Introduction

Transpulmonary thermodilution (TPTD) is increasingly used for hemodynamic evaluations in critically ill patients [14]. After injection of a cold indicator in the superior vena cava, TPTD allows the computation of cardiac output (CO) from a TPTD curve recorded by a thermistor-tipped femoral arterial catheter [4]. Additional physiological parameters can be derived from the dilution curve, such as global end diastolic volume (GEDV), a volumetric marker of cardiac preload [57], and extravascular lung water (EVLW) [710].

The TPTD method currently in clinical use and implemented in the PiCCO™ system (Pulsion Medical Systems, Munich, Germany) is based on mathematical models described in the 1950 s [11, 12]. A new and original method has recently been developed to derive GEDV and EVLW from a TPTD curve (VolumeView™; Edwards Lifesciences, Irvine, CA, USA). The aim of the present animal study was to compare the new VolumeView™ system with the PiCCO™ system, over a wide range up to extreme pathophysiological conditions.

Materials and methods

The study was approved for the use of swine by the Institutional Animal Care and Use Committee at the Edwards Lifesciences Biological Resource Center, and all experimentation was done in accordance with the Guide for the Care and Use of Laboratory Animals (1996; ILAR, NAP, Washington, DC, USA).

Eleven anesthetized and mechanically ventilated pigs (90 to 110 kg) were studied. Animals were premedicated with intramuscular midazolam (0.5 mg/kg) and atropine (0.5 mg) and were anesthetized with an injection of propofol (1 mg/kg) followed by continuous infusion of propofol (150 μg/kg/min) and sufentanil (2.5 μg/kg/h). After tracheal intubation, pigs were mechanically ventilated in a volume-controlled mode with a FiO2 of 50%, a respiratory rate between 12 and 16 breaths/minute (to maintain an end-expiratory partial pressure of carbon dioxide within the normal range), a positive end-expiratory pressure of 0 cmH2O and a tidal volume of 10 ml/kg.

All animals were instrumented with a right (PulsioCath™; Pulsion Medical Systems) and a left (VolumeView™; Edwards Lifesciences) 5F thermistor-tipped femoral arterial catheter. The correct position of femoral catheters was confirmed by radioscopy (Figure 1).

Figure 1
figure 1

Transpulmonary thermodilution curve. The assessment of global end-diastolic volume (GEDV) by the PiCCO™ system is based on the mean transit time (MTt) and exponential downslope time (DSt), while the assessment of GEDV by the new VolumeView™ method is based on MTt, maximum ascending slope (S1) and maximum descending slope (S2).

All animals were also instrumented with a pulmonary artery catheter (CCComboV™, 7.5F; Edwards Lifesciences) inserted through the right jugular vein and with a central venous catheter in the left jugular vein (Figure 1). The pulmonary artery catheter was used for continuous monitoring of CO (Vigilance II; Edwards Lifesciences) and pulmonary arterial pressures during the experimental protocol. The central venous catheter was used for cold indicator injections and for central venous pressure monitoring. Pulmonary artery pressures, continuous CO and central venous pressure data were used to guide therapy at various stages (as described below) but were not recorded nor analyzed.

The current transpulmonary thermodilution system

The right femoral catheter was connected to a PiCCO2™ monitor (Pulsion Medical Systems) and used to measure CO (CO1), GEDV (GEDV1) and EVLW (EVLW1) using the following equations [1, 7, 9, 10]:

CO 1 = V i ( T b T i ) k / AUC

where Vi is the injectate volume, Tb is blood temperature, Ti is injectate temperature, k is a constant proportional to the specific weights and specific heat of blood and injectate, and AUC is the area under the TPTD curve.

GEDV 1 = CO 1 × ( MTt DSt )

where MTt is the mean transit time of the cold indicator and DSt is the exponential downslope time (Figure 1).

