Background

Streptococcus pyogenes is a human-only pathogen with a diverse clinical spectrum including severe syndromes responsible for more than 500,000 deaths worldwide each year [1, 2]. Hope for a vaccine to prevent S. pyogenes infections and their complications predates the identification of the bacterium in the late nineteenth century, when the disease spectrum was loosely considered under the historical umbrella term of scarlet fever. While vaccination is viewed as the intervention most likely to achieve sustainable diseases control, development has been impeded by regulatory, commercial, and scientific obstacles, including critical knowledge gaps regarding host-pathogen interactions [3].

Controlled human infection (‘human challenge’) studies recruit carefully screened volunteers who are deliberately infected under highly controlled conditions with a well-characterised strain of an infectious agent. Human models contribute to vaccine development as platforms for direct evaluation of vaccine protective efficacy and for detailed studies of bacterial pathogenesis and human immunity, spotlighting potential correlates of protection [4]. A multinational collaborative group established the Controlled Human Infection for Vaccination Against Streptococcus pyogenes (CHIVAS) model of experimental human S. pyogenes pharyngitis in healthy adult volunteers in Melbourne, Australia. The first CHIVAS study (CHIVAS-M75, ClinicalTrials.gov number NCT03361163) sought to determine a dose of a carefully selected S. pyogenes M75 strain (GenBank CP033621) required to cause pharyngitis in at least 60% of participants following direct inoculation of the pharynx by a swab [5,6,7]. The M75 strain was isolated from a child with acute pharyngitis in Melbourne. It was selected for use as a challenge strain according to a rationale prioritising clinical relevance and participant safety, and was extensively characterised by whole genome sequencing, in vitro assays, and in an animal model of invasive infection [5, 8].

The CHIVAS model includes serial collection of throat swabs for detection and quantification of pharyngeal S. pyogenes colonisation following the challenge procedure. We aimed to develop a robust molecular method for quantifying colonisation by the challenge strain from serial throat swabs. To study the dynamics of infection and clearance, we developed a reliable, rapid, scalable, and highly sensitive and specific method to overcome some of the limitations of traditional culture-based methods [9]. Here, we describe the performance of a type-specific emm75 real-time quantitative PCR (qPCR) in testing DNA extracted from throat swabs collected during experimental human S. pyogenes pharyngitis in the CHIVAS-M75 trial.

Methods

Collection of throat swabs

Two throat swabs (FLOQswabs(R), Copan Italia SpA) were collected on the day before challenge, approximately every 12 h during an inpatient period of up to 6 days, and at outpatient follow-up visits (1 week, 1 month and 3 months after discharge). One swab was stored in 1 ml eSwab™ (Copan Italia SpA) solution for culture onto horse blood agar (HBA) and the other swab was placed into 2 ml eNat™ (Copan Italia SpA) nucleic acid preservation medium (guanidine thiocyanate-based solution) and stored at − 20 °C for up to 6 months.

Preparation of genomic DNA from different bacterial species

To assess the specificity of the emm75 qPCR assay to distinguish the S. pyogenes M75 challenge strain from other common upper respiratory tract pathogens, pellets of S. pyogenes, Streptococcus pneumoniae, Staphylococcus aureus and Haemophilus influenzae were harvested by centrifugation at 2,800 rcf from 25 ml mid-log phase cultures, and colony-forming units (CFU) were enumerated by plating. Bacterial cultures were centrifuged and genomic DNA (gDNA) was isolated from pellets using the DNeasy® Blood and Tissue kit (Qiagen) according to the manufacturer’s modifications for Gram-positive bacteria. Briefly, each bacterial pellet was incubated for 45 m at 37 °C with 180 μl of enzymatic lysis buffer (lysozyme 20 mg/ml, 20 mM Tris-HCl pH 8.0, 2 mM EDTA and 1.2% v/v Triton-X), then DNA from the lysate was purified on column as per manufacturer instructions. All DNA was eluted from purification columns in a final volume of 50 μl. Concentration of eluted DNA was adjusted to 104 Genome Equivalents (GE)/μl for all S. pyogenes strains (assuming one genome per CFU and S. pyogenes genome size of 1.8 Mb). DNA from S. pneumoniae (clinical isolate, serotype 5), S. aureus (ATCC 25923) and H. influenzae (ATCC 10211) was adjusted to 105 GE/μl according to their respective genome sizes (2 Mb, 2.8 Mb and 1.8 Mb) [10,11,12]. Control human pharyngeal DNA was extracted from a healthy adult donor, using the DNeasy® Blood and Tissue kit according to the instructions for mammalian cultured cells. For qPCR sensitivity testing, gDNA was prepared from a dilution series of M75 broth culture from 100 to 107 CFU/ml.

