Background

Approximately 280.000 heart valve prostheses are implanted worldwide each year; approximately 50% of them are mechanical and 50% bioprosthetic. Due to the medical development and increasing life expectancy, the number of surgeries and interventions is expected to rise [1, 2]. Currently, surgical aortic valve replacement with biological or mechanical prostheses is the standard of care in most cases of aortic stenosis [3, 4]. In patients with high risk to the surgical procedure, transcatheter aortic valve implantation (TAVI) can be considered [3, 5, 6]. Most prostheses have metal compounds in varying amounts and compositions [7].

After valve replacement, the evaluation of prosthetic function and potential prosthesis-related complications is essential [1, 6, 8]. The transthoracic echocardiography (TTE) is the first-line modality for this assessment [9, 10], but the validity of TTE may be limited due to the physical condition of the patient, especially in a postoperative setting. Furthermore, TTE is known to have a high observer dependency [1113]. Cardiovascular Magnetic Resonance (CMR) is independent of the patient’s anatomy and less observer dependent [13]. Flow measurement with phase contrast (PC) sequences in CMR is used for evaluation and quantification of heart valve disease; it is regarded as robust and valid [11, 14, 15]. In biological valve prostheses, the direct assessment of the orifice area and the analysis of the flow pattern using CMR have already been demonstrated [1618].

In patients with heart valve prostheses, the metallic compounds often evoke image artefacts, such as phase alterations, image distortion, or even signal loss due to the local distortion of the magnetic field [19].

The purpose of this study is to investigate, whether metal-containing heart valve prostheses have influence on PC based flow measurements.

Methods

The study was performed with a clinical 1.5 T MR system (MAGNETOM Avanto, Siemens Healthcare GmbH, Erlangen, Germany) equipped with a 6-channel body array coil in combination with a spine array coil.

Phantom setup

We used a custom-built closed-circuit flow phantom [20]. It consists of a semi-rigid tube system with a pump generating adjustable constant flow and an acrylic pipe resembling the dimensions of the native aortic root. This acrylic pipe could hold different prostheses. The system was filled with a mixture of glycerol (40%) and water (60%) to approximate blood viscosity and relaxation times.

The flow phantom was inserted in a water-filled container. The acrylic pipe was placed in the coronal plane 45° oblique between the head-foot and left-right axes to imitate the alignment of the ascending aorta in-vivo. The setting is illustrated in Fig. 1.

Fig. 1
figure 1

Setting of the MR flow phantom for in-vitro assessment of heart valve prostheses. a. Scheme of the experimental setting showing the closed circuit flow phantom as placed in the scanner. b. Scheme of the acrylic pipe and positioning of the measurement slices (S1-S12). c. Coronal magnitude image of the acrylic pipe, proximal and distal parts of the tube system and positioning of the measurement slices. The arrow demonstrates the direction of flow

To assess background phase errors, all imaging slices were obtained with the phase contrast sequence while the pump was switched off. To avoid unintended flow in the water-filled container, we performed the first scan after 15 min rest following each table movement. Flow measurements in the phantom were performed with the synthetic metal-free aortic valve, the prosthesis I, II and III. Subsequently, the measurements were repeated in the same order with the pump switched on at fixed submaximal power. The flow output varied due to the different orifice areas of the valves.

Synthetic aortic valve

A 3D-printed aortic valve shaped synthetic inlay (Fig. 2) served as the metal-free model. It approximated the prostheses with an orifice area of about 2.0 cm2 (Table 1). Flow measurements were compared between the metal-free aortic valve model and the metal-containing prostheses to identify metal-related influences.

Fig. 2
figure 2

Objects investigated in flow phantom. a. Metal-free aortic valve shaped inlay, containing solely plastics. b. Biological aortic valve prosthesis, containing metal-stent. c. Biological aortic valve prosthesis, containing metal-stent. d. Mechanical aortic valve prosthesis, containing carbon

Table 1 Detailed schedule of material components

Prostheses

We investigated three metal-containing aortic valve prostheses from different manufacturers (Fig. 2). Prosthetic sizes were chosen to yield similar orifice areas.

Prostheses I and II are biological valve prostheses consisting of bovine pericardium mounted on a stent. Prosthesis III is a mechanical bileaflet valve.

