Background

Cardiovascular magnetic resonance (CMR) is increasingly applied in clinical routine to determine myocardial perfusion [14] using contrast enhanced first-pass perfusion techniques [58]. Non-invasive characterization of myocardial microcirculation is thought to reflect myocardial tissue supply much better than mere luminographic detection and quantification of epicardial coronary stenosis, and has been shown to be useful for planning of revascularization procedures and cardiac risk stratification.

Blood Oxygen Level Dependent (BOLD) CMR is based on the paramagnetic properties of deoxyhemoglobin as an endogenous contrast agent with increased deoxyhemoglobin content leading to signal reduction on T2*- or T2-weighted images. Thus, BOLD CMR directly reflects myocardial oxygenation status [9, 10]. Coronary artery stenosis leads to poststenotic microvascular dilatation in a compensatory effort to maintain sufficient myocardial oxygen supply [11] and blood-oxygen level dependent imaging has been successfully introduced to determine capillary reserve [12, 13].

However, data on myocardial BOLD CMR during vasodilator stress in patients with coronary artery disease (CAD) is limited to experimental studies [1315] and small patient populations [11, 12]. The main challenge was low signal intensity differences between normal and pathologic areas of myocardium at 1.5T, and therefore BOLD CMR at 3T may benefit from the inherently higher signal-to-noise ratio (SNR). Hence, in the present study we determined the ability of BOLD CMR to detect stress induced myocardial ischemia and to differentiate between ischemic, non-ischemic, and normal myocardial segments in a population of patients with suspected or known CAD.

Materials and methods

Study Group

Forty-six consecutive patients (34 men; age 65 ± 9 years, range 40 to 81 years) referred for clinically indicated invasive coronary x-ray angiography due to chest pain syndromes were prospectively enrolled. Patients were eligible if they had suspected or known CAD (with or without prior percutaneous revascularization or a history of previous myocardial infarction). Patients with prior coronary surgery or typical contraindications for CMR (e.g. incompatible metallic implants, claustrophobia) and administration of adenosine (asthma, AV-block > grade I) were not considered.

All study participants were instructed to refrain from ß-blockers, antianginal medication, cigarettes, tea and coffee for at least 24 hours prior to CMR. Written informed consent was obtained from all subjects, and the Charité Institutional Review Board approved the study.

CMR Imaging Protocol

CMR was performed with the patient in the supine position using a 3T whole-body imager (Achieva 3T; Philips, Best, the Netherlands) equipped with a Quasar Dual gradient system (40 mT/m, slew rate 200 T/m/sec). A six element cardiac synergy coil was used for signal reception and cardiac synchronization was done with the use of a Vector-ECG. All acquisitions were performed during short end-expiratory breath-holds. After acquisition of cine standard cardiac geometries for the assessment of left ventricular function a fast-gradient-echo multi-echo sequence for BOLD CMR (3-slices of short axis geometry) was performed. Then, adenosine infusion (140 μg/kg/min; maximal total infusion duration of 6 minutes) was started and the identical BOLD CMR sequence was repeated after at least 3 minutes of adenosine infusion. After termination of adenosine infusion, a bolus of 0.2 mmol/kg of gadolinium-DTPA was administered followed by late gadolinium enhancement (LGE) imaging 10 minutes later in identical short axis geometry with full left ventricular coverage.

CMR Technique

Cine Imaging

Three short axis (apical, mid, and basal short axis views) and three long axis geometries (4-, 2-, and 3-chamber view) were acquired using an electrocardiogram-triggered balanced turbo field echo sequence (echo time 1.9 ms, repetition time 4.0 ms, flip angle 40°, spatial resolution 1.8 × 1.8 × 8 mm) during repetitive end-expiratory breath-holding.