Separate and combined sequential extraction of RNA and DNA

We compared RNA-only and DNA-only extractions using separate spin columns with a protocol for combined sequential RNA and then DNA extraction from the same spin column, as described by Kerllen Martins [13]. In the combined RNA + DNA method, RNA was initially isolated using a a single RNeasy Mini® kit column with RNase-free water (pH 4.5), then DNA was eluted off the same column with elution buffer from the DNeasy Blood and Tissue kit, pH 7.5 (Additional file 1: Method S1). RNA derived using either protocol was treated with the TURBO DNA-free kit (Invitrogen) to enzymatically digest contaminating DNA. The RNA was then assessed for purity and concentration by spectrophotometry using the NanoDrop™ 2000 system (ThermoFisher) and the 4200 TapeStation® (Agilent) prokaryotic RNA High sensitivity ScreenTape Assay.

Evaluation of extraction methods

To examine the effect of chemical pre-lysis prior to extraction, 200 μl volumes of eNat™ medium containing S. pyogenes M75 broth culture at three concentrations (9 × 102/104/106 CFU/ml) was stored at − 20 °C for 7 days. Using the combined RNA + DNA method, extraction was completed with and without a pre-lysis step. An additional extraction was done without pre-lysis using a larger 600 μl volume of spiked eNat™ at a single concentration (9 × 104 CFU/ml). To compare separate and combined extractions, a dilution series of S. pyogenes M75 broth culture (determined by spread-plate dilution) was combined with eNat™ medium for final concentrations of 3 × 101/102/104/106 CFU per ml. These eNat™ tubes were frozen at − 20 °C for 6 weeks and extraction was performed using the RNA + DNA method without pre-lysis.

Quantitative PCR

To assess the quality of simultaneously extracted RNA and DNA, the RNA was reverse transcribed using the Protoscript II Reverse Transcriptase kit (NEB) and the resulting cDNA and extracted gDNA were tested by qPCR. Relevant primer and probe sequences are listed in Table 1. The emm75 assay (IDT®) targets a 123 bp stretch of the emm75 hypervariable region (from 146 to 268 bp inclusive). Reactions using FAM-fluorescent probes (emm75 and speB) were performed using the GoTaq® Probe qPCR Master Mix (Promega). GoTaq® SYBR® green qPCR Master Mix (Promega) was used for gyrA, lytA, gltB, hpd3 and mammalian mitochondrial rRNA. Reactions were prepared using 4 μl of template DNA or cDNA in a final volume of 20 μl with primers and probes at a final concentration of 250 nM. Duplicate runs were completed on an AriaMX qPCR instrument (Agilent) with the following fast 2-step conditions: initial hot start of 1 cycle at 95 °C, followed by 40 cycles of denaturation at 95 °C for 3 s and combined annealing/extension at 60 °C for 30 s. No template control (NTC) and no reverse transcriptase (NRT) controls were included in each assay.

Table 1 Primers and probes used in this study

qPCR sensitivity and limit of detection

To assess the limit of detection (LOD) for the emm75 qPCR we prepared template using a 10-fold dilution series of S. pyogenes M75 gDNA (100–107 CFU/ml). Ct variation at the lowest template concentration (100) was high so these values were excluded from calculation of the standard curve. The lowest dilution to produce consistent Ct values was 101 CFU/ml, equivalent to 40 GE of S. pyogenes M75, considered as the LOD of the emm75 qPCR. The emm75 assay was compared to a previously published qPCR targeting speB, a virulence factor in the core chromosome of S. pyogenes and highly conserved across different S. pyogenes emm-types [8].

qPCR specificity

To assess the emm-type and species specificity of the emm75 qPCR assay, 14 other S. pyogenes strains were also tested, representing each of the 11 major emm clusters (E1–6, D2, D4, AC3–5), and two single protein clades [19]. Other bacteria commonly detected in the human oropharynx were also tested. The emm75 primer and probes were blasted against the online NCBI genome database (https://blast.ncbi.nlm.nih.gov/Blast.cgi) to confirm the absence of homologous sequences, besides the emm75 allele.

