Background

Positron emission tomography (PET) measurements using 18 F]-2-fluoro-2-deoxy-d-glucose (18 F]FDG) allow for quantitative determination of the cerebral metabolic rate of glucose utilization (CMRglu) [1, 2]. Although, initially, rate constants were measured [35] using the compartment model developed by Sokoloff and co-workers [6], thereafter, CMRglu measurements were increasingly based on the assumption of fixed rate constants in combination with static scans [7]. As a result, no recent estimates of the rate constants obtained using a state-of-the-art high-resolution scanner are available for human studies (estimates have only been reported for mice; see [8]).

Quantification of regional cerebral blood flow (CBF) based on 15O]H2O and PET also started around the same time period as metabolic assessments, i.e., around 1980. Due to hardware limitations, scanners were too slow for fast dynamic scans, leading to steady-state [9] and autoradiographic protocols [10]. Dynamic scan protocols were not introduced until the late 1980s [11, 12], but to date, relatively few methodological papers on fully dynamic implementations have been published. In a review by Silverman and Phelps, it was noted that CBF measurements showed more variability than CMRglu measurements with standard deviations within investigations of around 10% to 20% and up to 100% between investigations [13].

The high-resolution research tomograph (HRRT; CTI/Siemens, Knoxville, TN, USA) is a dedicated brain scanner that combines high spatial resolution (approximately 3 mm) with good sensitivity [14]. Previously, it has been shown that the increased spatial resolution allows for measurements of CMRglu with reduced underestimation due to partial volume effects [15, 16]. These glucose measurements, however, were based on a static FDG protocol together with population average-based values for the rate constants. In a recent study, HRRT image quality was shown to be good enough to allow for quantitatively correct CBF measurements [17].

The use of an arterial line (blood sampler derived input function (BSIF) to acquire an input function is the gold standard for dynamic data analysis of cerebral studies, but it is an invasive procedure. Therefore, use of an image-derived input function (IDIF) can be an interesting alternative [1820], but its utility needs to be validated for each tracer, each scanner, and each acquisition and data analysis protocol separately.

The main purpose of the present study was to derive human CBF and CMRglu values as measured using a current state-of-the-art high-resolution scanner. Data were analyzed by full kinetic analysis using nonlinear regression (NLR) of regional time-activity curve (TAC) data with a BSIF. A second objective was to assess the accuracy of high-resolution parametric images of CBF and CMRglu. In addition, CBF and K i values for automatically delineated regions were compared with manually drawn regions of interest and literature values. Based on earlier work on IDIFs [18, 20, 21] and given the high resolution of the HRRT, an additional objective was to assess whether a carotid artery-based IDIF could be used as a noninvasive alternative for arterial sampling in the case of both CBF and CMRglu measurements, thereby increasing clinical applicability of this methodology in humans.

Methods

Healthy subjects and study design

Thirteen healthy men (age 36.2 ± 13.2 years, body mass index 25.8 ± 3.7 kg/m2, arterial plasma glucose 5.5 ± 0.2 mmol/L) participated in this study. The study consisted of a screening visit and two visits for magnetic resonance imaging (MRI) and PET scan acquisition, respectively. All subjects were free of medical and psychiatric illness based on medical history, physical examination, and blood analysis. All subjects provided written informed consent prior to inclusion. The study was approved by the Medical Ethics Review Committee of the VU University Medical Center and the Central Committee on Research Involving Human Subjects; the study was conducted according to the Declaration of Helsinki.

Scan protocol

One week prior to the PET study, 3-D structural MRI images were acquired using a 3.0 T GE Signa HDxt MRI scanner (General Electric, Milwaukee, WI, USA) and a T1-weighted fast spoiled gradient echo sequence. Data consisted of 172 planes of 256 × 256 voxels with a voxel size of 0.94 × 0.94 × 1 mm3.

