The Integrity of Latex Gloves in Clinical Dental Practice

  • Clinical and Epidemiological Studies
  • Published:
Infection Aims and scope Submit manuscript

Summary

Background: The aim of the study was to assess the integrity of non-sterile latex gloves (Biogel® Diagnostic (A), Biogel® Dental (B), Gentle Skin® (C) and Manufix® (D)) in clinical dental practice.

Materials and Methods: After use, the gloves were collected and examined for perforations using the water inflation test according to the EN 455-1 Standard. Over a 12-month period 847 used gloves were collected and assessed. The mean duration of use for gloves A/B/C/D was 175/112/78/79 min respectively. In spite of these differences favoring gloves C and D, 16% of glove A, 14% of glove B, 21% of glove C and 29% of glove D were found to leak. The proportion of perforated gloves increased slightly if a hygienic handrub using an alcoholic hand disinfectant had been carried out immediately prior to gloving.

Conclusion: The results indicate striking differences between the quality of medical gloves. As all the gloves comply with the European regulations in terms of integrity, it is very difficult for the user to distinguish between gloves of superior or inferior quality. In a clinical setting it is therefore recommended that the appropriate glove should be determined for each situation. Hands must be thoroughly dry if they are disinfected prior to gloving, since hands still wet with alcoholic disinfectant may increase the risk of glove perforation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: September 15, 1999 · Revision accepted: April 10, 2000

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pitten, FA., Herdemann, G. & Kramer, A. The Integrity of Latex Gloves in Clinical Dental Practice. Infection 28, 388–392 (2000). https://doi.org/10.1007/s150100070011

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s150100070011

Navigation