CHIVAS-M75 study samples

The CHIVAS-M75 study protocol has been described in detail elsewhere [6]. The study was approved by the Alfred Hospital Human Research Ethics Committee (500/17). Twenty healthy adult volunteers were screened for S. pyogenes using culture and a rapid molecular point-of-care test (Abbott ID NOW Strep A 2, formerly Alere i) in the 28 days prior and then again in the 12 h before direct swab application of S. pyogenes M75 (at a dose of 105 CFU/ml) to the pharynx [20].

Throat swab cultures using eSwab™

Each eSwab™ collected in the CHIVAS-M75 trial (n = 192) was plated onto HBA at an external contract clinical laboratory according to the manufacturer instructions (Copan) within 48 h of collection and incubated at 37 °C in 5% CO2. The next day, growth of β-hemolytic colonies was visually estimated according to guidelines developed in our laboratory: 10-fold dilutions of S. pyogenes M75 in broth culture (range 0–105 CFU/ml) were prepared and an eSwab™ dipped into each dilution for 30 s then plated on HBA for single colonies following the eSwab™ protocol. Semi-quantitative growth scores were established based on the results of this test: 0 = no growth; 1 = light; 2 = moderate; 3 = heavy; 4 = profuse (Additional file 2: Figure S1). At the external laboratory, β-hemolytic colonies were selected and confirmed as S. pyogenes by MALDI-TOF mass spectrometry (Bruker Microflex) [21]. Colonies were re-plated for purity then stored at − 80 °C. Pure vials were sent to our laboratory and all were confirmed as emm75 by gene amplification and Sanger sequencing [22].

Throat swabs for molecular analyses using eNat™

Combined RNA + DNA extraction was done from 192 eNat™ swabs, with laboratory staff blinded to the clinical trajectory of the source participants in the CHIVAS-M75 trial. Genomic DNA was quantified by qPCR and compared to the eSwab™ semi-quantitative growth scores. Use of the extracted RNA to study gene bacterial and host expression (e.g., by reverse transcription-qPCR) will be the object of future work and is not explored in the methodology described herein. For determination of bacterial load in the eNat™ samples by emm75 qPCR: A set of standards were generated using 10-fold dilutions of S. pyogenes M75 (ranging from 101 to 107 CFU/ml) which was then used to spike eNat™ vials containing oropharyngeal swabs from two healthy adult donors. Standards were frozen for 7 days, thawed to mimic handling of samples from the clinical trial, and DNA extracted using the combined RNA + DNA protocol. These standards were included on each 96-well plate to generate a seven-point standard curve, along with a negative no template control (NTC). Each reaction was performed in duplicate. Cycle thresholds (Ct) were ascertained with Agilent AriaMX 1.7 software using ΔR fluorescence and automated thresholds. Reliable emm75 detection occurred at Cts ≤35, considered thereafter as the limit of detection (Additional file 3: Figure S2).

Data analysis

Analyses were performed using GraphPad Prism 8 software. Efficeincy of the qPCR was calculated with Agilent AriaMX software from a formula assigning 100% when 10-fold serial dilutions of template cause an increase of 3.3 cycles (i.e. linear regression slope = 3.3) [23]. M75 bacterial load was expressed as log10 emm75 genome equivalents (GE) per ml of eNat™ medium. Separate or combined column extractions were compared by paired t-tests and the non-parametric Spearman coefficient was used to consider correlation between plate growth scores and emm75 GE [8].

Results

qPCR sensitivity

The emm75 qPCR assay had superior efficiency of 99.2% based on a slope of 3.3 (R2 = 0.999) (Fig. 1). In comparison, the speB assay had 96.5% efficiency (slope = 3.4, R2 = 0.999). Ct values for the speB assay were marginally lower than the emm75 assay (mean difference 0.56 Ct, 95% CI [0.40, 0.73]), however the assays had similar levels of sensitivity, with lower limits of detection of 40 GE/ml when corrected for efficiency.

Fig. 1
figure 1

Performance of the emm75 qPCR. qPCR targeting the emm75 or speB gene was performed on genomic DNA isolated from 10-fold dilutions of S. pyogenes M75. The speB qPCR was used to provide a benchmark for the performance of the emm75 qPCR, using FAM-conjugated probes for both reactions. Each point represents the mean and standard deviation (SD) of duplicate reactions

qPCR specificity

The emm75 qPCR was 100% specific within the detectable concentration range, producing an amplification signal for assays containing emm75 template and no signal for assays containing non-emm75 S. pyogenes, S. pneumoniae, S. aureus, H. influenzae or DNA extracted from pharyngeal swabs from a healthy adult. The speB qPCR was 100% specific, being positive for all 15 S. pyogenes templates and negative for the other bacterial species tested (all 4 species were positive in relevant species-specific assays) (Additional file 4: Table S1).

