Retrieved Highly Crosslinked UHMWPE Acetabular Liners Have Similar Wear Damage as Conventional UHMWPE

  • Symposium: Papers Presented at the Hip Society Meetings 2010
  • Published:
Clinical Orthopaedics and Related Research®

Abstract

Background

Highly crosslinked UHMWPE is associated with increased wear resistance in hip simulator and clinical studies. Laboratory and case studies, however, have described rim fracture in crosslinked acetabular liners. Controversy exists, therefore, on the relative merits of crosslinked liners over conventional liners in terms of wear performance versus resistance to fatigue cracking.

Questions/purposes

We asked whether crosslinked liners would show less surface damage than conventional liners but would be more susceptible to fatigue damage.

Methods

We examined 36 conventional UHMWPE and 39 crosslinked UHMWPE retrieved implants with similar patient demographics and identical design for evidence of wear damage, including articular surface damage, impingement, screw-hole creep, and rim cracks.

Results

We observed no difference in wear damage scores for the two liners. Conventional liners more frequently impinged but were more often elevated with smaller head sizes. We observed creep in approximately 70% of both types of liners. Incipient rim cracks were found in five crosslinked liners, and one liner had a rim fracture. Only one conventional liner had an incipient rim crack.

Conclusions

Contrary to our expectation, damage was similar between crosslinked and conventional UHMWPE liners. Moreover, the 15% occurrence (six of 39) of incipient or complete fractures in crosslinked liners as compared with a 3% occurrence (one of 36) in conventional liners may have implications for the long-term performance of crosslinked liners. Longer-term studies will be necessary to establish the fate of rim cracks and thus the overall clinical fatigue performance of crosslinked liners.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2A–B
Fig. 3
Fig. 4A–B
Fig. 5

Similar content being viewed by others

References

  1. Bradford L, Baker DA, Graham J, Chawan A, Ries MD, Pruitt LA. Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners. J Bone Joint Surg Am. 2004;86:1271–1282.

    PubMed  Google Scholar 

  2. Bragdon CR, Barrett S, Martell JM, Greene ME, Malchau H, Harris WH. Steady-state penetration rates of electron beam-irradiated, highly cross-linked polyethylene at an average 45-month follow-up. J Arthroplasty. 2006;7:935–943.

    Article  Google Scholar 

  3. Calvert GT, Devane PA, Fielden J, Adams K, Horne JG. A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty. J Arthroplasty. 2009;4:505–510.

    Article  Google Scholar 

  4. Cole JC, Lemons JE, Eberhardt AW. Gamma irradiation alters fatigue-crack behavior and fracture toughness in 1900H and GUR 1050 UHMWPE. J Biomed Mater Res. 2002;5:559–566.

    Article  Google Scholar 

  5. Collier JP, Currier BH, Kennedy FE, Currier JH, Timmins GS, Jackson SK, Brewer RL. Comparison of cross-linked polyethylene materials for orthopaedic applications. Clin Orthop Relat Res. 2003;414:289–304.

    Article  PubMed  Google Scholar 

  6. Digas G, Karrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop. 2007;6:746–754.

    Article  Google Scholar 

  7. Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. The Otto Aufranc Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res. 2004;429:6–16.

    Article  PubMed  Google Scholar 

  8. Estok DM 2nd, Burroughs BR, Muratoglu OK, Harris WH. Comparison of hip simulator wear of 2 different highly cross-linked ultra high molecular weight polyethylene acetabular components using both 32- and 38-mm femoral heads. J Arthroplasty. 2007;4:581–589.

    Article  Google Scholar 

  9. Furmanski J, Anderson M, Bal S, Greenwald AS, Halley D, Penenberg B, Ries M, Pruitt L. Clinical fracture of cross-linked UHMWPE acetabular liners. Biomaterials. 2009;29:5572–5582.

    Article  Google Scholar 

  10. Gencur SJ, Rimnac CM, Kurtz SM. Fatigue crack propagation resistance of virgin and highly crosslinked, thermally treated ultra-high molecular weight polyethylene. Biomaterials. 2006;8:1550–1557.

    Article  Google Scholar 

  11. Halley D, Glassman A, Crowninshield RD. Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head. A case report. J Bone Joint Surg Am. 2004;86:827–830.

