Background

Acute ischemic strokes are commonly associated with unstable carotid atherosclerotic plaques that can be detected in-vivo by multicontrast CMR [13]. Morphology and composition of atherosclerotic plaques can be characterized and classified following the CMR-modified American Heart Association (AHA) scheme [4]. The strength of multicontrast CMR resides in its ability to detect the presence of different plaque components by discriminating their relative signal intensities on time-of-flight (TOF), T1-, T2- and PD-weighted (T1W, T2W, PDW) images. Identification of lipid-rich necrotic core (LRNC), calcification, intraplaque haemorrhage (IPH) and fibrous tissue by in-vivo multicontrast CMR has been thoroughly validated by histology in several clinical studies [49].

The size of LRNC and the presence of IPH were shown histologically to be strongly associated with plaque instability [10]. Relative to the adjacent sternocleidomastoid muscle, MR signal from LRNC is iso/hyper-intense on TOF, T1W and PDW images, and hypo-intense on T2W images [7]. IPH is an independent predictor of future cardiovascular events [11] and reflects intraplaque neovascularization and plaque vulnerability [12]. It may also stimulate the progression of atherosclerosis by fuelling the plaque core with plasma membrane lipid from extravasated, broken-down, blood cells to increase the LRNC size [8]. MR signal from fresh IPH (type I) is hyper-intense on TOF and T1W images, and hypo/iso-intense on T2W and PDW images, whereas MR signal from recent IPH (type II) is iso/hyper-intense on all images [9].

Despite the advantages of multicontrast CMR, its non-quantitative nature and the need for extensive post-acquisition interpretation are barriers to the widespread clinical application of in-vivo CMR plaque characterization. Comparison of results from in-vivo CMR studies of atherosclerosis using different CMR systems and parameters can be challenging because plaque characterization is affected by variability and inconsistency in the signal intensity of plaque components relative to the reference intensity of the adjacent sternocleidomastoid muscle, which can be particularly evident in T2W images [2]. Furthermore, image intensity inhomogeneity caused by the surface coil sensitivity can be difficult to correct and can thus affect the relative signal intensities of plaque tissues and consequently the accuracy of plaque segmentation, in particular when automated methods are applied [13]. Finally, T1W and PDW images acquired with the Fast-Spin-Echo (FSE) sequence (used in multicontrast carotid CMR protocols) are known to suffer from blurring along the phase-encoding direction, which is inherent to the k-space acquisition strategy and causes a significant reduction of the vessel edge sharpness [14].

In-vivo quantitative mapping of relaxation times could provide an alternative method for plaque characterization. This technique measures the MR properties of plaque tissues directly, thus addressing the need for an absolute physical measure that can be standardized among different CMR systems and widely adopted in multi-centre studies, while at the same time avoiding the problems of image intensity inhomogeneity and blurring. Given these characteristics, quantitative mapping has great potential for automated plaque segmentation and classification and may provide new opportunities for quantification of plaque composition in carotid atherosclerosis that will be useful in clinical studies to evaluate emerging drugs that directly target plaque biology [15]. T2 relaxation times of LRNC and fibrous tissue found in the literature were all measured ex-vivo[1623], except one study that compared in-vivo and ex-vivo T2 measurements [16]. To our knowledge, no in-vivo T2 measurements of IPH are reported in the literature.

The purpose of this study was thus to investigate the potential of quantitative T2 mapping as an in-vivo technique for atherosclerotic plaque characterization. In particular, we aimed to acquire in-vivo high-resolution quantitative T2 maps of carotid arteries, to study their use for AHA plaque type classification, and to measure in-vivo T2 relaxation times of the main plaque components at 3 T.

