THA With Delta Ceramic on Ceramic: Results of a Multicenter Investigational Device Exemption Trial

  • Symposium: Papers Presented at the Hip Society Meetings 2009
  • Published:
Clinical Orthopaedics and Related Research®

An Erratum to this article was published on 07 January 2010

Abstract

Although the published studies on the outcomes of total hip arthroplasty (THA) performed with currently available ceramic components show high survivorship and low bearing wear at midterm followup, concern over ceramic fracture and squeaking persist. For these reasons, the use of ceramic is limited. Recently, a new alumina matrix composite material (Delta ceramic) with improved material properties was developed to address these concerns. We report the early outcomes and complications of a prospective, randomized, multicenter trial of 263 patients (264 hips) at eight centers, comparing a Delta ceramic-on-ceramic (COC) articulation with a Delta ceramic head-crosslinked polyethylene bearing combination (COP). There were 177 COC hips and 87 COP hips. Complications were reported for all patients, whereas clinical and radiographic results were provided for the 233 patients with minimum 2-year followup (average, 31.2 months; range, 21–49 months). The Harris hip scores and clinical, radiographic, and survivorship outcomes were similar in both groups. There were four (2%) revisions in the COC group and two (2%) in the COP group. We encountered three intraoperative ceramic liner-related events. In addition, one patient receiving the COC underwent revision for chipping of the ceramic liner, and a second had ceramic fragmentation on followup radiographs but has not undergone revision. These liner related complications remain a concern. No patient reported squeaking in either group; this leaves us hopeful the new material will lessen the frequency of squeaking. In the short term, the Delta COC articulation provided similar functional scores and survivorship and complication rates with the ceramic head mated with crosslinked polyethylene.

Level of Evidence: Level I, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Anwar I, Bhatnagar G, Atrah S. Delayed catastrophic failure of a ceramic head in hybrid total hip arthroplasty. J Arthroplasty. 2009;24:158.e5–8.

    Google Scholar 

  2. Back DL, Dalziel R, Young D, Shimmin A. Early results of primary Birmingham hip resurfacings. An independent prospective study of the first 230 hips. J Bone Joint Surg Br. 2005;87:324–329.

    Article  CAS  PubMed  Google Scholar 

  3. Baek SH, Kim SY. Cementless total hip arthroplasty with alumina bearings in patients younger than fifty with femoral head osteonecrosis. J Bone Joint Surg Am. 2008;90:1314–1320.

    Article  PubMed  Google Scholar 

  4. Barrack RL, Burak C, Skinner HB. Concerns about ceramics in THA. Clin Orthop Relat Res. 2004;429:73–79.

    Article  PubMed  Google Scholar 

  5. Berry DJ, von Knoch M, Schleck CD, Harmsen WS. Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2005;87:2456–2463.

    Article  PubMed  Google Scholar 

  6. Capello WN, D’Antonio JA, Feinberg JR, Manley MT, Naughton M. Ceramic-on-ceramic total hip arthroplasty: update. J Arthroplasty. 2008;23:39–43.

    Article  PubMed  Google Scholar 

  7. Clarke IC, Gustafson A. Clinical and hip simulator comparisons of ceramic-on-polyethylene and metal-on-polyethylene wear. Clin Orthop Relat Res. 2000;379:34–40.

    Article  PubMed  Google Scholar 

  8. Colwell CW Jr, Hozack WJ, Mesko JW, D’Antonio JA, Bierbaum BE, Capello WN, Jaffe WL, Mai KT. Ceramic-on-ceramic total hip arthroplasty early dislocation rate. Clin Orthop Relat Res. 2007;465:155–158.

    PubMed  Google Scholar 

  9. D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.

    Article  PubMed  Google Scholar 

  10. Dorey F, Amstutz HC. Survivorship analysis in the evaluation of joint replacement. J Arthroplasty. 1986;1:63–69.

    Article  CAS  PubMed  Google Scholar 

  11. Dorlot JM, Christel P, Meunier A. Wear analysis of retrieved alumina heads and sockets of hip prostheses. J Biomed Mater Res. 1989;23:299–310.

    Article  CAS  PubMed  Google Scholar 

  12. Engh CA Jr, Claus AM, Hopper RH Jr, Engh CA. Long-term results using the anatomic medullary locking hip prosthesis. Clin Orthop Relat Res. 2001;393:137–146.

    Article  PubMed  Google Scholar 

  13. Engh CA Jr, Stepniewski AS, Ginn SD, Beykirch SE, Sychterz-Terefenko CJ, Hopper RH Jr, Engh CA. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty. 2006;21:17–25.

