Introduction

Recently, a time-of-flight (TOF) positron emission tomography (PET) system was introduced by Philips Healthcare, with digital photon counting technology using silicon photomultipliers. The replacement of conventional photomultipliers by digital detectors, including the implementation of single-photon avalanche photodiodes, provides true digital photon counting without the need of additional analogue-to-digital conversions [13]. Moreover, the detector elements and the scintillator crystals have equal sizes which enables one-to-one coupling. Acceptance tests on performance characteristics showed that this digital PET provides a higher timing resolution and improved spatial resolution, as compared to state-of-the-art analogue PET using conventional photomultipliers [4, 5]. In clinical practice, digital PET may provide a higher image quality and improved small lesion detection and quantification [6].

PET/computed tomography (CT) scanning, using fluor-18 fluordeoxyglucose (FDG), has an important role in tumour imaging for patients with cancer. There is a trend towards standardization in FDG-PET scanning to allow quantitative comparisons of FDG-uptake parameters across patients, scanners and medical centres [7]. To support standardization between scanners and medical centres, the European Association of Nuclear Medicine (EANM) has published guidelines on FDG-PET tumour imaging [7, 8]. Furthermore, the EANM launched the EANM research Ltd (EARL) to promote nuclear medicine research and multi-centre studies. EARL has developed an accreditation programme for tumour imaging with FDG-PET/CT [9].

In clinical practice, the EARL FDG-PET/CT accreditation specifications are widely implemented. These specifications, which are primarily about activity concentration recovery coefficient (RC) measurements on PET images, are based on analogue PET systems using conventional photomultipliers [7, 8]. Intrinsically, higher RCs may be expected using digital PET, due to improved spatial and time-of-flight resolution compared to other non-digital, but state-of-the-art systems [5]. Our aim was to evaluate if a recently introduced TOF PET/CT system with digital photon counting technology can fulfil EARL requirements as well.

Method

Phantom study

We have performed a phantom study using a NEMA IEC-61675-1 NU2-2001 image quality phantom (IQ phantom) with six fillable spheres (10, 13, 17, 22, 28 and 37 mm diameter). According to the EANM guidelines [7], the IQ phantom was filled with FDG-activity, with a sphere-to-background ratio of 10:1. Using a TOF PET/CT system with digital photon counting technology (Philips Healthcare) [6], we performed a PET scan of one bed position with a scan duration of 10 min. Additionally, a CT scan was acquired for attenuation correction. Prior to our measurements, the digital PET was calibrated with FDG and verified to be within an offset of 2%, using the method as described in [7].

PET reconstructions

We have performed six default TOF PET reconstructions, using blob-based ordered-subsets expectation maximization (OSEM) [10], with three voxel sizes and with/without point spread function (PSF) modelling, which corrects for partial-volume effects in PET images. When incorporating PSF modelling, we used a noise regularization kernel of 6 mm full-width at half-maximum and 1 PSF iteration. For each voxel size, we used a fixed number of iterations and subsets, as recommended by the manufacturer.

  1. 1.

    4 × 4 × 4 mm3, with 3 iterations and 15 subsets, without PSF

  2. 2.

    4 × 4 × 4 mm3, with 3 iterations and 15 subsets, with PSF

  3. 3.

    2 × 2 × 2 mm3, with 3 iterations and 17 subsets, without PSF

  4. 4.

    2 × 2 × 2 mm3, with 3 iterations and 17 subsets, with PSF

  5. 5.

    1 × 1 × 1 mm3, with 3 iterations and 9 subsets, without PSF

  6. 6.

    1 × 1 × 1 mm3, with 3 iterations and 9 subsets, with PSF

On each reconstructed PET dataset, we calculated mean and maximum activity concentration recovery coefficients (RCmean and RCmax) for all phantom spheres, according to EANM guidelines [7]. Next, we compared our RC results with EARL accreditation specifications [9].