At the day of the PET study, catheters were placed in the antecubital vein for tracer injection and in the radial artery for blood sampling. Next, subjects were positioned on the HRRT scanner bed such that the head was in the center of the field of view. Velcro tapes were used to minimize patient movement during the entire imaging procedure. Lights were dimmed, noise was minimized, and subjects were asked to close their eyes and stay awake during data acquisition. Prior to or immediately after the 15O]H2O scan, a 6-min singles-based transmission scan with a fan-collimated 137Cs moving point source was acquired. A 10-min dynamic emission scan was started 10 s prior to a bolus injection of approximately 800 MBq 15O]H2O. Radioactivity was contained within a volume of approximately 5 mL and was followed by a saline flush to give a total injected volume of 40 mL, administered at a rate of 2 mL/s using an infusion pump (Medrad Inc., Indianola, MS, USA). To allow for radioactive decay of 15O, the 18 F]FDG scan was acquired after an interval of at least 10 min following the 15O]H2O scan. A 60-min dynamic emission scan was started 30 s prior to a bolus injection of 186 ± 9 MBq 18 F]FDG. The administration protocol was identical to the one used for the 15O]H2O scan. During both emission scans, arterial blood concentrations were monitored continuously using a dedicated on-line blood sampler (Veenstra Instruments, Joure, Netherlands [22]). During the first 4 min of the 15O]H2O scan, blood was withdrawn at a rate of 450 mL/h and activity was read out once per second. For the remaining 6 min, the withdrawal rate was reduced to 150 mL/h and the readout sampling time increased to 10 s. In addition, three manual blood samples were taken at 5, 7.5, and 9 min post-injection. These samples were taken from the same arterial line by briefly interrupting continuous withdrawal. After each sample, the arterial line was flushed with heparinized saline to prevent clotting. Manual samples were used to measure whole blood radioactivity concentrations. The sampling procedure during the 18 F]FDG scan was similar to that of the 15O]H2O scan but with withdrawal rates of 300 mL/h during the first 5 min and 150 mL/h thereafter. During this scan, manual blood samples were taken 15, 35, and 55 min post-injection. These samples were used to measure both whole blood and plasma radioactivity concentrations, as well as arterial plasma glucose levels. This BSIF was used in all analyses except for analyses based on an IDIF.

Data reconstruction

Emission data were histogrammed into multiframe sinograms (15O]H2O 14 fames of 6 × 10, 2 × 30, 4 × 60, and 2 × 120 s; 18 F]FDG 19 frames of 6 × 10, 2 × 30, 3 × 60, 2 × 150, 2 × 300, and 4 × 600 s). Sinograms were normalized, corrected for randoms, dead time, and decay, and based on the transmission scan corrected for scatter and attenuation. Corrected sinograms were reconstructed using the iterative 3D ordinary Poisson ordered subset expectation maximization algorithm [23, 24] using eight iterations and 16 subsets. Reconstructed images consisted of 207 planes of 256 × 256 voxels with a voxel size of 1.22 × 1.22 × 1.22 mm3.

Regions of interest definition

The MRI image was co-registered with either a corresponding summed 18 F]FDG (15 to 60 min) or 15O]H2O (0 to 90 s) image using the software package VINCI [25]. Both PET and MRI images were rebinned and cropped into 128 × 128 × 63 matrices with an isotropic voxel dimension of 2.44 mm. Automatic delineation of regions of interest (ROIs) was performed using PVElab [26] resulting in a total of 17 gray matter regions, subdivided into their left and right constituents (cerebellar cortex, orbital frontal cortex, inferior medial frontal cortex, anterior cingulate cortex, thalamus, insula, caudate nucleus, putamen, superior temporal cortex, parietal cortex, inferior medial temporal cortex, superior frontal cortex, occipital cortex, sensory motor cortex, posterior cingulate cortex, enthorinal cortex, and hippocampus), a global white matter (WM) region, and a total gray matter (GM) region. To compare our data to literature values, two additional manually drawn ROIs were analyzed: a gray matter region (insular gray in four successive transversal planes) and a white matter region (centrum semiovale in two successive transversal planes), using Amide [27]. Corresponding TACs were generated by projecting these ROIs onto the dynamic image sequences.

Calculation of CMRglu

Sampler data were corrected for flushes and calibrated using the plasma concentrations derived from the three manual samples per subject to generate an arterial plasma input function. First, using NLR, TACs were fitted to the standard irreversible two-tissue compartment model, providing the three rate constants K 1, k 2, and k 3 as well as the blood volume fraction V B. Second, the validity of the Patlak linearization [28] was investigated by comparing regional values of the net influx rate constant K i with those obtained using NLR. Third, the Patlak method was used (without smoothing) on a voxel-by-voxel basis, and for each ROI, mean values extracted from parametric K i images were compared with those obtained from regional Patlak analyses. Additionally, parametric images were smoothed with a 6-mm Gaussian filter (typical resolution of the current-generation whole-body PET scanners) prior to analysis. CMRglu was calculated as K i times arterial plasma glucose divided by a lumped constant of 0.52 [29].