Methods

In-vivo CMR

The carotid arteries of 2 healthy volunteers and 15 patients with known atherosclerosis (11 males and 4 females, 71 ± 10 years, range 54-84) were imaged on two 3T scanners (12 patients on TIM Trio and 3 on Verio, Siemens Healthcare, both running VB17 software) using the Multiple-Spin-Echo (Multi-SE) sequence, called Spin-Echo-Multi-Contrast (SE_MC) on Siemens systems. Ethics approval from the local board was obtained and all subjects gave informed written consent. Black-blood cross-sectional 2D images of carotid arteries were acquired with the Machnet 4-channel phased-array carotid coil using Double-Inversion-Recovery (DIR) preparation and cardiac gating. Chemical shift selective fat saturation (FAT SAT) was used to suppress the signal from subcutaneous and perivascular fat, mainly composed of triglycerides, without affecting the signal from cholesterol and cholesteryl esters, the main lipids in atherosclerotic plaques [21, 22]. If necessary, a saturation band was positioned on the anterior region of the neck to reduce ghosting artefacts from breathing and swallowing. Low Specific-Absorption-Rate (SAR) pulses were used in order to keep the total dissipated radio-frequency energy below the SAR limit without changing the flip angle of the refocusing pulses. Multi-SE parameters were TE = 12.9, 25.8, 38.7, 51.6, 64.5, 77.4, 90.3, 103.2 ms, TR = 2 R-R intervals, field of view (FOV) = 160 × 128 mm2 and matrix size = 320 × 256. Image resolution after zero-padding was 640 × 512 (pixel size = 0.25 mm). Slice thickness (2 mm) and pixel bandwidth (130 Hz/pix) were the same for Multi-SE and FSE. FSE T1W (TE = 14 ms, TR = 1 R-R and ETL = 9), T2W (TE = 89 ms, TR = 2 R-R and ETL = 15) and PDW (TE = 14 ms, TR = 2 R-R and ETL = 9) images had FOV = 150 × 150 mm2 and matrix size = 320 × 320. Image resolution after zero-padding was 640 × 640 (pixel size = 0.234 mm). TOF angiography was acquired using 3D Fast Low Angle Shot (FLASH) with flip angle = 18°, TR = 45 ms, TE = 4.1 ms, FOV = 200 × 150 mm2, matrix size = 256 × 192 and slice thickness = 1 mm. 69 TOF and 11 DIR-FSE T1W consecutive slices centred on the carotid bifurcation were first imaged in order to localize the atherosclerotic plaque. Then a series of DIR-FSE T2W and PDW (and if necessary other T1W) images were acquired to obtain full plaque coverage. Finally, a single-slice DIR-Multi-SE image series was acquired through the plaque centre. Given the constraints imposed by other imaging parameters, the TE series used in Multi-SE was defined considering an expected T2 ~ 50 ms for the normal carotid wall, which corresponded to a transverse magnetization half-life time T2·ln(2) ~ 35 ms. The last TE used in this study was approximately equal to 3 half-life times and typical SNR values measured on normal carotid wall went from ≥ 30 at the first echo to ≥ 4 at the last echo. Partial Fourier imaging was used to collect 5/8 of k-space (160 phase-encoding steps) and reduce the acquisition time of single-slice Multi-SE to circa 320 R-R intervals. The quality of Multi-SE and multicontrast images was assessed after each scan by two reviewers (A.C.L. and L.B.) who decided by consensus when carotid wall and plaque were not clearly visible.

Pulse sequence

The Multi-SE sequence design is similar to the standard FSE that is widely used in carotid imaging, the main difference being the k-space acquisition strategy. Both are Carr-Purcell-Meiboom-Gill (CPMG) based sequences in which a 90° excitation pulse is followed by a train of 180° refocusing pulses to generate multiple spin echoes at every TR [24, 25]. Constant crusher gradients at the left and right of each refocusing slice selective gradient are used to select both primary and stimulated echoes, which sum up at the mid-point between two consecutive refocusing pulses, whereas all other unwanted signal pathways are filtered out. In FSE each echo is phase-encoded to a different k-space line in order to collect multiple k-space lines for a single image during every TR, whereas in Multi-SE the phase-encoding is the same for the entire echo train. In this way, each echo of the Multi-SE sequence fills a line of a series of independent k-space planes, so that a set of images with different TEs are collected (Figure 1). Accuracy and precision improvements for clinical T2 mapping using a CPMG sequence have been demonstrated by setting the slice thickness of the refocusing pulses larger than that of the excitation pulse [26]. This simple modification stabilizes the flip angle of the refocusing pulse across each slice and minimizes refocusing profile errors. A 1.5 ratio between refocusing and excitation slice thickness was used in this study.

