Background

In obese patients, the excessive adipose tissue around the thorax and the visceral organs reduce the functional residual capacity and expiratory reserve volume [1]. Obesity also leads to a low respiratory system compliance, early expiratory alveolar collapse with consecutive atelectasis, increased airway resistance [2] and increased risk for airway closure [3]. All these changes make mechanical ventilation in obese patients prone to respiratory complications [4, 5].

An emerging ventilation technique to linearize expiratory flow is flow-controlled ventilation (FCV), provided by the new ventilator Evone (Ventinova Medical B.V., Eindhoven, the Netherlands). This device provides a constant positive flow during inspiration and a constant negative flow during expiration. Thereby pressure increases linearly during inspiration [comparable to volume-controlled ventilation (VCV)] and decreases linearly during expiration. Recently, we demonstrated that linearizing the expiratory flow improved lung recruitment, the homogeneity of lung aeration [6, 7], gas exchange [8] and further attenuated experimental lung injury [9]. Since FCV is a new emerging technique comparative clinical studies in humans, particularly in patients with impaired respiratory system mechanics, are lacking.

We hypothesized that FCV improves regional ventilation distribution of the lung and respiratory system mechanics in obese patients. Therefore, we compared regional ventilation using electrical impedance tomography (EIT) and respiratory system mechanics during FCV and VCV in obese patients in a randomized controlled crossover trial.

Methods

Ethics, consent and permission

The study was approved by the Ethics Committee of the University Medical Center of Freiburg (Engelbergstr. 21, 79106 Freiburg, Germany, Ethical Committee N° 179/18) on 29th March 2018 (Chairperson Prof. Dr. R. Korinthenberg) and registered at the German Register for Clinical Trials (DRKS00014925). Please note that this study adheres to the CONSORT guidelines.

Study design and patient population

In order to cope with potential interindividual variability, the study was designed as a randomized controlled interventional crossover trial. After obtaining written informed consent, we studied twenty-three obese patients with body mass index (BMI) ≥ 30 kg∙m− 2. Patients eligible for enrolment were patients with physical status ASA ≤ III undergoing elective bariatric surgery. Exclusion criteria were ASA physical status >III, age < 18 years, pregnancy, emergency procedure, cardiac pacemaker and other active implants, chronic obstructive pulmonary disease classified as GOLD stage > II or refusal to participate. The trial was conducted at the University Medical Center Freiburg, Germany. Participants were enrolled and assigned by a study related anesthesiologist. Data were collected at the University Medical Center of Freiburg, Germany.

