Findings

In resource-limited settings where HIV-1 is a burden, the need for reliable incidence surveillance data to help access performance of interventions as well as monitor transmission patterns cannot be overestimated. It is however unfortunate that such settings are coupled with challenges such as equipment and technical expertise. Standard assays, like the 3A11-LS for determining recent HIV infections may be impossible to use in these settings which therefore calls for alternative reliable but cheaper methods to determine recent HIV-1 seroconversion and estimate incidence. Improved laboratory methods for determining recent human immunodeficiency virus type 1 (HIV-1) seroconversion and estimating incidence have been widely reported [19]. Improvements in these methods help eliminate the use of longitudinal studies, back calculations and other cohort studies which are not only costly but also require consistent follow up of clients and repeated testing which is problematic and much difficult to perform in resource-poor settings [10]. Methods for determining recent HIV-1 infections in single serum samples include minipool (MP) nucleic acid amplification testing (NAT) and individual donation (ID) NAT, detection of p24 antigen and the sensitive/less sensitive (S/LS) method for detecting recent HIV-1 seroconversion [3]. NATs and p24 antigen testing are costly and require specific laboratory equipments to perform. A report on the cost-effectiveness of including MP or ID NAT and or p24 antigen testing to blood screening in Ghana found that it was highly costly to perform such tests [11]. The sensitive/less sensitive strategy however seems simpler and 'less costly' to be used in resource-limited settings. Detection of recent seroconversion can provide information on serologically undetected residual infections, monitor transmission trends in a community and also determine incidence of recent HIV-1 infection. Studies have been performed in some African countries using the S/LS EIA to determine recent HIV-1 seroconversion and estimate incidence [1215]. It has however been reported that some of these assays give misleading results when used in Sub-Saharan Africa [16].

Recent HIV-1 seroconversion was determined among 76 blood donor plasma samples which tested repeatedly double reactive on the Genescreen Ultra HIV Ag-Ab ELISA which was used at the time of the study for routine HIV screening at the National Public Health & Reference Laboratory, Accra, Ghana. Of the 76 ELISA repeat reactive samples, 41 tested repeatedly reactive on the Determine™ HIV-1/2 (Abbott Laboratories, Abbott Park, IL) rapid test. These plasma samples (n = 41) were subsequently taken through the modified Determine S/LS (Det-S/LS) protocol as described elsewhere [8]. The protocol classified 23 as recent seroconversions and the remaining 18 as longstanding seroconversions. Due to previous reports of misclassifications, we further performed the INNO-LIA™ confirmatory assay to ascertain the HIV-1 antibody seropositivity on all the initial 76 ELISA double repeat reactive samples. After confirmation, only one of the 23 recent seroconversion samples turned out to be seropositive for HIV-1. It was thus concluded that only this was a true recent HIV-1 seroconversion. All the 35 negative samples as determined by the Determine™ HIV-1/2 rapid test initially were also negative by INNO-LIA™. All the longstanding seroconversions as determined by the modified Det-S/LS protocol were confirmed as seropositive for HIV-1. On the INNO-LIA™ strip, the one recent seroconversion showed intense (3+) bands for antibodies to five HIV-1 proteins (sgp120, gp41, p31, p24, & p17).

For blood donor screening, the use of a very sensitive assay is recommended. These assays though having high sensitivities are not very specific as observed in this study. In the absence of PCR, INNO-LIA™ is a good alternative gold method for confirming HIV-1 and HIV-2 seropositivity. This is employed in many resource-limited settings where PCR is not available. In this study, the existing systems were used to troubleshoot a local problem. It was observed that, screening samples with the Det-S/LS assay without confirmation was likely to result in an overestimation of recent HIV seroconversion. Any calculations therefore using the misclassified recent data would result in overestimation of incidence of HIV-1 infection. Though fourteen (14) of the initial recent seroconversions were indeterminate on INNO-LIA™, the classification algorithm for recent infections by INNO-LIA as reported by Schupbach and his colleagues [17] could not be used due to absence of clinical information on the blood donor samples used. This classification however is questionable by our results as recent HIV seroconversion did not correspond to faint banding patterns on INNO-LIA™ as reported. We will therefore like to stress that, incidence estimates calculated with recent seroconversion data obtained using S/LS and other antibody based recent seroconversion algorithms are likely to be overestimated if HIV seropositivity is not confirmed.

Our results also indicate that, the modified rapid tests may be a good alternative for determining recent HIV-1 seroconversion and estimating incidence in resource-limited settings. There is therefore a need extensive work into this area of research. Since INNO-LIA™ is used as a confirmatory test for HIV in Ghana and other settings, there is a need for further studies to characterize seroconversion status of individuals whose antibody status are indeterminate on INNO-LIA™ and clearly define an algorithm for classifying recent seroconversion as suggested [17]. By this study we propose a simple cost-effective algorithm for determining recent HIV-1 seroconversion in resource limited settings (figure 1). The likely positive samples are taken through the Det-S/LS protocol as described by Soroka et al [8]. Suspected recent seroconversions are subsequently confirmed for their HIV-1 seropositivity using INNO-LIA™. The number of tests for INNO-LIA™ at this stage thus will be reduced as compared to performing INNO-LIA™ from step 1. In bigger systematic studies, persons classified by this algorithm as false recent seroconvertions may be followed up or a PCR performed where the facility exists to ascertain whether they are really negative for HIV-1 or not.

Figure 1
figure 1

Algorithm for determining recent HIV-1 seroconversion using the modified Determine sensitive/less sensitive protocol.