FormalPara Key Summary Points

Why carry out this study?

A new category of intraocular lenses (IOLs) has been introduced that improves intermediate vision of pseudophakic patients and induces a comparable level of photic phenomena to a standard monofocal implant. Given that those IOLs are of recent introduction, the scientific literature on their performance is scarce.

Quality metrics derived from the optical transfer function offer a high correlation with clinical visual acuity, which can be used to predict the clinical performance of new monofocal IOLs with an extended depth of focus prior to implantation.

What was learned from the study?

Enhanced-monofocal IOLs, based on a higher-order aspheric design, provide a comparable far optical quality and extended intermediate range while producing a monofocal-lens halo type. However, the depth-of-focus extension varies between the models, which should be taken into account in preoperative counseling.

Although the enhanced-monovision approach offered an extended visual range, a − 1D defocus and spherical aberration effects might increase a binocular perception of photic phenomena.

Introduction

Monofocal intraocular lenses (IOLs) with an enhanced intermediate function can provide an extended range of vision with a far-dominant performance, which, however, does not meet requirements for extended-depth-of-focus (EDoF) lenses set by the American Academy of Ophthalmology [1]. A recent clinical trial assessing mono-EDoF technology demonstrated that patients with the Tecnis Eyhance ICB00 have significantly better vision at intermediate distances but similar at far compared to those with a standard Tecnis lens [2]. In another clinical investigation, higher spectacle independence was observed at the intermediate range with the ICB00 compared to a monofocal lens [3]. The performance of the ICB00 has also been documented in laboratory studies [4, 5]. In those reports, however, only one IOL model has been evaluated. But with increasing recognition of this technology, various novel mono-EDoF lenses are being introduced, the precise characteristics of which have yet to be tested in the laboratory before they become routinely used in surgery.

Objective metrics derived from in vitro assessment of IOLs are widely used to evaluate the performance of new lens models [4,5,6,7,8,9,10] or failed implants [11, 12]. One such metric is the optical transfer function (OTF) that represents the imaging ability of an optical system. This is a complex-valued function composed of the modulation transfer function (MTF) and the phase transfer function (PTF). The latest research indicates that optical-bench metrics offer a high correlation with clinical measurements of visual acuity (VA) and contrast sensitivity (CS) [7,8,9]. The optical quality of IOLs is typically assessed, during their manufacture, using the through-focus MTF, which is the internationally recognized standard for testing IOLs [6, 13]. However, the inclusion of the PTF component may provide a more accurate estimate of the IOL’s performance and its impact on postoperative vision [14].

Halos are dim circles of light surrounding a light source, which may occur due to refraction, diffraction, or ray aberration. In the case of IOLs, it often results from a simultaneous projection of multiple foci by an IOL. Thus, this postoperative complaint is typically associated with multifocal IOLs rather than monofocal ones [15, 16]. Recent clinical and laboratory studies demonstrated a comparable amount of dysphotopsia with the ICB00 and the control monofocal Tecnis IOL [2,3,4,5], which may be used as a reference to test the susceptibility of new mono-EDoF designs to produce such phenomena.

This research aimed to apply the MTF and the PTF to compare the optical quality and the halo projection of the latest mono-EDoF models to that of a standard monofocal lens.

Methods

Intraocular Lenses

In this laboratory investigation, we studied off-the-shelf IOLs; thus, ethics committee approval was not required. The following models were assessed.

  • Tecnis Eyhance ICB00 and Tecnis ZCB00 (both from Johnson & Johnson Surgical Vision, Inc., Santa Ana, USA).

  • AE2UV (Eyebright Medical Technology Inc, Beijing, China), distributed in Europe under the tradename ZOE Primus-HD (Ophthalmo Pro GmbH, Sankt Ingbert, Germany).

  • IsoPure 1.2.3 (PhysIOL sa/nv, Liège, Belgium).

  • RayOne EMV (Rayner Intraocular Lenses Limited, Worthing, UK).

Two IOL samples from each model were tested, each having the same refractive power of + 20D.