Figure 1
figure 1

Multi-SE sequence diagram with DIR and FAT SAT modules. The RF excitation pulse is followed by a train of RF refocusing pulses. GSS and GFE are the slice selection and the frequency encoding gradients. The moment of the phase encoding gradient GPE is the same along the pulse train and signal from consecutive echoes is encoded as lines of independent k-spaces. Phase encoding and rewinding gradients are applied before and after each signal acquisition window rather than a single phase encoding step applied before the first refocusing pulse. Constant crusher gradients at the left and right of the slice selective gradients are present in the sequence but are not shown in the diagram.

T2 Mapping and plaque segmentation

Quantitative T2 maps of manually drawn regions of interest (ROIs) including vessels and muscles were generated from the Multi-SE series of images. The first image was discarded because of the different nature of the first echo with respect to the others. Since the flip angles of refocusing pulses are not exactly 180° due to B1 field inhomogeneities, only the signal acquired at the first TE is a pure primary echo, whereas signals acquired at the subsequent TEs are composed by primary and stimulated echoes. The signal intensity (SI) of the first echo will thus not fit in the exponential decay curve followed by the SIs of subsequent echoes, which are enhanced by stimulated echoes. For every ROI voxel the T2 mono-exponential decay curve SI = β·e-TE/T2 was therefore fitted to 7 SIs collected at different TEs (β includes the effect of proton density, T1 weighting, coil sensitivity, signal amplification and other factors).

The regression model used for T2 mapping consisted of a nonlinear monoexponential fit (Figure 2) initialized by T2 and β resulting from a robust linear fit of ln(SI). Parameters estimated by linear least-squares regression are very sensitive to noise because the original error distribution becomes asymmetric after the logarithmic transform. To reduce this sensitivity to outliers, we used a robust regression method that iteratively minimizes the weighted sum of squares using bisquare weights, which depend on the residual of each data point [27]. To assess the statistical significance of fit parameter estimates, the two-tailed p-value of the ratio of each estimate to its standard error (derived from the diagonal elements of the estimated covariance matrix) was calculated on the t-distribution. Only linear fit parameter estimates rejecting the null hypothesis (p < 0.05) were used as initial values for the nonlinear fit in order to improve its convergence. Otherwise default initial values β = 500 and T2 = 50 ms were used. The nonlinear least-squares regression used the Levenberg-Marquardt algorithm [28]; data points at long TEs with SNR < 2 were discarded; 95% confidence intervals for the exponential curve (Figure 2) and for the fit parameters were estimated. Finally, only voxels with significant T2 and β estimates (p < 0.05) were accepted for generating T2 maps. The carotid arteries were subsequently segmented using a semi-automated method [29] that detected inner and outer wall boundaries on the second image (TE = 25.8 ms, TR = 2 R-R) of the Multi-SE series.

Figure 2
figure 2

Nonlinear fit of the T 2 relaxation curve of a normal carotid artery. The T2 mono-exponential decay curve (in red, with 95% confidence intervals) is fitted to the signal intensity (SI) of one voxel of the arterial wall in the Multi-SE image series. The voxel SI is represented by blue circles at 7 consecutive TEs.