The details of material composition based on manufacture’s data are listed in Table 1. All investigated prosthetic types are in clinical use for surgical aortic valve replacement and are regarded as MRI-conditional [10, 2123].

In-vivo example

To illustrate the effect in vivo, we have added single cases with identical types of aortic valve prostheses and a healthy volunteer. Echocardiography was performed within last 6 months and revealed an adequate valve function as well as a non-dilated ascending aorta.

Additionally one healthy volunteer was scanned as an in-vivo control reflecting the synthetic aortic valve of the phantom trial.

The protocol was similar to the phantom study.

MR measurements

The following scan protocol was applied in all prostheses. Localizer scans were performed to locate the aorta model. B0 maps were acquired in the coronal long axis of the model. Flow measurements were performed in 12 slices transversal to the flow direction.

B0 mapping

B0 maps in coronal planes were acquired in the absence of flow to reveal local magnetic field inhomogeneity. A multi-echo gradient echo sequence with a total of 5 echoes was used. Sequence parameters were: repetition time TR = 50.00 ms, echo time TE1 = 3.86 ms, TE2 = 9.12 ms, TE3 = 14.38 ms, TE4 = 19.64 ms, TE5 = 24.90 ms, slice thickness 5 mm + 1 mm gap, flip angle 15°, voxel size 2.0 × 2.0 × 5.0 mm3.

B0 maps were computed using MatLab (The MathWorks, Natick, MA, USA).

Quantitative B0 profile plots along the long axis of the flow phantom were generated. Therefore, an intraluminal long axis slice was set through the level of the prosthesis (demonstrated in Fig. 3 top right).

Fig. 3
figure 3

Results of B0 mapping. a. Coronal images of B0 maps, coloured scale in Hertz. b. Corresponding profile plots showing the quantitative magnetic field distortion from B0 in Hertz. c. Corresponding coronal magnitude images. White dashed line top right: Intraluminal positioning of slice for generating profile plots in MATLAB. d. Description of the magnetic field distortions according to quantitative score

To objectively assess the visual impressions from magnitude images and B0 maps, we used a quantitative score to describe the local magnetic field distortion. The score was defined by quantifying the distortions of the local magnetic field in Hertz (Hz) derived from the profile plots seen in Fig. 3. Magnetic field distortions from B0 of less than 50 Hz were graded as ‘none’, from 51 to 150 Hz as ‘mild’, 151 to 250 Hz as ‘moderate’ and more than 250 Hz as ‘severe’.

PC flow measurement

A conventional segmented gradient-echo-based PC cine sequence was applied. Sequence parameters were: temporal resolution 47.7 ms, TE 2.3 ms, TR 5.9 ms, slice thickness 5.5 mm + 1.1 mm gap, flip angle 30°, inplane resolution 1.9 × 1.9 mm2, GRAPPA reduction factor 2.

To maintain sequence parameters consistent throughout the measurements, but also to avoid possible phase aliasing due to the different maximum flow velocities in the prostheses, velocity encoding was set to 500 cm/s for all measurements.

A 60-beats-per-minute ECG simulator was used as acquisition trigger.

PC images were acquired at 12 positions: Slice 1 (S1) was set perpendicular to the tube and 40 mm proximal to the level of the prosthesis. This slice was defined as reference flow level, where constant laminar flow was assumed. S2 was set at the level of the prosthesis and was positioned to the isocenter of the scanner. Ten more positions were defined (S3-S12) based on slice thickness 5.5 mm +1.1 mm gap distal to the level of the prosthesis (see Fig. 1). S1 and S3 to 12 were not acquired in the isocenter to avoid table movement and therefore fluid motion in the water-basket.

Due to anatomical reasons we could not define a reference measurement point in-vivo similar to the phantom. So the in-vivo measurements start at the area of the aortic valve with slice 2 (S2) and amount 11 measurement positions.

PC measurements were repeated three times per position and flow results were averaged.

PC images were analysed using cvi42 Version 4.1.5 (Circle Cardiovascular Imaging, Calgary, Canada). Segmentation of the tube was done in the magnitude image. Contours were automatically propagated into all temporal phases of the phase image and manually adapted if necessary.