Procedure

After obtaining written informed consent, 23 patients were included in the study. After primary recruitment and preoperative evaluation, the patients received routine monitoring (electrocardiography, SpO2, noninvasive blood pressure measurement; Infinity Delta XL, Dräger Medical, Lübeck, Germany) and a 18–20-G intravenous catheter was established. After preoxygenation to an fraction of expired oxygen of 0.8, anesthesia was induced with 0.3–0.5 μg∙kg− 1 predicted body weight [10] iv sufentanil (Janssen-Cilag, Neuss, Germany) and 2–3 mg∙kg− 1 actual body weight iv propofol (Fresenius Kabi, Bad Homburg vor der Höhe, Germany). Tracheal intubation was facilitated with 0.6 mg∙kg− 1 predicted body weight iv rocuronium (Fresenius Kabi). If the patient required a rapid sequence induction, neuromuscular blockage was performed by the administration of 1.0 mg∙kg− 1 predicted body weight iv rocuronium. Neuromuscular blockage was monitored with a mechanomyograph (TOFscan; Dräger Medical). For tracheal intubation, we used tracheal tubes with low pressure cuffs (internal diameter of 7.0–7.5 mm for women and 8.0 mm for men; Mallinckrodt Hallo-Contour; Covidien, Neustadt an der Donau, Germany). After adequate placement of the tracheal tube, iv propofol was administered continuously (110–150 μg∙kg− 1∙min− 1). Potential hypotension (defined as mean arterial pressure < 65 mmHg) was treated with a continuous infusion of iv noradrenaline (0.03–0.2 μg∙kg− 1∙min− 1). Perioperative volume requirements were addressed with a crystalloid solution (8 ml∙kg− 1∙h− 1, Jonosteril; Fresenius Kabi). According to our local standard, mechanical ventilation was started as volume-controlled baseline ventilation (Fabius Tiro, Dräger Medical) with a tidal volume of 7 ml∙kg− 1 predicted body weight, inspiration-to-expiration ratio of 1:2, a positive end-expiratory-pressure (PEEP) of 9 cmH2O and ventilation frequency set to maintain an end-tidal carbon dioxide partial pressure between 4.7 and 5.1 kPa. These ventilation settings were based on our study protocol and in accordance with our clinical routine in obese patients. After 7 min of baseline ventilation, all patients were randomly allocated to one of two crossover groups to receive ventilation sequences either VCV-FCV or FCV-VCV for 7 min per ventilation mode. To avoid irritations due to the surgical procedure (e.g. impaired respiratory mechanics by the capnoperitoneum and electrical irritations of the measurement of Electrical Impedance Tomography), our study was performed prior to the surgical intervention. For adequate allocation, a computer generated randomization in blocks was used. Disclosure of the randomization was requested right after induction of anesthesia. A study related anesthesiologist conducted the randomization in blocks, enrolled participants and assigned participants to the interventions. During the study protocol, ventilation variables were kept constant as set during the baseline measurements. To prevent from the risks of extubation and reintubation, FCV was performed by introducing the narrow-bore tracheal tube (Tribute, Ventinova Medical B.V.) into the standard tracheal tube. Blocking the cuff of the Tritube in the lumen of the tracheal tube provided a sufficient seal. By controlling both tube’s markings, placement of the Tritube’s tip exceeding that of the standard tracheal tube by 2–5 mm was ensured, and the potential risk of bronchial intubation was avoided. Respiratory data were collected from both ventilators via the respective serial communication interface and analyzed offline. Electrical impedance tomography (EIT) was performed with PulmoVista 500 (Dräger Medical) in all patients to measure regional ventilation, changes in relative thoracic electrical impedance during the different ventilation phases, relative end-expiratory lung volume (ΔEELV) and to compare the expiratory decrease in intrapulmonary air [11,12,13].

Ventilation modes

Ventilation settings during baseline measurements and VCV were identical. In each patient, baseline measurements were performed prior to the intervention. During FCV, patients were ventilated with a constant positive flow during inspiration and a constant negative flow during expiration (Fig. 1). To avoid intrinsic PEEP, the intratracheal pressure is monitored continuously via a dedicated pressure measurement lumen of the Tritube. During FCV, the operator is able to adjust the inspiratory flow rate, inspiration to expiration ratio, peak inspiratory pressure, end-expiratory pressure and the inspiratory concentration of oxygen. In this special ventilation mode, there is no direct way to control minute volume via tidal volumes and/or respiratory rate. However, the respiratory rate depends on the peak inspiratory pressure, the set (positive) end-expiratory pressure, the set inspiratory flow rate, the inspiration to expiration ratio and the patient’s lung compliance [14]. The (end) expiratory pressure was kept constant in all conditions during the study procedure.

Fig. 1
figure 1

Comparison of flow-controlled ventilation (FCV) with conventional volume-controlled ventilation (VCV)