After this procedure, T2 maps of atherosclerotic arteries were segmented into 3 tissue types using a Bayes classifier [30] defined by the probability model P(C i |T 2 ) = P(C i )P(T 2 |C i ) combined with the maximum a posteriori (MAP) decision rule. P(C i |T 2 ) with i = 1,2,3 is the posterior conditional probability for a voxel with a specific T2 value to belong to one of 3 classes: LRNC, fibrous tissue or recent IPH. The MAP rule then assigned a particular voxel to the tissue type with the highest posterior probability. P(C i |T 2 ) is determined by the prior probability P(C i ), which was assumed to be equal for every class, and the normal probability distribution P(T 2 |C i ), which was estimated from a set of approximately 100 voxels for each class. These voxels were manually selected from all the T2 maps to represent a specific tissue using multicontrast CMR as a guide, i.e. they were chosen inside a plaque component identified on multicontrast CMR [4, 9]. In order to help matching the corresponding plaque features, multicontrast images were co-registered [31], so that inner and outer vessel boundaries detected on T1W images [29] could then be superimposed on PDW and T2W images. Calcification was identified by voxels inside the vessel wall boundaries with SNR < 2 at TE = 14 ms on the T2 decay curve (synthetic PDW image). Finally, mean ± SD of the T2 values of voxels classified as LRNC, fibrous tissue and recent IPH were calculated. The algorithms for quantitative T2 mapping, vessel wall segmentation and plaque classification were implemented in Matlab (Mathworks).

To compare atherosclerotic plaque characterization by T2 maps with conventional multicontrast CMR, two reviewers (A.C.L. and L.B.) independently classified the carotid arteries of 15 patients as normal or diseased (plaque type III, IV-V, VI, VII or VIII) following the CMR-modified AHA scheme [4, 9]. One reviewer analysed multicontrast CMR (T1W, T2W and PDW image intensity inhomogeneity was corrected [13]) and TOF, while the other relied on T2 maps and TOF at matching slice locations. Both were blinded to the identification and the clinical data of each patient.

Results

Out of 42 image series (including common, internal and external carotid arteries) acquired with single-slice Multi-SE in 15 patients, 5 were rejected due to poor image quality. The corresponding multicontrast (TOF, T1W, T2W and PDW) images were all accepted and the AHA plaque type classification performed on them by one reviewer found 14 normal and 23 diseased arteries (7 type III, 10 type IV-V, 2 type VI, 3 type VII and 1 type VIII plaques). AHA classification performed by the other reviewer on T2 maps and TOF found 11 normal and 26 diseased arteries (11 type III, 8 type IV-V, 3 type VI, 2 type VII and 2 type VIII plaques). The overall agreement on AHA plaque type between the reviewers was good (76% of matching classifications and Cohen’s κ = 0.68, Table 1).

Table 1 AHA plaque type classification by multicontrast CMR vs. T 2 maps + TOF

The T2 maps of the 10 arteries classified as normal by both reviewers, together with 4 common carotid arteries from 2 healthy volunteers, were used to measure T2 mean ± SD of normal carotid wall (T2 = 54 ± 13 ms, Table 2). The T2 maps of the 15 atherosclerotic arteries with advanced plaques (graded as type IV-V, VI, VII or VIII by both reviewers) were segmented by the Bayes classifier into 4 plaque components: calcification, LRNC, fibrous tissue and recent IPH. From a total of 11865 voxels, the T2 measurements of 668 voxels were rejected because not statistically significant (p > 0.05). 192 voxels were identified as calcification in one type IV-V, one VI and two VII plaques. 2394 voxels with T2 = 37 ± 5 ms were classified as LRNC by the segmentation algorithm, mostly detected in type IV-V and VI plaques (Table 2). 7511 voxels with T2 = 56 ± 9 ms were classified as fibrous tissue and normal intima/media (Table 2). Figure 3 shows an atherosclerotic artery with LRNC and a thick fibrous cap separating it from the lumen (type IV-V). Figure 4 shows a large complex plaque (type VI) at the carotid bifurcation composed of lipid and necrotic material mixed together with fibrous tissue. These features agreed with the multicontrast CMR classification, which revealed the presence of fresh IPH (hyper-intensity on TOF and T1W images).

Table 2 T 2 measurements (mean ± SD) of arterial wall and plaque tissues
Figure 3
figure 3

Multicontrast CMR and T 2 map of a type IV-V plaque in the right common carotid artery. Multicontrast CMR shows a hypo-intense region compared to the sternocleidomastoid muscle (SM) on the T2W image, indicative of LRNC (arrow). The T2 map and its segmentation show a thick fibrous cap separating the LRNC, characterized by shorter T2 values, from the lumen.