    Article  PubMed  Google Scholar 

  14. Garino JP. Modern ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop Relat Res. 2000;379:41–47.

    Article  PubMed  Google Scholar 

  15. Ha YC, Kim SY, Kim HJ, Yoo JJ, Koo KH. Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty. Clin Orthop Relat Res. 2007;458:106–110.

    PubMed  Google Scholar 

  16. Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L. Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study. J Bone Joint Surg Am. 2002;84:69.

    PubMed  Google Scholar 

  17. Hamouche D, Nich C, Bizot P, Meunier A, Nizard R, Sedel L. Fractures of ceramic bearings: history and present status. Clin Orthop Relat Res. 2003;417:19–26.

    Google Scholar 

  18. Heck DA, Partridge CM, Reuben JD, Lanzer WL, Lewis CG, Keating EM. Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty. 1995;10:575–580.

    Article  CAS  PubMed  Google Scholar 

  19. Heros RJ, Willman G. Ceramics in total hip arthroplasty: history, mechanical properties, clinical results, and current manufacturing state of the art. Semin Arthroplasty. 1998;9:114–122.

    Google Scholar 

  20. Hwang DS, Kim YM, Lee CH. Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty. 2007;22:468–471.

    Article  PubMed  Google Scholar 

  21. Koo KH, Ha YC, Jung WH, Kim SR, Yoo JJ, Kim HJ. Isolated fracture of the ceramic head after third-generation alumina-on-alumina total hip arthroplasty. J Bone Joint Surg Am. 2008;90:329–336.

    Article  PubMed  Google Scholar 

  22. Langdown AJ, Pickard RJ, Hobbs CM, Clarke HJ, Dalton DJ, Grover ML. Incomplete seating of the liner with the Trident acetabular system: a cause for concern? J Bone Joint Surg Br. 2007;89:291–295.

    Article  CAS  PubMed  Google Scholar 

  23. Lusty PJ, Tai CC, Sew-Hoy RP, Walter WL, Walter WK, Zicat BA. Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2676–2683.

    Article  CAS  PubMed  Google Scholar 

  24. Maher SA, Lipman JD, Curley LJ, Gilchrist M, Wright TM. Mechanical performance of ceramic acetabular liners under impact conditions. J Arthroplasty. 2003;18:936–941.

    Article  PubMed  Google Scholar 

  25. McCarthy MJ, Halawa M. Lining up the liner: 2 case reports of early ceramic liner fragmentation. J Arthroplasty. 2007;22:1217–1222.

    Article  PubMed  Google Scholar 

  26. McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.

    Article  CAS  PubMed  Google Scholar 

  27. Min BW, Song KS, Kang CH, Bae KC, Won YY, Lee KY. Delayed fracture of a ceramic insert with modern ceramic total hip replacement. J Arthroplasty. 2007;22:136–139.

    Article  PubMed  Google Scholar 

  28. Moore KD, Beck PR, Petersen DW, Cuckler JM, Lemons JE, Eberhardt AW. Early failure of a cross-linked polyethylene acetabular liner. A case report. J Bone Joint Surg Am. 2008;90:2499–2504.

    Article  PubMed  Google Scholar 

  29. Murphy SB, Ecker TM, Tannast M. Two- to 9-year clinical results of alumina ceramic-on-ceramic THA. Clin Orthop Relat Res. 2006;453:97–102.

    Article  PubMed  Google Scholar 

  30. Park YS, Hwang SK, Choy WS, Kim YS, Moon YW, Lim SJ. Ceramic failure after total hip arthroplasty with an alumina-on-alumina bearing. J Bone Joint Surg Am. 2006;88:780–787.

    Article  PubMed  Google Scholar 

  31. Poggie RA, Turgeon TR, Coutts RD. Failure analysis of a ceramic bearing acetabular component. J Bone Joint Surg Am. 2007;89:367–375.

    Article  PubMed  Google Scholar 

  32. Ranawat AS, Ranawat CS. The squeaking hip: a cause for concern—agrees. Orthopedics. 2007;30:738–743.

    PubMed  Google Scholar 

  33. Restrepo C, Parvizi J, Kurtz SM, Sharkey PF, Hozack WJ, Rothman RH. The noisy ceramic hip: is component malpositioning the cause? J Arthroplasty. 2008;23:643–649.

    Article  PubMed  Google Scholar 

  34. Rhoads DP, Baker KC, Israel R, Greene PW. Fracture of an alumina femoral head used in ceramic-on-ceramic total hip arthroplasty. J Arthroplasty. 2008;23:1239.e25–30.

