Skip to main content
Log in

Radial Head Reconstruction in Elbow Fracture-Dislocation: Monopolar or Bipolar Prosthesis?

  • Symposium: Traumatic Elbow Instability and its Sequelae
  • Published:
Clinical Orthopaedics and Related Research®

Abstract

Background

Monopolar and bipolar radial head prosthetic arthroplasties have been used successfully to treat elbow fracture-dislocation with unsalvageable radial head fractures. The relative stability of these two designs in different clinical situations is a topic of ongoing investigation.

Questions/purposes

We tested the effects of monopolar and bipolar fixed-neck prosthetic radial head implants on improvement in elbow coronal and axial plane laxity in a terrible triad biomechanical model that accounted for lateral collateral ligament integrity and the presence of a transverse coronoid fracture.

Methods

Kinematic data were collected on six fresh-frozen cadaveric upper extremities tested with passive motion throughout the flexion arc. Varus and valgus gravity stress were applied with the wrist in neutral position. A lateral collateral ligament reconstruction was simulated. We assessed instability after radial head resection and reconstruction with either a monopolar or bipolar implant in the presence of a transversely fractured (Regan and Morrey Type 2) or fixed coronoid process.

Results

With collateral ligament integrity, no difference was detected, with the numbers available, in valgus laxity between implants under valgus stress (p = 1.0). Laxity improvement with each prosthesis was higher when the coronoid was fractured (mean ± SD: monopolar: 7.4° ± 1.6°, p < 0.001; bipolar: 6.4° ± 1.6°, p = 0.003) than when it was fixed (monopolar: 4.0° ± 1.6°, p = 0.02; bipolar: 4.2° ± 1.6°, p = 0.01). With the numbers available, there was no difference in external rotation laxity between implants under valgus stress (p = 1.0). The greatest stabilizing effect of the prostheses occurred when the coronoid was fractured (monopolar: 3.3° ± 1.2°, p = 0.15; bipolar: 3.3° ± 1.2°, p = 0.17). Radial head arthroplasty offered no substantial stability under varus stress for varus or internal rotation laxity.

Conclusions

In our terrible triad cadaveric model, coronoid fixation was effective in improving varus laxity with a monopolar or bipolar prosthesis in place. Also, both types of prostheses were effective in improving valgus and external rotation laxity to the elbow, regardless of coronoid status. With collateral ligaments reconstructed, no large kinematic differences were noted between implants regardless of the varus-valgus position or whether the coronoid was fractured or fixed.

Clinical Relevance

The data from our cadaveric model support the use of either implant type in terrible triad injuries if the collateral ligaments are intact or reconstructed.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Athwal GS, Frank SG, Grewal R, Faber KJ, Johnson J, King GJ. Determination of correct implant size in radial head arthroplasty to avoid overlengthening: surgical technique. J Bone Joint Surg Am. 2010;92(suppl 1, pt 2):250–257.

    PubMed  Google Scholar 

  2. Berschback JC, Lynch TS, Kalainov DM, Wysocki RW, Merk BR, Cohen MS. Clinical and radiographic comparisons of two different radial head implant designs. J Shoulder Elbow Surg. 2013;22:1108–1120.

    Article  PubMed  Google Scholar 

  3. Burkhart KJ, Mattyasovszky SG, Runkel M, Schwarz C, Küchle R, Hessmann MH, Rommens PM, Lars MP. Mid- to long-term results after bipolar radial head arthroplasty. J Shoulder Elbow Surg. 2010;19:965–972.

    Article  PubMed  Google Scholar 

  4. Chanlalit C, Shukla DR, Fitzsimmons JS, An KN, O’Driscoll SW. The biomechanical effect of prosthetic design on radiocapitellar stability in a terrible triad model. J Orthop Trauma. 2012;26:539–544.

    Article  PubMed  Google Scholar 

  5. Chanlalit C, Shukla DR, Fitzsimmons JS, Thoreson AR, An KN, O’Driscoll SW. Radiocapitellar stability: the effect of soft tissue integrity on bipolar versus monopolar radial head prostheses. J Shoulder Elbow Surg. 2011;20:219–225.

    Article  PubMed  Google Scholar 

  6. Doornberg JN, Parisien R, van Duijn PJ, Ring D. Radial head arthroplasty with a modular metal spacer to treat acute traumatic elbow instability. J Bone Joint Surg Am. 2007;89:1075–1080.

    Article  PubMed  Google Scholar 

  7. Harrington IJ, Sekyi-Otu A, Barrington TW, Evans DC, Tuli V. The functional outcome with metallic radial head implants in the treatment of unstable elbow fractures: a long-term review. J Trauma. 2001;50:46–52.

