Introduction

Breast cancer remains a significant cause of morbidity and mortality in women internationally [1]. Proton magnetic resonance spectroscopy (1H-MRS) represents a non-invasive and non-ionising method of characterising a user-selected volume of tissue on the basis of the metabolic (chemical) content. 1H-MRS has been evaluated as an adjunct modality to breast MRI for increasing the specificity in differentiating malignant from benign tumours, in addition to its potential in monitoring responsiveness to chemotherapy.

This article aims to review choline related 1H-MRS in breast assessment, and to discuss the qualitative and quantitative approaches to 1H-MRS in breast cancer investigation. Both approaches to the technique will be described, along with a summary of the reported studies.

Methods

English language studies investigating single-voxel (SV) or multi-voxel evaluation of the total choline containing resonance using in vivo 1H-MRS assessment of pre-operative local or advanced breast cancer in human adult females, with or without additional assessment of benign or normal breast tissues, were sought. MEDLINE and EMBASE were searched for the period January 1990 to February 2012 using English language restrictions. The reference lists of relevant prior reviews were examined. Authors were contacted to identify relevant additional studies. Only full peer-reviewed articles were included. Sensitivity and specificity values in this review are expressed in terms of the percentage of correctly diagnosed malignant and benign lesions, respectively.

Breast cancer physiology and 1H-MRS

1H-MRS assessment of the human breast demonstrates a number of distinct resonances attributable to choline, glycerides (esters of fatty acids and glycerols), saturated and unsaturated fatty acid, and water [2]. Numerous breast 1H-MRS studies performed in vivo have reported the association of a resonance at approximately 3.2 parts per million (ppm) with malignancy [28]. High-resolution high-field strength ex vivo analysis has revealed that a number of chemical compounds contribute to this single peak, including major contributions from free choline (3.19 ppm), phosphocholine (3.21 ppm), and glycerol-phosphocholine (3.22 ppm), as well as minor contributions from phosphoethanolamine (3.23 ppm), glucose (3.26 ppm), taurine (3.25 ppm), and myoinositol (3.27 ppm) [9]. With in vivo analysis these peaks undergo line-broadening and superposition and are generally represented by a single combined resonance at clinical field strengths as high as 4 Tesla (T). Due to the major contribution of cholinecontaining metabolites to the composite signal at 3.2 ppm, it is commonly referred to as the total choline containing resonance (tCho). Figure 1 demonstrates a breast cancer on MRI and the corresponding SV 1H-MRS spectrum acquired.

Figure 1
figure 1

(a) A post-contrast gradient echo (TR/TE = 4.4/1.1 ms) 1.5 T MRI from a 51-year-old woman with invasive ductal carcinoma demonstrating single-voxel 1H magnetic resonance spectroscopy (1H-MRS) voxel positioning. (b) The corresponding water and lipid suppressed single-voxel 1H-MRS spectrum acquired from the voxel, indicating the resonances present. Spectroscopy parameters: PRESS single voxel 15 × 17 × 20 mm, TR/TE = 3,000/125 ms, 128 averages, CHESS water suppression, MEGA/BASING lipid suppression. ppm, parts per million; tCho, total choline-containing resonance.

The relative concentrations of certain cellular metabolites change when cells transform from the normal state to a malignant form. Specifically, metabolic variations in the cell membrane resulting from proliferation, principally involving phosphocholine, manifest as variations in the concentration of choline-containing molecules. The tCho amplitude as determined by in vitro 1H-MRS has shown positive correlation with the metabolic proliferative activity of malignant cells [10]. In vivo investigation has reported that at least an order of magnitude more phosphocholine is present in breast cancer cells than normal mammary epithelial cells [11]. However, two in vitro studies [12, 13] show that choline-containing metabolite concentrations remain low in cultured normal human mammary epithelial cells when they proliferate at a similar rate to tumour-derived lines. This suggests that proliferative changes alone cannot completely account for the elevated choline metabolic activity seen in tumours.

Single-voxel spectroscopy

SV 1H-MRS is considered the most suitable method for the assessment of distinct individual lesions. Adequate shimming has been shown to be more feasible with this approach. A major disadvantage with SV versus multivoxel spectroscopy is that SV acquisitions permit only a single lesion to be evaluated at a time. In addition, there is the inherent inability of SV methods to demonstrate spatial variability in biochemistry over a large area of heterogeneous tissue [14]. SV spectroscopy also requires lesion location to be known for accurate voxel placement; therefore, spectroscopic sequences should follow contrast injection and require a radiologist or technician to decide on appropriate voxel placement before 1H-MRS sequences.

