Intra- and interobserver reliability of transorbital sonographic assessment of the optic nerve sheath diameter and optic nerve diameter in healthy adults

  • Original Article
  • Published:
Journal of Ultrasound Aims and scope Submit manuscript

Abstract

Purpose

Transorbital optic nerve sonography (TOS) can measure the optic nerve inclusive the sheath diameter (ONSD) and the optic nerve diameter (OND), which are useful parameters in the diagnosis of several neurological disorders. Data on the reproducibility of TOS are, however, required to use B-mode sonography for clinical purposes. The aim of this study was to assess intra- and interobserver reliability of ultrasound-based evaluations of both OND and ONSD in healthy subjects.

Methods

Using a 4-11-MHz linear array transducer, the OND and ONSD of 20 healthy subjects were independently measured by two expert investigators.

Results

Depicting the optic nerve and its sheath was possible in all subjects. The intra- and interobserver reliability was high for both ONSD and OND measurements. Intraobserver agreement, analyzed with Cronbach´s Alpha, was higher for ONSD (range: 0.69–0.72) than for OND measurements (range: 0.55–0.65). No differences in interobserver reliability between ONSD and OND measurements were found (p = 0.83 for right and 0.47 for left eye).

Conclusions

Transorbital B-mode sonography is a feasible method to assess both ONSD and OND with a high intra- and interobserver reliability. Technical difficulties in differentiating the optic nerve from its sheaths may explain the lower intraobserver agreement for OND than that for ONSD measurements.

Sommario

Obiettivo La sonografia del nervo ottico per via transcranica (TOS) è in grado di misurare il diametro del nervo ottico inclusi i suoi rivestimenti (ONSD) o il diametro del solo nervo ottico (OND), parametri utili nella diagnostica di molte patologie neurologiche. È necessario tuttavia disporre di dati sulla riproducibilità di tale metodica, ottenuta in sonografia B-mode, per poterla applicare clinicamente. Obiettivo di questo studio è stato quello di determinare il grado di riproducibilità intra- e inter-osservatore nella determinazione di questi due parametri in soggetti sani.

Metodi

Due esaminatori hanno determinato in maniera indipendente valori di ONSD ed OND in 20 soggetti sani utilizzando una sonda ecografica a 4–11 MHz.

Risultati

In tutti i soggetti è stato possibile visualizzare il nervo ottico e i suoi rivestimenti. Sono stati ottenuti elevati valori di riproducibilità intra- e inter-osservatore nella determinazione dell'ONSD e dell'OND. La riproducibilità intra-osservatore, analizzata con il metodo del Cronbach´s Alpha, era maggiore per l'ONSD (range: 0.69–0.72) che per l'OND (range 0.55–0.65). Non sono state riscontrate differenze nella riproducibilità inter-osservatore tra ONSD ed OND (p = 0.83 per l'occhio destro e 0.47 per quello sinistro).

Conclusioni

La TOS è un metodo in grado di ottenere i valori di ONSD ed OND con elevato grado di riproducibilità intra- ed inter-osservatore. Difficoltà tecniche nel distinguere il nervo ottico dai suoi rivestimenti possono spiegare il minor grado di riproducibilità intra-osservatore per l'OND rispetto all'ONSD.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Lochner P, Nardone R, Tezzon F, Coppo L, Brigo F (2013) Optic nerve sonography to monitor treatment efficacy in idiopathic intracranial hypertension: a case report. J Neuroimaging 23:533–534

    Article  PubMed  Google Scholar 

  2. Lochner P, Mader C, Nardone R, Tezzon F, Zedde ML, Malferrari G, Brigo F (2014) Sonography of the optic nerve sheath beyond the hyperacute stage of intracerebral hemorrhage. J Ultrasound. doi:10.1007/s40477-014-0069-6

    PubMed Central  PubMed  Google Scholar 

  3. Carraro N, Servillo G, Sarra VM, Bignamini A, Pizzolato G, Zorzon M (2012) Optic nerve and its arterial-venous vascularization: an ultrasonologic study in multiple sclerosis patients and healthy controls. J Neuroimaging. doi:10.1111/j.1552-6569.2012.00758.x

    PubMed  Google Scholar 

  4. Fernández-Domínguez J, García-Rodríguez R, Mateos V (2012) Transorbital echography for assessment of optical nerve atrophy in demyelinating diseases: a pilot study. Rev Neurol 54:587–592

    PubMed  Google Scholar 

  5. Titlić M, Erceg I, Kovacević T, Gabrić N, Karaman K, Zuljan I, Orsolić K, Kalajzić J (2005) The correlation of changes of the optic nerve diameter in the acute retrobulbar neuritis with the brain changes in multiple sclerosis. Coll Antropol 29:633–636

    PubMed  Google Scholar 

  6. Karakitsos D, Soldatos T, Gouliamos A, Armaganidis A, Poularas J, Kalogeromitros A, Boletis J, Kostakis A, Karabinis A (2006) Transorbital sonographic monitoring of optic nerve diameter in patients with severe brain injury. Transplant Proc 38:3700–3706

    Article  CAS  PubMed  Google Scholar 

  7. Geeraerts T, Newcombe VF, Coles JP, Abate MG, Perkes IE, Hutchinson PJ, Outtrim JG, Chatfield DA, Menon DK (2008) Use of T2-weighted magnetic resonance imaging of the optic nerve sheath to detect raised intracranial pressure. Crit Care 12:R114

    Article  PubMed Central  PubMed  Google Scholar 

  8. Geeraerts T, Merceron S, Benhamou D, Vigué B, Duranteau J (2008) Non-invasive assessment of intracranial pressure using ocular sonography in neurocritical care patients. Intensive Care Med 34:2062–2067

    Article  PubMed  Google Scholar 

  9. Malayeri AA, Bavarian S, Mehdizadeh M (2005) Sonographic evaluation of optic nerve diameter in children with raised intracranial pressure. J Ultrasound Med 24:143–147

    PubMed  Google Scholar 

  10. Bäuerle J, Lochner P, Kaps M, Nedelmann M (2012) Intra- and interobsever reliability of sonographic assessment of the optic nerve sheath diameter in healthy adults. J Neuroimaging 22:42–45

    Article  PubMed  Google Scholar 

  11. Brzezinska R, Schumacher R (2002) Diagnosis of elevated intracranial pressure in children with shunt under special consideration of transglobe sonography of the optic nerve. Ultraschall Med 23:325–332

    Article  CAS  PubMed  Google Scholar 

Download references

Conflict of interest

Piergiorgio Lochner, Lorenzo Coppo, Roberto Cantello, Raffaele Nardone, Andrea Naldi, Maurizio A. Leone, Francesco Brigo declare that they have no conflict of interest.

Human and animal studies

The study described in this article does not contain studies with human or animal subjects performed by any of the authors.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Francesco Brigo.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lochner, P., Coppo, L., Cantello, R. et al. Intra- and interobserver reliability of transorbital sonographic assessment of the optic nerve sheath diameter and optic nerve diameter in healthy adults. J Ultrasound 19, 41–45 (2016). https://doi.org/10.1007/s40477-014-0144-z