The Tecnis ZCB00 and ICB00 are made of the same hydrophobic acrylic material with a refractive index of 1.47 at 35 °C and the Abbe number of 55. The two models were designed to correct a spherical aberration (SA) of 0.27 µm at 6 mm [17]. Despite a similar appearance, the Eyhance’s anterior surface differs (by approximately 15%) in its central, optical area from that of the ZCB00—having an increase in lens power through utilizing higher-order aspheric components [4]. Thus, the Eyhance ICB00 aims to enhance intermediate vision compared to the ZCB00 while maintaining identical correction of primary SA [4].

The AE2UV/ZOE is a hydrophobic-acrylic lens with a refractive index of 1.47 and the Abbe number of 57. The IOL features an aspheric design to lower the primary SA of the cornea by − 0.20 µm at 6 mm. A high-order aspheric surface enhances intermediate vision with increased SA aberration in the IOL center that gradually decreases towards the periphery. A smooth and continuous higher-order aspheric surface aims to minimize photic phenomena that may result from abrupt profile changes, such as observed in diffractive IOLs.

The IsoPure 1.2.3 is a hydrophobic-acrylic IOL with a refractive index of 1.52 and an Abbe number of 42. The IsoPure features anterior and posterior aspheric surfaces with high-order aspheric terms to extend the visual range compared to a monofocal IOL while maintaining a good far-focus performance. A posterior surface of the lens has a conic profile to correct − 0.11 μm of SA.

The RayOne EMV is made of (26%) hydrophilic-acrylic material with a refractive index of 1.46 at 35 °C and an Abbe number of 56. The RayOne EMV is offered as enhanced monovision IOL where a plano target is maintained in the dominant eye, and a power-offset is applied in the non-dominant eye. According to the manufacturer, a 1D offset offers a 2.25D depth-of-focus extension in binocular vision. The inner part of the IOL includes positive SA to improve patients' distance vision in the non-dominant eye.

Optical-Metrology Analysis

We used an OptiSpheric IOL PRO2 (Trioptics GmbH, Wedel, Germany) device in the assessment of the optical performance of the study IOLs. This setup was built based on the ISO standard configuration that includes a corneal model with 0.28 µm of SA at 5.15 mm [13]. The device’s spectral property mimics the spectral sensitivity of the human eye with a peak intensity of 555 nm. The OTF components (i.e., the MTF and the PTF) were derived from the line spread function projected by the device through the tested IOL [18]. The longitudinal chromatic aberration of the model eye (without an IOL) was approximately 1D between 480 and 644 nm [19]. We assessed the optical quality parameters for the aperture size of 3 mm and 4.5 mm at the IOL plane [20]. After finding the best focus using the MTF criterion, the IOL’s tolerance to defocus was tested within a + 0.50D to − 2.50D range (0.25-D resolution). At each focus-point, sagittal and tangential MTFs and PTFs were acquired, and the two meridians were averaged. Besides, at the 3-mm pupil, we recorded the 1951 USAF resolution test chart.

The weighted optical transfer function (wOTF) was calculated as described by Alarcon et al. [7], who in a later study also applied this metric to evaluate the optical quality of the ICB00 lens [4]. The wOTF was derived from the measured MTF and PTF at 3 mm and weighted by the neural CS function (CSF) using the following formula:

$$\mathrm{wOTF}=\frac{d}{150} {\sum }_{f=1}^\frac{150}{d}\mathrm{MTF}\left(fd\right)\mathrm{cos}\left(\mathrm{PTF}\left(fd\right)\right){\mathrm{CSF}\left(fd\right)}_{\mathrm{neural}},$$

where f is the spatial frequency expressed in lp/mm with a resolution of d = 1 lp/mm. The neural CSF data were obtained from the work of Campbell and Green [21]. The wOTFb change with defocus (where b is − 0.36) has been reported to correlate strongly with clinically measured VA, and we compared this change between models [7, 22].

A polychromatic point spread function (PSF), i.e., an image of a 0.1-mm pinhole, was used to assess the size of halos projected by the IOLs at the 4.5-mm aperture [4, 16]. A long transformation of each recorded image was applied, and the background noise was removed. The halo size was defined as the total area (2-D images) with the assumption of circularity.

Binocular Summation

Given that the RayOne EMV is implanted bilaterally through monovision, we performed the binocular summation of recorded USAF-chart and PSF images. We applied a quadratic summation proposed by Legge and Rubin for a pair of images corresponding to the left and right eye with the defocus offset of 1D [23, 24].