Multi-voxel spectroscopy

Known either as chemical shift imaging or magnetic resonance spectroscopic imaging (MRSI), this technique enables the simultaneous acquisition of a grid of multiple spectroscopic voxels using gradient encoding. A major benefit of this method is the capacity to acquire a matrix of multiple spectra within a slice of anatomical structure such as the brain or breast, thereby enabling 'mapping' of spatial variations of in vivo metabolites. This may be advantageous in demonstrating cancer infiltration or margins. The capability to evaluate multiple lesions simultaneously is significant given that dynamic contrast enhanced MRI (DCE-MRI), which is generally performed immediately prior to 1H-MRS, can demonstrate previously unknown additional lesions. The greater spatial coverage volume afforded orded by MRSI, as well as the ability to retrospectively shift the voxel grid to better align with a lesion of interest, makes it a more amenable technique for pre-contrast 1H-MRS acquisitions where the lesion location is uncertain. MRSI also makes it possible to simultaneously assess normal breast parenchymal tissue (perhaps with the potential for referencing). In practice, however, several issues present significant challenges to the installation of MRSI. One of the most important such factors is the difficulty associated with achieving adequate shimming over such a large volume of tissue, to the extent required for robust simultaneous spectroscopic acquisitions from each voxel within the matrix [14]. The spatial localisation is typically not as precise as in SV, which increases partial volume errors. Further-more, the quantitative analysis approaches (described below) are more difficult to implement for MRSI than SV due to greater receive coil variation and the time needed to acquire an internal water reference spectrum.

Several breast MRSI studies have been conducted to date [1522]. Potential implementations of the technique are highlighted with the following two publications. A 2008 1.5 T MRSI study by Hu et al. [15] involving ten women with known or suspected breast cancer reported the use of a tCho signal-based colour-coded map of a breast lesion to detect an intra-lesion choline 'hot spot'. Such a demonstration highlights the potential for MRSI to be used as a method for selecting the optimal site for breast biopsy.

A 2006 1.5 T MRSI study by Su et al. [16] involving 14 breast cancer patients investigated the correlation between tCho signal-to-noise ratio (SNR) and DCE-MRI vascular attributes. They reported a significant linear correlation of tCho SNR with the percentage of enhancement two minutes following contrast administration (P = 0.002), as well as with the pharmacokinetic properties Ktrans (dynamic contrast enhanced derived transfer coefficient; P = 0.003) and Kep (exchange rate constant; P = 0.002). These results suggest that there is a relationship between choline metabolic activity and angiogenic activity. As choline is involved in cellular proliferation, it is logical that angiogenesis increases to support tumour metabolic requirements.

Qualitative approach to tCho

Roebuck et al. [3] suggested the potential of utilising tCho as a biomarker of breast malignancy in 1998. The qualitative approach to tCho, first reported for use in in vivo breast assessment by this team, involves the subjective determination by an observer as to whether a distinct resonance at approximately 3.2 ppm is present. No objective statistical analysis of the spectrum, in terms of SNR or tCho signal amplitude, is performed within the studies using this method. To date, 13 studies [35, 18, 2331] using this approach to breast lesion evaluation have been published. All of these investigations were per-formed using 1.5 T machines. A summary of the reported qualitative literature is documented in Table 1.

Table 1 Summary of qualitative 1H-MRS literature published to date

Analysis by Katz-Brull et al. [11] of three of the qualitative studies [3, 24, 25] examined the sensitivity of cancer diagnosis with regard to lesion size. The study data were stratified into three groups by the size of the examined lesion (<2.5, 2.5 to 4.9, and ≥5 cm). It was noted that the sensitivity of qualitative tCho detection, and hence cancer diagnosis, significantly increased with increasing lesion size from 72% to 90% to 100% (P = 0.025). Hence, smaller cancers have a greater tendency to be diagnosed as benign due to insufficient tCho signal for detection.

In order to improve the sensitivity for qualitatively detecting smaller cancers, and likewise with the utility of the quantitative methods discussed later, it is necessary to increase the SNR. In order to increase the tCho signal, the use of clinical MR machines with field strengths of greater than 1.5 T, which have increasing installation in centres worldwide, could be used. The 2003 4 T study by Bolan et al. [32] was the first to report the technique at higher field strength. In terms of additional hardware factors, innovations in specialised MR breast coil design and sensitivity might also improve the detectability of tCho. With regard to 1H-MRS parameters, the signals within the chemical shift range of tCho can be detected with greater sensitivity using optimised MR pulse sequences.