Figure 4
figure 4

Multicontrast CMR and T 2 map of a large type VI plaque at the left carotid bifurcation. Multicontrast CMR reveals the presence of fresh IPH (arrow), characterized by hyper-intensity on TOF and T1W images compared to the sternocleidomastoid muscle (SM). The T2 map and its segmentation illustrate that the plaque is composed by fibrous tissue, pools of lipid and necrotic material (LRNC) and a small calcification. Some sparse voxels were classified as recent IPH due to T2 overestimation.

Finally, large areas of recent IPH were detected on T2 maps in two type VI plaques for a total of 528 voxels measuring T2 = 105 ± 25 ms. However, only the plaque with recent IPH in Figure 5 (236 voxels with T2 = 107 ± 25 ms) was confirmed as type VI by multicontrast CMR. In the other 13 plaques, sparse voxels and small regions (for a total of 572 voxels with T2 = 86 ± 9 ms) were classified as recent IPH by the segmentation algorithm. Additionally, since most clinical CMR studies of atherosclerosis used the sternocleidomastoid muscle as a signal intensity reference for plaque tissue classification in T2W images [4, 7, 9], we generated T2 maps of sternocleidomastoid muscles in volunteers and patients and measured T2 = 39 ± 6 ms.

Figure 5
figure 5

Multicontrast CMR and T 2 map of an occlusive type VI plaque in the right internal carotid artery. Multicontrast CMR reveals the presence of recent IPH, characterized by iso/hyper-intensity on TOF, T1W and T2W images compared to the sternocleidomastoid muscle (SM). The T2 map and its segmentation show a large region with long T2 values, corresponding to recent IPH, surrounded by fibrous tissue. Due to the borderline quality of multicontrast CMR, it was not possible to co-register the images and segment the vessel wall, therefore a circle was manually placed to indicate the occluded artery.

Discussion

The objective of this study was to investigate the use of T2 mapping for plaque characterization. We acquired in-vivo high-resolution T2 maps of carotid arteries in 15 atherosclerotic patients using the Multi-SE sequence and demonstrated good agreement between AHA plaque type classified by T2 maps and TOF images and by conventional multicontrast CMR. The present study measured the T2 of LRNC, fibrous tissue and recent IPH and showed the potential of T2 mapping for plaque segmentation and classification. Our in-vivo T2 measurements (Table 2) were slightly longer but broadly in agreement with the only in-vivo T2 study of carotid atherosclerosis available in the literature [16]. The T2 values of tunica intima and media measured in 14 normal carotid arteries fell within the T2 range reported by ex-vivo studies of carotid and coronary atherosclerosis [1620]. With the exception of one study that reported much longer values [18], the T2 measurements of fibrous tissue and LRNC in 15 advanced plaques were comparable with those obtained ex-vivo[16, 17, 19, 20]. Recent IPH was characterized by long T2, consistent with signal hyper-intensity on T2W images [9], and was detected in one complex (type VI) plaque (confirmed by multicontrast CMR). In other plaques the classification of some sparse voxels and small regions as recent IPH was most likely caused by T2 overestimation errors: their T2 values were longer than their neighbouring voxels but shorter than the complex plaque and TOF images showed no signs of IPH. The measured T2 of sternocleidomastoid muscles agreed with the range of values reported in literature for muscle tissue (30-50 ms) at 1.5 T and 3 T [32]. It is slightly longer than the T2 of LRNC but shorter than that of normal carotid wall and fibrous tissue, confirming it as a suitable intensity reference in T2W images [4, 7, 9]. However, intensity inhomogeneity introduced by the sensitivity profile of surface coils, which strongly affects the sternocleidomastoid muscle due to its skin proximity, must be corrected before it can be used as reference.

Table 2 shows the ex-vivo studies of atherosclerosis that were found in literature [1623]. Despite their methodological differences (i.e. field strength, imaging parameters and arterial locations from which the plaques had been excised), the common feature of these ex-vivo studies is a shorter T2 in LRNC than in fibrous tissue and normal intima and media, which was confirmed by our in-vivo T2 measurements. However, care must be taken when comparing in-vivo and ex-vivo T2 values, because the effect of temperature and fixatives on tissue specimens can be significant [18]. Five ex-vivo CMR studies were performed at body temperature [16, 18, 20, 21, 23] and three at room temperature [17, 19, 22].