Data Analysis

The analysis of optical-quality data and images were performed with custom-made software developed in MATLAB (MathWorks, USA).

Results

The MTF curves of all IOL samples measured at the best focus through the 3- and 4.5-mm apertures are presented in Fig. 1 with 2D MTF visualization shown in Fig. 1S (Supplementary Material).

Fig. 1
figure 1

Modulation transfer function (MTF) levels of the studied IOLs at the best focus for 3- and 4.5-mm apertures. The dotted lines show the values of each lens separately; the solid lines refer to the average of two samples

At 3 mm, the ZCB00 demonstrated the highest MTF values of the five models, particularly at higher spatial frequencies, with an average level of 0.50 ± 0.00 at 50 lp/mm. The AE2UV/ZOE produced the average MTF @ 50 lp/mm of 0.37 ± 0.02. At the same frequency, the ICB00’s MTF was comparable, and that was 0.36 ± 0.01. The RayOne EMV’s MTF was 0.40 ± 0.00, which was reduced dramatically after the 1D offset (MTF = 0.04 ± 0.01). The IsoPure’s performance at 50 lp/mm (MTF = 0.41 ± 0.01) was minimally better than found in the other mono-EDoF models.

At 4.5 mm, the ZCB00’s high-frequency performance (MTF@50 lp/mm = 0.39 ± 0.01) was better than that of the mono-EDoF IOLs. The ICB00’s MTF was 0.31 ± 0.01, comparable to that measured with the AE2UV/ZOE (0.29 ± 0.00) at 50 lp/mm. For the IsoPure, it was 0.23 ± 0.00. The RayOne EMV’s MTF at zero-defocus was 0.17 ± 0.00, but a simulation of a − 1D refractive target resulted in a poor far-focus performance at 4.5 mm (MTF@50 lp/mm = 0.02 ± 0.00).

Figure 2 reports the wOTFb change with defocus for the five IOL models. The wOTFb of the ZCB00 was minimally better at the best far-focus than that of the other models. However, at about − 0.5D, the ICB00, the AE2UV/ZOE, and the IsoPure demonstrated improved tolerance to defocus compared to the monofocal IOL. The RayOne EMV’s and the ZCB00’s performance under defocus was comparable with the simulated plano refractive target. Although the 1D offset resulted in a worse wOTFb level of the RayOne EMV at the far focus, in binocular vision, the defocus tolerance may be further extended.

Fig. 2
figure 2

Defocus tolerance of the weighted optical transfer function to the power of b = − 0.36 (wOTFb) assessed at 3 mm and visual acuity (VA) predictions presented on a secondary axis. The dotted lines show each lens′ values separately; the solid lines refer to the average of two samples. The vertical dashed line indicates the position of the far focus

The resolution-test images confirm the wOTFb results. Figure 3 presents the quadratic summation results for the RayOne EMV and the recorded (single) images for the other models. Figure 2S (Supplementary Material) shows the comparison between the USAF-chart photographs taken with the simulated plano and − 1D refractive target and the corresponding binocular image.

Fig. 3
figure 3

US Air Force resolution targets recorded at a defocus range of + 0.5D to − 2.5D and the 3-mm aperture. *Quadratic summation of two images with the defocus difference of 1D

The PSF projection confirmed a similar halo profile among the ZCB00, the ICB00, and the AE2UV/ZOE (Fig. 4). The ZCB00 produced the halo radius of the smallest size (1.73 ± 0.06 arcmin), which was closely followed by the ICB00’s 1.88 ± 0.08 arcmin and AE2UV/ZOE’s 2.01 ± 0.04 arcmin. The IsoPure created a halo with an average size of 2.64 ± 0.00 arcmin. The RayOne EMV’s binocular halo size was 3.67 ± 0.18 arcmin, which was the largest among the studied IOLs. Figure 3S (Supplementary Material) shows the PSF images of the RayOne EMV at zero, and − 1D defocus with the corresponding quadratic summation.