Quantitative approaches to tCho

Several different quantitative approaches to tCho processing have been reported. Compared to qualitative approaches, quantitative results can be more objectively compared between different sites and systems, allow for stronger statistical analyses, and can be used to measure change in longitudinal studies, such as response to chemotherapy. There is only one study investigating tCho amplitudinal scan-rescan reproducibility, reporting non-significant variability, though no publications to date have reported multi-site consistency for any of the qualitative or quantitative approaches [27]. There is a great need to assess the reproducibility of the modality, especially with regard to the absolute values of choline concentration obtained using the quantitative approaches. Quantitative approaches using the tCho peak integral, SNR and referencing methods are discussed below.

tCho peak integral

A 2009 1.5 T SV study by Sardanelli et al. [33] reported the use of the tCho peak integral for discriminating malignancy from benign tissue. Using spectra from 45 lesions, 89.5% sensitivity and 92.3% specificity were attained for the differentiation of malignant from benign breast lesions. Whilst this approach demonstrates an impressive level of diagnostic accuracy, the authors note that, due to the expression of the tCho peak interval in terms of arbitrary units, it may not be possible to use the reported thresholds in differing technical and clinical situations.

tCho signal-to-noise ratio approach

This methodology involves the determination of the SNR of the spectral region about which the tCho peak would be expected (approximately 3.2 ppm). This entails firstly quantifying the tCho peak signal amplitude, then sampling a region of the spectrum at which no signal would be expected (for example, >6 or <0 ppm) to determine noise intensity. If the resulting SNR is higher than a study defined cut off value (for example, SNR ≥2), then the resonance at approximately 3.2 ppm is assumed to correspond to tCho.

The qualitative and SNR techniques are both based upon the premise that if sufficiently large concentrations of choline-containing metabolites are present within a given voxel of interest (VOI), then a tCho peak is more likely to be observed/detectable. However, several important issues confound this assumption, since the sensitivity to detect tCho signals vary, depending on the following factors. The sensitivity of MR signal detection is directly proportional to the volume of the VOI and approximately proportional to increasing magnetic field strength (B0). SNR also depends upon the specific breast coil design used (widely variable according to model), the VOI positioning in relation to the breast coil elements, and the variable coil loading resulting from diverse patient habitus [34].

Distinct from the sensitivity of detection, several other factors also influence the acquired signal intensity of a tCho resonance, including the intrinsic T1 and T2 relaxation times of the various chemical components contributing to the resonance, the sequence repetition time (TR) and echo time (TE) values, the capability of shimming to optimise the homogeneity of B0 over the VOI, as well as acquisition parameters such as spectral width and receiver gain.

Nine studies investigating the SNR method in breast cancer assessment have been published to date. All of these investigations were performed using 1.5 T machines. Table 2 summarises the reported SNR method literature.

Table 2 Signal-to-noise ratio approach results published to date

Externally phantom referenced tCho approach

The external phantom reference method (EPRM) involves the quantification of tCho, to produce a molar concentration value, based upon the ratio of the tCho signal amplitude acquired from an in vivo VOI to that from a solution of known phosphocholine concentration within a 'phantom' container positioned outside the patient's body. The advantages of this method compared to the use of an internal reference such as endogenous water or lipid are with regard to it being a more reliable referencing method with less of the operator dependence (VOI positioning to avoid inclusion of adipose tissue) and physiologic dependence (tissue water content can vary with the menstrual cycle) [35] inherent to internal water referencing. One disadvantage of this approach is the additional time required to perform 1H-MRS separately on the external standard phantom (eight minutes in the Bakken study) [36] with clear repercussions on patient throughput and cost-effectiveness. Additionally, this method may require the acquisition of a sensitivity map, which also increases the time required.

To date, two studies have reported use of the EPRM in the evaluation of breast cancer in vivo. The first study to look at the quantification of tCho in breast cancer in this manner was performed in 1998 at 1.5 T by Roebuck et al. [3]. This group used an external phantom quantification strategy, which involved performing 1H-MRS on a phantom solution of known choline concentration (1 mmol/L) following the in vivo spectroscopic evaluation of the lesion of interest. The tCho SNR was then calculated both in vivo and in the phantom solution. The concentration of choline in each VOI was determined after voxel volume and full width at half maximum (FWHM) values were adjusted for. A quantifiable tCho peak was detected in 7 out of 10 malignancies, with the in vivo choline concentration ([Cho]) ranging from 0.7 to 2.1 mM using VOI volumes ranging from 1 to 1.8 ml in the cancerous lesions. However, a tCho resonance was also detected in a patient with a rare benign lesion, tubular adenoma. This particular lesion demonstrated a [Cho] of 5.8 mM from a 9.8 ml VOI.