Statistical analysis

Calculations were performed using SAS 9.2 (SAS Institute Inc., Cary, NC, USA). Figures were prepared using R Version 3.2.1 [24] with packages ggplot2 [25] and RColorBrewer [26]. Least-square mean flow deviations from reference flow were calculated along with 95% confidence intervals (CI) in a statistical linear mixed model with the flow at the reference level (S1) as co-factor. The sensitivity analyses based on non-parametric models supported the outcomes of the linear mixed model and are not further displayed. Other factors were the distance and the type of prosthesis. Due to the small sample size, the statistical linear mixed model was repeated based on ranks to account for any non-normality of the data as sensitivity analysis to check for the robustness of the assumption of normality. The flow deviations were presented graphically per type of prosthesis. In addition, the distance to the level of the prosthesis was displayed along with equivalence limits of ±15% and ±20%. The equivalence limits were calculated as mean flow at the reference level ± the defined percentage limit of clinical irrelevance. As no validated or established limit of equivalence is published, two reasonable limits are reported here.

Significance in this setting is defined as a deviation of the flow values from reference flow (S1) of less than 15% or 20% respectively (range of equivalence). Positions in that range are described as significant, i.e. showed equivalence referred to the reference flow. Conversely no equivalence can be concluded for values out of this range. Convergence of flow values is defined as mean flow difference of less than 15% to reference flow and nearly identical flow values beyond that distance.

Results

All PC image series were successfully acquired and evaluable.

B0 mapping

The B0 maps in Fig. 3 demonstrate that depending on the type of valve, the local magnetic field distortions varied. In the synthetic aortic valve, no relevant distortion was detected. In contrast, the B0 maps of the prostheses showed moderate and severe distortions. The quantitative aberration from B0 magnetic field as an expression of distortion is shown in Fig. 4. It was 27 Hz in the synthetic aortic valve, which was classified as not relevant according to the score. In contrast, the aberration was 230 Hz in biological prosthesis I and 170 Hz in the mechanical valve III, which were classified both as moderate. In biological prosthesis II, the aberration was 280 Hz, which was classified as severe distortion.

Fig. 4
figure 4

Quantitative aberration from B0 magnetic field in all investigated objects. Deviation in Hertz (Hz) derived from B0 maps and profile plots. Corresponding table showing positive and negative maximum of field aberration in Hz over the whole investigated distance. The numbers within the bars indicate the absolute range of deviation in Hz, as it is used for the quantitative score

PC flow measurement - flow phantom

As seen in Table 2 and Fig. 5, the synthetic aortic valve showed a maximum absolute flow deviation over the whole distance of 46 ml/s, corresponding to 15% of the baseline flow. In bioprosthesis I, the flow deviation amounted 88 ml/s, corresponding to 29% of the baseline flow. Bioprosthesis II showed 154 ml/s flow deviation, corresponding to 53% of the baseline flow. In the mechanical prosthesis, a flow deviation of 140 ml/s was observed, corresponding to 35% of the baseline flow. There was a convergence to baseline flow after 19.8 mm (3 slice distances) distal to the bioprosthesis I, in bioprosthesis II and the mechanical prosthesis after 13.2 mm (2 slice distances).

Table 2 Results of Phase contrast-based flow measurement
Fig. 5
figure 5

Flow deviation. Deviation of the assessed flow from reference flow over distance. Dark grey: range of equivalence 15%. Light grey: range of equivalence 20%. Black dashed lines: 95% confidence interval

PC flow measurement - in-vivo example

As seen in Fig. 6b) the volunteer showed a maximum absolute flow deviation over the whole distance of around 18 ml/s with a maximum mean flow of 154 ml/s in the area of the native aortic valve (S2 compared to phantom) and a minimum mean flow of 136 ml/s 33 mm distal to the area of the aortic valve (S7 compared to phantom).

