Background

An inherent problem of artificial ventilation is the difficulty to precisely assess individual lung mechanics from ventilation parameters and displayed measurements, knowing that setting the ventilator based on body weight does not adequately address variations in lung mechanics. This fact is especially critical in severely ill patients such as patients suffering from SARS-CoV-2 associated acute respiratory distress syndrome (ARDS) [1,2,3]. Flow-controlled ventilation (FCV) is a novel ventilation mode which, due to its constant flow and direct intratracheal pressure measurement, enables accurate measurement of dynamic compliance. In the following article we would like to demonstrate by means of a case report how individualized ventilation with compliance guided pressure settings can be performed with FCV, thereby reducing the risk of ventilator induced lung injury on the one hand and improving ventilation efficiency on the other.

Case Presentation

A 50-year-old woman with confirmed SARS-CoV-2 infection was admitted to a peripheral hospital three weeks prior to admission to our hospital. Because of deterioration of respiratory function, the patient had to be intubated and mechanically ventilated. A CT-scan was performed at this stage revealing spacious areas of ground glass opacities including consolidation and air bronchogram (Fig. 1). As the patient’s situation worsened, she was transferred to the University Hospital Innsbruck, a tertiary medical center, for evaluation of extracorporeal membrane oxygenation (ECMO) therapy.

Fig. 1
figure 1

Computed tomography four days prior to admission showing spacious areas of ground glass opacities including consolidation and air bronchogram (a, coronal plane; b, axial plane, with kind permission of the Department of Radiodiagnostics, Central Hospital Bolzano, Bozen/Bolzano, Italy)

Evaluation of the pulmonary state revealed a moderate, SARS-CoV-2-associated ARDS (paO2/FiO2 ratio of 124) according to Berlin definition [4]. Initially, pressure-controlled ventilation (PCV) was applied with an Evita® Infinity® V500 respirator (Dräger, Lübeck, Germany) with a peak inspiratory pressure (Ppeak) of 27 cm H2O, a positive end-expiratory pressure (PEEP) of 12 cm H2O, an inspiration to expiration ratio (I:E-ratio) of 1:1, and a respiratory rate of 21 /min, resulting in a tidal volume of 5.8 ml/kg predicted body weight (PBW) and a compliance of 24 ml/cm H2O.

In accordance with ELSO guidelines [5], ECMO therapy was declined. Prone position improved the paO2/FiO2 ratio to 130–150 and was performed for the next 16 h. After this period compliance worsened again in supine position and invasiveness of ventilation had to be increased (Ppeak 28 cm H2O, PEEP 10 cm H2O, respiratory rate of 25 /min, calculated mechanical power (MP) [6] of 16.9 J/min). Nevertheless, oxygenation was still compromised with a paO2 of 78.2 mmHg (10.4 kPa) at an FiO2 of 0.6 (paO2/FiO2 ratio of 130).

In an attempt to reduce the invasiveness of ventilation and its deleterious effects on lung tissue flow-controlled ventilation (FCV) was initiated with the Evone® respirator (Ventinova Medical B.V., Eindhoven, The Netherlands). This mode of ventilation can be applied via any conventional endotracheal tube and establishes a continuous, constant flow during inspiration as well as expiration, which is monitored and controlled by direct intratracheal pressure measurement. First, identical PCV pressure settings were adopted to FCV with a Ppeak of 28 cm H2O, a PEEP of 10 cm H2O, an I:E-ratio of 1:1 and the flow set to 12 l/min, otherwise deep sedation (Richmond Agitation-Sedation Scale of -4) was continued with midazolam 0.4–0.5 mg/kg/h, morphine 0.4 mg/kg/h and esketamine 2.1 mg/kg/h without the use of neuromuscular blocking agents.

The continuous, constant flow in FCV creates a continuous, stable pressure drop throughout inspiration and expiration from the set and measured tracheal pressure to the alveolar space and vice versa, as there is no pause during the ventilation cycle [7]. As a flow of gas only occurs if a pressure gradient exists, this gradient can be calculated if the flow and the resistance of the system are known. This relationship is based upon Ohm’s law, where the pressure drop (ΔP) equals the flow (\(\dot{V}\)) multiplicated by the resistance (R) of the system. Translated to the described patient, a pressure drop of 2.3 cm H2O between the trachea and the alveolar space can be calculated (ΔP = \(\dot{V}\) x R; flow = 12 l/min = 0.2 l/s; resistance = 11.6 cm H2O/l/s). During inspiration when gas flow is directed to the alveolar space this fact reduced alveolar Ppeak by 2.3 cm H2O (from 28 to 25.7 cm H2O), whereas during expiration when flow changes direction, PEEP was increased by 2.3 cm H2O (from 10 to 12.3 cm H2O). Therefore, driving pressure was reduced by 4.6 cm H2O compared to PCV (from 18 cm H2O to 13.4 cm H2O; Fig. 2b). In PCV however, airway pressure is presumed to be equal to alveolar pressure due to a zero flow phase at the end of inspiration as well as expiration, which should result in an equilibrium pressure phase.