BOLD CMR

BOLD CMR imaging was carried out in the identical three short axis geometries of the left ventricle using an electrocardiogram-triggered, spoiled segmented gradient-echo sequence with six echo times (first echo at 2.7 ms with echo spacing of 1.7 ms, repetition time 13 ms, flip angle 35°, slice thickness 8 mm, reconstructed spatial resolution 1.2 × 1.2 × 8 mm, bandwidth 776 Hz/pixel). Image data acquisition was restricted to end-diastole in order to minimize cardiac motion related artefacts. A black blood dual inversion prepulse was used to nullify signal from flowing blood in the left ventricular cavity. Figure 1 shows a representative example of BOLD CMR at rest and during vasodilator stress.

Figure 1
figure 1

BOLD CMR at rest (upper row) and during adenosine stress (bottom row) showing the apical (a), midventricular (b) and basal slice (c).

Late Gadolinium Enhancement Imaging

Late gadolinium enhancement imaging was carried out in short axis orientation with full left ventricular coverage using a three-dimensional inversion prepared spoiled gradient echo sequence (echo time 1.8 ms, repetition time 3.7 ms, flip angle 15°; spatial resolution 1.5 × 1.5 × 5 mm). The prepulse delay was determined from an inversion prepared cine scan (Look-Locker) and individually adjusted to optimally suppress signal from normal myocardium.

Quantitative Coronary Angiography

All patients underwent invasive coronary angiography in standard Judkins technique within 48 hours after CMR. The procedure was done according to angiographic guidelines using a simultaneous biplane, multidirectional and isocentric x-ray system. At least two orthogonal views of every major coronary vessel and its side branches were acquired. Quantitative coronary angiography (Philips Inturis CardioView, QCA V3.3, Pie Medical Imaging) was performed off-line by an independent observer being unaware of the results of CMR. The severity of coronary stenosis was derived from one single view showing the maximal reduction in absolute luminal diameter and a significant coronary stenosis was defined as ≥50% luminal diameter reduction in vessels with ≥2 mm diameter; significant left main stenosis was considered double-vessel disease. In addition, myocardial segments were assigned to the supplying coronary artery based on a consensus read of the interventionalist and the CMR imager taking the respective coronary dominance type into account.

Image analysis

BOLD CMR

The BOLD CMR data sets were evaluated on a per segment basis according to the standardized 16-segment model [16]. For each segment, the time constant of the signal intensity decay over all echoes was derived. Moreover, the global T2* value per patient averaged over all segmental T2* values was provided at rest and during stress. Image quality of T2* CMR measurements was graded visually on a per patient basis at rest and during stress on a four-point scale as excellent (4 = clear delineation of the left-ventricular myocardium with sharp endocardial contours, no motion artifacts), good (3 = clear delineation of the left-ventricular myocardium but mildly blurred endocardial contours, no motion artifacts), moderate (2 = moderately blurred endocardial contours, occurrence of respiratory/cardiac motion artifacts) or poor (1 = severely reduced delineation of left-ventricular endocardial contours and severe motion artifacts). Occurrence of susceptibility artifacts caused by the heart-lung interface and cardiac veins was scored between 0 and 4 (0 = none, 1 = minor, 2 = moderate, 3 = severe, 4 = non-diagnostic).

Late Gadolinium Enhancement Imaging

LGE images were analyzed visually on a per segment basis regarding the presence of myocardial scar ≥25% transmurality with each myocardial segment being classified as normal or scarred.

Statistical Analysis

Statistical analysis was performed by using the SPSS software package release 17 (Chicago, Ill, USA). The paired Student's t test or Wilcoxon test and independent-samples t test or Mann-Whitney test were used to test for differences within and between groups. All tests were two-tailed. The Kolmogorov-Smirnov test was used to test for normality. A P value of less than 0.05 was considered significant. For ROC analysis and the comparison of the area under the curve Medcalc® release 9.2.1.0. (MedCalc Software, Belgium) was used. The areas under the curves were compared using the method of DeLong et al.[17]

Results

Patients Characteristics and Hemodynamic Data

A total of 46 consecutive patients were enrolled in this study. In 4 patients, CMR scans could not be completed: 2 patients were claustrophobic, and 2 did not tolerate adenosine. Hence, 42 (91%) of 46 patients of this study cohort successfully completed the CMR examination and were included in the final analysis.Patient characteristics and hemodynamic data are summarized in table 1 and 2, respectively. Significant CAD was found in 23 patients (54.8%) with ten patients (24%) having single-vessel and 13 patients (31%) multi-vessel disease. According to the results of QCA and LGE imaging a total of 672 myocardial segments were classified as follows: 272 (40.5%) normal segments, 224 (33.3%) non-ischemic segments, 152 (22.6%) ischemic segments. In addition, 24 (3.6%) scarred segments were excluded from further analysis.