Fig. 6
figure 6

PC flow measurements in-vitro and in-vivo including a synthetic aortic valve model, two biological and one mechanical aortic valve prosthesis. a. Flow measured in-vitro over the distance to the valve. b. Flow measured in-vivo over the distance to the valve. c In-vivo: Long axis view (Cine-imaging, steady state free precession) showing the left ventricular outflow tract and the position of the aortic prosthesis (white arrow). d In-vivo: cine imaging of the aortic valve location (left) and ascending aorta in the area of the suspected restored flow 26.4 mm distal to the aortic valve (right) for each subject (columns)

In bioprosthesis I, the flow deviation amounted 28 ml/s with a maximum mean flow of 71 ml/s 13.2 mm distal to the area of the prosthesis (S4 compared to phantom) and a minimum mean flow of 42 ml/s 59.4 mm distal to the area of the aortic valve (S11 compared to phantom).

Bioprosthesis II showed 65 ml/s flow deviation with a maximum mean flow of 123 ml/s in the area of prosthesis (S2 compared to phantom) and a minimum mean flow of 59 ml/s 26.4 mm distal to the area of the aortic valve (S6 compared to phantom).

In the mechanical prosthesis III a flow deviation of 64 ml/s was observed with a maximum mean flow of 100 ml/s 6.6 mm distal to the prosthesis (S3 compared to phantom) and a minimum mean flow of 37 ml/s in the area of the prosthesis (S1 compared to phantom).

The flow volume curves of patient and flow phantom are showing a similar shape as seen in Fig. 6. A normalization of flow volume could be identified approximately 26 mm distal to the prostheses level in all in-vivo tests. At this position no artefacts could be identified anymore in the corresponding anatomical images.

The biological valve prosthesis II showed the most severe artefacts and flow miscalculation, which meets also the result of the in-vitro trial. A comparison between in-vivo cine imaging (Fig. 6c) and d)) and in-vitro magnitude pictures (Fig. 3) shows the accordance clearly.

Background offset error detection in the absence of flow

Statistical analyses of the range of equivalence were not meaningful for the measurements without flow generated by the pump, as the reference flow values were too low. In this case, significance is defined as a relevant deviation from zero flow in ml/s with a two-sided p-value below or equal 0.05.

Flow measurements without flow revealed no differences between reference level and the slices distal to the prosthesis in all investigated objects. As depicted in Fig. 7, there is no significant deviation of flow values over the entire distance (p > 0.05), meaning no relevant background offset errors.

Fig. 7
figure 7

Background offset error detection. Measurements without flow. Deviation of estimated flow from reference flow over the entire distance in all investigated objects. S2-12 = positions of image acquisition in flow phantom

Discussion

This phantom study investigated the influence of three different metal containing heart valve prostheses on PC flow measurements at 1.5 T. The main finding is that the flow measurements were corrupted in the vicinity of the tested valve prostheses. The extent of flow miscalculation depended on the composition of the prosthesis and on the distance of the measurement slice to the prosthesis. In contrast, PC measurements even in close proximity to a synthetic metal-free aortic valve model or a native aortic valve yielded nearly constant flow values.

PC measurements are expected to be B0-insensitive, as two images containing the same flow-unrelated phase errors are subtracted from each other [27]. In this study, we could reproduce this finding in measurements where no flow was present – notwithstanding all tested metal containing prostheses markedly altered the local magnetic field. Beyond that, imaging and flow encoding gradients induce eddy currents in the metal stent that evoke superimposing local magnetic field distortions.

In case of no flow, i.e. the pump was switched off, it could be shown that the phase contrast measurement yielded a net phase of zero after subtraction of the flow-compensated from the flow-encoded measurement even in the presence of these additional local magnetic field distortions.

In case of flow, however, the results showed that in the proximity of the metal stented prostheses severe flow measurement errors occurred. As for the static magnetic field related distortions and the imaging gradients related distortions, the flow-encoded measurement and the flow-compensated measurement experience identical field alterations. However, as the flow-encoding gradient scheme and the flow-compensating gradient scheme are dissimilar, these gradient schemes also induce different local and temporal magnetic field distortions. Analysis of the B0 maps in the proximity of the heart valve prostheses revealed that the severity of local magnetic field inhomogeneity was coincident with the severity of flow miscalculations. Coherence between the extent of the flow miscalculation and the material composition of the prosthesis seems to be likely as different material compositions inherit different magnetic susceptibilities and thus induce different magnetic field distortions. Analogous to the fact that the interaction with more susceptible material leads to stronger B0 inhomogeneity, the interaction of the gradients with the more susceptible material also leads to stronger net phase errors. Beyond that, also flow velocity influences the amount of additionally accumulated phase and thus the extent of flow miscalculation.