End points and data collection

ΔEELV was the primary endpoint of this study. EIT recordings were analyzed using software developed in Matlab (R2014, The MathWorks Inc.). We derived ΔEELV from adjusting end-expiratory impedance changes by tidal volume and tidal impedance changes as described before [7, 11]. As a first step, the lung area estimation method was applied to all EIT recordings to estimate the relevant lung area [15]. Afterwards, global tidal impedance curves were calculated. These curves represent the sum of impedance of all pixels per frame over time. To scale the absolute impedance values to milliliters, the relation between tidal impedance change and tidal volume was used. Changes of the baseline of these tidal impedance curves were determined as estimates for changes of the end-expiratory lung volume. ΔEELV was then calculated as the difference of end-expiratory lung volume during the different ventilation phases [11]. Secondary endpoints were the respiratory system variables: plateau pressure (PPlat), mean tracheal pressure (Pmean), mean tracheal pressure during expiration (Pmean expiration), peripheral oxygen saturation (SpO2), fraction of inspired oxygen (FiO2) and quasi-static respiratory system compliance (CRS). To calculate CRS during FCV, the plateau pressure was determined from a short (approximately 0.1 s) end-inspiratory pause. This pause is performed automatically by the Evone ventilator (Ventinova Medical B.V.) with every ten breaths and used to calculate CRS. Non-invasively collected hemodynamic variables included mean systolic blood pressure, mean diastolic blood pressure, mean arterial pressure and heart rate. To compare relative intrapulmonary air distribution, baseline tidal impedance curves for ventral and dorsal lung areas were determined and compared as described before [7, 12]. The differences in mean lung volume (ΔMLV) between baseline ventilation and VCV and FCV were calculated, respectively. Further, the decrease in global thoracic electrical impedance during each ventilation mode was separated into four equal sections (ΔEI25, ΔEI50, ΔEI75 and ΔEI100), then matched with the correlating decrease in tidal volume and compared successively.

Pressure data from the Evone are based on direct tracheal pressure measurement via a dedicated lumen of the Tritube. To allow for comparability of pressure data from both ventilators and to calculate quasi-static compliance of the respiratory system, airway pressure data from the Dräger Fabius Tiro were generally converted into tracheal pressure data by calculating the flow dependent pressure drop across the respective tracheal tube and pointwise subtracting this value from airway pressure [16]. Thus all pressure data in the following refer to the respective tracheal pressure.

The datasets used and analyzed during the current study are available from the corresponding author on request. Please note that EIT data files require large memory.

Sample size calculation and statistical analysis

In regard to previous investigations on gas exchange during FCV in a porcine model of ARDS [17] and the crossover design (paired test conditions) we assumed a standardized effect size of the primary endpoint of 0.7 (being the quotient of differences in means and SD). To reach a test power of 0.8 and a desired level of significance of 0.05, 19 patients were required. To compensate for potential incomplete data sets, 23 patients were included in the study. Lilliefors tests were used to confirm that the assumed normal distribution cannot be rejected.

Values are presented as mean ± standard deviation, unless indicated otherwise. Statistical analysis was done using Matlab (R2014, The MathWorks Inc., Natick, MA, USA). Linear mixed effects model analyses were performed to check for differences between respiratory variables and variables resulting from EIT measurements during the ventilation phases using R based software (jamovi project (2018), jamovi (Version 0.9.2.3), retrieved from https://www.jamovi.org). For each measured primary and secondary endpoint (dependent variable), the influence of the ventilation mode (baseline ventilation, VCV and FCV) and the ventilation sequence (baseline-VCV-FCV, baseline-FCV-VCV) (factors) was investigated. P < 0.05 was considered statistically significant.

Results

In total, 23 consecutive patients presenting for elective bariatric surgery were included and 19 complete data sets could be recorded. Patients were recruited from 30th July 2018 to 23rd October 2018. One patient had to be excluded due to limited size of the EIT belt, three other patients due to incomplete data collection (Fig. 2). There were no adverse events during the study procedure. The study was ended regularly after the last subject was included. Age, gender, ASA physical status, predicted and actual body weight and BMI were comparable between the two interventional groups (Table 1).

Fig. 2
figure 2

Flow diagram of the study population

Table 1 Patients characteristics (n = 19)

During mechanical ventilation, end-expiratory lung volume decreased generally (Fig. 3). ΔEELV between baseline ventilation and FCV (− 126 ± 207 ml) was lower than between baseline and VCV (− 316 ± 254 ml, p < 0.001). ΔMLV between baseline and FCV (− 108 ± 198 ml) was lower than between baseline and VCV (− 315 ± 252 ml, p < 0.001) (Fig. 4). Pmean and Pmean expiration was higher during FCV. No significant differences in tidal volume, ventilation frequency, PPlat, SpO2 and CRS were found between FCV and VCV. All hemodynamic variables were comparable during FCV and VCV (Table 2).

