Background

Breast-conserving therapy followed by whole breast irradiation (WBI) equals radical surgery in terms of overall survival with limited long-term toxicity [13]. Even though the role of radiotherapy is well established, its use is sometimes challenged owing to accessibility, equipment and cost issues. Hypofractionated radiotherapy, such as accelerated partial breast irradiation (APBI), could be a response to certain WBI drawbacks. APBI allows to shorten treatment time and to limit the exposure of organs at risk (OAR) with a putative equivalent efficacy compared to standard fractionated WBI in patients with early breast cancer [412]. APBI is of special interest in elderly or frail patients who barely tolerate standard-course radiotherapy. A wide variety of APBI techniques are used such as brachytherapy (interstitial needles or balloon-based), intraoperative or external beam radiotherapy, especially 3-dimensional conformal radiotherapy (3D-CRT) [13]. The latter is mainly performed thanks to its availability, but the optimal delivery technique remains to be determined [6]. Recent studies have raised concerns regarding possible higher toxicity rates after APBI. In particular, after balloon catheter brachytherapy, a significant higher risk of rib fracture, breast pain and fat necrosis have been reported [14]. After external beam radiotherapy (intensity-modulated radiation therapy [IMRT] and/or 3D-CRT), a fair aesthetic outcome has also been noted, in specific clinical, technical and dose conditions [15, 16]. On the contrary, another group has published improved toxicity results when using IMRT APBI as compared to WBI plus tumor bed boost [17, 18]. These results call for new clinical studies on other delivery techniques of APBI, which might entail different types and rates of toxicity.

Volumetric-modulated arc therapy (VMAT) is an attractive IMRT technique that enables a fast delivery and an improved efficiency. It has been evaluated in a broad spectrum of tumours but has not been extensively studied for APBI [1921]. Qiu et al. reported dosimetry feasibility of the VMAT approach compared to 3D-CRT treatment planning in 8 patients with breast cancer and showed that VMAT was more efficient, with equivalent or improved dose conformity and lower doses to OAR [22]. Essers et al. compared VMAT and 3D-CRT in a larger series of 37 patients [23] No clinical data has been yet reported with VMAT APBI. Here, we present the feasibility and early clinical results of BC patients treated with VMAT APBI.

Methods

Patient selection

Our study design was validated by our institutional ethical board (Comité d'éthique) and informed consent was obtained from all patients before treatment.

Nine patients with early breast cancer (ten lesions: 5 right-sided, 3 left-sided and 1 bilateral breast cancer) were prospectively recruited from May 2011 to July 2012. Patients’ inclusion criteria were based on the guidelines of both work-task forces ASTRO and GEC/ESTRO for APBI indications [24, 25]. All patients were older than 50 years and had undergone breast-conserving surgery with free margins followed by APBI using VMAT. They all had pT1N0 invasive ductal carcinoma, grade 1–2 SBR score, positive hormonal receptor, and negative HER2 expression. The patients were excluded in case of metastatic disease. Patients’ characteristics are listed in Table 1. All patients were elderly (above 70) or presented with serious comorbidities including cardiovascular disease, chronic respiratory insufficiency or major scoliosis.

Table 1 Patients characteristics

Acquisition and simulation

The patients underwent computed tomography (CT)-based virtual simulation (CT Simulator, General Electric, Cleveland, OH) with 2.5 mm thick slices obtained at 2.5 mm intervals. Patients were in supine position with arms above the head (arms and knee support, Sinmed, The Netherlands) and within a personalized immobilization device (Mold Care, Bebig). The isocenter was set in the middle of the surgical tumour bed.

Contouring and volume definition

Target volumes and OAR were delineated on CT scan slices. The gross tumour volume and clinical target volume (CTV) were considered as equal and included the surgical bed, the surgical clips placed within the lumpectomy cavity, and/or the seroma. The planning target volume (PTV) was defined as an isotropic circumferential margin of 18 mm surrounding the CTV and excluded the first 5 mm under the skin surface, thoracic wall, ribs and pectoral muscles. OAR were automatically or manually contoured: ipsilateral and contralateral lungs, heart, ipsilateral and contralateral breast (using wires), thyroid, oesophagus, and humeral heads.

Treatment planning using RapidArc®

A maximum dose rate of 600 MU/min and 6 MV photon beams were used. The optimization process started with the constraints obtained with the IMRT plans. RapidArc® (RA, Eclipse software version 10.0.28, Helios, Varian, Palo Alto, CA) was delivered with two partial coplanar arcs (less than 240° of rotation) sharing the same isocenter and optimized independently and simultaneously. These two arcs were delivered with opposite rotations (clockwise and counterclockwise) so that the off-treatment time between the two beams was minimized to about 25 s. The field size and collimator rotation were determined using the automatic tool from Eclipse software to encompass the PTV. The first clockwise arc used a 45° collimator rotation in order to avoid the tongue-and-groove effect. The second arc rotated counterclockwise with a collimator rotation of 360° - X° (X° corresponding to the rotation of the collimator for the first arc). To improve results, we modified optimization constraints and priority factors of RA plans during optimization. These parameters were modified with regard to the DVH results for each patient.