EVLW 1 = ( CO 1 × MTt ) ( 1.25 × GEDV 1 )

The new transpulmonary thermodilution system

The left femoral catheter was connected to the EV1000™ monitor (Edwards Lifesciences) and used to measure CO (CO2), GEDV (GEDV2) and EVLW (EVLW2). CO was derived from the dilution curve using the same Stewart Hamilton equation:

CO 2 = V i ( T b T 1 ) / AUC

GEDV, however, was derived from a different equation as follows:

GEDV 2 = CO 2 × MTt × f ( S 2 / S 1 )

where S1 and S2 are respectively the maximum ascending and descending slopes of the thermodilution curve (Figure 1) and f is a proprietary function.

Finally, EVLW was assessed using the equation:

EVLW 2 = ( CO 2 × DSt ) ( 0.25 × GEDV 2 )

The same cold saline bolus injected through the central venous catheter was used to compute simultaneously the two transpulmonary curves: one with the right femoral catheter PiCCO2™ (Pulsion Medical Systems), the other with the left femoral catheter (EV1000™; Edwards Lifesciences). The average of three bolus measurements was considered for analysis and is reported in Results.

Experimental protocol

The experimental protocol is summarized in Figure 2. Measurements were performed: at baseline; during dobutamine infusion (DOBU, starting at 7.5 μg/kg/minute and titrated to induce a 30 to 50% increase in continuous CO); 5 minutes after stopping dobutamine infusion; after inducing hypovolemia (HYPO, controlled hemorrhage to decrease mean arterial pressure (MAP) around 50 mmHg); after blood restitution and fluid loading (2/3 blood + 1/3 serum saline); and after fluid overloading (HYPER, 75% serum saline + 25% gelatin in order to increase MAP up to 130 mmHg and/or central venous pressure up to 20 mmHg). At each stage, a 10-minute stabilization period was observed before doing the measurements. Finally, additional measurements were performed after inducing acute lung injury (ALI) by injecting intravenously oleic acid (O1383, 100 mg/kg/hour).

Figure 2
figure 2

Flow chart of the experimental protocol. *Multiple measurements. IV, intravenous.

At this stage, several measurements were performed successively in order to capture high EVLW values. Oleic acid-induced pulmonary edema was confirmed by the occurrence of arterial hypoxemia (drop in PaO2/FiO2 and SaO2), a drop in the compliance of the respiratory system (increase in airway pressures while the tidal volume was maintained constant or even decreased) and lung infiltrates on chest X-ray scan (Figure 1). At this point, FiO2 and the positive end-expiratory pressure were respectively increased up to 100% and 15 cmH2O when necessary to maintain SaO2 > 90%. Oleic acid may induce a dramatic increase in pulmonary artery pressures and a decrease in CO (right ventricular failure). Phenylephrine and dobutamine were therefore also administered when necessary to maintain MAP >50 mmHg and continuous CO >5 l/minute as long as possible. When it was no longer possible to maintain SaO2 > 90% and MAP >50 mmHg, data collection was stopped and animals were sacrificed (with pentobarbital and phenytoin).

Statistical analysis

Results are expressed as the mean ± standard deviation (SD), unless specified otherwise. Percentage errors for CO, GEDV and EVLW comparisons were calculated as twice the SD of the bias over the average CO, GEDV or EVLW value, respectively [13]. All bias, SDs, limits of agreement (2SD) and percentage errors reported in the manuscript have been corrected for multiple measurements according to the method proposed by Bland and Altman [14].

Reproducibility of TPTD measurements was assessed by calculating the standard deviation/mean ratio of triplicate measurements and is expressed as a percentage. The effect of each intervention (DOBU, HYPO, HYPER, ALI) versus the previous stage was assessed using a parametric test (paired t test) or nonparametric test (paired Wilcoxon test) when appropriate. Values obtained using both methods were also compared at each stage using unpaired tests (parametric or nonparametric as appropriate).

Several measurements were performed at the latest stage (ALI) in order to capture high EVLW values. At this stage, only measurements corresponding to the maximum EVLW1 (the reference method in the present study) have been selected for comparisons with EVLW2. For the linear regression analysis, however, all measurements were taken into account. P < 0.05 was considered statistically significant.