Calculation of CBF

Whole blood input functions as well as regional TACs were derived as described above. First, TACs were fitted to the standard single-tissue compartment model, fixing delay and dispersion to the values obtained from a fit to the whole brain TAC [12], providing CBF as well as V T, distribution volume. Second, after smoothing with a 6-mm Gaussian filter, this analysis was repeated, and in addition, parametric CBF images were generated using a basis function method (BFM) implementation of the blood flow model [30]. For each ROI, mean parametric CBF was compared with the corresponding value from the regional analysis. The FDG extraction fraction was calculated from combined CBF and CMRglu data as K 1 = E · CBF.

Image-derived input functions

BSIFs are invasive, and therefore, the use of IDIFs was investigated. For FDG, arterial activity was clearly seen after smoothing an early frame (approximately 20 to 30 s post-injection) with a 6-mm Gaussian filter. Starting three planes below the circle of Willis to avoid contamination with activity from the brain, ten successive planes were combined into an ROI representing a carotid artery [18, 19]. The four hottest pixels per plane were identified and combined in a carotid artery ROI. The average time-activity curve obtained from the two carotid artery ROIs, scaled to the manual samples, was taken as the whole blood IDIF. A plasma IDIF was obtained by multiplying the whole blood IDIF with the average plasma-to-whole blood ratio derived from the samples. For 15O]H2O, an identical procedure was followed.

Results and discussion

Results

[18 F]FDG scans were acquired in all 13 subjects and [15O]H2O scans in 12 subjects (at one occasion, no [15O]H2O was available). Analysis of one [18 F]FDG scan failed due to sampling problems, and two [18 F]FDG scans were discarded because of movement artifacts. Due to technical problems, [18 F]FDG and [15O]H2O data could not be obtained for one subject. Overall, 9 complete [18 F]FDG and 11 complete [15O]H2O data sets were available for analysis, and 8 combined data sets could be used for the calculation of the FDG extraction fraction.

CMRglu

No significant differences between left and right CMRglu values were observed for any of the regions delineated automatically. Average (left and right) CMRglu values are listed in Table 1. Total gray matter CMRglu was 0.29 ± 0.03 μmol/cm3 per minute, and white matter CMRglu was 0.19 μmol/cm3 per minute (ratio GM/WM = 1.5). Excluding the entorhinal cortex, an average coefficient of variation (CoV) of 10.0% was observed. The high CoV in the entorhinal cortex is probably due to the small volume of this region (3.4 ± 0.4 mL), and this region was excluded from the further analyses. A typical blood volume fraction, V B, of 0.05 ± 0.01 was obtained.

Table 1 Regional CMR glu , K 1 , CBF, V T , and E values obtained in healthy males

Manually drawn ROIs resulted in CMRglu values of 0.33 and 0.11 μmol/cm3 per minute for gray matter and white matter, respectively (ratio GM/WM = 3.0). In Table 2, mean values of the separate rate constants for these manual ROIs are listed as well as a comparison with literature data in which similar types of ROIs were used; no recent papers were available for comparison of these separate parameters.

Table 2 Rate constants of FDG parameters for manually drawn gray and white matter ROIs and literature values

Figure 1 shows the (linear) relationship between Patlak- and NLR-derived K i values for the 16 (automatic) combined gray matter regions and a white matter region. Linear regression provided a slope of 0.96 and an r2 of 0.98.

Figure 1
figure 1

Correlation between Patlak- and NLR-derived K i values. Data of all 16 gray matter regions and a white matter brain region of nine healthy subjects are presented. Data points for each individual subject are shown with a separate symbol. The solid line indicates the line of identity.

Based on the good correlation between Patlak and NLR results, parametric (Patlak) images were generated without smoothing. Figure 2 shows the relationship between average parametric and ROI-derived K i values, both before and after smoothing of the parametric images. Without smoothing, a slope of 1.04 and an r2 of 0.99 were obtained. After smoothing, these values were 0.97 and 0.90, respectively.