A second group employed the EPRM at 1.5 T in a single patient with known breast carcinoma [36]. Table 3 summarises the reported EPRM literature.

Table 3 In vivo external phantom reference method breast cancer literature published to date

Internal water referenced tCho approach

A recent approach in quantitative breast 1H-MRS first reported by Bolan et al. [32] involves the use of the non-suppressed water resonance from the same in vivo VOI used for the suppressed tCho detection acquisition. This technique shall further be referred to as the internal water reference method (IWRM). Variations of this internal referencing approach are performed routinely in the United States for brain 1H-MRS studies. When implementing the IWRM, the [Cho] is calculated by means of a ratio of the tCho amplitude to the un-suppressed water amplitude. This ratio can be interpreted directly as a molal quantity (moles per mass of water), or can be converted to a molar concentration (moles per volume) by assuming a water density. Adjustments are made for differences in receiver gain settings, the number of signal averages (NSA), and MR signal relaxation rates. A large NSA is required for the water and lipid suppressed 1H-MRS sequence acquiring the tCho resonance due to the need to resolve the relatively low amplitude tCho resonance from the baseline noise amplitude. A low NSA is required for the 1H-MRS sequence acquiring the unsuppressed water spectrum due to the dominating signal amplitude of water, which is normally resolvable from the baseline noise with a single excitation.

The IWRM gives slightly higher but relatively similar [Cho] results compared with the EPRM. One of the benefits of the IWRM is that it is more straightforward to implement, with no VOI repositioning required and all 1H-MRS acquisitions being carried out during the patient examination protocol. It also intrinsically adjusts for radiofrequency transmission efficiency and radio-frequency receive sensitivity, in addition to VOI size, B0 shim effects, and the partial volume effect resulting from the inclusion of adipose tissue [37].

The IWRM has been used to investigate cancerous breast lesions in seven studies at a range of clinical grade field strengths. A summary of the IWRM literature is detailed in Table 4.

Table 4 In vivo internal water reference method breast cancer literature published to date

Breast lesion diagnosis when combining MRI and SV 1H-MRS

In 2005 Meisamy et al. [38] reported a four reader blinded observer performance study (n = 55) performed at 4 T evaluating the addition of the IWRM to DCE-MRI for breast lesion assessment. The sensitivity and specificity of diagnosis using tCho concentration alone, with a cut off threshold of 1.05 mmol/kg (as determined by receiver operating characteristic (ROC) curve analysis) were 69% and 90%, respectively. When the DCE-MRI and 1H-MRS findings were used in conjunction, the sensitivity and specificity were reported to be 94% and 57%, respectively. The sensitivity and specificity of DCE-MRI increased from 87% to 94% and 51% to 57%, respectively, once 1H-MRS results were accounted for.

The 2004 SNR paper by Huang et al. [6] also reported the sensitivity and specificity of breast cancer detection when DCE-MRI, SNR method SV 1H-MRS and T2* weighted perfusion MR imaging results were combined within the examination. DCE-MRI alone demonstrated 100% sensitivity and 62.5% specificity. Specificity improved to 87.5% when 1H-MRS findings were integrated and increased further to 100% once perfusion MR results were considered. This clearly highlights the benefit of incorporating secondary MR modalities into the routine breast MR examination with regard to increasing specificity in breast lesion diagnosis.

1H-MRS in assessing tumour responsiveness to chemotherapy

Neoadjuvant chemotherapy (NACT), consisting of systemic agents provided preoperatively, increases the likelihood of breast conserving surgery in lieu of mastectomy when used in a patient with a chemoresponsive tumour. A therapeutic response associated with a reduction in tumour size may permit downstaging of the disease and consequently enable excision of what previously were inoperable lesions. In addition, this mode of treatment enables the tumour response to be monitored in vivo. The total disappearance of tumour at the time of resection is associated with the most favourable overall survival [39].