    Article  CAS  PubMed  Google Scholar 

  8. Hartzler RU, Llusa-Perez M, Steinmann SP, Morrey BF, Sanchez-Sotelo J. Transverse Coronoid Fracture: When Does It Have to Be Fixed? Clin Orthop Relat Res. 2014 March 1 [Epub ahead of print].

  9. Herald J, O’Driscoll S. Complete dissociation of a bipolar radial head prosthesis: a case report. J Shoulder Elbow Surg. 2008;17:e22–e23.

    Article  PubMed  Google Scholar 

  10. Leigh WB, Ball CM. Radial head reconstruction versus replacement in the treatment of terrible triad injuries of the elbow. J Shoulder Elbow Surg. 2012;21:1336–1341.

    Article  PubMed  Google Scholar 

  11. Moon JG, Berglund LJ, Zachary D, An KN, O’Driscoll SW. Radiocapitellar joint stability with bipolar versus monopolar radial head prostheses. J Shoulder Elbow Surg. 2009;18:779–784.

    Article  PubMed  Google Scholar 

  12. Moungondo F, El Kazzi W, van Riet R, Feipel V, Rooze M, Schuind F. Radiocapitellar joint contacts after bipolar radial head arthroplasty. J Shoulder Elbow Surg. 2010;19:230–235.

    Article  PubMed  Google Scholar 

  13. Pollock JW, Brownhill J, Ferreira L, McDonald CP, Johnson J, King G. The effect of anteromedial facet fractures of the coronoid and lateral collateral ligament injury on elbow stability and kinematics. J Bone Joint Surg Am. 2009;91:1448–1458.

    Article  PubMed  Google Scholar 

  14. Pomianowski S, Morrey BF, Neale PG, Park MJ, O’Driscoll SW, An KN. Contribution of monoblock and bipolar radial head prostheses to valgus stability of the elbow. J Bone Joint Surg Am. 2001;83:1829–1834.

    PubMed  Google Scholar 

  15. Popovic N, Lemaire R, Georis P, Gillet P. Midterm results with a bipolar radial head prosthesis: radiographic evidence of loosening at the bone-cement interface. J Bone Joint Surg Am. 2007;89:2469–2476.

    Article  PubMed  Google Scholar 

  16. Schneeberger AG, Sadowski MM, Jacob HA. Coronoid process and radial head as posterolateral rotatory stabilizers of the elbow. J Bone Joint Surg Am. 2004;86:975–982.

    PubMed  Google Scholar 

  17. Wu G, van der Helm FC, Veeger HE, Makhsous M, Van Roy P, Anglin C, Nagels J, Karduna AR, McQuade K, Wang X, Werner FW, Buchholz B; International Society of Biomechanics. ISB recommendation on definitions of joint coordinate systems of various joints for the reporting of human joint motion. Part II. Shoulder, elbow, wrist and hand. J Biomech. 2005;38:981–992.

    Article  CAS  PubMed  Google Scholar 

  18. Yian E, Steens W, Lingenfelter E, Schneeberger AG. Malpositioning of radial head prostheses: an in vitro study. J Shoulder Elbow Surg. 2008;17:663–670.

    Article  PubMed  Google Scholar 

  19. Zunkiewicz MR, Clemente JS, Miller MC, Baratz ME, Wysocki RW, Cohen MS. Radial head replacement with a bipolar system: a minimum 2-year follow-up. J Shoulder Elbow Surg. 2012;21:98–104.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Bernard F. Morrey MD.

Additional information

Each author certifies that he or she, or a member of his or her immediate family, has no commercial associations (eg, consultancies, stock ownership, equity interest, patent/licensing arrangements, etc) that might pose a conflict of interest in connection with the submitted article.

All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research ® editors and board members are on file with the publication and can be viewed on request.

Clinical Orthopaedics and Related Research ® neither advocates nor endorses the use of any treatment, drug, or device. Readers are encouraged to always seek additional information, including FDA approval status, of any drug or device before clinical use.

Each author certifies that his or her institution approved the human protocol for this investigation and that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.

This work was performed at Mayo Clinic, Rochester, MN, USA.

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hartzler, R.U., Morrey, B.F., Steinmann, S.P. et al. Radial Head Reconstruction in Elbow Fracture-Dislocation: Monopolar or Bipolar Prosthesis?. Clin Orthop Relat Res 472, 2144–2150 (2014). https://doi.org/10.1007/s11999-014-3672-0

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11999-014-3672-0

Keywords

Navigation