Results

A total of 137 paired measurements were available for comparisons. Sixty-six paired measurements were collected from stages 1 to 6 (6 stages × 11 pigs) and 71 additional paired measurements (6.5 ± 2.1 per pig) were collected at the final lung injury stage. No data were discarded. The reproducibility of hemodynamic parameters is reported in Table 1.

Table 1 Reproducibility of transpulmonary thermodilution measurements

Overall, CO1 and CO2 ranged from 3.1 to 15.4 l/minute and from 3.4 to 15.1 l/minute, respectively. CO1 and CO2 were closely correlated (r2 = 0.99), with mean bias (± SD) of 0.20 ± 0.30 l/minute and percentage error of 7% (Figure 3). GEDV1 and GEDV2 ranged from 701 to 1,629 ml and from 774 to 1,645 ml. GEDV1 and GEDV2 were closely correlated (r2 = 0.79), with mean bias of -11 ± 80 ml and percentage error of 14% (Figure 4). EVLW1 and EVLW2 ranged from 507 to 2,379 ml and from 495 to 2,222 ml. EVLW1 and EVLW2 were closely correlated (r2 = 0.97), with mean bias of -5 ± 72 ml and percentage error of 15% (Figure 5).

Figure 3
figure 3

Cardiac output comparison. Left: correlation between cardiac output (CO) measured by the PiCCO™ system (CO1) and the VolumeView™ system (CO2). Right: Bland-Altman representation depicting the agreement between both methods. SD, standard deviation.

Figure 4
figure 4

Global end-diastolic volume comparison. Left: correlation between global end-diastolic volume (GEDV) measured by the PiCCO™ system (GEDV1) and the VolumeView™ system (GEDV2). Right: Bland-Altman representation depicting the agreement between both methods. SD, standard deviation.

Figure 5
figure 5

Extravascular lung water comparison. Left: correlation between extravascular lung water (EVLW) measured by the PiCCO™ system (EVLW1) and the VolumeView™ system (EVLW2). Right: Bland-Altman representation depicting the agreement between both methods. SD, standard deviation.

Changes in CO2, GEDV2, and EVLW2 were closely correlated with changes in CO1, GEDV1, and EVLW1, respectively (Figure 6).

Figure 6
figure 6

Correlations between changes in hemodynamic parameters between the two measurement methods. Correlations between changes in cardiac output (CO), changes in global end-diastolic volume (GEDV) and changes in extravascular lung water (EVLW) measured by the PiCCO™ system (CO1, GEDV1 and EVLW1) and by the VolumeView™ system (CO2, GEDV2 and EVLW2).

The effects of each intervention are summarized in Table 2. Inotropic stimulation (DOBU) was achieved by administering an average 23 μg/kg/minute dose of dobutamine, hypovolemia (HYPO) by an average 1.2 l controlled hemorrhage, and hypervolemia (HYPER) by the average infusion of 4.5 l serum saline and 1.5 l gelatin. Both GEDV1 and GEDV2 decreased significantly during bleeding and increased significantly after blood restitution and fluid loading (Table 2). EVLW1 and EVLW2 increased slightly but significantly during fluid overload and dramatically (+110%) during ALI (Table 2). At each stage, values measured with the new VolumeView™ and with the current PiCCO™ method were comparable (Table 2).

Table 2 Transpulmonary thermodilution parameters over the study period

Discussion

In animals, and over a wide range of values, the present study demonstrates that GEDV and EVLW derived from the new VolumeView™ method and from the current PiCCO™ method are interchangeable.

Both methods derive CO from the TPTD curve using the Stewart-Hamilton principles and the same equation [4] so, not surprisingly, the agreement was extremely good with a percentage error of 7%, far below the clinically acceptable threshold value of 30% proposed by Critchley and Critchley [13].