Fig. 2
figure 2

demonstrates the course of respiratory parameters during pressure-controlled ventilation (PCV; blue) and flow-controlled ventilation (FCV; red) since hospital admission. The grey area indicates prone position of the patient, the white area supine position. For FCV the (effective) alveolar pressures are calculated based on the measured resistance and the set flow according to Ohm’s law. a, peak inspiratory pressure (Ppeak) and positive end-expiratory pressure (PEEP). b, driving pressure. c, ratio of arterial partial pressure of oxygen (paO2) to fraction of inspired oxygen (FiO2)

After 30 min, blood gas analysis revealed an improvement of oxygenation resulting in a paO2 of 110 mmHg (14.7 kPa) (paO2/FiO2 ratio of 183; Fig. 2c). In order to decrease invasiveness, ventilation settings were then adjusted to dynamic lung mechanics with compliance-guided PEEP and peak pressure settings as follows:

First, PEEP level was determined by an incremental (or decremental) PEEP titration, which represents the point of the highest tidal volume or consecutive dynamic compliance at a fixed driving pressure. Subsequently, driving pressure was stepwise increased, as long as tidal volume showed an overproportional increase based on measured dynamic compliance. This strategy allows for a precise determination of the (so-called) lower and upper inflection point during ventilation and leads to an almost linear relationship of pressure and volume during the ventilation cycle within the individual limits of dynamic lung mechanics [8] (Fig. 3a + b). Finally, gas flow was set to achieve a paCO2 level < 60 mmHg (8.0 kPa).

Fig. 3
figure 3

Pressure-volume loops obtained from intratracheal pressure measurement. a, represents the pressure-volume loop at the beginning of FCV with a peak inspiratory pressure (Ppeak) set to 28 cm H2O and a positive end-expiratory pressure (PEEP) of 10 cm H20. Considering the measured resistance of 11.7 cm H2O/l/s and the set flow of 12 l/min and applying Ohm’s law, the alveolar Ppeak calculates to only 25.7 cm H2O and the alveolar PEEP to 12.3 cm H2O. b, after compliance-guided pressure settings Ppeak was reduced to 23 cm H2O (alveolar 21 cm H2O) and PEEP 7 cm H2O (alveolar 9 cm H2O) resulting in an almost linear relation of pressure and volume during in- and expiration. The steeper course of the PV loop in B compared to A indicates the increased compliance after individualized pressure settings. c, two hours after first compliance-guided pressure setting, re-evaluation showed a slight decline in lung mechanics. Therefore, Ppeak was adjusted to 22 cm H2O (alveolar 20 cm H2O) and PEEP to 5 cm H2O (alveolar 7 cm H2O). The initial kinking of the inspiratory pressure volume curve in C is an indicator for intratidal recruitment

Calculation of alveolar pressure according to Ohm’s law revealed a Ppeak of 21 cm H2O and a PEEP of 9 cm H2O, thus a driving pressure of 12 cm H2O, resulting in a compliance of 27 ml/cm H2O and a calculated MP [9] of 8.6 J/min with individualized ventilation settings. Thirty minutes after compliance-guided adjustment paO2/FiO2 ratio was 140 (Fig. 2b).

Re-evaluation of lung mechanics after two hours allowed for a further reduction of Ppeak to 20 cm H2O and a PEEP of 7 cm H2O (Fig. 2c) resulting in a MP of 7.7 J/min. However, during a period of four hours after initial assessment of the patient’s lung mechanics, the compliance worsened, and subsequently tidal volume decreased from 5.3 to 4.7 ml/kg PBW with a decline of paO2 to 67.6 mmHg (9.0 kPa) at an FiO2 of 0.65 (paO2/FiO2 ratio of 104).

Ventilation of the patient was then continued with PCV at significantly increased pressure settings (Ppeak 28 cm H2O, PEEP 10 cm H2O, respiratory rate 25 /min, MP 17.3 J/min) and the patient was put in prone position. In the further course ongoing invasive ventilation led to ECMO therapy and unfortunately the patient died two months later due to therapy refractory B-cell depletion and no ability to form antibodies against SARS-CoV-2.

