reduction (MAR) for planning CT scans that improve definitions of device boundaries all have had variable success rates for reducing systematic errors and uncertainties [20, 21]. This study did not use MAR but as we said we manually corrected contour geometry of components. Our results were similar to studies that have shown that accuracy of dose calculation improves when using material-specific energy kernels for applying depth-hardened photon beam in CCCS algorithms [1, 6, 22, 23]. The shape and distribution of energy deposition depends on the composition of interaction medium and the surrounding materials. Calculation algorithms that take into account for lateral heterogeneities, have acceptable accuracy for calculating inhomogeneous media [6, 24], nevertheless, in Rayplan CCCS calculations the energy deposition kernels are based on photon interactions and scatter in water, and these water-based kernels are scaled in dimension based on the local density defined by HU values of the neighboring voxels. Therefore, a water-based kernel does not correctly describe the shape of the true energy deposition in heterogenous media [24]. The difference between dose to water calculations and more accurate calculations that apply dose to medium with radiation transport in the medium (i. e. Monto Carlo simulation) can be largely depending on the mean ionization number of the consisting materials [13]. Default density scaling of CCCS water kernels may be inadequate, and a worse approximation encounters when the difference between its effective atomic number and water is larger [25]. Zirconium containing implants can induce more pronounced streaking artifacts when compared to pure titanium in CT images [3]. Different algorithms of TPSs including Rayplan have shown agreement with radiochromic film measurements in inhomogeneous low densities in a phantom of thorax [26]. For media including higher densities such as those with metal implants which the density and physical dimensions of the implant are used, CCCS methods could accurately calculate dose attenuation. This has also been investigated in Pinnacle CCCS versus Monaco Monte Carlo algorithms in series of head and neck CT set of real patients. When the material composition was used to modify the dose kernels, calculations improved. The authors suggested that density corrections should be considered for cases in which their PTV includes many slices affected by streaking artifacts [25]. Moreover, beam orientation, device location, and technique of radiotherapy affect the intensity of errors, as a study showed that calculation errors were less intense when an implant in a cylindrical phantom was exposed by VMAT rather than another IMRT plan for both uncorrected and corrected planning CT for metal artifacts [20]. Another study provided mixed results for the role of different MAR algorithms and metal kernels for reducing IMRT dose calculation errors. Interestingly some MAR algorithms increased errors remote from the metal device. The authors showed that unlike calculations in a titanium phantom with simple geometry, metal kernels of CCCS algorithm did not substantially improve dose calculation accuracy in anthropomorphic phantoms of clinical cases (<1% in spine phantom containing titanium rods and dental phantom with amalgam fillings). The authors explained that messy artifacts at metal interfaces had possibly reduced effects of kernels [10]. These data should be taken into account as although there are some surrogate calculations in interfaces but they might not surpass all errors such as contouring and global systematic errors. Knowledge of the geometry and material of device components provided by the implant dentist can be practically implicated by the radiotherapy team and enhance calculations. However, streaking artifacts that are remote from the implant may still affect dose predictions.

This study has some limitations. Firstly, point-to-point comparison is prone to errors since CT scans are reconstructed and contoured films may not replicate their actual volumes. Secondly, only a single lateral field was used to evaluate the effect of implant, that is so simplified for radiotherapy planning. There is less known for the yield of calculation errors on case-by-case basis. It is supposed that complex plans will amend effects of shadow and back scatter dose while the relative effects of each field remain. In future, the effect of high-density materials in adaptive simulations of real cases. (i. e. multiple dental fillings and implants) on dose calculations should be critically evaluated in more sophisticated plans.

Conclusion

In conclusion, the results demonstrate that high Z materials when included in the radiotherapy field of cancers of oral cavity could cause dose perturbation that interferes with radiotherapy planning. The results of dose measurements of the oral cavity and backscatter suggests that the composition and structural dimension of dental restorations whenever available may be provided for the radiotherapy team and considered in density corrections for oral cancer radiotherapy planning.