Our in-vivo T2 measurement method is more accurate than the previous in-vivo study [16], which calculated T2 of 7 carotid arteries by collecting and fitting only 2 echoes with relatively short TE (20 and 55 ms). In that study [16] the high sensitivity to noise due to the limited number of echoes constraining the T2 exponential decay curve was aggravated by the use of short TR (1 R-R),which causes more T1 weighting, and very thick slices (5 mm), which can mix T2 values of different tissues (partial volume averaging). We used a CPMG sequence (Multi-SE) to acquire more echoes (7 TEs ranging from 25.8 to 103.2 ms) designed to obtain the best possible sampling of the T2 relaxation decay curve of target tissues. The Multi-SE acquisition parameters were more appropriate for T2 quantitation than those used by the previous in-vivo study, since TR was longer (2 R-R) and slices thinner (2 mm). We also implemented a nonlinear regression method that is more robust to outliers, and we ensured that the T2 estimates were statistically significant by accepting only results with p < 0.05. Finally, the present study imaged 14 normal and 15 atherosclerotic carotid arteries, more than those analysed previously (Table 2).

In some cases the first image of the Multi-SE series showed artefacts due to residual signal from slowly flowing blood not suppressed by DIR. We speculated that the main cause was inaccurate shimming of B0 inhomogeneities. Flow artefacts did not produce T2 errors because the first contrast image was never used to estimate T2, as explained in the methods.

Limitations

The main limitation of this study is the absence of plaque histology, since only two of the participating patients underwent surgery and their endarterectomy specimens were of insufficient quality for detailed analysis. However, plaque characterization performed on in-vivo T2 maps agreed with results of multicontrast CMR, which has been previously validated by histology [49]. A physical limitation of T2 mapping is that the presence of fresh IPH (type I) cannot be detected by T2 alone. Fresh IPH infiltrating the LRNC is characterized by short T2[8, 9], so it can only be detected using T1 information, e.g. independent identification on TOF images (Figure 4).

Finally, imperfections of the refocusing pulses represent the main source of errors in T2 quantitation [33], although the CPMG sequence used in this study partially compensates for them. Stimulated echoes produced by non-ideal refocusing pulses are acquired together with the primary echoes, thus introducing T1 weighting into the signal and altering the pure T2 relaxation decay. After discarding the first echo, which is purely primary, the contribution of stimulated echoes to the remaining 7 echoes may have caused T2 overestimation. These errors depend on the amount of T1 weighting introduced by the use of TR = 2 R-R. A very long TR (~5·T1 of target tissue) would limit the T1 effect, but it would be absolutely impractical under in-vivo high-resolution imaging conditions. Finally, the elimination of the first echo may have affected the estimation of short T2 components in the LRNC (~35 ms), which have a transverse magnetization half-lifetime (~24 ms) as long as the second TE. In general, this problem can be alleviated by the use of shorter echo spacing (ESP). This study used the shortest ESP possible, given the requirements for matrix size, receiver bandwidth and the duration of low-SAR refocusing pulses. Future developments will aim at improving in-vivo T2 measurement accuracy, decreasing acquisition time and validating plaque characterization with histology.

Conclusions

This study draws attention to the potential use of in-vivo quantitative T2 mapping for carotid plaque characterization. On 37 carotid arteries imaged in-vivo in 15 atherosclerotic patients, AHA plaque type classification using high-resolution T2 maps and TOF images showed good agreement with conventional multicontrast CMR. T2 relaxation times measured in-vivo at 3 T for tunica intima and media in 14 normal arteries and for LRNC, fibrous tissue and recent IPH in 15 arteries with advanced plaques demonstrated the ability to discriminate 4 major plaque components including calcification, thus supporting the application of plaque segmentation and classification based on T2 maps.