Fig. 3
figure 3

Exemplary relative global thoracic electrical impedance (EITh) of one patient during the study protocol. BL, baseline (volume-controlled) ventilation; VCV, volume-controlled ventilation; FCV, flow-controlled ventilation. The first slope represents the insertion of the Tritube® into the standard tracheal tube. The second slope represents the remove of the Tritube and re-connecting to the Dräger Fabius Tiro ventilator. Arrows indicate the switch between the respective ventilation modes

Fig. 4
figure 4

Alteration of end-expiratory lung volume ΔEELV (a), mean lung volume ΔMLV (b) and comparison in percentage air distribution between ventral and dorsal lung areas (c). BL = volume-controlled baseline ventilation, VCV = volume-controlled ventilation and FCV = flow-controlled ventilation. On each box, the central mark indicates the second quartile, the bottom and top edges indicate quartiles (25th percentile and 75th percentile). * = p ≤ 0.001 for FCV vs. VCV [linear mixed effect model analyses were used to check for differences between the ventilation phases using R based software (jamovi project 2018, version 0.9.2.3)]. The randomization had no significant effect on the measured difference in end-expiratory lung volume between the ventilation phases

Table 2 Respiratory and hemodynamic variables

FCV was characterized by a more even decay of impedance throughout the expiration phase (Fig. 5). ΔEI25, ΔEI50, ΔEI75 and ΔEI100 showed a more even decrease during FCV compared to VCV (Fig. 6). ΔEI25 decreases about 45% during baseline ventilation and VCV and 25% during FCV. ΔEI50 showed no differences between the ventilation modes. ΔEI75 and ΔEI100 showed a lower decrease in global thoracic electrical impedance during baseline ventilation and VCV compared to FCV (Fig. 6).

Fig. 5
figure 5

Exemplary global thoracic electrical impedance (EITh) during two tidal breathes of flow-controlled ventilation (FCV) and volume-controlled ventilation (VCV) in one obese patient. For further comparison, decrease in impedance during expiration was separated into four equal sections (ΔEI25, ΔEI50, ΔEI75 and ΔEI100) and matched with simultaneous tidal changes (comp. Fig. 6).

Fig. 6
figure 6

Relative expiratory decrease in tidal volume during the previously defined sections using the electrical impedance tomography (EIT) for volume-controlled baseline ventilation (BL), volume-controlled ventilation (VCV) and flow-controlled ventilation (FCV). In brief: the decline in global electrical thoracic impedance was separated into four equal sections (ΔEI25, ΔEI50, ΔEI75 and ΔEI100) (compare Fig. 5) and matched with the tidal changes simultaneously. On each box, the central mark indicates the second quartile, the bottom and top edges indicate quartiles (25th percentile and 75th percentile). On each box, the whiskers indicate the most extreme data points. Outliers are plotted individually (‘+’). * = p < 0.001 for baseline vs. FCV and VCV vs. FCV, x = p > 0.05 for baseline vs. FCV and VCV vs. FCV. Linear mixed effect model analyses were used to check for differences between the ventilation phases using R based software (jamovi project 2018, version 0.9.2.3). The randomization had no significant effect on the measured difference in end-expiratory lung volume between the ventilation phases

Discussion

In this study, we compared respiratory system mechanics and regional ventilation in obese patients during short application of FCV and VCV. The main findings of our study are that in obese patients, ΔEELV and mean lung volume decreased less during FCV than during VCV – even with identical respiratory and hemodynamic variables.

These effects were comparable to the effects one would expect from a PEEP increase and/or a tidal volume increase. However, minimal and maximal airway pressure and tidal volume remained unchanged. Our results are consistent with and enlarge upon earlier findings on the implications of a linearized expiratory pressure decrease in lung-healthy patients, lung healthy pigs and a porcine lung-injury model [6,7,8,9].