Evaluation of extraction methods

Effective extraction of DNA from Gram-positive bacteria such as S. pyogenes usually requires an enzymatic peptidoglycan digestion step to lyse the cell wall. The eNat™ medium contains lysing agents, so that a pre-lysis incubation step or addition of supplemental lytic agents may not be required. We performed DNA extraction with and without an enzymatic pre-lysis step to assess whether this influenced DNA yield in the emm75 qPCR. At each dilution tested, pre-lysis generated lower Cts than extraction without pre-lysis (mean difference 0.9458 Ct, P value = 0.03, 95% CI [0.28, 1.61]; Fig. 2a). When a higher volume of eNat™ medium was processed through the column without pre-lysis (600 μl versus 200 μl, both 9 × 104 CFU/ml) the mean Ct was 26.39 (Fig. 2a,◆)- lower than the extrapolated 27.27 for the equivalent bacterial load with pre-lysis (‘pre-lysis’ linear regression: Ct = [− 2.91 × 4.73] + 41.05). Although not tested across a range of input bacteria CFUs, this finding suggests that increasing the processing volume of eNat™ is more effective for increasing DNA yield than applying a pre-lysis step.

Fig. 2
figure 2

Optimisation of the method for isolation of DNA and RNA from throat swabs in eNat™. a Pre-lysis versus No pre-lysis extractions using the combined RNA + DNA approach. eNats™ were spiked with S. pyogenes M75 at three concentrations and 200 μl aliquots (1.8 × 102, 1.8 × 104, 1.8 × 106 total GE) underwent DNA extraction using a chemical Pre-lysis step (●) or No pre-lysis (■). A larger 600 μl aliquot (3x vol) of eNat™ medium spiked with M75 (5.4 × 104 total GE) was processed without pre-lysis (◆). Data shown in total GE to account for differences in input bacteria. Each point represents the mean and SD of two eNats™ tested in duplicate qPCRs. b eNats™ containing a range of M75 were used to test separate column extractions (RNA-only ■ or DNA-only ●) versus combined RNA + DNA isolation (RNA and DNA). The emm75 qPCR was used to assess DNA whilst RNA was converted to cDNA and tested in a SYBR Green qPCR for the housekeeping gene, gyrA. Each point represents the mean and SD of three eNats™ tested in duplicate qPCRs targeting the emm75 gene

DNA yield from the combined RNA + DNA extraction protocol was lower than the DNA-only method as determined by UV absorbance and qPCR (mean difference 2.4 Ct, P value = 0.03, 95% CI [0.38, 4.3]; Fig. 2b, Additional file 5: Table S3). As expected, for RNA (and resulting cDNA) both extraction protocols achieved similar Ct values and limits of detection in a qPCR assay targeting the S. pyogenes housekeeping gene gyrA (mean difference 0.08 Ct, P value 0.93, 95% CI [− 2.57 to 2.73]; Fig. 2b). NanoDrop™ and TapeStation® systems did not produce consistent estimations of RNA concentration and quality due to low concentration (Additional file 6: Table S3). As results were similar and the combined extraction protocol reduced both time and cost of extraction, this method was used for the CHIVAS-M75 study samples.

Detection of emm75 bacterial load in vivo by qPCR

During the CHIVAS-M75 study, 20 participants were challenged with 105 CFU/ml of M75, and 192 paired throat swabs (between 8 and 14 time points per participant) were collected for culture (eSwab™) and molecular testing (eNat™). There was high correlation between culture and qPCR results (Spearman correlation R = 0.8058, P < 0.001; Fig. 3).

Fig. 3
figure 3

Correlation between culture and qPCR results. The number of emm75 genome equivalents (GE), extrapolated from qPCR Ct values, was compared to plate growth scores derived by culture. The median log10 emm75 GE of each score is represented by the dotted line (---), whilst the width of the shape reflects the density of data at the respective values

Culture and qPCR results were concordant (positive or negative) for 171 of 192 evaluable sample pairs (89.1%) (Table 2). All paired swabs taken prior to challenge were negative by both qPCR and culture. One culture-positive eSwab™ was negative by qPCR, giving a sensitivity of 98.8% (95% CI [97.3, 100]) for the qPCR assay. Swabs for 20 time points were qPCR positive and culture negative, including 16 collected after antibiotic treatment and 4 from the first 48 h after challenge in participants whose subsequent swabs were culture positive (Additional file 7: Table S4). All culture-positive swabs were verified as S. pyogenes M75 by qPCR and emm typing.