Fig. 4
figure 4

Logarithmic images of the recorded polychromatic point spread function (PSF, left panels) at 4.5 mm with color-coding. The right panels present the linear cross-section of the PSF’s intensity profile. *Quadratic summation of two images with the defocus difference of 1D

Discussion

This study provides further evidence that mono-EDoF IOLs have the potential to improve the patient’s visual function at the intermediate range while producing the optical performance that is close to that of standard monofocal IOLs at their best focus. Although the RayOne EMV showed a more extended range of vision than that of the other mono-EDoF models, the impact of binocular summation is challenging to predict in a laboratory setting.

Given that the studied IOLs are still relatively new to clinicians, we were not able to identify any peer-reviewed publication on the in vivo or in vitro functioning of the AE2UV/ZOE, IsoPure, and the RayOne EMV. This gap underlines the importance of our research. In a recent publication, Mencucci et al. made a clinical comparison of the ZCB00 and ICB00 [3]. Various aspects of the visual function were assessed, including VA, CS, and the perception of photic phenomena. They found that monocular corrected distance VA differs only by 0.01 in favor of the ZCB00 group. However, monocular distance corrected intermediate VA measured at 66 cm (~ 1.5D) was improved in patients with the mono-EDoF lens by one line [3]. Those findings are in agreement with the results of our in vitro evaluation, as we also found that ZCB00 had a minimally higher wOTFb than the ICB00 at far-point, but worse at intermediate by 0.12. Following the conversion of the wOTFb to logMAR VA using a model proposed by Alarcon et al. [7, 22], a 1.3 line difference between the two Tecnis IOLs can be predicted at − 1.5D.

In a paper by Auffarth et al., the European multicenter study results on the comparison between the ICB00 and ZCB00 IOLs were presented [2]. They found that the mean corrected distance VA was − 0.02 ± 0.01 logMAR in the mono-EDoF group and − 0.06 ± 0.01 logMAR in the standard-monofocal one. In the current in vitro study, we also found a minimally decreased wOTFb of the ICB00 IOLs at the far focus with the estimated VA effect of 0.02 logMAR in favor of the ZCB00 group. At the intermediate range (66 cm), the distance corrected VA was 0.19 ± 0.02 logMAR and 0.31 ± 0.02 logMAR in the mono-EDoF and monofocal patients, respectively. They reported the mean improvement of 0.11 ± 0.02 logMAR, which is close to the predicted level from our in vitro assessment. At 1.5D defocus, the AE2UV/ZOE’s wOTFb only slightly differed from that of the ICB00, which would result in a minimally lower VA by 0.01 logMAR and 0.03 logMAR at 2.5D according to the model [7, 22]. The AE2UV/ZOE provides a clear advantage over the standard monofocal lens, as at 66 cm and 40 cm, one line VA improvement (0.1 logMAR) is expected at both distances compared to the ZCB00. For the IsoPure it is less, about 0.07 logMAR at 1.5D of defocus. Although RayOne EMV showed < 0.02 logMAR difference over the monofocal lens in the ‘dominant eye’ simulation, in the ‘non-dominant eye’ a nearly 3-line (0.27 logMAR) improvement would be expected. However, these laboratory results should be confirmed in a clinical study.

The ICB00 and the ZCB00 have also been compared in a laboratory setting. Alarcon et al. studied their optical quality by measuring the MTF and the OTF in polychromatic light using their optical bench [4]. Despite expected differences in our and the J&J Vision group’s optical setup, they also predicted a 0.12 logMAR improvement at the intermediate range after the implantation of their mono-EDoF lens compared to the standard monofocal model. Furthermore, Alarcon and associates reported a negligible impact of higher-order aberrations and pupil size on their comparison. However, Vega et al. found a shift of the through-focus MTF area by − 0.50D at 2 mm, indicating some level of pupil dependency [5]. Although Auffarth et al. did not see clinically significant changes for pupils larger than 2.5 mm [2], the comparison for smaller pupils was not performed due to insufficient sample size. More research is needed to determine the impact of pupil size on the visual quality of patients with the new mono-EDoF IOLs.