Given the wide range of response to chemotherapeutic agents indicated for breast cancer, it is useful to identify the e ectiveness of a particular regimen at an early stage in the treatment. The methods used to determine breast cancer responsiveness to NACT can give variable results and, at present, the traditional methods of clinical examination, mammography and ultrasound, are used routinely in the clinic. Clinical MRI is playing an increasing role in the evaluation of invasive breast cancer, particularly in the trial setting.

With respect to NACT monitoring, there is a correlation between clinical response and changes in MRI-derived characteristics such as lesion size and dynamic contrast enhancement [40, 41]. Unfortunately, however, such alterations cannot be identified until several weeks into a regimen, thus inhibiting the ability to optimally tailor treatment to patients on the basis of demonstrable antitumour activity [42, 43].

As alluded to, the capability to determine tumour responsiveness to a particular chemotherapeutic regimen in a timely manner would be highly preferable, enabling superior tailoring of a NACT regimen with the objective of achieving a full pathological response. The use of 1H-MRS for this purpose has been the subject of relatively small-scale investigation to date.

The first study to report a change of the tCho signal in breast cancer in response to NACT was conducted in 1999 by Kvistad et al. [4] at 1.5 T. The qualitative SV methodology used demonstrated that a single patient with an observable tCho resonance from their invasive ductal carcinoma at pre-treatment baseline 1H-MRS examination no longer had a detectable tCho signal following that regimen. This finding was accompanied by a clinical response marked by a significant reduction in tumour diameter (5.7 to 3.7 cm) as determined using DCE-MRI.

The next study, by Jagannathan et al. [5] at 1.5 T in 2001, reported sequential qualitative SV 1H-MRS findings both at pre-treatment baseline and within the first NACT regimen. Ten patients demonstrated an observable tCho at pre-treatment baseline, but subsequently had either absent or significantly reduced tCho one week into the regimen. These spectral findings correlated with both a clinical and histopathological response in seven of these ten cases. Both of these initial studies suggest that the qualitative assessment of tCho may have a role in monitoring the chemoresponsiveness of a tumour to a NACT regimen.

The first SV 1H-MRS study to investigate the quantification of tCho as a biomarker of therapeutic response to NACT was performed using the IWRM at 4 T in 2004 by Meisamy et al. [44]. The small cohort involved, comprising 13 participants, consisted of women with breast cancer scheduled to undergo a doxorubicin (anthracycline)-based NACT regimen. The difference in [Cho] from pre-treatment baseline to one day following the initial doxorubicin administration demonstrated a statistically significant (P = 0.001) positive correlation with the difference in tumour size after four treatment cycles compared to baseline.

A 2009 1.5 T SV IWRM study by Baek et al. [45] assessed 35 breast cancer patients from pre-NACT baseline to the first follow-up assessment following one to two cycles of doxorubicin and cyclophosphamide (DC) and a second assessment following either two further cycles of DC or a taxane-based regimen. They reported that, at second follow-up, patients who went on to have a pathologically complete response were more likely to demonstrate a significantly greater reduction in tCho concentration relative to the change in tumour size than those who had an incomplete response.

A small-scale 2010 1.5 T SV EPRM study by Tozaki et al. [46] assessing the response of 16 patients with breast cancer after two cycles of anthracycline-based NACT reported that there was a statistically significant difference (P = 0.004) in normalised choline signal between pathological responders and non-responders. A 2008 in vivo EPRM study by the same group [47] compared the ability of the integral tCho signal to evaluate early responsiveness of breast cancer to NACT compared with use of the standardised uptake value with 18F-fluoro-deoxyglucose positron emission tomography (PET). They reported that 1H-MRS is significantly correlated with the peak standardised uptake value both during and following NACT cycles (P = 0.03 and P < 0.001, respectively), demonstrating that 1H-MRS may represent a viable alternative modality to sequential PET examinations for this purpose.

The first study to investigate the use of MRSI in assessing chemoresponsiveness was reported in 2010 by Danishad et al. [20]. The SNR approach was implemented in 25 patients at 1.5 T. Using ROC analysis, they reported that a baseline to post-third NACT cycle reduction in tCho SNR threshold of 53% showed a sensitivity and specificity of 85.7% and 91%, respectively, for differentiating clinical responders from non-responders. Clinical response was represented by a 50% reduction in tumour volume (baseline to post-third NACT cycle), as assessed by lesion size on palpation. The mean reduction in tCho SNR following one cycle of NACT was 27.8 ± 7.8% for clinical responders and 8.6 ± 3.9% for non-responders.