Table 2 Test results for culture and qPCR from throat swabs

Discussion

We have developed a sensitive and specific quantitative PCR assay for measurement of S. pyogenes M75 from throat swabs in a human infection study of S. pyogenes pharyngitis. We have shown that the qPCR is specific for the emm75 gene encoded by the challenge strain with a high level of analytical performance for efficiency, specificity, and sensitivity. A lysis step prior to extraction of DNA from throat swabs stored in eNat™ was not critical for DNA extraction, attributable to bacterial lysis by the proprietary reagents in the eNat™ medium. There were only minor differences in yield between separate RNA-only and DNA-only extractions and combined RNA + DNA isolation using a single RNeasy column. The combined approach was simpler, faster, and enabled processing of the entire 2 ml eNat™ volume using a single column, reducing cost. Although not the primary aim of this study, the ability to extract quality cDNA from RNA, in addition to DNA for bacterial load quantification, increases the potential value of this technique. Finally, we compared qPCR results to semi-quantitative culture scores ('no growth' score of 0 is quantitative), showing the performance of the assay for its intended use- to measure bacterial load from throat swabs collected in a human challenge study.

Several studies have shown that qPCR is highly sensitive compared to culture and can be used to accurately quantify bacterial load during natural S. pyogenes pharyngitis [8, 24,25,26,27].

The emm75 qPCR assay we developed is rapid, scalable, and type-specific. The entire process from DNA extraction to qPCR can provide a fully quantitative result in 4 h and can be performed on swabs stored in eNat™ solution for at least 6 months without loss of DNA yield. The contract clinical trials facility for the CHIVAS-M75 study was offsite, with limited local laboratory capabilities, so that stability during storage and transport was critical. In comparison, the combined processes for S. pyogenes M75 enumeration and emm-typing by traditional methods requires at least 7 days (initial culture, purity plating, speciation, group typing, extraction of DNA, emm gene sequencing and analysis).

Historically, human infection studies for bacterial pathogens (including S. pyogenes pharyngitis studies in the 1970s) have used traditional bacterial culture methods to measure bacterial load during experimental infection [28,29,30,31]. Several non-human primate models of S. pyogenes pharyngitis have performed PCR for sample analysis (mainly for gene expression or to confirm the emm-type), however plate enumeration was the principal method to quantify bacterial load [32,33,34,35].

We found a strong correlation between emm75 qPCR Ct values and plate growth scores. Discrepancies were noted in 20 swab pairs that were qPCR positive and culture negative, early after challenge as bacterial load was increasing or after initiation of antibiotic treatment when S. pyogenes M75 colonisation was waning. These data likely reflect the lower limit of detection and higher sensitivity of qPCR compared to culture, and detection of non-culturable bacterial elements [8, 36]. Reduced sensitivity of bacterial culture following antibiotic treatment is common [37]. Molecular methods have also shown higher sensitivity in studies using other human pathogens, especially following antibiotic treatment [38, 39]. There was a single pair of discordant qPCR negative and culture positive swabs, collected after antibiotic treatment. Subsequent swabs were negative by both techniques. This discrepancy is likely explained by variation in swab collection. All volunteers returned a negative qPCR result within 7 days of discharge, showing the usefulness of qPCR in differentiating persistent colonisation from relapse or re-infection during the period of outpatient follow up.

This study has several limitations. The categorical and semi-quantitative plate growth scoring system clouds direct comparison to continuous qPCR data. The eSwabs™ for culture were collected at the off-site commercial trials facility and transferred to a contracted laboratory for culture according to diagnostic protocols and workflows with semi-quantitative enumeration as an additional step. The eNat™ swabs were immediately frozen and later transported to our research laboratory for batched molecular testing. This variation could conceivably have contributed to differences between culture and qPCR results. Detection of bacterial DNA from non-viable bacteria may cause false-positive results by qPCR. For the CHIVAS-M75 trial, qPCR detection was at least equivalent to culture. Even more precise bacterial quantification could be overcome if desired in future studies using a marker of viability. As discussed, further work remains to be done with RNA extracted from throat swabs collected in the CHIVAS-M75 trial.

Conclusions

We have developed a reliable, rapid, scalable, and highly sensitive and specific molecular method to quantify pharyngeal colonisation by S. pyogenes M75 in a controlled human infection model.