The European multicenter clinical study concluded that the occurrence of photic phenomena is comparable between the ICB00 and ZCB00 IOLs [2]. Mencucci et al. also reported insignificant differences between the two groups in the perception of glare and halo [3]. The two laboratory studies have shown that the ICB00 and the ZCB00 produce a similar halo pattern with a slightly lower intensity profile for the latter model [4, 5]. Our analysis of the PSF may further reinforce those findings, as we also reported that the observed halo was only 8% larger in the ICB00 than that recorded through the monofocal lens. This difference was slightly increased but still comparable for the AE2UV/ZOE, which was 14%. Our results indicate that the AE2UV/ZOE’s contribution to photic phenomena is minimal, and it is comparable to that of a monofocal lens.

The IsoPure and the RayOne EMV produced a halo pattern that was larger than that of the ZCB00 by 34% and 53%, respectively. One potential explanation of this observation is a lower SA correction. In the current study, the cornea model exerts 0.28 µm of SA while the IsoPure corrects − 0.11 µm, but the RayOne EMV is considered ‘aberration neutral.’ Although the choice of this SA correction might have been motivated by the broad range of SA distribution in the population, in this in vitro model, we simulated the average SA value of the cornea [17], which favors the lens design of the Tecnis IOLs. In such a setting, however, the IsoPure and RayOne EMV performance at the scotopic pupil could be compromised, which might also result in an increased halo-pattern due to a higher level of SA. Two other factors may also contribute to the increased halo size in the RayOne EMV.

  1. 1.

    The Rayner model induces additional positive SA to optimize the far vision of the non-dominant eye. Thus, we expect that increased SA effects may spread the RayOne EMV’s halo-profile and reduce the optical quality, as shown in Fig. 1.

  2. 2.

    Similarly, defocus blur also has the potential to increase the halo perception [25, 26].

Figure 4 shows the quadratic summation of the focus and unfocused (1D offset) PSF. Although both log images (Fig. 3S, Supplementary Material) show an extended halo pattern around the central disk, defocus appears to produce higher-intensity effects in the simulation of the 1-D offset. Johannsdottir and Stelmach found that subjective complaints about photic phenomena under low-light conditions are more common in monovision compared to binocular emmetropia [25]. Still, the monovision approach provides fewer halo symptoms than the implantation of a diffractive-bifocal IOL, as Zhang et al. have demonstrated [26]. Thus, whether the observed increased halo profile in the RayOne EMV may induce functionally significant photic phenomena is a question that must be addressed in a clinical study.

In a healthy emmetropic eye, the summation of the visual information from both eyes results in enhanced visual perception compared to monocular vision. Campbell and Green estimated that binocular summation improves the visual quality by a factor of \(\sqrt{2}\) [27]. However, a recent meta-analysis by Baker and Lygo indicates that the ratio between binocular and monocular perception may be higher, implying a multifactorial mechanism underlying this process [28]. Conversely, the binocular summation may also be hampered, e.g., under monovision correction, with the ratio close to or less than 1 indicating the deterioration of binocular vision [29, 30]. If we assume that the binocular summation is 1 for the RayOne EMV and \(\sqrt{2}\) for other IOLs, the far-focus wOTFb increases for all models but one (Fig. 5). The binocular summation may also minimize the image quality difference between the RayOne EMV and the other mono-EDoF models at the intermediate distance, which may be a good option for patients who may not tolerate well dissimilarity in the blur perception between the eyes. More research is needed to determine the impact of enhanced monovision on patients’ binocular performance.

Fig. 5
figure 5

Simulations of wOTFb binocular summation at 3 mm and under defocus. The dotted lines show each lens′ values separately; the solid lines refer to the average of two samples. The vertical dashed line indicates the position of the far focus

The use of one pupil size (i.e., 3 mm) might not fully represent a clinical situation [23], which may be considered a limitation of the current study. However, the wOTF metric was computed by Alarcon et al. from laboratory data obtained at 3 mm [7]; thus, its implementation for other pupil sizes requires validation. This should be a subject of further research.

Conclusions

We found that the mono-EDoF models' far-point MTF was minimally lower than that of the standard monofocal IOL, but the expected impact on clinical VA is negligible. All mono-EDoF models have the potential to extend the patient’s intermediate vision beyond the range of a standard monofocal lens. Our laboratory results for the two Tecnis IOLs appear to conform to the published results from other researchers. We found that the AE2UV/ZOE and the ICB00 produced a halo pattern comparable to that of the ZCB00 indicating a low potential of these IOLs to induce photic phenomena.