Introduction

Normal knee function relies on smooth, uninterrupted motion, which in turn depends on stable, well-lubricated, low-friction articular surfaces. Knee replacement involves compromises between stability and flexibility, and historically has included removal of one or both cruciate ligaments. As an alternative to this philosophy, a bicruciate-retaining knee prosthesis was developed in 1971 [24]. The design emphasized preserving the ligaments, minimizing bone resection, and limiting constraint with the goal of allowing more natural movement of the knee compared with other prostheses. AP stability was provided by the preserved cruciate ligaments rather than the shape of the polyethylene.

Although the ACL is functional in more than 60% of patients undergoing a TKA [1, 7, 8] and kinematic studies support the concept of bicruciate-retaining (that is, ACL-preserving) TKA [14, 23], few studies have evaluated patients treated with bicruciate-retaining TKA implants [1619, 21]. Because of these perceived advantages, I used a bicruciate-retaining TKA implant in my practice for all patients with an intact ACL and acceptable deformity during a 3-year period, and because of the limited number of studies available at longer term [21], I sought to evaluate their performance and durability.

I asked: (1) what is the long-term (minimum 20-year) survivorship, (2) what are the functional results, and (3) what are the reasons for revision of a bicruciate-retaining knee arthroplasty?

Patients and Methods

From January 1989 to September 1992, I performed 639 total knee replacements in 537 patients. Of these, 489 (77%) were performed in 390 patients using a bicruciate-retaining, minimally constrained device (Townley Anatomic, Biopro Inc, Port Huron, MI, USA) (Fig. 1). Patients selected for a TKA with a bicruciate-retaining prosthesis had no history of ACL insufficiency. During the period in question, this prosthesis was used for all patients observed intraoperatively to have an intact, functional ACL and limited deformity. There were 156 men (40%) and 234 women (60%) with a mean age at surgery of 65 years (SD, 8.6; range, 42–84 years). The primary diagnosis was osteoarthritis in 347 (89%) knees and inflammatory arthritis or osteonecrosis in 43 (11%). At the time of the surgery, the ACL was functionally intact as shown by testing with a probe and by negative Lachman, pivot shift, and anterior drawer tests. The appearance and integrity of the ACL also was tested before and after placing the implants. For patients without an ACL and in patients with greater than 15° valgus or 15° varus, I used a posterior cruciate or posterior stabilized prosthesis. The final decision regarding implanting a bicruciate-retaining prosthesis was made intraoperatively.

Fig. 1
figure 1

These drawings show the prepared femur and tibia including the retained tibial eminence and both cruciate ligaments. The femoral and two-piece tibial component is shown with the prepared slot in the tibia for the keel and pilot holes for the spikes of the tibial component.

The surgical technique followed the basic principles of ligamentous balancing. Neither the ACL nor the PCL was recessed. The tibia was prepared with 6° of posterior slope. Initially, alignment and balance in extension were achieved by correcting the coronal deformity with appropriate capsular and ligamentous releases. The femur was prepared first and the tibia was not subluxed forward on the femur during surgery. The tibial osteotomy was performed in flexion in 2° to 3° varus with respect to the mechanical axis of the knee [12, 24]. The tibial spines and insertions of the cruciate ligaments were left in continuity with the rest of the tibia. The tibial component was placed in slight external rotation following the orientation of the fibers of the ACL [21]. A complete ROM test was performed with trial implants in place. There were two intraoperative complications consisting of two instances of tibial eminence fracture as the knee was moved from flexion to extension during trial ROM testing. This occurred because the knee was too tight owing to insufficient distal femur or proximal tibial resection, or incomplete ligament balancing. The eminence was repaired with a screw, the knee was rebalanced, and the procedure was completed. There were no ligament ruptures during surgery. These procedures were performed by one surgeon (JWP) with many years of experience performing TKAs with implantation of a bicruciate-retaining prosthesis.

The femoral component is cobalt-chromium with an asymmetric trochlear groove requiring right and left components. The tibial component consists of a flat symmetric polyethylene (PE) insert on a single-piece horseshoe-shaped titanium tibial tray with two round fixation pegs. The prosthesis is minimally constrained. One-piece tibial PE inserts 8 to 11 mm thick were used for all patients. Conventional 1050 ram-extruded PE sterilized in ethylene oxide was used. Patellar resurfacing was performed for all patients with a dome-shaped PE patellar component. Implants were placed with cement [24].

Range of knee movement was assessed with a goniometer, laxity was assessed manually, and American Knee Society knee scores were completed at each visit [13]. Patients were followed every 2 years with The Knee Society Score©. The knee score at the end of the second year was used to compare with the preoperative score.

Standing AP (Fig. 2A) and lateral (Fig. 2B) radiographs of the affected knee were taken at each followup to look for signs of implant failure such as wear and loosening. The orientation of the prosthesis relative to the anatomic axis of the knee was measured on the AP and lateral radiographs.