Figure 2
figure 2

Correlation of average K i values derived using parametric and regional Patlak analyses. Data of all 16 total gray matter regions and a white matter brain region are presented for nine subjects. Parametric values represent the mean of all voxels within an ROI. The solid line indicates the line of identity. Results for both images without smoothing (black dots) and those smoothed with a 6-mm Gaussian filter (white dots) are shown.

The correlation between average regional NLR and parametric values had a slope of 1.0 and r2 of 0.90 (data not shown). A representative parametric image is shown in Figure 3.

Figure 3
figure 3

Representative parametric images of a single subject . The CBF image (upper panel) and the CMRglu image (lower panel) of the same subject are presented. The parametric CBF image was generated after smoothing with a 6-mm Gaussian filter.

CBF

Average delay and dispersion values were 12 ± 3 and 10 ± 2 s, respectively. CBF values for the various automatically delineated regions using NLR analyses are given in Table 1. Total gray and white matter CBFs were 0.35 ± 0.05 and 0.25 ± 0.04 mL/cm3 per minute, respectively, (GM:WM = 1.4). Table 3 shows CBF values obtained from the manually drawn gray and white matter regions (ratio GM/WM = 3.3) as well as a comparison with literature data, recently obtained using another HRRT scanner [17]. Figure 4 shows the relationship between CBF values derived from BFM and NLR analyses on smoothed data. Linear regression provided a slope of 1.02 and an r2 of 0.73. In Figure 3, a representative parametric image is presented.

Table 3 Rate constants of H 2 O parameters for manually drawn gray and white matter ROIs and literature values
Figure 4
figure 4

Correlation of CBF values derived using parametric (basis function method) and regional (NLR) analyses. Data of 16 gray matter regions and a white matter region are shown for 11 subjects. Parametric values represent the mean of all voxels within an ROI. Data points for each individual subject are shown with a separate symbol. The solid line indicates the line of identity.

FDG extraction fractions for each ROI are shown in Table 1. For the total gray matter region, the FDG extraction fraction was 18 ± 1%.

Image-derived input functions

In order to scale the IDIF derived from the [18 F]FDG scan to the manual samples, a multiplication factor was derived (average value 2.2 ± 0.5). Figure 5 shows the relationship between IDIF- and BSIF-derived K i values for regional NLR ROI analyses. In two out of nine subjects, a slope of 0.76 ± 0.09 and r2 of 0.40 ± 0.65 were observed. For the other seven patients, slope and r2 were 1.03 ± 0.05 and 0.98 ± 0.01, respectively.

Figure 5
figure 5

Correlation between IDIF- and BSIF-based NLR-derived K i values. The correlation is for 16 gray matter regions and a white matter region. Data points for each individual subject (n = 9) are shown with the same symbol. The solid line indicates the line of identity.

Figure 6 shows the results of a similar comparison for Patlak-derived K i values. In this case, no apparent outliers, either in single fits or complete subjects, were apparent. Slope and r2 were 0.99 ± 0.06 and 0.99 ± 0.007, respectively.

Figure 6
figure 6

Correlation between IDIF- and BSIF-based Patlak-derived K i values. The correlation is for 16 gray matter regions and a white matter region. Data points for each individual subject (n = 9) are shown with the same symbol. The solid line indicates the line of identity.

In order to scale the IDIF derived from the [15O]H2O scan to the manual samples, a multiplication factor of 1.8 ± 0.4 (averaged over 11 subjects) was needed. However, after scaling, a BSIF/IDIF peak ratio of 3.0 ± 1.2 was observed. Therefore, the shape of the IDIF and the simultaneously measured BSIF did not match, and thus, an IDIF approach was not possible in this case.

Discussion

To date, no full kinetic analysis of [18 F]FDG data based on arterial sampling and a dynamic HRRT scanning protocol in humans has been reported. In the present study, average CMRglu values of 0.33 and 0.11 μmol/cm3 per minute were obtained for manually drawn gray and white matter regions, respectively. Automatically delineated regions yielded values of 0.29 and 0.19 μmol/cm3 per minute for gray matter and white matter, respectively, averaged over nine patients with a CoV of less than 15%. However, especially the white matter estimate based on the latter method was contaminated with some gray matter spill in.

