A 28-year-old woman with familial hypertrophic cardiomyopathy caused by a mutation in the gene encoding MYH7 presented with an implantable cardioverter defibrillator (ICD) shock to the emergency department. In the past she suffered from non-sustained ventricular tachycardias as well as atrial flutter but she had not experienced an ICD shock before. She cycled home and felt dyspnoeic and shortly thereafter experienced a bang and intense burning sensation retrosternally. ICD interrogation revealed an inappropriate shock because of P wave oversensing. Fig. 1a shows her electrocardiogram with large P waves due to atrial enlargement. Fig. 1b shows inappropriate sensing with detection of both QRS complexes and P waves (arrows indicate sensed P waves). This triggered the ICD to deliver a shock as the calculated frequency exceeded the conditional shock zone. To prevent future shocks, the vector was manually selected and the SMART Pass filter was turned on. During exercise testing appropriate QRS sensing was observed up to 130 beats per minute.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Mulder, B.A., Maass, A.H. & Blaauw, Y. Inappropriate shock caused by P wave oversensing in an entirely subcutaneous ICD. Neth Heart J 26, 411–412 (2018). https://doi.org/10.1007/s12471-018-1099-z
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12471-018-1099-z