    Google Scholar 

  35. Rosneck J, Klika A, Barsoum W. A rare complication of ceramic-on-ceramic bearings in total hip arthroplasty. J Arthroplasty. 2008;23:311–313.

    Article  PubMed  Google Scholar 

  36. Squire M, Griffin WL, Mason JB, Peindl RD, Odum S. Acetabular component deformation with press-fit fixation. J Arthroplasty. 2006;21:72–77.

    Article  PubMed  Google Scholar 

  37. Stewart TD, Tipper JL, Insley G, Streicher RM, Ingham E, Fisher J. Long-term wear of ceramic matrix composite materials for hip prostheses under severe swing phase microseparation. J Biomed Mater Res B Appl Biomater. 2003;66:567–573.

    Article  PubMed  Google Scholar 

  38. Sugano N, Nishii T, Miki H, Yoshikawa H, Sato Y, Tamura S. Mid-term results of cementless total hip replacement using a ceramic-on-ceramic bearing with and without computer navigation. J Bone Joint Surg Br. 2007;89:455–460.

    Article  CAS  PubMed  Google Scholar 

  39. Tateiwa T, Clarke IC, Williams PA, Garino J, Manaka M, Shishido T, Yamamoto K, Imakiire A. Ceramic total hip arthroplasty in the United States: safety and risk issues revisited. Am J Orthop. 2008;37:E26–E31.

    PubMed  Google Scholar 

  40. Toran MM, Cuenca J, Martinez AA, Herrera A, Thomas JV. Fracture of a ceramic femoral head after ceramic-on-ceramic total hip arthroplasty. J Arthroplasty. 2006;21:1072–1073.

    Article  PubMed  Google Scholar 

  41. Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2212–2217.

    Article  PubMed  Google Scholar 

  42. Walter WL, Insley GM, Walter WK, Tuke MA. Edge loading in third generation alumina ceramic-on-ceramic bearings: stripe wear. J Arthroplasty. 2004;19:402–413.

    Article  PubMed  Google Scholar 

  43. Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22:496–503.

    Article  PubMed  Google Scholar 

  44. Willmann G. Ceramic femoral head retrieval data. Clin Orthop Relat Res. 2000;379:22–28.

    Article  PubMed  Google Scholar 

  45. Willmann G. Fiction and facts concerning the reliability of ceramics in THR. In: Zippel H, Dietrich M, eds. Bioceramics in Joint Arthroplasty. Berlin, Germany: Steinkopff Verlag; 2003:193–196.

  46. Yang CC, Kim RH, Dennis DA. The squeaking hip: a cause for concern—disagrees. Orthopedics. 2007;30:739–742.

    PubMed  Google Scholar 

  47. Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ. Alumina-on-alumina total hip arthroplasty. A five-year minimum follow-up study. J Bone Joint Surg Am. 2005;87:530–535.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

We thank additional surgeons and study coordinators (listed respectively by site) who were critical to the study design, enrollment, and followup of this patient cohort: William L. Bargar, MD, and Andrea Hankins (Sacramento, CA); C. Anderson Engh, Jr, MD, Charles A. Engh, Sr, MD, and Supatra Sritulanondha, MPH (Alexandria, VA); Theodore Firestone, MD, and Wayne McBride (Scottsdale, AZ); Robert Molloy, MD, and Lisa Hegemier (Cleveland, OH); Charles Nelson, MD, Craig Isrealite, MD, and Karunya Manikonda (Philadelphia, PA); and David Pollock, MD, and Charlene Cash (Winston-Salem, NC). Study coordinators affiliated with the authors include Donnis Rafferty, CCRC (Denver, CO) and Cheryl Seeley, RN, CCRC (Columbus, OH). Independent radiographic review was performed by Jonathan P. Garino, MD, and Denise Knox (Philadelphia, PA).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to William G. Hamilton MD.

Additional information

One or more of the authors (WGH, TJB, JP) are consultants for DePuy Orthopaedics; one or more authors (JPM, DAD) are consultants for DePuy Orthopaedics and receive royalties; one author (JAM) is an employee of DePuy Orthopaedics; and one author (DAD) receives institutional support from DePuy, Zimmer, and CeramTec. Each author certifies that his or her institution has approved the human protocol for this investigation, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.

An erratum to this article can be found at http://dx.doi.org/10.1007/s11999-009-1214-y

About this article

Cite this article

Hamilton, W.G., McAuley, J.P., Dennis, D.A. et al. THA With Delta Ceramic on Ceramic: Results of a Multicenter Investigational Device Exemption Trial. Clin Orthop Relat Res 468, 358–366 (2010). https://doi.org/10.1007/s11999-009-1091-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11999-009-1091-4