We observed the changes of respiratory mechanics during the VCV and FCV phases. Since baseline measurements were performed prior to the following ventilation sequence, we attribute the observed differences between baseline and VCV to the general tendency of the respiratory system to continued derecruitment during mechanical ventilation [18] which may be more pronounced in obese patients. The implications of obesity on respiratory system mechanics are well known: chest wall mechanics are impaired, and respiratory system compliance is reduced. Obese patients have an increased risk for early expiratory alveolar collapse and potential consecutive atelectrauma [3, 5, 17, 19, 20] and thus for decreased functional residual capacity and expiratory reserve volume [5, 13, 21, 22]. Therefore, besides low tidal volume and optional recruitment maneuvers, lung-protective ventilation strategies include the application of adequate PEEP in these patients. However, the ideal adjustment of applied tidal volume, and PEEP — with respect to the potential injurious effects of alveolar overdistension — in obese patients still remain obscure [20]. In this regard, FCV improved lung recruitment without modifying PEEP or tidal volume. Further, it should be noted that because of the controlled end-expiratory pressure during FCV, intrinsic PEEP is nearly excluded. Caused by the active control of the expiration phase, FCV is able to overcome the usually passively driven expiration during conventional ventilation.

The mechanisms behind this recruiting effects may be time dependent: when the lung volume falls below the closing capacity airway closure can occur within the expiration [5, 23]. In obese patients, this airway closure can be observed frequently [3]. The overall delayed expiration during FCV delays the time point at which the lung volume falls below the closing capacity. Consequently, the time until the lung volume exceeds closing capacity within the next inspiration is reduced and thus the risk of airway closure may be lowered [22]. The characterization and correlation between the expiratory decrease in global electrical impedance and expiratory decrease in intrapulmonary air and the increased mean airway pressure during expiration support this conjecture.

Theoretical and clinical observations predict that the linearized decrease in expiratory airway pressure has a beneficial impact on the intrapulmonary inhomogeneity [6, 7, 9, 17, 24]. However, the comparison of tidal impedance variation revealed no differences in intrapulmonary gas distribution during the different ventilation phases. The reduced accessibility of EIT images in obese and morbidly obese patients was described earlier and may be caused by the excessive volume of fat tissue around the chest wall. In horizontal supine position, this fat tissue moves laterally and may create potential shortcuts for the electrical currents of the EIT [25]. Therefore, the resolution of the EIT is limited, which may have masked differences in intrapulmonary inhomogeneity in our patients.

CRS did not differ significantly between the investigated ventilation conditions. Reduced CRS in obese patients may be caused mainly by excess adipose tissue around the chest wall and poor posture caused by thoracic kyphosis and lumbar hyperlordosis, aggravated through excessive abdominal fat tissue [19]. Under these conditions, the recruiting effect of FCV may have influenced CRS only to a minor extent. This hypothesis is supported by earlier investigations that demonstrated that the intratidal recruitment state might differ without affecting CRS [26, 27]. To investigate potential effects of FCV on the CRS in obese patients, longer application of FCV may be necessary.

Limitations of the study

We did not perform arterial blood gas analyses to examine the effects of FCV on gas exchange in our patients. In preclinical [17] and clinical studies [8], the controlled expiration improved oxygenation and CO2 elimination. However, in contrast to other centers, placing an arterial line is not part of our standard treatment in this patient group. Therefore, we felt that such invasive approach was not justified for our study. Further, it should be stated that duration of ventilation of 7 min in each ventilation mode is too short to fully evaluate the effects of FCV on regional ventilation. It follows that further studies are required to investigate the long-term effects of FCV on measurements of regional ventilation, respiratory and hemodynamic variables in obese patients.

Conclusion

This is the first study to investigate the influence of FCV on respiratory mechanics and regional ventilation and in obese and morbidly obese patients. Utilizing measurement of regional ventilation, we could demonstrate that the linearized expiratory flow during FCV provided better maintenance of lung aeration with comparable tidal volume, PPlat and PEEP, compared to VCV. The recruiting effect caused by the linearized expiratory air flow and the elevated Pmean during FCV may help prevent atelectasis and hypoxemia during mechanical ventilation in obese patients.