Dose prescription, dose constraints, dosimetric evaluation and dose delivery

The radiotherapy was prescribed in fractions of 4.0 Gy “bis in die” (b.i.d.) over 5 consecutive days (40.0 Gy in total), with a minimal 6-hour interval between each fraction. 99 % and 95 % of PTV were to receive 38.0 Gy and 40.0 Gy, respectively. The total dose delivered wasn’t to exceed 110 % of the prescribed dose. Volumes receiving more than 110 % of the prescribed dose (D) were considered as “hot spots”. The homogeneity index (HI) was defined as: HI = (D2–D98 %)/D median. Vx% was defined as the proportion of the total structure volume that received x% of the prescribed dose. To limit the ipsilateral lung exposure, volumes receiving 20.0 Gy (V20Gy) had to account for less than 3 % of the total structure volume; less than 10 % and 20 % for V10Gy and V5Gy, respectively. Regarding the contralateral lung exposure, respective thresholds for V20Gy, V10Gy and V5Gy were 1 %, 2 % and 3 % of the lung volume. Moreover, to limit heart exposure, V20Gy and V5Gy were not to exceed 1 % and 70 % of the heart volume. Finally, exposure of the ipsilateral breast was expressed as V50% and V100%. The mean and maximal doses delivered to the contralateral breast were also determined.

Image-guided radiotherapy with cone-beam computed tomography (CBCT) using soft-tissue matching was performed before each fraction. The γ-index methodology was used to validate the planned delivery with a minimum of 95 % of the points meeting a 3 %/3 mm criterion.

Toxicity assessment

A clinical examination was performed before APBI. Acute and late adverse events were assessed 1, 2, and 6 months after APBI completion, then every 6 months until 5 years, and then annually, according to the National Cancer Institute Common Terminology Criteria for Adverse Events (NCI-CTCAE) version 4.0. The main expected toxicities included: breast pain, breast oedema, skin erythema, skin desquamation, radiation pneumonitis, telangiectasia, fat necrosis, skin pigmentation, skin atrophy, and breast fibrosis.

Results

Treatment planning

Mean PTV and breast volumes were 100.0 cm3 (range 38.9 cm3 to 219.5 cm3) and 899.7 cm3 (range 390.7 cm3 to 1932.3 cm3), respectively. The mean dose encompassing 95 % of the PTV accounted for 99.7 % of the prescribed dose (range 99.4–99.9 %). Hot spots accounted for 0.3 % of the PTV (range 0.0–1.4 %). Mean HI was 5.6 % (range 4.0–8.5 %).

The Fig. 1 shows typical dose distribution using RA for right breast cancer (a) and bilateral breast cancer (b). Mean dose-volume histograms are presented in Fig. 2 (n = 10). The main dosimetric results are presented in Table 2 for ipsilateral breast, contralateral breast, heart, and ipsilateral and contralateral lung. An average of 580 monitor units (MU) was delivered with RapidArc® (range 473 MU to 655 MU). The mean treatment time was 3.2 min.

Fig. 1
figure 1

Typical dosimetric results obtained with accelerated partial breast irradiation using volumetric-modulated arc therapy on two patients: one right breast cancer (a) and one bilateral cancer (b). a (upper left) dose distribution in axial view, dose colorwash from 38.0 Gy to maximal dose at 45.3 Gy; (upper right) corresponding dose-volume histograms for this patient: planning target volume (PTV) in red, right breast minus 3 mm above the skin in blue, right breast in yellow, right lung in green, heart in purple, left lung in red; (lower left) dose distribution in coronal view, dose colorwash from 38.0 Gy to maximal dose at 45.3 Gy; (lower right) dose distribution in sagittal view, dose colorwash from 38.0 Gy to maximal dose at 45.3 Gy. b (upper left) dose distribution in axial view, dose colorwash from 40.0 Gy to maximal dose at 45.8 Gy; (upper right) corresponding DVH for this patient: left PTV in red, right PTV in purple, spinal cord in orange, heart in yellow, right lung in green, left lung in red; (lower left) dose distribution in coronal view, dose colorwash from 40.0 Gy to maximal dose at 45.8 Gy; (lower right) dose distribution in sagittal view, dose colorwash from 40.0 Gy to maximal dose at 45.8 Gy

Fig. 2
figure 2

Mean dose-volume histograms for the ten lesions treated: planning target volume in violet, ispilateral breast in pink, contralateral breast in blue, homolateral lung in cyan, controlateral lung in yellow, heart in brown

Table 2 Main dosimetric results regarding the protection of organs at risk

Acute and late toxicity

As summarized in Table 3, acute toxicities were of grade 1 or less in all patients at any early time point (i.e. at APBI completion, 1 month and 2 months). Most observed symptoms were breast pain, oedema and erythema. There were no grade 2 or more adverse events. No treatment discontinuations occurred.