Discussion

FCV enables a continuous, constant flow not only during inspiration, but also during expiration, which is a novelty in artificial ventilation [10, 11]. Steady, slow changes in pressure and volume minimize the mechanical impact of ventilation (i.e. stress and strain) or – in terms of physics – mechanical power and dissipated energy. Previous preclinical [12, 13] as well as clinical trials [14,15,16] with FCV already demonstrated improved lung recruitment and gas distribution, which was hypothesized to be related to flow control and a linearized airway pressure decline during expiration. In our case, this may explain the initial increase in paO2/FiO2 ratio by 30 % (from 130 to 183) with identical ventilation settings as in preceding PCV.

Strictly speaking, due to physical reasons, pressure settings were not the same in PCV and FCV. The continuous, constant flow in FCV results in a pressure drop from the trachea to the alveolar space during inspiration and reversely during expiration. The pressure measured in the trachea and shown on the monitor of the ventilator therefore does not reflect the “true” (= effective) alveolar pressure, which according to Ohm’s law must be lower during inspiration and higher during expiration, given that gas flow and resistance of the system are kept constant. This specific physical characteristic must be kept in mind when assessing pressure settings during ventilation with FCV and thus higher driving pressures and lower PEEP levels than generally accepted in PCV [17,18,19,20] may be applicable in FCV.

The only uncertainty is the amount of tissue resistance (resulting from e.g. inertia, friction, viscoelasticity) not contributing to the flow-related difference between tracheal and alveolar pressures. In FCV constant in- and expiratory flow allows for more precise resistance measurements at Ppeak (representing airway as well as tissue resistance) as well as PEEP (mainly representing airway resistance) and a well-founded estimation of tissue resistance (as the difference of both values). To keep it simple, in our patient only the resistance measured at Ppeak level was used for conversion of tracheal pressures into (global) alveolar pressures.

Compliance-guided pressure settings aim to reduce both the occurrence of atelectasis and overdistension of lung tissue by adjusting the patient’s ventilation between the (so-called) upper and lower inflection point of the individual pressure-volume curve (Fig. 2a + b). Because in FCV the degree of freedom is limited to settings of Ppeak, PEEP and gas flow, the administered tidal volume and respiratory rate will automatically adjust depending on the amount of functional lung tissue. This fact is of high importance especially in COVID-19 patients, as they can show quick alterations in lung compliance.

An issue to be discussed is the low PEEP level determined during compliance-guided pressure adjustments, which apparently did not fit the demands of our ARDS patient. Initial rapid improvement of gas exchange after switching from PCV to FCV was probably related to a slight increase in PEEP (with the same PEEP set at the ventilator as during PCV) due to the nature of FCV as explained above. However, two hours later PEEP was decreased to 5 cm H2O (corresponding to a PEEP of approximately 7 cm H2O in the alveolar space) after re-evaluation of the patient. Finally, the patient’s very poor compliance (< 20 ml/cm H2O) and significantly reduced PEEP level (5 cm H2O) resulted in a tidal volume not capable of maintaining sufficient oxygenation (4.7 ml/kg PBW). This fact could have been anticipated by examination of the corresponding pressure-volume (PV) loops, which upon detailed analysis indicated alveolar collapse at the end of expiration and subsequent initial alveolar recruitment during inspiration (initial kinking of the inspiratory pressure volume curve, Fig. 3c).

Unfortunately, the PV loop feature is not implemented in the respirator’s software yet which undoubtedly would be very helpful to monitor undesired changes in dynamic lung mechanics in critically ill patients.

Conclusions

In FCV due to constant flow, monitored tracheal pressure does not reflect the pressure at the alveolar space and thus higher driving pressures and lower PEEP levels than generally accepted may be applicable. Additionally, in FCV individualization of ventilation settings based on dynamic compliance measurements is possible, even though it needs constant reevaluation of the patient’s lung mechanics and a prompt adjustment of ventilator settings if it becomes clear that the severity of the pulmonary disease impedes a lung-protective ventilation strategy. Under these circumstances increased driving pressures or PEEP levels above lung mechanic limits and/or an increased fraction of inspired oxygen must be accepted. This case may demonstrate the limits of a lung-protective ventilation strategy in ARDS and highlights the complexity of this disease. However, FCV might offer a clinical applicable approach to reduce applied and dissipated energy of artificial ventilation in order to improve the outcome in COVID-19 patients requiring prolonged ventilation.