Such preliminary results highlight a promising method by which to approximate the eventual clinical response of a tumour to NACT. Larger-scale trials are necessary in order to further evaluate the utility of 1H-MRS for this purpose. Results from the recently completed North American multicentre ACRIN 6657 trial, which used 1H-MRS and DCE-MRI to assess NACT response, are still pending.

Impact of gadolinium-based contrast agents on 1H-MRS

The majority of previous in vivo 1H-MRS studies involved the identification and localisation of one or more enhancing lesions based upon post-contrast imaging. However, this approach assumes the impact of gadolinium-based contrast agents on spectroscopic results is negligible.

A 2009 study performed at 3 T by Lenkinski et al. [48] investigated the spectroscopic effect of six commercial gadolinium-based contrast agents (Dotarem®, Magnevist®, MultiHance®, Omniscan®, OptiMark® and ProHance®), all of which are currently approved for intravenous human use during MR examinations in the UK. This study examined the impact each of these agents had on the tCho peak using both phantom and rat cancer models.

Three of these agents (Omniscan®, OptiMark® and ProHance®) possess an overall neutral charge under physiological pH conditions, whilst the other three (Magnevist®, MultiHance® and Dotarem®) have a net negative (ionic) charge. Only the three ionic agents were found to bind, in an ionic manner, with choline. The interaction of these three particular agents with choline resulted in a mean reduction in the in vivo spectral tCho peak area of 40%, whilst the neutral agents showed minimal or no effect. There was a statistically significant difference in the average decrease in tCho peak area between the two agent groups in vivo (P < 0.001). The use of the aforementioned negatively charged agents might therefore result in underestimated choline levels in patients with underlying breast cancer. On this basis, the authors went on to recommend only the administration of neutral agents for breast MR examinations involving post-contrast image localised 1H-MRS protocols.

A 2011 in vivo study by Baltzer et al. [30] compared ionic (Gd-DTPA; Magnevist®) and neutral (Gd DTPA-BMA; Omniscan®) contrasts, each at a concentration of 0.1 mmol/kg of bodyweight. All 1H-MRS sequences were performed 10 to 12 minutes post-administration. Significantly (45%) lower tCho signal intensity was reported in women who were administered ionic contrast, reflecting an agreement with the Lenkinski et al. breast cancer-bearing rat results reporting a 40% decrease in tCho signal. No increase in false negative rates with qualitative 1H-MRS was reported; however, it is important to emphasise the significance of this contrast agent charge related bias with regard to the implementation of quantitative lesion assessment and inter-study comparisons.

A 2005 in vivo study by Joe et al. [27] reported disadvantageous effects of the neutral charge Omniscan® on breast cancer lesion tCho resonance when 1H-MRS was performed pre- and post-contrast. There was a significant (P < 0.0001) decrease in tCho amplitude 15 minutes following a 20 ml bolus of Omniscan®, with an average decrease of 20% (range 3.3 to 52%). This study also reported non-significant variations in tCho amplitude from examinations performed on separate days over a seven-day period, suggesting a greater impact of contrast on tCho amplitude compared to the variability of repeated measurements.

Given that usage of certain gadolinium-based contrast agents can be associated with a reduced tCho peak, there could be the potential for failing to detect malignancy or overestimating tumour responsiveness to NACT. A 2011 3 T study by Kawai et al. [49] investigated whether the SV 1H-MRS VOI could be positioned accurately on breast lesions before gadolinium-based contrast agent administration, using a combination of diffusion weighted imaging (DWI) and fat suppressed T2 weighted imaging (T2WI). Using DCE-MRI performed later in the same MR session as the reference, the VOI was positioned entirely within the target lesion using DWI/T2WI in 64 (65%) of 98 lesions. This group also compared the tCho resonances from the breast cancer lesions both before and after administration of Magnevist® (gadopentetate dimeglumine) in 44 of the patients with good precontrast VOI positioning. They reported that the integral of the tCho peak was significantly greater in the precontrast versus post-contrast 1H-MRS acquisitions (P < 0.001).

Conclusion

1H-MRS using the tCho biomarker can provide clinically beneficial additional information in the context of breast lesion diagnosis and remains an area under active research. This adjunct modality involves the appending of at least an additional 10 minutes to the existing breast MR examination duration, with obvious repercussions on patient comfort and suitability. A growing number of higher field strength commercial clinical MR systems featuring 1H-MRS capabilities are being implemented. Further large-scale multi-centre and model-based research geared towards examining reproducibility is imperative. Once robustly established, the technique has the potential to increase the clinical value of the breast MR examination.