Fig. 2A–B
figure 2

(A) AP and (B) lateral radiographs show the bicruciate-retaining total knee prosthesis 20 years after insertion. There is a cemented all polyethylene patellar prosthesis and the femoral and tibial prostheses are cemented. The preserved tibial eminence is seen in the center of the radiograph. The spikes and keel of the tibial component are shown. There is no sign of component loosening or polyethylene wear.

Patients were followed every 2 years for a minimum of 20 years after the surgery (range, 20–24 years; mean, 23 years). At the time of this review, 199 (51%) patients had died (237 knees) and 31 (8%) (37 knees) were lost to followup, leaving160 (41%) (214 knees) with a mean age of 84 years (range, 65–101 years) at followup.

The implant survival rate was determined by a competing-events and Kaplan-Meier survivorship analysis with 95% CIs with revision for any reason as the endpoint [4, 9, 10]. Comparisons were tested using the Mann-Whitney U test and chi-square analysis where appropriate. A p value less than 0.05 was considered statistically significant. SPSS Version 20 (SPSS Inc, Chicago, IL, USA) was used for the statistical analyses.

Results

The Kaplan-Meier survivorship analysis at a mean followup of 23 years was 89% (95% CI, 82%–93%) with revision for any reason as the endpoint (Fig. 3). Competing-risk survivorship at 23 years was 94% (95% CI, 91%–96%) (Fig 4).

Fig. 3
figure 3

A Kaplan-Meier survivorship curve with revision for any reason as the endpoint with 95% CIs (dotted lines) is shown.

Fig. 4
figure 4

A competing-risk survivorship curve with revision for any reason as the endpoint with 95% CIs (dotted lines) is shown.

Patients showed improvements in preoperative to postoperative flexion and Knee Society scores (Table 1). Flexion improved from a preoperative mean of 104° to 117° postoperatively (p < 0.001) and Knee Society scores improved from a preoperative mean of 42 (range, 26–49) to 91 (range, 61–100) postoperatively (p < 0.001).

Table 1 Mean pre- and postoperative functional scores for 160 patients (214 knees)

Twenty-two knees in 21 patients were revised (Table 2). The mean time from implantation to revision was 12 years (range, 5–22 years). No difference was found between the group of patients with revised implants and the group with surviving implants for primary diagnosis, mean age, mean BMI, mean postoperative tibiofemoral axis, mean posterior slope of the tibial tray, mean ROM, and stability.

Table 2 Indication(s) for revision of 22 knees (21 patients)

The most common reason for revision was PE wear and this was the sole reason for revision in seven knees. The PE was exchanged in these seven knees. Two knees that were unstable and one that was stiff also were treated by PE exchange. The revision procedure consisted of exchange of only the PE insert in 10 patients. Femoral and tibial component revision was performed in four knees, tibial component revision was performed in three knees, and the patellar component was revised in one knee. The mean PE thickness in revised knees was 11 mm (range, 8–14 mm). Osteolysis was seen in seven knees during revision surgery. The components in three knees were revised to stemmed components. In four knees the areas of osteolysis at surgery were 4 to 7 mm and in three knees the areas were 8 to 12 mm. The ACL and PCL were present during revision surgery in all but one knee.

Discussion

The ACL is commonly sacrificed during TKA, but there is some evidence that retention of the ACL results in superior kinematics after surgery [14, 22, 23]. During the early development of total knee implants, several prostheses were designed specifically to retain the ACL and PCL [21, 24]. Between 1989 and 1992, I used a bicruciate-retaining implant for patients undergoing TKA who had an intact ACL and acceptable deformity. Because of the paucity of long-term data regarding bicruciate-retaining total knee implants, I sought to assess the results of procedures at least 20 years later. I posed the following questions: (1) What is the long-term survivorship of this bicruciate-retaining knee replacement? (2) What are the functional results of this bicruciate-retaining knee replacement? (3) What are the reasons for revision of bicruciate-retaining knee replacement?

This study has limitations. As with other studies of TKAs with followups greater than 20 years, the number of patients who died is high yet similar to the mortality rate of a comparable population without knee problems [6, 15, 20, 21]. At followup, the mean age of the patients was 84 years and these patients were likely more sedentary and less demanding of their knees for activity. The current knee scoring systems lose their validity when measuring patients in the third decade after surgery [15]. A TKA in which a bicruciate-retaining prosthesis is implanted is a demanding surgical procedure. If the knee is not balanced, the ACL will rupture or the tibial eminence will fracture during surgery as the knee’s ROM is tested moving from flexion to extension. These procedures were performed by one surgeon (JWP) with many years of experience implanting bicruciate-retaining prostheses during TKAs; thus, the results described in this report may not be reproducible in every center. Further, because of patient selection at the time of surgery, it is possible that patients with either less disease burden or with secondary stability were favored to receive the bicruciate-retaining implant. Although followup of the patients receiving implants other than the bicruciate-retaining device during the same time is not complete, almost 80% of patients did receive the implant in question. Finally, in this study, joint stability was assessed manually and may not be accurate. More sophisticated techniques to determine joint stability during and after surgery are being developed, such as sensor-guided technology [11].