To evaluate reconstruction settings for digital PET that meet the requirements for EARL accreditation, we performed additional reconstructions. We selected the reconstructed PET dataset whose RCs fitted best to the EARL accreditation specifications and added a 3D Gaussian post-smoothing filter with varying kernel widths of 1–7 mm, using standard vendor software. Again, RCmean and RCmax were measured for all phantom spheres and compared to EARL accreditation specifications.

Results

Figure 1 shows RCmean and RCmax results for all spheres, for each of the six default PET reconstructions. For each reconstruction, at least one RCmax value was above EARL accreditation specifications. The PET reconstruction without PSF modelling and using 4 × 4 × 4 mm3 voxels showed RCs nearest to EARL requirements. For this specific reconstruction, only RCmax for the 10-mm sphere was above the EARL limit.

Fig. 1
figure 1

RCmean and RCmax values for all phantom spheres, as compared to EARL minimal and maximal accreditation specifications. a RCmean for six default TOF PET reconstructions. Only for a PET reconstruction using 4 × 4 × 4 mm3 voxels without PSF modelling, RCmean values were all within accreditation specifications. b RCmax for six default TOF PET reconstructions. For all reconstructions, at least one RCmax was above maximal accreditation specifications

Table 1 shows the impact of an additional Gaussian post-smoothing filter with 1 to 7 mm kernel widths, on RCmean and RCmax in a PET reconstruction using 4 × 4 × 4 mm3 voxels, without PSF modelling. As shown in Fig. 2, EARL accreditation specifications for RCmean and RCmax can be achieved with digital PET using filters with kernel widths of 2 to 4 mm.

Table 1 The impact of a Gaussian post-smoothing filter with a kernel width of 1–7 mm on RCmean and RCmax for a TOF PET reconstruction with 4 × 4 × 4 mm3 voxels, without PSF modelling. RCs within EARL accreditation specifications are marked in italic style
Fig. 2
figure 2

Impact of a Gaussian post-smoothing filter using a kernel width of 2, 3 and 4 mm on RCmean (a) and RCmax (b) for a TOF PET reconstruction with 4 × 4 × 4 mm3 voxels without PSF modelling. For all phantom spheres, both RCmean and RCmax fulfilled EARL accreditation specifications

Conclusion

PET with digital photon counting technology typically shows an activity concentration recovery coefficient above EARL specifications, especially for small objects. To meet EARL standards, a TOF OSEM reconstruction without PSF modulation, with 3 iterations, 15 subsets, 4 × 4 × 4 mm3 voxels and a Gaussian post-smoothing filter with a kernel width of 2 to 4 mm can be used.

Discussion

To meet EARL standards for PET with digital photon counting technology, the use of relatively large 4 × 4 × 4 mm3 voxels and a post-smoothing filter is recommended. With smaller voxel sizes and/or PSF modelling, RCs in our study were above EARL specifications. This has been demonstrated for state-of-the-art analogue PET systems as well [11, 12]. With the introduction of advanced reconstruction algorithms (e.g. using small voxels or incorporating PSF modelling), eventually combined with new digital PET technologies, EARL specification updates may be needed in the future. Under the assumption that the availability and presence of PET scanners using older technology will decrease, a way to maintain the uniformity across modern PET cameras is to increase both lower- and upper RC EARL specifications, especially for small spheres. Furthermore, the use of smaller phantom spheres, for example as available in a micro phantom that we used in a previous study [11], may be warranted to be able to compare reconstruction algorithms for smaller sphere sizes and to harmonize the quantification of small lesions across scanners.

Besides, given the high RCs that can be achieved with digital photon counting technology combined with advanced reconstruction settings, it might be appropriate to perform multiple PET reconstructions for different purposes. Next to an EARL-approved reconstruction to perform quantitative analyses, a high-resolution small-voxel PET reconstruction could be made for visual evaluation and optimal small lesion detection [8, 11, 13].

This short communication focused on determing PET reconstruction settings to fulfil EARL RC specifications. However, to obtain the EARL accreditation, these reconstruction settings should be chosen to meet both RC requirements and specifications for the calibration QC [7], and EANM guidelines should be fully implemented in clinical practice [7, 8].