Table 3 Acute and late toxicitiesa that occurred within three months after APBI with VMAT (acute) and up to 26 months of follow-up (late) (n = 10)

Late toxicities were collected with a median follow-up of 26 months (range 6–37 months). Only three grade 1 fibrosis and one grade 1 breast pain were observed. No toxicities of greater grade were reported.

Oncological outcomes

With a median follow-up of 26 months, no local or distant relapses were diagnosed.

Discussion

VMAT for APBI has not been extensively studied in the literature. We present here the first feasibility study including clinical results in this setting. VMAT, as a rotational IMRT technique, has the potential to deliver highly conformal and homogeneous dose to the targeted volumes. However, because of its rotational nature, it is thought to be less efficient at low and medium doses, thereby irradiating a larger OAR volume. On the one hand in our study, such dispersion was not observed and we obtained a dose distribution to the ipsilateral lung, contralateral lung and heart similar to the one reported in 3-dimensional (3D) conformal APBI [16, 26]. On the other hand, we should remain aware of the low but not negligible dose delivered to the contralateral breast with this technique (Table 2). Indeed, this effect has not been reported when using tangential field 3D conformal APBI. Moreover, the ipsilateral breast exposure to a significant radiation dose seems to be a significant risk factor of unacceptable toxicity and impaired cosmesis [15]. The ipsilateral breast is considerably less irradiated with VMAT as compared to other external beam radiotherapy techniques: mean V50% and V100% being respectively of 25.8 % and 10.9 % in our study versus 47.9 % and 27.2 % in the IMRT study by Jagsi et al.[15], and 44.1 % and 23.8 % in the 3D-CRT study by Bourgier et al. [16, 26] Our results remain competitive when compared with those obtained in other studies on VMAT for APBI published by Qiu et al. [22] and Essers et al. [23].: they reported ipsilateral breast V50% of 45.9 and 19.7 % and V100% of 20.9 % and not reported, respectively. However, in our study, a greater maximal dose was delivered to the contralateral breast: 3.0 Gy [1.3–5.8 Gy] versus 2.56 Gy [0.46–4.83 Gy] in the study by Qiu et al., but Essers et al. reported a greater maximal dose of 4.6 Gy [0.1–9.4 Gy]. Regarding the ipsilateral lung, we report lower V10Gy and V20Gy than Qiu et al. (respectively 1.6 % [0.0 %–10.4 %] vs. 2.0 % [0.0 %–5.0 %] and 0.27 % [0.00 %–2.67 %] vs. 0.5 % [0.0 %–2.0 %]), but a higher V5Gy (6.2 % [0.0 %–19.9 %] vs. 5.8 % [0.0 %–11.2 %]). Finally, Essers et al. reported a higher ipsilateral lung V5Gy (10.4 % [0.0 %–40.3 %]). Another factor that might be involved in the onset of late fibrosis and retraction is the treated volume and/or the PTV to whole breast ratio. In the study by Jagsi et al., the mean PTV was 185.8 cm3 compared to 123 cm3 in the study by Livi et al., 117 cm3 in the study by Bourgier et al. and 100.0 cm3 in our study [17, 18]. These differences in PTV can be explained by different surgical and remodelling techniques, entailing large variations in surgical bed and seroma cavity volumes. Indeed, APBI may not be the most suitable method to treat patients with large lumpectomy cavities.

Finally, PTV homogeneity is believed to be an important factor for cosmetic results [27]. Our homogeneity index was 5.6 [4.0–8.5], which compares favourably to 3D APBI studies (e.g. 9.7 [6.2–15.1] in the study by Bourgier et al.).

Regarding exposure of OAR, especially lung, contralateral breast and heart, no comparison has yet been made between VMAT and static field IMRT. Whether the higher modulation possibilities of VMAT could improve OAR protection remains to be proven.

Acute toxicity was acceptable as all patients had grade ≤ 1 toxicities. With a median follow-up of 26 months, late toxicity was low, most of the patients experiencing none; and no grade 2 late toxicities occurred. Nevertheless, a longer follow-up and a larger cohort of patients are warranted to consider VMAT as a safe APBI modality.

Our study has some limitations, including the low number of patients and the limited follow-up time. This is mainly due to the fact that APBI is not a standard treatment outside clinical trials in France; therefore we mainly recruited patients presenting clinical features that would not allow standard radiotherapy, such as advanced age, respiratory insufficiency or cardiovascular disease. The treatment tolerance seemed to be especially good in this population. This limited follow-up allowed us to consider the clinical toxicity only without drawing any conclusion regarding tumour control. If longer-term studies confirm its efficacy and tolerability, VMAT APBI might become a therapeutic alternative for patients otherwise not treated.

Conclusions

In this study performed in nine patients with breast cancer, VMAT offered a good OAR sparing while maintaining PTV coverage within acceptable levels for APBI. The early evaluation of oncological outcomes was promising.