In the patients in the current study, there was a high rate of implant survivorship and a low rate of complications using a bicruciate-retaining total knee prosthesis. Although implant survivorship is not the same as wellness, the functional results achieved by the patients are comparable or superior to results described in other reports using the total condylar, press-fit condylar, and meniscal-bearing prostheses [6, 15, 20]. The ACL in the osteoarthritic knee often is degenerated macroscopically and histologically [1, 7, 8]. However, I found that with a functional ligament, degeneration was not a barrier to implanting a bicruciate-retaining prosthesis during TKA. Sixty percent or more of my patients, and as reported in other studies [1, 7, 8], had a functionally intact ACL at presentation. The results presented here indicate a long-term survival rate for TKAs with this bicruciate-retaining prosthesis is similar to rates for TKAs with posterior cruciate-retaining and posterior-stabilized prostheses [3, 13, 18]. In the current study, the survivorship rate is similar to the 91% survival rate in 46 knees after 12 years with the meniscal-bearing bicruciate-preserving implant [6]. The Cloutier bicruciate-retaining prosthesis showed a survivorship of 82% at 20 years [21]. Kaplan-Meier survival analysis is commonly used to estimate cumulative incidence of revision after a joint arthroplasty. Because of concern in the literature regarding overestimation of revision [9], competing risk analysis is recommended in studies where death is common [4, 10]. Since a revision cannot occur when there is death, survivorship can be overestimated with other statistical methods; therefore competing risk and Kaplan-Meier analyses were performed here.

In this study, patients achieved knee scores comparable to those reported in 20-year studies of other designs [6, 15, 20, 21]. However, some studies indicate that saving both cruciate ligaments during a TKA results in more normal kinematics compared with using posterior cruciate-retaining medial pivot and posterior-stabilized prostheses [14, 22, 23]. The mean flexion of 117° at followup is consistent with flexion achieved with implants available more than 20 years ago [6, 15, 21]. Randomized comparison studies in which patients received a bicruciate-retaining knee prosthesis on one side and a posterior cruciate-retaining or substituting prosthesis on the other side have shown that 76% of patients prefer the bicruciate-retaining prosthesis [16, 18]. Comparison studies I have done also show there is less noise, less heat generated, and less bone loss with a bicruciate-retaining prosthesis compared with other designs [17, 19]. A TKA with a bicruciate-retaining prosthesis necessitates close approximation of the prosthesis to the anatomic contours, meticulous alignment technique, and preservation of well-balanced cruciate and collateral ligaments. A TKA in which a bicruciate-retaining prosthesis is used is a demanding surgical procedure compared with a TKA with either a posterior-cruciate or posterior-stabilized prosthesis. Preserving the ACL requires the correct ligament tension and relies on preserving the joint line, ligament balance, and alignment in all planes.

PE wear was the main mode of failure in this series (nine of 22 knees). There was a low rate of osteolysis in this series, femorotibial instability was seen twice, and late rupture of the ACL occurred once. Although the tibial component only had two stabilizing pegs and a small keel, loosening and bone loss around the prosthesis were rare in this series. The minimal constraint of the prosthesis may limit the shear on the prosthesis. Other series have shown that good long-term survival with a low incidence of osteolysis and loosening is possible using pegged tibial trays and limited constraint [3, 5]. Patellar complications were uncommon in the current patients. It is possible that preserving the ACL is helpful for patellar stability. I did not perform a lateral patellar retinacular release in any patient. The more anatomic trochlea of the femoral component used may result in reduced forces in the patellofemoral joint [2].

I found satisfactory survivorship and function for patients receiving a bicruciate-retaining prosthesis after more than 20 years of use. Many more patients are included in the current study than in another study using bicruciate-retaining prostheses [21]. This bicruciate-retaining prosthesis showed improved function, ROM, and a lower revision rate than other bicruciate-retaining prostheses that also had satisfactory long-term outcomes [6, 21]. The bicruciate-retaining prosthesis used in this long-term study is no longer commonly used and has been replaced by more contemporary prostheses. Sacrifice of the ACL may be an unnecessary concession during TKA. However, a TKA with implantation of a bicruciate-retaining prosthesis is a demanding surgical procedure. Meticulous knee balancing is necessary to avoid intraoperative ACL rupture or tibial eminence fracture when the ROM is tested moving the knee from flexion into extension. Exposure can be difficult as the tibia is not subluxed forward on the femur and less invasive techniques are not possible with a TKA with implantation of a bicruciate-retaining prosthesis. Future studies using validated patient-reported outcomes should be done to determine if retention of both cruciates during TKA is appropriate in selected patients.