In contrast, GEDV was derived from two different equations. The PiCCO™ equation is based on time characteristics of the TPTD curve (mean transit time of the cold indicator and exponential downslope time) while the new VolumeView™ equation additionally relies on the ascending and descending slopes of the dilution curve (Figure 1). The present results show that both methods are interchangeable to assess GEDV even when significant changes in cardiac preload are induced by bleeding and fluid loading. They also confirm that GEDV is not affected by dobutamine-induced changes in CO, and hence that there is no mathematical coupling between both parameters [3, 5]. The GEDV has been shown to be a reliable indicator of cardiac preload [5], varying in the same direction as echocardiographic preload indices [6]. A goal-directed strategy based on the optimization of GEDV has been shown to be useful to improve the postoperative outcome of cardiac surgical patients [15].

Both methods were also interchangeable for the assessment of EVLW; not only during slight modifications induced by fluid overload, but also during dramatic increases related to capillary leak as those observed during the ALI phase (Table 2). Assessing EVLW may be useful for clinicians treating patients with ALI or left ventricular failure [16]. EVLW has been shown to be more sensitive and specific than chest X-ray and ALI criteria to diagnose pulmonary edema [17, 18]. EVLW is also a prognostic parameter since it has repeatedly been shown to be correlated with mortality in patients with ALI as well as in the general intensive care unit population [1922]. Moreover, it has been suggested in critically ill patients that goal-directed strategies based on the measurement of EVLW may be associated with a decrease in the duration of mechanical ventilation and length of hospital stay [15, 23, 24].

Surprisingly, EVLW1 increased slightly but significantly during dobutamine infusion and decreased slightly but significantly during bleeding, while EVLW2 did not change (Table 2). From a pathophysiological point of view, no change in lung water is expected during inotropic stimulation or hypovolemia, particularly over such a short period of time [25]. Since our study was not designed to compare the PiCCO™ method and the VolumeView™ method with a third reference method (such as gravimetry), however, we cannot draw any definitive conclusions regarding the superiority of one method over the other.

Our study also confirms the very good reproducibility of TPTD measurements. These findings are in line with previous studies [5, 26] reporting reproducibility of CO, GEDV and EVLW of 4 to 7%, 5 to 8% and 11%, respectively.

Study limitations

The gravimetric method in animals and the double indicator (cold green dye) dilution method in humans are considered gold standard methods to quantify EVLW [9, 10]. The goal of the present study was to compare the new VolumeView™ system with the TPTD system currently in clinical use - this is why the PiCCO™ system has been selected as the reference method in our study. A clinical validation is necessary to investigate whether the new VolumeView™ system is also comparable with the PiCCO™ system in critically ill patients. The new VolumeView™ algorithm was originally developed to decrease the sensitivity of TPTD to recirculation and thermal baseline drifts. The present study was not designed to investigate this potential advantage over the existing TPTD technology, but instead to ensure that the new VolumeView™ system and the PiCCO™ system are interchangeable in clinical-like conditions where CO, blood volume and lung water vary significantly. Further studies are therefore required to compare both systems in situations where technical (thermal baseline drift) or other clinical challenges (for example, valvular regurgitation-induced recirculation) are encountered.

Conclusions

In animals, and over a very wide range of values, the new TPTD VolumeView™ system is comparable with the current PiCCO™ system to assess CO, GEDV and EVLW during inotropic stimulation, acute hemorrhage, fluid overload and severe acute lung injury.

Key messages

  • TPTD is increasingly used for hemodynamic evaluations in critically ill patients.

  • The TPTD method currently in clinical use and implemented in the PiCCO™ system (Pulsion Medical Systems) is based on mathematical models described in the 1950 s.

  • A new and original method has recently been developed to derive GEDV and EVLW from a TPTD curve (VolumeView™; Edwards Lifesciences).

  • In animals, and over a very wide range of values, the new transpulmonary thermodilution VolumeView™ system is comparable with the current PiCCO™ system to assess CO, GEDV and EVLW during inotropic stimulation, acute hemorrhage, fluid overload and severe acute lung injury.