Table 1 Patient Characteristics
Table 2 Left Ventricular Function at rest and Hemodynamic Data

Image Quality

On BOLD CMR at rest, 15 patients showed excellent (35.7%), 21 patients good (50.0%), five patients moderate (11.9%) and one patient poor image quality (2.4%); under adenosine stress image quality was excellent in 11 patients (26,2%), good in 21 patients (50.0%), moderate in eight patients (19.0%) and poor in two patients (4.8%). The mean visual score of BOLD CMR was 3.2 ± 0.8 at rest and 3.0 ± 0.8 under adenosine stress (p = 0.005). Though all studies were of sufficient quality for analysis, segments demonstrating poor image quality were excluded from analysis (17 of 648, 2.6%). The overall artifact score was significantly higher under adenosine stress than at rest (1.6 ± 0.6 vs 1.4 ± 0.5; p = 0.047; for an imaging example see figure 2).

Figure 2
figure 2

Example of rest (top rows) and stress (bottom rows) BOLD CMR showing the apical (a), mid-ventricular (b) and basal slice (c) with six echo times. Susceptibility artifacts (white arrow) occurred at long echo times predominantly in the inferolateral segment caused by the heart-lung interface.

BOLD CMR

BOLD measurements at rest revealed significantly lower values for ischemic segments (n = 148; 26.7 ± 11.6 ms) compared to normal segments (n = 260; 31.9 ± 11.9 ms; p < 0.0001) and non-ischemic segments (n = 223; 31.2 ± 12.2 ms; p = 0.0003). During adenosine stress T2* values demonstrated a significant increase in normal segments only (37.2 ± 14.7 ms; p < 0.0001). On the contrary, T2* values of non-ischemic (32.8 ± 14.9 ms; p = 0.19) and ischemic segments (27.2 ± 12.3; p = 0.06) did not significantly differ for the comparison of stress versus rest. When comparing stress T2* values of normal versus non-ischemic and ischemic segments significant differences were found (p = 0.0007 and p < 0.0001, respectively). Corresponding error bar charts are given in figure 3. Figure 4 shows the relationship between the changes of T2* values at rest and under adenosine stress and the degree of coronary artery stenosis.

Figure 3
figure 3

Error bar charts of T2* at rest and under adenosine stress comparing the mean values and corresponding 95% confidence intervals.

Figure 4
figure 4

Relationship between T2* values (mean values and corresponding 95% confidence intervals) at rest and under adenosine stress and the degree of coronary artery stenosis.

The global T2* value of patients without significant CAD was 32.2 ± 3.5 ms at rest and 37.4 ± 6.3 ms under stress (p = 0.0003). In patients with significant CAD the global T2* value was 30.3 ± 5.3 at rest and 30.7 ± 5.8 under stress (p = 0.78). Imaging examples are provided in figure 5.

Figure 5
figure 5

Imaging examples: Rest and stress BOLD CMR showing a patient with lateral ischemia (A) and a patient without coronary artery disease (B).

Diagnostic Performance of BOLD CMR

The area under the curve (AUC) of the ROC analysis (Figure 6) for the ability of T2* CMR imaging to detect significant CAD on a per patient basis differed significantly between rest (0.61, 95%-CI [0.44-0.75]) and stress imaging (0.82, 95%-CI [0.67-0.92], p = 0.015). Using a cut-off value of 33.8 ms, sensitivity and specificity for rest and stress were 78% and 21%, and 78% and 68%. AUC of the ROC analysis for the detection of patients with multivessel disease (≥2 diseased vessels) under stress was 0.80 (95% CI 0.63-0.92). Using a cut-off value 33.8 ms, sensitivity and specificity were 77% and 68%.