Previously, Heiss and co-workers have acquired CMRglu data on an HRRT scanner [15]. Using fixed rate constants from the literature, they found similar gray and white matter CMRglu values with a GM/WM ratio of 2.7, which is only slightly lower than the present ratio for manually drawn regions. Therefore, it seems that a static scanning protocol, together with fixed rate constants, is a valid approximation of full kinetic modeling in the case of healthy volunteers. Nevertheless, it should be noted that the assumption of fixed rate constants may not be valid in certain clinically relevant patient populations, such as those having diabetes [31], characterized by a.o. hyperglycemia and hyperinsulinemia which both affect the blood vessel wall and the permeability surface area product [32], and decreased glucose metabolism [33]. Furthermore, normal fixed rate constants are not automatically applicable to flow-limited states [34].

The good correlation between Patlak and NLR results (slope 0.96, r2 0.98) implies that this linearization is a valid approach, as shown previously [7]. The underestimation of 4% is likely due to the blood volume (5%) that is taken into account in the NLR analysis, but not in the Patlak linearization. Parametric K i images without smoothing showed good correlation with regional K i values (r2 of 0.99) with a slope of 1.04, probably induced by differences in noise and tissue heterogeneity present in regional- versus voxel-based TACs. Smoothing the dynamic 18 F]FDG images resulted in a lower slope and poorer correlation, and this is therefore not recommended.

Average CBF values of 0.43 and 0.13 mL/cm3 per minute were obtained for manually drawn gray and white matter regions, respectively. These values are in line with the recent data acquired on an HRRT by Walker and co-workers [17], who found values of 0.44 and 0.15 mL/cm3 per minute (Table 3). The good correlation between BFM and NLR results implies that this linearization is a valid approach.

We found an FDG extraction fraction of 18%, which is in line with earlier reported values of approximately 20% [35], determined using the double-indicator method [36]. To our knowledge, this has not been derived from a combined CBF and CMRglu (K 1) measurement before.

The use of a noninvasive input function, derived from the carotid arteries, allowed for quantitatively correct estimates of regional CMRglu when Patlak linearization was used. In the case of NLR, however, the more stringent requirements placed on the input function (i.e., in addition to the area under the curve, the detailed shape of the peak is needed) resulted in good CMRglu estimates for only seven out of nine subjects and incorrect (with an average slope of 0.76) estimates in the other two. Nevertheless, clinical dynamic [18 F]FDG studies without the need for an arterial line can be performed (on a voxel-by-voxel basis) by analyzing data using the Patlak linearization.

Unfortunately, the IDIF approach was not applicable to the analysis of CBF scans. Scaling to the manual samples yielded a similar factor as for CMRglu scans, but an underestimation of the peak was observed (by a factor of 3). Although 15O has a higher positron energy (leading to an effective spatial resolution of 3.4 in the case of 15O, if it were 3.0 for 18 F), this is not likely to be the main reason for the underestimation of the peak. Possible explanations will probably include the implementation of scatter correction as well as the performance of the reconstruction algorithm for highly localized activity distributions, as prevailing in the blood just after the [15O]H2O bolus. Therefore, optimization of the frame definition and the use of a point spread function reconstruction may lead to a better match between BSIF and IDIF.

It is of importance to note that, although the images used were smoothed 6 mm at FWHM, the IDIF applicability cannot be automatically used for data generated at other PET scanners. This still needs a validation for every single tracer, every single scanner, and each acquisition and data analysis protocol separately. The success of the use of an IDIF depends on the intrinsic spatial resolution of the scanner, which is often lower than that of the scanner used in our study, and on the iterative reconstruction algorithm used (e.g., whether or not priors are included and whether the partial volume effect is implemented in the reconstruction algorithm), determining the signal-to-noise ratio as well as the minimum frame duration that can be applied. Furthermore, the duration of tracer injection needs to be optimized since the number of frames that can be acquired during the bolus should not be too low. The exact effect of these different options cannot be predicted and needs to be tested in the way as was described in this paper.

Conclusions

NLR-based regional CMRglu and CBF values can be obtained using a high-resolution PET brain scanner in healthy humans. Good correspondence between the nonlinear and linear models was observed for the regional data, which formed the basis for the calculation of accurate parametric images. Although no test-retest study was performed, average inter-subject variability of approximately 15% was observed, which implies a good reproducibility. For studies that would only require measurements of CMRglu, regional Patlak-based K i values can be derived without arterial sampling by using a carotid-based image-derived input function.