The strongest local magnetic field inhomogeneity and maximum flow miscalculation were observed in prosthesis II, with a maximum flow overestimation of around 150 ml/s in phantom, which is 53% deviant to the reference flow. In-vivo around 65 ml/s flow deviation was seen.

Our findings are related to the dedicated material of these prostheses. The influence on the magnetic field depends on the material. The induced local magnetic field distortions have different impact on flow measurements [19, 28].

The impact of heart-valve-prostheses-related magnetic field distortions depended on the distance of the measurement slice to the level of the prostheses. Maximum miscalculation is seen in the immediate proximity of the prostheses. For the investigated prostheses, the flow values converged in-vitro beyond a distance of about 20 mm and in-vivo about 30 mm distal to the level of the prosthesis. We used a range of equivalence of 15% as a marker for significant deviation from reference flow. The background phase offset errors are within this range; so we deduce a flow-related impact of the metal components.

Although it could be shown that heart valve prostheses are MR-safe up to 4.7 T, the magnetically induced electric currents can hamper the diagnostic performance [2931]. Both biological prostheses cause magnetic field distortions and signal loss to various extents as seen in Fig. 3. Prosthesis I is constructed onto a flexible, tender stent, whereas prosthesis II comprises a semi-rigid, three-base frame. The extent of magnetic field distortion is related to the amount of metal. Another trial demonstrated metal-stented aortic valve prostheses inducing artefacts in CMR in terms of signal loss in the whole left-ventricular outflow tract. In contrast, the stent-less valve prostheses did not cause artefacts [32].

Among the research on heart valve prostheses, magnetic field distortions due to metallic compounds were investigated in several studies, but flow-measurement-related data are rare [33]. In our study, prosthesis II consists of Elgiloy, an alloy recommended for biomedical implants. As seen in Figs. 3 and 4, it evokes local magnetic field distortions. The mechanical prosthesis III, composed mainly of pyrolythic carbon, was deemed to be nonmagnetic, but showed magnetic field distortions.

The interaction between both mechanical and biological valve prostheses and magnetically induced electric currents has been demonstrated [34].

It has already been described that elements from stainless steel cause stronger artefacts than Nitinol-based products [19, 28]. Our in vitro results are supported by single case examples, but have to be evaluated in a larger cohort.

Assessment of valvular prosthesis is of high impact in clinical cardiology and CMR has an increasing influence. Therefore these results could have an impact on applicability of flow quantification in this patient group.

Strengths and limitations

Our study has the strength that we faced a potential obstacle for clinical decision-making when using phase-contrast techniques in patients with valve prostheses. Our results are limited by the fact that our setting did not include MR-independent flow quantification as a reference. However, the setting has been validated previously [35] and the flow was kept constant to meet the requirements for this study.

In our study, we investigated the behaviour of PC based flow quantification using a flow-encoded and a flow-compensated measurement. As other techniques exist - such as the measurement with two flow-encoded measurements - our findings cannot be generalized for all existing flow measurement techniques in the same manner.

We have investigated three often-used prostheses. Our findings cannot be directly extrapolated to all other types of prosthesis. Especially the recommended distance to achieve correct flow measurements needs careful studies covering all types of prostheses. Further investigation and analysis of all different valve prostheses have to be conducted to address this need.

Conclusions

Our findings are of clinical interest as they may impact clinical routine. Currently, specific distances are defined for flow measurement of native valves and valve prostheses [36]. But the awareness of magnetic field distortion-related flow miscalculations is limited. However, it can be concluded that flow measurements in the proximity of stented valve prostheses are not reliable.

When performing PC based flow measurements in patients with heart valve replacement, one should be aware of potential flow and volume miscalculation due to prosthesis-related distortions of the magnetic field.

Further investigations regarding other available prostheses are needed, as clinical outcome after valve replacement is based on a patient-matched prosthesis, which has to be determined by reliable measurements.

Further studies are needed to provide guidance for different field strengths and all types of prosthesis.