Figure 6
figure 6

ROC analysis showing the diagnostic performance of rest and stress BOLD CMR to identify the presence of significant coronary artery disease.

Discussion

In the present study, BOLD CMR at 3T was shown to be feasible at 3T resulting in an overall good image quality at rest and during vasodilator stress. BOLD CMR at rest and under stress conditions distinguished between normal, ischemic and non-ischemic myocardial segments. In addition, adenosine stress BOLD CMR gave a good diagnostic performance with regard to the detection of angiographically defined coronary stenosis.

The non-invasive detection of ischemic myocardium is one of the major strengths of CMR. BOLD imaging represents a promising alternative to first-pass CMR perfusion techniques, which rely on the application of exogenous contrast agents. BOLD imaging allows to directly assess the myocardial oxygenation status taking advantage of the paramagnetic properties of intravascular deoxyhemoglobin [11]. During adenosine stress, myocardial segments being supplied by normal coronary arteries show an increased oxygen content while deoxyhemoglobin concentration is low resulting in a higher signal on T2*- or T2-weighted images. On the contrary, myocardial segments being supplied by a stenosed coronary artery exhibit a dilated capillary bed already under resting conditions in order to maintain sufficient blood supply. Hence, additional vasodilatation under stress is minimal and the hyperaemic response is abolished.

In the present study, ischemic myocardial segments showed significantly reduced T2* values under resting conditions, while during adenosine stress T2* values increased in normal myocardial segments only. Our findings were consistent with previous reports by Wacker et al. and Niemi et al. [11, 18], but somewhat contrasted the results by Friedrich et al. [12] even demonstrating a signal decrease during vasodilator stress in myocardial territories supplied by coronary arteries with >75% luminal narrowing.

Interestingly, in the present study non-ischemic myocardial segments did not show a significant increase of T2* values during stress. These findings may be explained by microvascular disease [19], which is present in myocardial segments of patients with CAD despite the absence of significant epicardial coronary artery narrowing. The T2* measurements of the current study showed a relation between the degree of coronary stenosis and the T2* value at rest and under adenosine stress.

The T2* CMR values at rest in patients without CAD were within the range of previously published data by Story et al. (33.3 ± 8.3 ms, outliers were excluded) [20] and O'Regan et al. (27.3 ± 6.4 ms) [21] using a single-slice image acquisition. However, in our study, the standard deviation of T2* values was substantially larger, which can be explained by the imaging strategy employing a 3-slice acquisition per patient in order to achieve coverage of all 16-standard myocardial segments. Susceptibility artifacts appeared predominantly in the inferolateral segments of the myocardium on images with the longest echo times and resulted mainly from interference with cardiac veins [22] and the heart-lung interface [23]. In order to reduce these artifacts an end-expiratory breath hold was used as proposed by Wacker et al.[9] in conjunction with a black blood prepulse to reduce possible sources of artifacts arising from the blood pool. Furthermore, the use of only six echo times and a relatively short echo spacing and repletion time of 13 ms in our study prevents susceptibility artefacts caused by longer echo times [9]. Recently, O'Regan et al. [21] reported that black blood breath-hold multiecho T2* imaging can be adequately performed at 3T and resulted in only minor susceptibility artifacts in the inferolateral region of the heart.

Importantly, the current study population consisted of consecutive patients with a high proportion of known CAD and prior myocardial infarctions but such patients were usually not considered in previous studies [11, 12, 21]. However, a direct comparison to first pass CMR perfusion or SPECT imaging would have been desirable and may be considered a limitation of the present study.

Conclusions

Rest and stress BOLD CMR at 3T proved feasible and may be used to differentiate between ischemic, non-ischemic, and normal myocardial segments in a clinical patient population. BOLD CMR during vasodilator stress identified patients with significant CAD.