    PubMed  Google Scholar 

  12. Hermida JC, Bergula A, Chen P, Colwell CW Jr, D’Lima DD. Comparison of the wear rates of twenty-eight and thirty-two-millimeter femoral heads on cross-linked polyethylene acetabular cups in a wear simulator. J Bone Joint Surg Am. 2003;85:2325–2331.

    PubMed  Google Scholar 

  13. Holley KG, Furman BD, Babalola OM, Lipman JD, Padgett DE, Wright TM. Impingement of acetabular cups in a hip simulator: comparison of highly cross-linked and conventional polyethylene. J Arthroplasty. 2005;7(Suppl 3):77–86.

    Article  Google Scholar 

  14. Hood RW, Wright TM, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res. 1983;5:829–842.

    Article  Google Scholar 

  15. Kelly NH, Rajadhyaksha AD, Wright TM, Maher SA, Westrich GH. High stress conditions do not increase wear of thin highly crosslinked UHMWPE. Clin Orthop Relat Res. 2010;2:418–423.

    Article  Google Scholar 

  16. Kligman M, Furman BD, Padgett DE, Wright TM. Impingement contributes to backside wear and screw-metallic shell fretting in modular acetabular cups. J Arthroplasty. 2007;2:258–264.

    Article  Google Scholar 

  17. Lazzarini AM, Cottrell JM, Padgett DE, Wright TM. Remelting of highly cross-linked polyethylene worn under laboratory conditions. Clin Orthop Relat Res. 2007;465:128–132.

    PubMed  Google Scholar 

  18. McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.

    Article  PubMed  Google Scholar 

  19. Mendenhall S. Hospital resources and implant cost management—a 2008 update. Orthop Network News. 2009;20:15.

    Google Scholar 

  20. Muratoglu OK, Bragdon CR, O’Connor D, Perinchief RS, Estok DM 2nd, Jasty M, Harris WH. Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty. 2001;8(Suppl 1):24–30.

    Article  Google Scholar 

  21. Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award. J Arthroplasty. 2001;2:149–160.

    Article  Google Scholar 

  22. Muratoglu OK, Greenbaum ES, Bragdon CR, Jasty M, Freiberg AA, Harris WH. Surface analysis of early retrieved acetabular liners. J Arthroplasty. 2004;19:68–77.

    Article  PubMed  Google Scholar 

  23. Muratoglu OK, Ruberti J, Melotti S, Spiegelberg SH, Greenbaum ES, Harris WH. Optical analysis of surface changes on early retrievals of highly cross-linked and conventional polyethylene tibial inserts. J Arthroplasty. 2003;7(Suppl 1):42–47.

    Article  Google Scholar 

  24. Muratoglu OK, Wannomae K, Christenson S, Rubash H, Harris WH. Ex vivo wear of conventional and cross-linked polyethylene acetabular liners. Clin Orthop Relat Res. 2005;438:158–164.

    Article  PubMed  Google Scholar 

  25. Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232.

    CAS  PubMed  Google Scholar 

  26. Pruitt LA. Deformation, yielding, fracture and fatigue behavior of conventional and highly cross-linked ultra high molecular weight polyethylene. Biomaterials. 2005;8:905–915.

    Article  Google Scholar 

  27. Schmalzried TP, Callaghan JJ. Current Concepts Review: wear in total hip and knee replacements. J Bone Joint Surg Am. 1999;81:115–136.

    CAS  PubMed  Google Scholar 

  28. Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty: a study of retrieved acetabular components. J Arthroplasty. 2005;4:427–435.

    Article  Google Scholar 

  29. Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;10:2212–2217.

    Article  Google Scholar 

Download references

Acknowledgments

We thank Stephen Kim, MD, for his assistance with the radiographic alignment measurements and Joseph Nguyen, MPH, for his assistance with the statistical analysis.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Timothy M. Wright PhD.

Additional information

One or more of the authors (MLP) has or may receive payments or benefits from a commercial entity related to this work.

About this article

Cite this article

Schroder, D.T., Kelly, N.H., Wright, T.M. et al. Retrieved Highly Crosslinked UHMWPE Acetabular Liners Have Similar Wear Damage as Conventional UHMWPE. Clin Orthop Relat Res 469, 387–394 (2011). https://doi.org/10.1007/s11999-010-1552-9

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11999-010-1552-9

Keywords

Navigation